VST COFFEE User manual

Sample Well
5 Character
LCD
Read Key
Calibrate Key
Mode Key
Battery Door
VST
COFFEE
Refractometer
User Guide
INTRODUCTION:
VST, Inc. would like to extend our appreciation to you for purchasing
this coffee refractometer. It is our mission to provide a quality product
at a very competitive price to our end users and we feel this instru-
ment will prove its value time and again to you.
FEATURES:
CLEANING:
Cleaning of the measurement surface and well should be performed
immediately after each sample reading. Never immerse the instru-
ment in any liquid. When the measurement surface and well have
been completely cleaned no residue should be present.
To properly clean the sample well and glass measurement surface
use a mild soap and water solution, Isopropyl Alcohol or Methanol
followed by a distilled water rinse and then thoroughly dry with a soft
lint and residue free cloth or a product such as KimTech tissues.
To clean the refractometer’s body use a soft cloth with a mild clean-
ing solution like window cleaner applied to it and wipe it clean. Never
use any harsh cleaning agents that will damage the instrument.
Never use any harsh solvents on the instrument that will adversely
affect the material that it is made from.
AUTOMATIC TEMPERATURE COMPENSATION:
Automatic temperature compensation corrects readings over a range
of temperatures. As an example: Samples taken within the working
temperature range of the instrument (15-30 Deg C) are corrected to
a reference temperature of 20 degrees C. Temperature correction is
essential because refractive index varies inversely with temperature.
This feature is used in lieu of temperature control of the prism and
sample. For this feature to be effective however it is necessary to al-
low the sample to temperature equilibrate to the ambient temperature
of the prism.
Finally, for the most accurate and consistent results, coffee TDS
readings should be taken within +/- 2 Deg C of the CALIBRATION
temperature using distilled water.
SIMPLE SET UP:
Install Batteries
1. Remove battery door by rotating it the counter-clockwise to the
OPEN position until it stops and lift it off. Do not apply excessive
force.
2. Install batteries, negative side into cavity rst, positive facing
outward.
3. Re-install battery door by rotating the cover in the clock wise
direction until it is in the position shown below.
PRECAUTIONS:
1. Do not use Acetone, Acetone based products or other harsh sol
vents when cleaning the body or it’s components.
2. Using strong solvents such as N-Nimethylacetamide, cresols,
phenols and other tar acids should be avoided, Tetrahydrofuran
or some other lacquer thinners and their components will deterio
rate the prism seal and should not be used.
3. Do not apply excessive force when removing or installing the
battery door.
4. Do not store refractometer in extremely warm or cold environ-
ments.
OPEN
CLOSED
CALIBRATION USING DISTILLED OR DEIONIZED WATER:
1. Clean the measuring surface and well as outlined under “Cleaning”.
2. Apply approximately 0.2-mL DI water to the measurement surface.
3. Allow the water sample to temperature equilibrate to the tempera-
ture of the instrument. (20-25 Deg C, approx 1-2 minutes).
4. Press and hold the “CAL” key for at least 5 seconds.
5. While holding the “CAL” key press the “READ” key, then release
both keys.
6. Successful calibration will be indicated by the word “END” being
displayed. Dry the sample well using KimTech or similar tissues.
7. Select nD using the Mode Key, and take a measurement. Check
the nD at Temperature indicated against the DI Water Table, below.
If the nD does not agree, clean and dry the sample well and glass,
and place another sample of DI water, and wait 1-2 minutes to
equibrilate, then CALIBRATE. When END is reached, take another
nD reading. The reading should agree to the table, within +/-
0.0002 nD for the temperature indicated.
DISTILLED WATER REFRACTIVE INDEX TABLE:

SPECIFICATIONS:
RI Range: 1.3330 - 1.4465
TDS Range: 0.00 - 9.99 %
Accuracy: Typical +/- 0.06 % TDS
Warranted +/- 0.12 % TDS
+/- 0.0002 nD
Resolution: 0.01 % TDS
0.0001 nD, 0.1ºC
ATC: 20 Deg C for TDS, 15-30 Deg C
Calibration: Distilled Water
Prism: Glass
Illumination: 589nm LED
Dimensions: 54 x 27 x 100 mm / 2.13 x 1.06 x 3.9 inches
Power: Qty 2 - AAA Batteries, included
Battery life: 10,000 readings, Auto Off Sleep Mode
Ratings: IP65 Dustproof/Water Resistant, CE, RoHS, and
WEEE compliant.
Reading % TDS = Percent Total Dissolved Solids coffee.
Modes: TC-nD = Refractive index temperature corrected
to 20 degrees C.
nD = Refractive index and Temperature
Warranty: One year against manufacturing defects. Evidence
of tampering voids warranty. Contact VST directly
for warranty service.
This instrument is distributed exclusively by:
VST inc
www.mojotogo.us
13940010-101 REV D
ERROR CODES:
Err01 - No sample present. Add sample.
Err02 - Inadequate sample. Add additional sample.
Err03 - Sample exceeds the refractive index reading range.
Err04 - Sample temperature has not equilibrated. Allow more time for
temperature equilibration.
Err05 - Excessive ambient light. Cover sample well when reading.
Err06 - Excessive ambient light. Cover sample well when reading.
Err07 - Positive calibration error. Re-calibrate with distilled water.
Err08 - Negative calibration error. Re-calibrate with distilled water.
Err09 - Poor sample condition. Sample may not be able to be read.
Err10 - Index of sample is out of range of instrument.
Err5X - For any errors in this series contact VST Inc for technical
assistance.
SAMPLE METHOD:
Brew batch and stir. Draw a 4-5 mL sample. Transfer to cooling glass.
Allow 1 minute for sample
to cool.
Transfer cooled sample to
refractometer sample well.
Wait 30 seconds for
sample to equibrilate, then
press Read to take a %
TDS measurement.
MEASUREMENT METHOD:
Press Mode to change
measurement scale to nD.
Press Read for native mode
measurement as shown:
1.3354 nD at 19.8ºC.
Other VST Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Amprobe
Amprobe AM-530 user manual

Bender
Bender ISOMETER isoHR685W Series manual

Eastron
Eastron SDM72D quick start guide

RLS Wacon
RLS Wacon Aqua Inform Sycon 2501 Fe user manual

Martini Instruments
Martini Instruments MI 306 instruction manual

Endress+Hauser
Endress+Hauser Smartec CLD18 operating instructions