5.4.5 Method of Setting Linearization Data
Thelinearizationfunctionmeansafunctionthatchangestheslopeofstraightlinesinthe
relationshipbetweentheinputandindicationbycorrectingtherelationsatarbitrarypoints.
Linearizationdataaresetusingtheinputvalue(indicatedvaluebeforecorrection)andthe
output value(indicated valueafter correction) at each arbitrary point.
Note: The setup conditions are N-1 < N-2 … N-15 < N-16.
6. Control Functions
6.1 Hold Function
The Hold function temporarily retains the indication. The hold function is enabled by
shortcircuiting the HOLD and COM terminals or setting both terminalsto the same
voltagelevel. As a result thedisplay unit retains theindication givenat thatmoment.
6.2 Digital Zero Function
The DigitalZero function zerosthe indication given at an arbitrarytiming. There-
after, the function shows the amount of change from the point of zeroing. How-
ever, this function serves as an indication resetting function for a frequency mea-
suring unit. Thus, the Digital Zero function can be used to reset the indication
when there is no input signal at all.
Note that, the on/off control of theDigital Zero function can be achieved by means
of terminal control or front panel keys. In the case of terminal control, the Digital
Zero function is turned on byshortcircuiting theDZ and COM terminals or setting
both terminals to the same voltage level. The indication at that moment is zeroed.
In the case of control with the front panel keys, hold down the Mode key and press
the Increment keyfor about 1 second to zero the indication at that moment.
Note: Operation with the control terminals takes priority over operation
with the front panel keys. The Digital Zero function is disabled if the
control terminals are made to go through the off-on-off sequence with
the function enabled by means of the front panel keys.
Instruction Manual forA5000 Series
8/12
(1) Press the Mode key and the Enter key during
measurement.
(2) Press the Shift key a few times to move to the
linearization data menu.
(4) Press the Shift key (change digit) and the
Increment key (change numeric value) and then press
the Mode Key to after the number of data to be
corrected has been set.
&
Note: The decimal point in the selected digit
flashes.
(3) Press the Mode key to move to the setup for the
number of data to be corrected.
&
(10) When all the settings have been made, press the
Enter key to return to the measurement mode.
(8) Press the Mode key to display the setup for output
values of the data to be corrected.
Note: For a single display unit, the unit
automatically returns to N-01 indication if there is
no key operation for about 8 seconds. In this case,
press the Mode key to return to the numeric value
indication.
&
(5) For a single display unit, press the Mode key to
change to the parameter information indication.
(6) Press the Shift key (change digit) and Increment
key (change numeric value) and then press the Mode
key after the N-01 input value has been set.
RE
Note: For a single display unit, RE flashes when
the input value is set and DZflashes when the
output value is set.
DZ
&(7) Press the Shift key (change digit) and Increment
key (change numeric value) and then press the Mode
key after the N-01 output value has been set.
RE
Note: For a single display unit, the unit is
automatically returned to N-02 indication if there is
no key operation for about 8 seconds. In this case,
press the Mode key to return to the numeric value
indication.
(9) Press the Shift key (change digit) and Increment
key (change numeric value) and then press the Mode
key after the N-02 input value has been set.
Repeat steps (5) to (9) until all the settings have
been made.
Multi-display unit Single display unit
6.3 Peak Hold Function
The Peak Hold function retains one of the maximum(peak hold)/minimum (valley
hold)/maximum - minimum (peak-valley hold) values and provides output for that
value. Selection from these values is made using the condition data. The peak
hold function is enabled by shortcircuiting the PH and COM terminals or setting
both terminals to the same voltage level.
7. Output Function
7.1 Comparison Output Function
The A5000 series of unit meters is designed so that the two judgment values HI
and LO can be set for the measured (indicated) value to provide the results of
judgment as relay contact output. (This function is effective when the meter is
equipped with a comparison output unit.) For details on the contact ratings and
other specifications, refer to the section “Output Specifications.”
7.2 Analog Output Function
The A5000 series of unit meters can output an analog signal for an indicated
value (when the meter is equipped with an analog output unit). There are four
output ranges, 0 to 1 V/0 to 10 V/1 to 5 V/4 to 20 mA, from which a selection
can be made using the condition data. In addition, the analog output of the
A5000 series allows for arbitrary output scaling. This scaling can be achieved
by setting the indication value for an output of the maximum scale value (20 mA
for 4–20 mAoutput range) in the AOHI parameter of the scaling data.
7.3 RS-485 Interface Function
The A5000 series can be equipped with an RS-485 interface (when the meter is
provided with an RS-485 unit). For details on the RS-485 function, see the
separate manual on communication functions.
7.4 RS-232C Interface Function
TheA5000 series can be equipped with an RS-232C interface (when the meter is
provided with an RS-232C unit). For details on the RS-232C function, see the
separate manual on communication functions.
8. Specifications and External Dimensions
8.1 Input Specifications
8.1.1 DC Voltage Measuring Unit (range 11)
8.1.2 DC Voltage Measuring Unit (ranges 12 to 15)
8.1.3 DC Current Measuring Unit
Range Measurement
range Indication Highest
resolution Input
impedance
Maximum
permissible
input Accuracy
11 ±99.99mV Offset:±9999
Fullscale:0to±9999 10μV About
100MΩ ±100V ±(0.1%ofFS)
Inputcircuit:Singleendedtype
Maximumsamplingrate:12.5timespersecond
Noiserejectionratio:NMR(normalmoderejection)50dBormore(50or60Hz)
Operatingsystem:ΔΣconversion
Range Measurement
range Indication Highest
resolution Input
impedance
Maximum
permissible
input Accuracy
12 ±999.9mV
Offset:±9999
Fullscale:0to±9999
100μV About
100MΩ ±100V
±(0.1%ofFS)
Inputcircuit:Singleendedtype
13 ±9.999V 1mV About
1MΩ ±250V
±(0.15%ofFS)
±250V
±600V
About
10MΩ
About
10MΩ
10mV
100mV
14
15
±99.99V
±600V
Operatingsystem:ΔΣconversion
Maximumsamplingrate:12.5timespersecond
Noiserejectionratio:NMR(normalmoderejection)50dBormore(50or60Hz)
Range Measurement
range Indication Highest
resolution Input
impedance
Maximum
permissible
input Accuracy
23 ±9.999mA 1μA About
10Ω ±100mA
Offset:±9999
Fullscale:0to±9999
±(0.2%ofFS)
±(0.3%ofFS)
±500mA
±3A
About
1Ω
About
0.1Ω
10μA±99.99mA
±999.9mA 100μA
24
25
Inputcircuit:Singleendedtype
Operatingsystem:ΔΣconversion
Maximumsamplingrate:12.5timespersecond
Noiserejectionratio:NMR(normalmoderejection)50dBormore(50or60Hz)