Wells Johnson Company ELITE User manual

1
USER'S MANUAL
ELITE™
20-5061-00 (110V)
20-5066-00 (220V)
8000 South Kolb Road
Tucson, AZ 85756
Phone 800 528 1597
Fax 520885 1189
www.wellsgrp.com

2
Table Of Contents
General Description/Safety 3
Specifications 4
Checklist 4
Component Descriptions 5
Infusion Pump Head Loading 5
Pneumatic Foot Switch 5
Infusion Tubing 5
Infusion Handles 5
Infusion Needles and Cannulas 5
IV Pole 6
Infusion Controls 6
Vacuum Pumps 6
Vacuum Control 7
Volume Control 7
Foot Switch 7
Overflow Trap 8
Circuit Breaker 8
Set-up and Operation 9
Quick Fit™ Canister System 10
Sorenson™ Canister System 11
Baxter™ Canister System 12
Troubleshooting - Aspiration 13
Troubleshooting - Infusion 14
Maintenance 15
Parts List 16
Warranty Info 17

3
General Description
The ELITESeries™Aspirator and Infiltration System is a combination of the 2 most reliable
machines in the medical field - the Hercules™and the Wells Johnson Infiltration Pump.
Aspiration in the ELITE™is a high-powered suction system with three (3) oil-less piston vacuum
pumps. The system is designed to pull a vacuum on any closed container or any open process
system. Vacuum is accomplished by mounting the vacuum pumps in parallel. When the system
is connected, the air flow will be from the tip of the cannula through the tubing, collection
system, overflow trap, bacterial filter and finally to the pumps evacuate the air from the lines,
creating a vacuum pulling aspirated material in the same direction as the air flow. Aspirated
material continues to the collection system under normal suction until the cannula is exposed
to air again.
Infiltration in the ELITE™ incorporates a totally enclosed, permanent magnet D.C. motor that
has a long brush life, with permanent lubrication and quiet operation. The pump control unit
features a belt speed reduction system. Motor control is via SCR circuitry featuring indirect
velocity feedback for good load/line regulation which assures dependable speed (flow) control.
Power on/off and motor speed adjustment is accomplished using push buttons, with tactile and
auditory feedback, located on the front panel. “Soft-start” and smooth, gradual acceleration to
speed setting is provided when power is switched remotely from the unit.
SAFETY INFORMATION
Precautions:
Read all instructions carefully before operating equipment.
Motors are thermally protected and will automatically restart when protector resets.
Current leakage tests are performed and are certified within the AAMI maximum 300uA
limit for medical devices.
Cautions:
Avoid ingesting aspirated material into the pumps. Severe damage can occur to the
pumping system upon contact with aspirated material.
Federal Law restricts this device to sale by or on the order of a physician.
Warnings:
Always disconnect the power supply to the ELITE™ before servicing.
Safety checks should be scheduled according to purchaser’s policies.
Additional warnings and power requirements, which are part of the labeling, are located
on the back door of the ELITE™.

4
SPECIFICATIONS
Item number 20-5061-00 (110V)
20-5066-00 (220V)
Weight 118 lbs.
Dimensions 47.5” x 21” x 18”
Maximum Vacuum - Suction 29+ inches of Hg
Maximum Vacuum - Infusion 24 inches
Maximum flow rate - Suction 6.8 cubic feet per minute
Maximum flow rate - Infusion 1200ml/min (with specified tubing)
Display Unit - Suction inches Hg
Display Unit - Infusion no units
Overflow Protection Auto cutoff and in-line filter
Pump Type - Suction 2 cylinder piston
Number of pumps 3
Horsepower 1 hp
Pump Type -Infusion Peristaltic
Number of Rollers 3
Checklist
Carefully unpack the ELITE™from its shipping container and check that the following
accessories are included in the package and received in good condition:
ELITE™ (1) ●Canister Holder Ring (if needed)
Power Cord (1) ● Canister Adapter (if needed)
Aspirator Foot Switch (1) ●10 pcs Sterile, Single-Use Suction Tubing
Pneumatic Infusion Foot Switch (1) ●10 pcs Sterile, Single-Use Infusion Tubing
IV Pole (1) ● 10 pcs Single-Use Canister Liners
2 Bacterial Filters (1 installed in machine) ●2 Infiltration Handles
1 Overflow Trap Assembly (Pre-Installed) ●2 Infiltration Cannulas
Note: *** You will receive either 1 or 2 canisters, depending on which collection system you
ordered:
1. Baxter (1500cc's) = 1 Canister
2. Sorenson (1900cc's) = 2 canisters
3. Quick Fit (3000cc's) = 1 Canister
Call Customer Service Immediately to report any damaged or missing items: 1-800-528-1597

5
COMPONENT DESCRIPTIONS
Infusion Pump Head Loading
Sterile solution may be pumped through Infusion Tubing using the following technique to load
the pump head:
1. Spike IV bag
2. Clamp Infusion Tubing
3. Rotate the Pump Head lever counter clockwise to open the head.
4. Insert the enlarged center section of the infusion tubing into the pump head opening
over the rollers. The flow direction should be from left to right (when facing the side of
the machine) or from back to front (when facing the front of the machine)
5. Rotate the pump head lever clockwise to close the pump head and clamp the tubing.
a. **Make sure the tubing is not punctured by the black teeth when the
pump head is closed
Pneumatic Foot Switch
The foot switch tubing is connected to either barb (1 or 2) located below the Infusion Pump
Head on the side of the unit.
Infusion Tubing
Use sterile Infusion Tubing only. Manufacturer suggests only using approved Wells Johnson
tubing. Please refer to parts list. This tubing is precision-extruded and optically inspected to
insure top performance with the pump head. Other commercial tubing may not meet, or
perform, to specifications. Warning: Infusion tubing is sterile, single-use and not intended to
be re-sterilized under any circumstances to prevent cross-contamination.
Infusion Handles
Infusion Handles are designed to accept infusion tubing via luer lock. Standard luer lock needles
or infusion cannulas are connected to the handle via conventional luer lock. The Infusion
Handle may be gassed, autoclaved, or soaked. When sterilizing, handles cannot exceed 315
degrees.
Infusion Needles and Cannulas
The infusion technique is based on using large volumes of a very dilute concentration of local
anesthetic, which produces extremely efficient anesthesia, while minimizing blood loss.
All infiltration needles and cannulas are manufactured from surgical grade stainless steel with
luer lock fittings. Specify length and diameter when ordering.

6
IV Pole
Place the IV Pole into the bracket located on back of the ELITE™. The pole can be adjusted up or
down. When pole is in desired position, tighten the knob to hold it in place. Make sure the knob
is tight and the pole is secure before hanging a bag.
Infusion Controls
The variable speed drive provides ease of operation. After installing tubing as referred to in
"Infusion Pump Head Loading" section, press the "on" button located on the Infusion side of
the panel display. Make sure the ELITE™is plugged into the wall outlet. The infusion display
should be illuminated. Adjust the infusion flow rate by pressing the (+) or (-) buttons until the
desired flow rate is displayed.
Note: The display numbers are for reference only and do not directly
correspond to a unit of flow.
The ELITE™should remain outside the sterile field. The peristaltic type pump head assures the
sterile fluids remain within the tubing, never allowing direct contact with either the pump head
or pump control unit.
CAUTION: To prevent air infusion, always prime tubing prior to
administering any fluid to the patient.
Vacuum Pumps
The ELITE™uses three (3) 1/3 hp piston pumps
Oil-less, non-lubricated piston and cylinder
Permanently lubricated ball bearings
Stainless steel valves
Light weight aluminum components
Long-life, high performance piston seals
Twin fan cooling system
UL recognized motors
Thermal protected
Balanced for low vibration operation
Capacitor

7
Vacuum Control
Operate Vacuum Control by pressing the (+) or (-) button located on the right hand side of the
panel. Adjust up or down until the desired vacuum level is shown on the display.
Note: The ELITE™system has been calibrated at 2,500 feet above sea level to
26+ in Hg. The maximum vacuum level will decrease by 1" Hg for each 1,000
foot increase in elevation. Conversely, the maximum vacuum level will increase
by 1" Hg for each 1,000 foot decrease in elevation.
Volume Control
Operate the volume control by turning the knob located on the right hand side of the unit
between High and Low.
Note: Do not leave the Volume Control Knob in the off or Low position when first turning on
the aspirator pumps. This restricts the volume of air to nearly zero and since the pumps are
so powerful, the High Vacuum Pump could stall.
Foot Switch
The ELITE™ can be operated with or without a foot switch.
OPERATION WITH FOOT SWITCH
Plug the foot switch cord into the unit. Turn the Foot Switch and Power switches
to the "ON" position. Press the pedal and the ELITE™ will start. Press the pedal
again to stop the power of the unit.
OPERATION WITHOUT FOOT SWITCH
Turn the Power toggle switch to the ON position, leaving the toggle for the Foot
Switch OFF.
Note: The power switch on the back of the unit must be in the ON position before the unit
will start.

8
Overflow Trap
The Overflow Trap is designed to catch aspirated material which has by-passed, or overflowed,
from the main collection canister. The fluid entering the trap will cause the ball inside the
plastic jar to float up into the bell shaped tube, which will cause the vacuum to stop.
Note: The plastic ball only acts as safety mechanism. There is no guarantee that the ball will
completely stop the power of the unit.
If there is a problem with the overflow, please check the following:
1. The liner may be collapsing, causing material to overflow into the trap.
2. The collection system may be full.
3. The collection system may not be set up properly.
NO ASPIRATED MATERIAL SHOULD BE ENTERING THE OVERFLOW
TRAP AT ANYTIME.
Read the troubleshooting section of this manual for instructions to correct an overflow
situation.
Circuit Breaker
To protect the equipment and reduce shock hazard, a re-settable circuit breaker is installed on
the back of the unit. This is labeled as "RESET". If there is a voltage surge or the pump motor
overheats, the circuit breaker will pop out. To reset the ELITE™, turn the unit off, push in the
circuit breaker, and turn the unit on. If the unit fails to start, the unit may require a cool-down
period or technical repairs may be required.
WARNING: DO NOT LET TECHNICIANS BY-PASS THE CIRCUIT BREAKER

9
Setup and Operation
1. Carefully remove the ELITE™from the shipping container. Be sure to save the box and
packing material for convenient maintenance returns.
2. Read all instructions carefully before operating the unit.
3. After you have removed the ELITE™, open the accessory box and locate the power cord,
both foot switches and the IV pole. Plug the Infusion foot switch and aspirator foot
switch into the appropriately labeled ports. The aspiration footswitch should be pushed
in and twisted lightly clockwise to lock it into place. The aspiration foot switch is
optional as the vacuum pumps will operate without the foot switch. Install the IV pole
into the bracket on the back of the unit, tighten knob so pole is secure.
4. Plug the power cord into the back of the ELITE™, then into an appropriate power outlet.
Power on the machine by pressing the power switch located on the back of the unit.
5. To power each component of the ELITE™, press "ON" on the corresponding Infusion or
Aspiration side. Note: The controls are separate and control each component
separately.
6. To make sure the vacuum is working properly, place a finger over the tubing coming
from the side of the machine labeled "Vacuum Port" and remember the reading of the
vacuum display. Note: Keep your finger over the tubing until the vacuum level becomes
constant. Also, keep in mind that vacuum changes according to elevation.
7. After checking the vacuum, turn the ELITE™off.
NEXT:
8. The Overflow Trap and HEPA Filter come pre-installed in the ELITE™. You will need to
unscrew the canister from the Overflow Trap and remove the ball from the packing
material. Replace the canister with the ball inside and be sure to tighten!
Note: Change the HEPA Filter after 10 procedures or every 24hrs of use.
9. Locate the Collection System. Determine if the system is a Baxter™, Sorenson™or Quick Fit™
canister system, then read the applicable instructions included in this manual to set up the
collection system.
NEXT:
10. Set up the collection system as described on the following pages.
11. At this point, the following accessories should now be connected:
a. Power cord and foot switches
b. IV Pole - which is securely installed in the bracket
c. Canister –placed in the canister holder with liner intact
The unit is now ready for operation.

10
Setup and Operation of Collection System
Quick Fit™Canister System (3000cc's)
1. Place canister in the canister bracket of the ELITE™on the right side of the unit.
2. Insert a Quick Fit™liner into the canister and secure the lid.
Warning: Quick-Fit™ liners are single-use and not intended to be re-used under any
circumstances to prevent cross-contamination.
3. Connect the piece of tubing that is connected to the bracket to the top of the lid. (as
shown on picture)
4. Connect the 10” piece of silicon tubing from the Vacuum Port on the ELITE™, to the
bottom of the bracket on the Canister.
5. The sterile aspiration tubing will connect to the liner lid on the port labeled: Patient
Warning: Aspiration tubing is sterile, single-use and not intended to be re-sterilized
under any circumstances to prevent cross-contamination.
THE CANISTER SYSTEM IS NOW READY FOR OPERATION.
Caution: Aspirated material should not enter the overflow trap at any time.
If an overflow situation occurs, please do the following:
1. Turn off the power immediately.
2. Check the connections between accessories to make sure the unit was set up correctly.
3. Check for a crack in the canister.
4. Check the volume setting (should be on high).

11
Sorenson™ Canister System (1900cc's)
1. Place the Sorensen Adapter into the canister bracket of the ELITE™on the right side of
the unit. Next place the canister into the slot of the adapter.
2. Insert a Sorenson™liner into the canister and secure the lid.
Warning: Sorenson™ liners are single-use and not intended to be re-used under any
circumstances to prevent cross-contamination.
3. Connect the piece of tubing that is connected to the liner lid to one end of the Tee
connector.
4. Connect the 10” piece of silicon tubing from the Vacuum Port on the ELITE™, to the
bottom of the bracket on the Canister.
5. The sterile aspiration tubing will connect to the liner lid on the port labeled: Patient
Warning: Aspiration tubing is sterile, single-use and not intended to be re-sterilized
under any circumstances to prevent cross-contamination.
THE CANISTER SYSTEM IS NOW READY FOR OPERATION.
Caution: Aspirated material should not enter the overflow trap at any time.
If an overflow situation occurs, please do the following:
1. Turn off the power immediately.
2. Check the connections between accessories to make sure the unit was set up correctly.
3. Check for a crack in the canister.
4. Check the volume settings (should be on high).

12
Baxter Canister System (1500cc's)
1. Place canister in the canister bracket of the ELITE™on the right side of the unit.
2. Snap the red Baxter liner lid onto a soft-shell liner. Insert the hard-shell canister.
Warning: Baxter liners and lids are single-use and not intended to be re-used under
any circumstances to prevent cross-contamination.
3. Plug the red hose on the canister into the top of the liner lid on the port marked
Vacuum.
4. Attach the 9” piece of silicon tubing from the Vacuum Port on the ELITE™, to the red
spout on the side of the canister.
5. The sterile aspiration tubing will connect to the port labeled: Patient
Warning: Aspiration tubing is sterile, single-use and not intended to be re-sterilized
under any circumstances to prevent cross-contamination.
THE CANISTER SYSTEM IS NOW READY FOR OPERATION.
Caution: Aspirated material should not enter the overflow trap at any time.
If an overflow situation occurs, please do the following:
1. Turn off the power immediately.
2. Check the connections between accessories to make sure the unit was set up correctly.
3. Check for a crack in the canister.
4. Check the volume settings (should be on high).

13
Troubleshooting - Aspiration
No Suction
1. Turn Volume Control Knob to High. If this knob is on Low, there will be no noticeable
suction.
2. Check for a collapsing liner.
3. Disconnect tube from canister only that attaches to ELITE™vacuum port. Place finger
over the open end of the tube. If there is no suction, turn the unit off, remove finger
from hose, wait a moment, then turn the unit on again.
Precaution: Before starting any suction procedure, let the ELITE™ go to full vacuum when
switched on. During operation, if the ELITE™ is momentarily turned off, let the vacuum gauge
indicator return to “0” before restarting the machine. Air must fill the vacuum lines before all
the pumps will start again. Failure to follow this procedure can result in the loss of vacuum.
Overflow Situation
1. Monitor the collection system for fluid level. Do not overflow.
2. A collapsing liner can cause the aspirated material to overflow into the overflow trap.
3. Check for a crack in the canister. This can cause the liner to collapse, resulting in an
overflow situation. A collapsing liner bag does not mean that there is something wrong
with the ELITE™.
Caution: With proper operation and monitoring, at no time during any procedure,
should an overflow situation occur.
Broken Canister
1. Do not use a damaged canister. A damaged canister will cause the liner bag to
collapse, causing the aspirated material to overflow into the overflow trap.
Caution
Ingesting aspirated material into the pumping system can severely damage the pumps and void
the warranty. If material is going into the overflow trap, check the collection system
immediately to see if the liner bag is collapsing or is too full. If further assistance is needed, call
technical support at 800-528-1597.
Refer to the collection system section and instructions for proper set up.

14
Troubleshooting - Infusion
Infusion Pump Head
The black lever on each side of the pump head can be pressed inward, allowing it to slide up or
down to adjust to tubing size.
Sliding button upward loosens pump head grip on the infusion tubing. Tubing will have a
tendency to “walk”, or bunch-up, in the pump head if the buttons are adjusted too high.
Adjusting the button downward tightens the pump head grip on the tubing. Adjusting the
buttons toward the bottom of their excursions may restrict the flow rate.
NOTE: Adjustment of the buttons in the center of their excursion provides an optimum setting.

15
Maintenance - Aspiration
All Aspirators are thoroughly checked by the Quality Assurance Department before
being released for shipping. Inspect the unit upon receipt to insure the equipment was
not damaged during shipping. The vacuum power should also be checked.
Do not lubricate pumps. The oil-less components are sealed and packed with an
optimum level of lubricant, requiring no lubrication.
Do not let liquid or aspirated material enter the pumping system.
The mufflers (Part No. 16-5073) should be changed after approximately 80 hours of
operation, or whenever the white filter inside the muffler becomes black.
No scheduled maintenance is recommended. Follow the procedures established by your
engineering department.
Warnings are part of the labeling and are indicated on the ELITE™. The serial number is
stamped on a small metal tag, located on the back of the unit. Record the serial number.
In-services on new equipment can be provided by telephone with either your sales
representative, or our technical support staff. If additional information is needed, call
Customer Service or Technical Support at 800-528-1597.
Change HEPA Filter after 10 procedures.
Maintenance - Infusion
The speed control circuit has solid-state components which do not require service. An
excessive load on the system may, however, cause the printed circuit board mounted
fuse to blow. Indication of a broken fuse is when the power is applied to the drive, and
the speed control on/off button is ON and illuminated, however the motor does not
operate when the footswitch is pressed.
Exact motor brush and commutator life will depend on duty cycle and motor operation
speed. Brushes should be inspected every six months, or after 2,000 hours of operation,
whichever occurs first. Replace when less than 0.300” long. The commutator should be
periodically inspected and cleaned if necessary. If preventative maintenance is not
performed, excessive commutator wear or “bridging” between commutator segments
will cause excessive current through the controller circuit.

16
PARTS LIST
ITEM PART NO.
ELITE™ (110V) 20-5061-00
ELITE™ (220V) 20-5066-00
Power Cord (AC 110V) 20-5100-00
Power Cord (AC 22V) 20-5101-00
Aspiration
Foot Switch (Single) 20-5102-00
Vacuum Gauge 16-5030-00
HEPA Filters (case of 10) 20-5201-00
Overflow Trap (plastic) with 6” and 9” silicone tubing 17-5100-00
Overflow Ball 16-5107-00
6” Silicone Tubing 17-5101-00
9” Silicone Tubing 17-5106-00
Canister, Baxter™ 20-5153-00
Disposable, Single-Use Soft-Shell Liners, Baxter™ 20-5154-00
Baxter™ Canister Holder 20-5159-01
Canister, Sorenson™ 20-5157-00
Disposable, Single-Use Liners, Sorenson™ 20-5158-00
Sorenson™ Canister Holder 20-5159-00
Canister, Quick-Fit™ 20-5156-00
Disposable, Single-Use Liners, Quick-Fit™ 20-5155-00
Quick-Fit™ Canister Holder 20-5159-03
Standard Tubing, 9ft, sterile, single-use Baxter™ 24-5103-00
MB Tubing, 9ft, sterile, single-use Quick-Fit™ and Sorenson™ 24-5104-00
Flex B Aspiration Tubing, 10ft, sterile, single-use, Baxter™ 24-5102-02
Flex A Aspiration Tubing, 10ft, sterile, single-use Quick-Fit™ and Sorenson™ 24-5102-01
Vacuum Pumps, 1/3 hp, piston, hi-vac (2 ea) 18-5057-01
Vacuum Pump, 1/3 hp, piston, hi-vac (1 ea) 18-5058-01
Muffler (MIN) (3per unit) 16-5073-00
Tubing (Internal PVC) 17-5003-00
Infusion
Infusion Tubing, sterile, single-use disposable 24-6008-00
Multi Hole Infusion Cannulas (specify size) 20-1218-00
Spinal Needles, sterile, disposable
18g x 6” 24-6002-00
25g x 3” 24-6003-00
20g x 3 ½” 24-6004-00
Infusion Pump Head, single 18-4102-00
Infusion Pump Head, double 18-4102-01
Pneumatic Foot Switch, single 20-5102-00
Pneumatic Foot Switch, dual 20-5103-00
IV Pole 18-4200-01

17
WARRANTY
The Wells Johnson Company will void warranty on any products indicating negligence,
or noncompliance with operating and maintenance instructions.
Products are warranted against defect in material and workmanship for a period of one
(1) year from date of purchase. If repair or adjustment is necessary, and are not the
result of abuse or misuse, please return, freight prepaid, and correction will be
performed without charge.
All returns require prior authorization from the Wells Johnson Quality Assurance
Department.
Any aspirated material found to have been ingested into the pumping system will void
the warranty.
Retain the custom-made packing box with foam for returns.
For questions about the warranty, call the Quality Assurance Dept: 800-528-1597
For technical assistance call
Customer Service or Technical Support:
520-298-6069
800-528-1597
Table of contents
Popular Medical Equipment manuals by other brands

HeartSine
HeartSine SAM 350P AED Setup

Weinmann
Weinmann ULM CASE II Instructions for use

Welch Allyn
Welch Allyn Spot Vital Signs LXi Service manual

Skytron
Skytron ST9 Series OPERATORS, MAINTENANCE, PARTS

Stryker Medical
Stryker Medical Renaissance Series Maintenance manual

Siemens
Siemens MAMMOMAT 1000 Nova Maintenance Instruction

HEBU medical
HEBU medical ECO Accu HB 8870 Operating and service manual

Atmos
Atmos S 201 Thorax operating instructions

Symmetry Surgical
Symmetry Surgical The BOOKWALTER IMA manual

orlvision
orlvision FS2 Instructions for use

Getinge
Getinge MAQUET X'TEN user manual

Otto Bock
Otto Bock Genium X3 Instructions for use