1. Watch the video
2. Unpack
3. Remove from carry case
4. Insert Pad-Pak
5. Press On/Off, confirm
6. Check that status indicator
EN
Set-up
1. Removefrom packaging.
2. PlaceHeartSine samaritan PAD (AED) face
upon a flat surface and slide in Pad-Pak or
Pediatric-Pak.
3. Listenfor “double click” to indicate it is
properlyinserted.
4. Verifythat green status indicator is blinking,
indicatingthe AED is ready for use.
5. Turnon AED. You should hear expected
languageand should not hear a warning
message.Turn off AED.
Disposal after use
6. RemovePad-Pak. Cut wires. Stick pads together.
Disposeof pads as infectious waste. Recycle
lithiumbattery separately per local regulations.
Pad-Pakand Pediatric-Pak contain:
• Pairof disposable single use defibrillation pads
• LiMnO2(18V– 1500mAh) non-rechargeable
battery
Intended purpose
EachPad-Pak and Pediatric-Pakaccessor y
containsthe battery to power the Hear tSine
samaritanPAD and two electrode pads to provide
theelectrical connection to the patient’s chest for
deliveryof a defibrillation shock.
Indications for use
ThePad-Pak and Pediatric-Pak,when used with
theHeartSine samaritan PAD, are indicated for
useon victims of cardiac arrest who are exhibiting
thefollowing signs: unconscious, not breathing
andwithout circulation.
• Pad-Pak:Intended for use on adults or children
>8 years or > 25 kg (55 lb).
• Pediatric-Pak:Intended for use on children
betweenthe age of 1-8 years or < 25 kg (55 lb).
Contraindications
Donot use if patient is responsive or conscious.
Battery capacity
>60shocks or 6 hours battery use.
Warnings
DONOT USE if expiry date is exceeded.
DONOT USE if the pouch is opened or damaged.
Thismay result in the electrode gel being dry.
Cautions
Singleuse ONLY.Reuse may cause AED to be
unableto deliver therapy leading to a failure to
resuscitate.It may also lead to cross-infection
betweenpatients.
DONOT PULL the green tab on the Pad-Pak during
set-up.
Ifyou hear a warning message when you tur n
onyour HeartSine samaritan PAD, consult
thetroubleshooting section of the HeartSine
samaritanPAD user manual.
Forfull instructions for use and safety
informationrefer to your Hear tSine samaritan
PADuser manual. This also can be found at
https://heartsine.com
Notes
Pediatric-Pakis only suitable for use with
HeartSinesamaritan PADs with the symbol .
TheTSO/ETSO-cer tified Pad-Pakmust be
usedon board commercial fixed wing aircraft.
DONOT STORE the Pediatric-Pak next to
magneticallysensitive storage media as it contains
amagnetic component (surface strength 6500
gauss).
Pleaserepor t any serious incident that occurs
withthis device to Hear tSine Technologies,
Ltd.and to your national competent authority
orother local regulator y authority as per local
regulations.