WoMaster ES100 Series User manual

Quick Installation Guide
Outdoor Environment Air Quality
Inspection System
ES100 Series
RS485 Connection
1) Connect the cable to the device, connect the yellow wire to RS485-A,
and connect the blue wire to RS485-B.
2) When wiring the RS485 cable, pay attention to A/B the two lines cannot
be reversed.
3) Make sure the addresses of multiple devices on the Modbus setting do
not conflict.
*Note: RS485A line and RS485B line are strictly forbidden to contact with
the power supply line, otherwise the communication of the electrode will
be permanently damaged.
ES100 is an intergraded device and designed f or a l l t y pes o f
environmental monitoring. It can be used to detect multiple parameters of
temperature, humidity, PM2.5, PM10, wind speed and wind direction in
the environment. Each parameter is independent and high sensitivity,
users can freely integrate monitoring parameters. ES100 has the
characteristics of high precision and good stability, which is suitable for
various environmental monitoring.
The ES100 is equipped with a standard RS485 interface and supports the
Modbus RTU protocol, which can be integrated into the Internet of Things
and cloud interfaces, such as smart city boxes and wireless routers, for
real-time monitoring and analysis via remote smartphones or computers.
・Overview
www.womaster.eu
Louver Box Installation
The device is equipped with three mounting
screws at the bottom. The customer can use
screws to install the louver box to the bracket.
The schematic diagram is shown on the right.
・ES106 Installation
Power Input Connection
1) Connect the power cable to the device. Brown cable is the positive pole
of the power supply and black cable is the negative pole of the power
supply.
2) Connect the power wires to suitable DC Switching type power supply.
The input range is 10-30VDC.
Model
Description
ES102
Outdoor Environment Air Quality Inspection System. Integrated
external Wind Direction, Wind Speed sensors, Output: RS485
ES104
Outdoor Environment Air Quality Inspection System. Integrated
embedded Hum., Temp., PM2.5, PM10 sensors, Output:
RS485
ES106
Outdoor Environment Air Quality Inspection System. Integrated
embedded Hum., Temp., PM2.5, PM10 sensors, and external
Wind Direction, Wind Speed sensors, Output: RS485
ES102
ü1 x Wind Direction Sensor (Including cable)
ü1 x Wind Speed Sensor (Including cable)
ü1 x QIG
ü2 x Mounting Bracket
ES104
ü1 x Shutter
ü1 x QIG
ü1 x Mounting Bracket
ES106
ü1 x Shutter
ü1 x Wind Direction Sensor (Including cable)
ü1 x Wind Speed Sensor (Including cable)
ü1 x QIG
ü3 x Mounting Bracket
Optional Accessory (for detailed information please refer to the
Datasheet)
・Package Checklist
Wind Speed and Direction Installation
Install the wind speed and direction on the mounting
bracket. Four mounting holes are provided on the
bracket and attach the sensor to the bracket then
screw the bolts tightly on the bracket to make the
whole set of the sensors are installed properly. To
maintain the best level and ensuring the accuracy of
the wind speed and direction data, the mounting
bracket is easy to use and can withstand large
pressures.
Louver Box Installation
The device is equipped with three mounting
screws at the bottom. The customer can use
screws to install the louver box to the bracket.
The schematic diagram is shown on the right.
・ES104 Installation
RS485 Connection
1) Connect the cable to the device, connect the yellow wire to
RS485-A, and connect the blue wire to RS485-B.
2) When wiring the RS485 cable, pay attention to A/B the two lines
cannot be reversed.
3) Make sure the addresses of multiple devices on the Modbus
setting do not conflict.
*Note: RS485A line and RS485B line are strictly forbidden to
c o n t a c t w i t h t h e p o w e r s u p p l y l i n e , o t h e r w i s e t h e
communication of the electrode will be permanently damaged.
Power Input Connection
1) Connect the power cable to the device. Brown cable is the
positive pole of the power supply and black cable is the negative
pole of the power supply.
2) Connect the power wires to suitable DC Switching type power
supply. The input range is 10-30VDC.
Install the wind speed and direction on the mounting
bracket. Four mounting holes are provided on the
bracket and attach the sensor to the bracket then
screw the bolts tightly on the bracket to make the
whole set of the sensors are installed properly. To
maintain the best level and ensuring the accuracy of
the wind speed and direction data, the mounting
bracket is easy to use and can withstand large
pressures.
・ES102 Installation
NOTE: Attention to the installation, the
arrow on the device is pointing must
be consistent with the north direction,
so as to avoid measurement errors.
North
RS485 Connection
1) Connect the cable to the device, connect the yellow wire to RS485-A,
and connect the blue wire to RS485-B.
2) When wiring the RS485 cable, pay attention to A/B the two lines
cannot be reversed.
3) Make sure the addresses of multiple devices on the Modbus setting
do not conflict.
*Note: RS485A line and RS485B line are strictly forbidden to
c o n t a c t w i t h t h e p o w e r s u p p l y l i n e , o t h e r w i s e t h e
communication of the electrode will be permanently damaged.
Power Input Connection
1) Connect the power cable to the device. Brown cable is the positive
pole of the power supply and black cable is the negative pole of the
power supply.
2) Connect the power wires to suitable DC Switching type power supply.
The input range is 10-30VDC.
NOTE: Attention to the installation, the
arrow on the device is pointing must
be consistent with the north direction,
so as to avoid measurement errors.
North

www.womaster.eu
・Safety Precaution
ØThe equipment should be placed horizontally as much as possible to
ensure that the installation is perpendicular to the horizontal plane.
ØTry to stay away from high-power interference equipment to avoid
inaccurate measurement, such as inverters, motors, etc.
ØBefore connecting the DC power module to the device, make sure all
connections are established and then turn on the DC power.
V1.00 Mar 4, 2020 3160-0ES1000-01
・Protocol
© WoMaster Inc. All rights reserved.
Parameter
Description
Protocol
Modbus RTU
Data Bits
8 bit
Parity Bit
No
Stop Bit
1
Error Detection Code
CRC
Baud Rate
2400bps/4800bps/9600bps, default setting
is 9600bps
•Basic Parameter
•Data Format Definition
Adopt Modbus-RTU communication protocol, the format is as follows:
Initial structure ≥ 4 bytes of time
Address code = 1 byte
Function code = 1 byte
Data area = N bytes
Error check = 16-bit CRC code
End structure ≥ 4 bytes of time
Address code: for the function indication of the transmitter, this transmitter
only uses the function
Code 0x03 (read register data).
Data area: The data area is a specific address and is unique in the
communication network.
(Factory default 0x01).
Function code: command communication data sent by the host, pay
attention to 16bits data high
The byte is in front!
CRC code: Two-byte check code.
Adr.
Func.
Register
Start Adr.
Length
Low
Checksum
High
Checksum
1 byte
1 byte
2 byte
2 byte
1 byte
1 byte
•Inquiry Frame
Adr.
Func.
Effective
Number
First
Data
Area
Second
Data
Area
。。
N Data
Area
Low
Checksu
m
High
Checksu
m
1
byte
1
byte
1
byte
2
byte
2
byte
。。
2
byte
1
byte
1 byte
•Response Frame
•Device Stack List
・Appearance
Louver Box
Wind Speed Sensor
Wind Direction Sensor
・Support
At WoMaster, you can use the online service forms to request the
support. The submitted forms are stored in server for WoMaster team
member to assign tasks and monitor the status of your service. Please
feel free to write to help@womaster.eu if you encounter any problems.
WoMaster reserves the right to make changes to this QIG or to the
product hardware at any time without notice. It is the user’s responsibility
to determine whether there have been any such updates or amendments
herein.
Defects, malfunctions, or failures of the warranted Product(s) caused by
damage resulting from unforeseeable incidents (such as lightings, floods,
fire, etc.), environmental and atmospheric disturbances, other external
forces such as power line disturbances and surge, host computer
malfunction and virus, incorrect power input, or incorrect cabling, incorrect
grounding and damages caused by misuse, abuse and unauthorized
alteration or repair are not warranted.
・Disclaimer
・Warranty
1-year Global warranties are available for WoMaster products assuring
our customers that the products shall remain free from defects in
workmanship or materials and conform in all material respects to
WoMaster specifications, or Purchaser’s supplied and accepted
specifications. The warranty is limited to the repair and/or replacement, at
WoMaster' sole discretion, of the defective product during its warranty
period. The customer must obtain a Return Merchandise Authorization
(RMA) approval code prior to returning the defective Product to WoMaster
for service. The customer agrees to prepay shipping charges, to use the
original shipping container or equivalent, and to insure the Product or
assume the risk of loss or damage in transit. Repaired or replaced
products are warranted for ninety (90) days from the date of repair or
replacement, or for the remainder of the original product's warranty period,
whichever is longer.
Slave
ID
Register
Address
PLC or
Configurati
on Address
Content
Desc.
1
01F4 H
40501
Humidity
Range:0-100.0%RH;
Upload data 10 times real
data
1
01F5 H
40502
Temperature
Range:-40-80℃;
Upload data 10 times real
data
1
01F7 H
40504
PM2.5
Range:0-1000ug/m3;
Uploaded data is actual
value
1
01F8 H
40505
PM10
Range:0-1000ug/m3;
Uploaded data is actual
value
2
0000 H
40001
Instantaneous
Wind Speed
Range:0 ~ 70m/s
Upload data 10 times real
data
3
0001 H
40002
Wind
Direction
Integer angle value (0-359)
Uploaded data is actual
value
07D0 H
42001
Slave ID
1 ~ 254 (Factory Default 1)
07D1 H
42002
Baud Rate
0=2400 1=4800
2=9600(Default)
This manual suits for next models
3
Popular Measuring Instrument manuals by other brands

Multiplex
Multiplex MX8120 operating instructions

Keysight
Keysight E5063A Configuration guide

Primes
Primes Cube L Original instructions

Electro Industries
Electro Industries Shark 200 Installation & operation manual

FlowLine
FlowLine EchoSpan LU81-5101 Series quick start guide

Brigade
Brigade MDR 600 Series Installation and operation guide