
SI-BAM21-LC
World Precision Instruments 5
3. Connect the AC power adapter to the
Power Input
(labeled
12V DC, 2.5A
) on the back
panel of the
SI-BAM21-LC
.
4. Turn the
Power
switch on the back panel on (I).
5. Allow 30 minutes warm up time for the system to stabilize with the transducer.
OPERATING INSTRUCTIONS
Calibrating the SI-BAM21-LC
Before taking measurements, the
SI-BAM21-LC
must be calibrated. Each
SI-KG
force transducers shown responds linearly within its respective measurement ranges.
Consequently, the
SI-BAM21-LC
can be calibrated using only two reference points.
Force
Transducer
Force
Range
Range
(g)
Noise
(µN)
Compliance
(nm/mN)
Resonance
Frequency
SI-KG2 0-2 N 0-200 250 150 1.3 kHz
SI-KG2B 0-0.2 N 0-20 80 590 Hz
SI-KG4 0-50 mN 0-5 15 0.5 1.2 kHz
SI-KG4A 0-20 mN 0-2 4 1 1.2 kHz
SI-KG7 0-5 mN 0-0.5 0.2 10 250 Hz
SI-KG7A 0-5 mN 0-0.5 0.4 5 500 Hz
SI-KG7B 0-10 mN 0-1.0 1 1.5 550 Hz
SI-KG20 0-0.2 N 0-20 80 590 Hz
Under ideal conditions, use a model of
SI-KG
transducer that has a full-load range that is
no more than 120% of the maximum force that is anticipated. For example, if the greatest
force to be measured is 4g, use a transducer that has a full-load range of 5g, like the
SI-KG4
transducer. To use the transducer at its full-load range, set the gain of the
SI-BAM21LC
to
X1. Higher resolutions are possible using the other gain settings (X2, X5, or X10). However,
using a gain of X10 allows only a tenth of the full-load range of the transducer to be displayed
as an output. In general, it is best to choose a gain factor that does not need to be changed
during an experiment, since each gain factor can have slight variances in its oset. If it is
necessary to switch between gain ranges during an experiment, check the osets in each of
the ranges after the calibration and before conducting the experiment. Then, use the
Oset
Adjustment
switch to set the minimum average oset between the ranges.
NOTE
: Before calibrating the
SI-KG
transducer, position the tissue mount being used on the
actuator rod of the transducer. During the calibration, place the weight on the tissue mount
at the same position where the tissue will be attached.
The basic procedure for calibrating the
SI-BAM21-LC
involves:
1. Setting a zero reference point with the force transducer un-loaded.
2. Applying a load with a known mass to the tissue mount on the transducer.
3. Choosing one of the two calibration methods that best serves the application. Use the