
17
Release 01/2009
photoLab®Series
5 Analytical Quality Assurance (AQA)
5.2.1 Checking the Photometer
Test for the overall system includes checking the working equipment and
checking the handling operations.
The overall system can be checked using standard solutions of a known
content, preferably with the Spectroquant®CombiCheck; this corresponds
with the AQA 2 mode in the photometer.
Spectroquant®CombiCheck are ready-to-use standard solutions that in
terms of the analyte concentration are finely adjusted to the individual test
kits. They contain a mixture of several analytes that do not interfere with
each other. The standard solution (R-1) is used in the same way as a
sample. A double determination is recommended as a measure to
diagnose any random errors.
The desired values with the permissible tolerances are already electroni-
cally stored with the method in the photometer.
In addition to the CombiCheck, it is also possible to use CertiPUR®stan-
dard solutions for this checking procedure. These contain 1000 mg of the
respective analyte per liter of solution.
They can be diluted to different final concentrations, which should pre-
ferably lie approximately in the middle of the measuring range of the
respective test kit. The table presented in Appendix 2 provides an over-
view of the available CombiCheck and ready-to-use standard solutions.
5.2.2 Checking the Overall System
As soon as the photometer is activated it is running a Self-Check. This
means the hardware and the software of the photometer is checked and
compared with internal standards.
As soon as the photometer is activated it is running a Self-Check. This
means the hardware and the software of the photometer is checked and
compared with internal standards.
The photometer itself is checked in the AQA 1 mode with the
Spectroquant®PhotoCheck: the pack includes round cells containing
stable
test solutions (secondary
standards) for checking the photometer
at the
445, 525, and 690 nm wavelengths. The test solutions
are measured
in a reference photometer monitored with primary standards, and the
certificate stating the absorbance values is enclosed with the package
unit. These desired values with the permissible tolerances are entered into
the photometer or else handwritten into the control chart. For the meas-
urement the cell is placed in the compartment for the round cell and identi-
fied by the photo-meter via the bar code, and the measured absorbance is
compared with the desired value. The absorbance is shown on the display
and can be entered into the corresponding control chart.
The measurement of four cells for a given wavelength tests – in addition to
the wavelength accuracy – also the linearity of the absorbance over the
effective range.
The verification of the instrument, as it is required by DIN/ISO 9000 or
GLP, can be easily performed by using the Spectroquant®PhotoCheck.
The PhotoCheck hence offering the possibility to check the instrument. All
of the corressponding documentation, required by these certification guide-
lines, is done by the photometer automatically.
Mischen
Küvettewirdheiß,
amVerschluss
anfassen
2ml Probelösung
inein Reaktions-
küvettegeben
imThermoreaktor
erhitzen
148 C,120 min
mind.10min
abkühlen
Abkühlenauf
Raumtemperatur
(mind
.30min
)
Mischen Messen
C2/25 CSB1500ChemischerSauerstoffbedarf
Me§bereich1001500mg/l CSB
1
4
.
7
2
5
8
0
3
6
9
C
14mm