YaliTech RTR505B User manual

Area
Battery Installation
Recording starts automatically by inserting the battery with the default
or previous settings.
Thank you for purchasing our product.
To ensure safe and proper operation, please read this guide thor-
oughly before use.
Overview
RTR505B is a data logger designed to measure and record
different items depending on the input module to be con-
nected: temperature (Thermocouple, Pt), analog signal
(4-20mA, DC voltage), and pulse.
Recorded data is then automatically collected by the Base Unit via
wireless communication and downloaded for archive and analysis.
Battery
Cover
Screw
Rubber
Packing
Default Settings
Recording Mode
Endless
Recording Interval
10 minutes
Recording Start
Method
Immediate Start
•
Make sure to use the proper type and
size screwdriver. (Phillips head #1
screwdriver is recommended.)
•
Insert the supplied battery with the
tube attached. When using a CR2 lithi-
um battery, the tube is not necessary.
•
Before closing the cover, check the rub-
ber packing for dust or scratches, as
they can reduce the water resistance of
the rubber.
•
Be sure to completely close the cover.
Make sure not to over tighten the
screws.
•
Appropriate Tightening Torque:
Lithium Battery 20N·cm to 30N·cm (2Kgf·cm to 3Kgf·cm)
Wireless Data Recorder RTR505B is referred to as the “(data) logger” or
“device” in this manual.
tandd.com
© Copyright T&D Corporation. All rights reserved.
2021.11 16508210006 (2nd Edition) Printed on recycled paper.
Package Contents
•
RTR505B or RTR505BL (Large Battery Type)
•
Lithium Battery LS14250 (or Large Capacity Battery Kit RTR-500B1 for L type model)
•
Strap (Not included with L type model)
•
Manual Set (Warranty Included)
Part Names
Antenna
LCD Display
O
C
o
p
m
ti
c
m
a
l
unic
a
tion
Module Jack
Battery Replacement
When it is time for the battery to be
replaced, a battery warning mark will ap-
pear. Please change the battery as soon
as possible if this mark appears.
If you continue to use the logger without changing the battery, the
current temperature and [bAtt] will be displayed alternately and wireless
communication will stop. (Recording will continue.)
Alternating
Display
•
If the battery is further left unchanged, the display will automatically shut
off and all previously recorded data will be lost.
•
Although the logger continues to work for a period after the battery has
been removed, leaving the device without a battery until the LCD display
goes blank will cause all recorded data to be lost.
•
After the battery installation, the battery warning mark may not disappear
from the logger’s LCD for 10 to 60 minutes. This is due to the characteristic
of the lithium battery and is not a malfunction of the logger or the battery.
Remote Unit Registration and Settings
Via Software and Optical Communication
Connect the Base Unit to a PC with a USB cable, and place the data log-
ger face down to align the communication areas as shown below.
<F
r
ont>
<Side> (
<
L
S
t
i
y
d
p
e
e
>
)
RTR500BM,
Input Modules (Sold Separately)
RTR500BW,
RTR-500NW/AW
RTR-500DC
RTR-500MBS-A
RTR500BC,
RTR-500
•
Before using the pulse input cable, it is necessary to set the measurement item to “pulse
type” in the Remote Unit Settings of the application.
•
In order to change the measurement item, initialize the Remote Unit without connect-
ing the input module and then redo the Remote Unit registration and settings.
Make sure that the module is completely
inserted until you hear a “click” sound.
Via Mobile App and Bluetooth
®
Communication
When the Base Unit is a RTR500BW or RTR500BM, it is possible to make
Base Unit and Remote Unit settings from nearby mobile devices using
Bluetooth.
Wireless Data Recorder
RTR505B User’s Manual
RTR505B requires a Base Unit to carry out wireless communication.
(Compatible Base Units: RTR500BC, RTR500BW, RTR500BM, RTR-
500MBS-A, RTR-500DC, RTR-500, RTR-500NW/AW)
For the operation and configuration of the Base Unit, refer to
the instruction manual attached to the Base Unit or the RTR500B
Series Help available on T&D Website.
Measurement Item
Input Module
LCD Displayed Items
Temperature
(Type K, J, T, S)
Thermocouple Module
(TCM-3010)
Measurement, Unit of Measurement,
Sensor Type, Operational Status
Temperature
(Pt100, Pt1000)
PT Module
(PTM-3010)
Measurement, Unit of Measurement,
Sensor Type, Operational Status
Voltage
Voltage Module
(VIM-3010)
Measurement, Unit of Measurement,
Operational Status
4-20mA
4-20mA Module
(AIM-3010)
Measurement, Unit of Measurement,
Operational Status
Pulse
Pulse Input Cable
(PIC-3150)
Measurement, Unit of Measurement,
Operational Status
How to Read the LCD Display
[REC] Mark Shows the recording status.
ON: Recording in progress
BLINKING: Waiting for programmed start
OFF: Recording stopped
[ONETIME] Mark Shows the recording mode.
ON: One Time
OFF: Endless
Other Indications on Display
Logging Capacity FULL
When Recording Mode has been set to “One Time” and
the logger reaches its logging capacity of 16,000 readings,
recording will automatically stop and in the LCD the
measurement and the word [FULL] will alternately appear.
Estimation of time until [FULL] is displayed
Rec Interval
1 second
30 seconds
1 minute
10 minutes
60 minutes
Time Period
About 4
hours
About 5 days
About 11
days
About 111
days
About 1 year and
10 months
Data Transmission via Wireless Communication
The measurement and the word [SEnd] will alternately
appear when data is being sent via wireless communica-
tion to the Base Unit. Recording will continue during
wireless transmission.
Input Module Unrecognized (factory default)
This will appear if, after purchasing, the input module has
Battery Warning
Mark
Indicates it is time for the battery to be replaced. never been connected to the logger. (No unit displayed)
Measurement and Measurements or operational messages are shown
Input Module Unconnected or Damaged
Message Area
Sensor Type
here.
The type of sensor connected to or set in the logger is
shown here.
Thermocouple: Type K, J, T, S
Platinum Thermal Resistance Sensor: Pt (Pt100), PtK
(Pt1000)
This will appear if the device cannot confirm a connection
with the input module after having recognized it. (Unit of
Measurement displayed)
•
If nothing is displayed after reconnecting the module to the device, there is a possibili-
ty that the module or the device has been damaged.
Indicates that the logger (Remote Unit) has not been
[Ir] Mark registered to the Base Unit or the wireless communi-
cation is stopped (inactive radio).
[COM] Mark Indicates that the device is in Bluetooth communication.
Sensor Unconnected or Damaged
This will be displayed when a sensor has not been
connected to the module or the wire has been broken.
Recording is in progress and so is battery consumption.
Unit of Measure-
ment
Shows the unit of measurement.
•
If nothing appears on display after reconnecting the sensor to the device, there is a
possibility that the sensor or the device has been damaged.
•
When being used in very hot or cold environments the display may become difficult to
read. This is not a malfunction.
Basic Measurement Display
Display varies depending upon the input module connected.
Temperature (Thermocouple / Pt100 / Pt1000)
Temperature measurement (Unit: °F / °C) will be
displayed. Sensor type will be displayed under the
measurement; the factory default setting is Type K
and Pt (Pt100). The sensor type setting can be
changed in the Remote Unit Settings of the
application.
Measurement Range Exceeded
[OL] will appear if a measurement exceeds the measure-
ment range.
Display Range Exceeded
When measuring voltage in mV range, the measurement
in the LCD display will flash if it exceeds the display range
of the device.
Specifications
Voltage
4-20mA
Voltage measurement (Unit: V / mV) will be displayed.
Due to the wide measurement range, the device has
been set by default to adjust the decimal point
automatically to display the measurement in V. The
unit of display can be changed in the Remote Unit
Settings of the application.
4-20mA measurement (Unit: mA) will be displayed.
Pulse
There are two display methods for the pulse measurement. The display
method can be changed in the Remote Unit Settings of the application.
Pulse Rate (Max: 61439)
The most recent pulse count (Unit: P) for the recording
interval period will be displayed. The display will be
refreshed every one-sixtieth of the recording interval
(at minimum of every one second). 50,500 pulse count
will be displayed as [50.50P] with “K“ below the
measurement value. The display is in units of 10 pulses.
Total Pulse Count
The cumulative number of pulses (Unit: P) will be
displayed from 0 to 9999. The displayed count will be
refreshed every one second, and upon exceeding
9999, the count will start over again from 0.
*1: Measurement item depends on the input module (sold separately).
*2: Only “Endless” is available when using the RTR500BW, RTR500BM, RTR-500NW/AW or
RTR-500MBS-A as a Base Unit.
*3: Bluetooth is available when using the RTR500BW or RTR500BM as a Base Unit and
making device settings in the mobile app (T&D 500B Utility).
*4: When using RTR-500B1 it is necessary to purchase Lithium Battery (LS26500). For
details, contact your local authorized distributor.
*5: The listed battery life is based on the following usage conditions: Recording at 10
second (or longer) intervals, Current Readings Transmission every 10 minutes, and
Recorded Data Transmission once a day. Battery life also varies depending on ambient
temperature, radio environment, frequency of communication, etc.
*6: Input module (sold separately) is not water resistant.
*7: A firmware update is required to a RTR500B series compatible version.
*8: A software update is required to a RTR500B series compatible version.
The specifications listed above are subject to change without notice.
Measurement Item
Temperature, Voltage, 4-20mA, or Pulse Count (*1)
Logging Capacity
16,000 readings
Recording Interval
1, 2, 5, 10, 15, 20, 30 sec. 1, 2, 5, 10, 15, 20, 30, 60 min.
Recording Mode(*2)
Endless (Overwrite oldest data when capacity is full) or One
Time (Stop recording when capacity is full)
Communication
Interfaces
Short Range Wireless Communication
Frequency Range: 902 to 928MHz
RF Power: 7mW
Transmission Range: Approx. 150 meters (500 ft) if direct and
unobstructed
Bluetooth 4.2 (Bluetooth Low Energy) (*3)
Optical Communication
Power
Lithium Battery LS14250 x 1
L Type: Large Capacity Battery Adaptor Kit (RTR-500B1) (*4)
External Power Adaptor Kit RTR-500A2
Battery Life (*5)
Approx. 10 months / L Type: About 4 year
Dimensions
H 62 mm x W 47 mm x D 19 mm
L Type: H 62 mm x W 47 mm x D 46.5 mm
(excluding protrusions and Input Module)
Antenna length: 24 mm
Weight
Approx. 50 g / L Type: Approx. 65 g
Operating Environ-
ment
-40 to 80℃
-30 to 80℃ during wireless communication
Waterproof Capacity
IP64: Splash proof (rated for use in daily life) (*6)
Compatible Base
Units
RTR500BC, RTR500BW, RTR500BM
RTR-500DC, RTR-500MBS-A, RTR-500NW/AW (*7)(*8)
RTR-500 (*8)
Rio Refugio 9648 - Parque de Negocios ENEA, Pudahuel, Santiago ~ CHILE
Tel. +56 2 28988221 • www.yalitech.cl

Popular Data Logger manuals by other brands

Fleet Complete BigRoad DashLink DL-200 user manual

Fleet Complete

Fleet Complete BigRoad DashLink DL-200 user manual

ProSoft Technology PLX51-DL user manual

ProSoft Technology

ProSoft Technology PLX51-DL user manual

Fourtec DaqPRO 5300 user guide

Fourtec

Fourtec DaqPRO 5300 user guide

LogTag USRIC-8 quick start guide

LogTag

LogTag USRIC-8 quick start guide

inelo TachoReader Combo Plus Tutorial

inelo

inelo TachoReader Combo Plus Tutorial

McElroy DataLogger DL3 Quick reference card

McElroy

McElroy DataLogger DL3 Quick reference card

Symmetricom ATS-6501 Tflex user guide

Symmetricom

Symmetricom ATS-6501 Tflex user guide

Meier-NT ADL-MXmini manual

Meier-NT

Meier-NT ADL-MXmini manual

EasyLog EL-USB-RT quick start guide

EasyLog

EasyLog EL-USB-RT quick start guide

METER EM60 Series manual

METER

METER EM60 Series manual

MadgeTech Volt101A Product information

MadgeTech

MadgeTech Volt101A Product information

SIRETTA SENTRY-G-LTE4 user manual

SIRETTA

SIRETTA SENTRY-G-LTE4 user manual

Giacomini GE552Y051 user guide

Giacomini

Giacomini GE552Y051 user guide

Lascar Electronics EasyLog 9902-2004 quick start guide

Lascar Electronics

Lascar Electronics EasyLog 9902-2004 quick start guide

Racelogic VBOX Micro quick start guide

Racelogic

Racelogic VBOX Micro quick start guide

EROAD ELD user manual

EROAD

EROAD ELD user manual

Dickson SM300 Getting started

Dickson

Dickson SM300 Getting started

MadgeTech EggTemp-RH Product Information Card

MadgeTech

MadgeTech EggTemp-RH Product Information Card

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.