Yeti IQ Waxer NT light 230 V Instruction Manual

Betriebsanleitung IQ Waxer NT light
YETI Dentalprodukte GmbH ·Industriestr. 3 ·D78234 Engen ·Tel: 07733/9410-0 ·Fax: 07733/941022
www.yeti-dental.com · info@yeti-dental.com
Das Wachsportal: www.dentalwax.com
Technische Betriebsanleitung
Instructions of use
Mode d’emploi
Istruzioni per l’uso
Instrucciones de uso
_____________________________________________________
für / for / pour / per / para
IQ Waxer NT light 230 V Art.-Nr. 410-9000
IQ Waxer NT light 115 V Art.-Nr. 410-9115

Operating Instructions IQ WAXER NT light
Operating Instructions
Status of technical operating instructions: 1/2007
and of CE sign: 1/07
1. General term: Electric wax knife
1.1 Identification of product:
IQ WAXER NT Light Art.: 410-9000 230 V
IQ WAXER NT Light Art.: 410-9115 115 V
1.2 Delivery includes: IQ WAXER NT light with Continous and Touch-Control function,
handpiece and modelling tip ….420-0200…. and clamp for handpiece
410-1002
1.3 Manufacturer: YETI Dentalprodukte GmbH - Industriestr. 3 D-78234 Engen
Tel. 07733-9410-0, Fax 07733-9410-22
www.yeti-dental.com www.dentalwax.com info@yeti-dental.com
1.4 Technical Features:
Mains voltage: 207...244V or /90 ... 145 V
(gem. EN 50160)
Mains Frequency: AC 50 Hz /AC 50/60 Hz
Mains Plug: EURO-Schuko /US Schuko
Power consumption: max. ca. 30 W
Tip temperature: ca. 50 ... 230°C
Type of protection: IP 40/protective earthing
1.5 Storage: - store dry at room temperature
- protect from acid and lye
2. Use according to determinations:
Our waxers are to be used exclusively for the processing of sculpturing waxes used by dental
technicians. Any use on human beings is not allowed. Any claims on guarantee or liability will not
be accepted if unit is not used according to determinations.
3. Risks and dangers during application, application regulations:
1. Ensure that unqualified persons (e.g. children) do not use these units. Use only for dental technicians
and authorised persons.Operating instructions must be read before putting into operation. For uncleared
questions contact manufacturer.
2. Unit to be operated in dry area, clean dry and do not immerse in water.
3. Caution: risk of burns when touching the heating unit or the inserts.
4. Warming up of the unit (hand-warm), specially at the back of the unit, is normal. Permanent operation
with higher tip temperatures increases warming up: this is not dangerous. In this case change to Touch
Control or reduce temperature.
5. Any modification of these units or grips is exclusively on the user´s own risk. Do not sharpen or bend tip
(picking-up wax with too cold tip). Integrated heating or temperature sensor could be damaged.
6. Mains wire is non-flammable, but sensitive to temperature. Avoid contact with modelling tip or other hot
objects.

Operating Instructions IQ WAXER NT light
7. Handpiece-wire is non-flammable and more stable to temperature. Nevertheless avoid contact with hot
modelling tip or open flame (Bunsen burner !).
8. If wire is damaged stop usage immediately and send in for repair.
9. Unit only to be opened by skilled personnel. Any misuse of the unit willl lead to loss of warranty and
liability.
10. Pull out the plug before opening the unit !
11. Do not pull the plug out of the socket by means of the wire.
12. Do not use any solvents for cleaning of the unit.
13. Ensure safe packaging when returning a unit (e.g. for repair)
4. Operating instructions:
4.1. Before putting into operation:
Our Waxer follow the latest technical developments and the standard regulations of safety. Nevertheless
dangerous situations may occur.
♦Usage of the unit only in perfect conditions and according to operating instructions !
♦Observe applicable regulations for save work. Usage by skilled personnel only.
♦Unit to be checked for perfect technical condition, wires included !
♦Turn the required modelling tip onto the heating element of the handpiece.
♦Lay down handpiece on a safe place (Art. No. 410-0000 is not appropriate). If foot pedal is used connect
pedal switch into unit back.
♦Put the plug into the socket (adapt only to licensed sockets - observe type registration)
♦The Waxer does contain a special programm which through it, deffected tips can be detected and
prevented from booting the unit.
4.2 Function:
Switch on the unit at the back. The LED optical sensor lights on 2 times in orange when switching on the
unit and lights afterwards green. The unit is under “Touch Control” and heats up in seconds by touching the
metal ring. For optical control of the heating the color of the optical sensor LED changes to orange until the
contact with the metal ring is interrupted. The unit stops heating and cools down to room temperature. The
optical sensor lights green again. The Touch Control function can be released by finger touch of the metal
ring on the handpiece or optional by using the pedal. The connector for the pedal is on the back of the unit.
If you prefer the “Continous” modus just press the button (C) under the light sensor. The color of the LED
changes to orange and works in “Continous”. By pressing again the button the unit changes again to the
Touch Control function and the LED color changes to green.
The temperature of the unit can be controlled with the regulator for both functions. We recomed to reduce
the temperature on the continous mode.
Attention:
We recomend to read the instructions of use. In case you have lost the instructions please
download it under: www.yeti-dental.com .
4.3 Adjusting the ring of the handpiece

Operating Instructions IQ WAXER NT light
Because of the different anytomy types of the hands (big, small, long or short finger), we organized the
metallic ring of the handpiece (B) in order to be regulated by modifying the positionning of the rubber rings
(C) of the handpiece.
These rubber rings (C) must be pulled from the handpiece by using an instrument, so that the free
movement of the metal ring allows the technician an individual adjustment of the contact ring in function with
the distance to the modelling tip in order to suit the hand anatomy.
In order to avoid any burning risks, this operation must be done when the unit is Off and the tip is cold.
We suggest that such modification can be done after working for a certain time with the standard
positionning in order to be able to judge on the indispensability of that operation.
5. Disposal: Dispose according to applicable local, state and federal law.
6. Warranty: Our IQ WAXERS are under warranty for 2 years from the date of delivery.
Modelling tips and handpieces are excluded from this warranty.
7. Spare parts & accessories:
Following tips are available :
Art.: 410-1002 Holder for handpiece
Art.: 409-9500 Pedal
From left to right :
PKT 1 extra fine ø 0,8 mm. Art.-Nr. 420-0101
PKT 1 fine ø 1,0 mm. Art.-Nr. 420-0100
PKT 2 standard ø 1,2 mm. Art.-Nr. 420-0200
Hollenback vertical Art.-Nr. 420-0300
Spear shaped Art.-Nr. 420-0400
Do not use the old tips of the IQ Contact Waxer with this
new units. It is recommanded to use the tips mentionned
above.
This manual suits for next models
1
Table of contents
Popular Dental Equipment manuals by other brands

Tecno-gaz
Tecno-gaz Hydra Evo Instructions for use

Air Techniques
Air Techniques CamX Triton HD Installation and operating instructions

Arkray
Arkray SillHa LH-4912 operating manual

Renfert
Renfert Sympress 6000-0000 instruction manual

FINO
FINO VITESSE SF User information

Castellini
Castellini SKEMA 6 Series operating instructions

Aseptico
Aseptico AEU-14CF OPERATION, SERVICE MANUAL & PARTS LIST

DeguDent
DeguDent SOLARIS Instructions for use

Fona
Fona MyCrown Mill Service manual

KaVo
KaVo SURGmatic S11 L Instructions for use

Instrumentarium
Instrumentarium Orthopantomograph OP200 D user manual

DentalEZ
DentalEZ StarDental Concentrix FX instruction manual