User Manual: Zopec Explore Travel Series CPAP/BiPAP Batteries (Ver. 02.05.2022)
Page 2 of 9
IMPORTANT – BEFORE FIRST USE
1. Initial Charge/Battery Reset: Please charge Zopec Explore battery (~8 hours) until
100% charged (LED = FF) before first use. Future charging will be faster. This is to
reset the battery capacity counter. It is normal if the capacity does not change for the
first 3 - 5 hours. See troubleshooting section if charging issues occur.
2. Heated Humidifier and Heated Tube Impact on CPAP Running Time:
• Longest Running Time: NOT using Heated Humidifier and Heated Tube.
• Shortest Running Time: Using Both Heated Humidifier and Heated Tube will
reduce about 75 - 80% of the CPAP running time.
• Heated Humidifier: With Heated Humidifier ON, it will reduce CPAP running
time by about 50 - 60%. It takes a lot of power to evaporate water.
• Heated Tube: We do not recommend running Heated Tube because it will
reduce CPAP running time by an additional 50 - 60%. In the US, water
shouldn’t pool in the CPAP hose. Please contact us if rain out is an issue.
3. Instruction for Turning Off the Heated Humidifier and Heated Tube:
• ResMed AirSense 10:
i. Go to “MyOptions” and change “ClimateControl” to “Manual”. NOTE:
this must be done first. Otherwise, the automatic climate control will
turn the heated humidifier and heated tube back ON without notice.
ii. Next, turn “Humidity Level” and “Tube Temp” to “OFF”. If you don’t
have a heated tube, just turn the “Humidity Level” to “OFF”.
iii. Exit “MyOptions” and re-enter “MyOptions” to verify that “Climate
Control” is in “Manual” and “Humidity Level” and “Tube Temp” are
“OFF”.
• Philips Respironics DreamStation, detach the heated humidifier and plug
the CPAP hose directly to the back of the CPAP Unit.
• All Other Machines, detach the heated humidifier if possible and just run the
CPAP unit. Please call or email us if you don’t know how to turn off your
heated humidifier and heated tube.
4. Determine the Exact CPAP Run Time: Test the run time with your machine and
exact settings first without sleeping with it. At the end of the CPAP hose, remove
your mask and put paper towel inside the CPAP hose. Turn the machine on and just
let a little air through the CPAP hose. This imitates breathing. Simply run the CPAP
with a fully charged battery during the day and record the run time.
5. Maintenance: Fully charge battery to 100% (LED = FF) monthly for maintenance.
6. Contact Us: We are proud to provide best-in class customer service so please
contact us with assistance any time. Below is our contact information:
• Zopec Medical Website: www.zopec.com
• Phone: Please see our website “Contact Us” for the current phone numbers.