Accuniq BC 380 User manual

User Manual

The device bears the CE label in accordance with the provisions of Medical Device Directive 93/42/
EEC.
THE PERSONS RESPONSIBLE FOR PLACING DEVICES ON THE EC MARKET UNDER MDD 93/42/EEC.
SELVAS Healthcare, Inc.
155, Sinseong-ro, Yuseong-gu, Daejeon, 34109 Republic of Korea
TEL: +82 42 879 3000, FAX: +82 42 864 4462
Obelis s.a.
Bd. Général Wahis, 53, B-1030, Brussels, Belgium
Table of Content_3
TABLE OF CONTENTS
TABLE OF CONTENTS 3
INTRODUCTION 4
Intended Use ............................................................ 4
Contraindications ...................................................... 4
Intended patient population ...................................... 4
Word Definitons ........................................................ 5
Classification and Compliance ................................. 5
Safety Symbols and Information .............................. 6
Safety Precautions .................................................... 8
Guidance for Electromagnetic Compatibility (EMC).... 12
ABOUT BODY COMPOSITION 13
SAFETY MANAGEMENT 15
NAME AND FUNCTION OF EACH PART 17
Main Components ..................................................17
Options .................................................................... 18
Device Appearance ............................................... 20
Front Part .......................................................20
Base Part ........................................................ 20
Rear Part ......................................................... 21
INSTALLATION 22
Power Connection .................................................. 22
External Connection ............................................... 22
PC Connection - Wired (USB Cable) ........... 23
PC Connection - Wireless (Bluetooth).........24
How to Connect Printer .................................24
Blood Pressure Monitor Connection ............25
How to Replace Thermal Paper in the
Thermal Printer ............................................... 26
SYSTEM SETUP 28
Enter the ‘System Setup’ Screen ......................... 28
Type ................................................................. 28
Setup ....................................................................... 30
Basic Settings .................................................30
Data Management .......................................... 31
Printer / Printing ............................................. 31
Results ..............................................................32
Communication ................................................32
Weight...............................................................33
Management Options ......................................33
Display ..............................................................34
Customer Support ...........................................34
MEASUREMENT AND ANALYSIS 35
Precautions for Measurement ............................... 35
Correct Measuring Posture ................................... 36
How to Stand on a Scale With a Foot
Electrode ......................................................... 36
How to Hold the Hand Electrode ................. 36
Correct Measuring Posture ........................... 37
Measurement ...........................................................38
Basic Analysis .................................................38
View Results ....................................................41
Logging in and taking measurements with the
ACCUNIQ Connect app ..................................... 42
To install ACCUNIQ Connect ............................. 42
Loading information into ACCUNIQ Connect . . 43
Analysis Using the Ultrasonic Height Meter .... 44
Analysis Using a Blood Pressure Monitor ....45
RESULT SHEET INTERPRETATION
46
Body Composition Result Sheet ........................... 46
Result Sheet for Child (Option) ............................. 50
STORAGE AND MAINTENANCE 53
PROBLEM SOLVING 54
Error Occurrence and Actions .............................. 54
Error Code Occurrence & Actions ....................... 55
Measurement Error Occurrence and Actions ..... 56
Problems in Use ..................................................... 57
Packing and Transportation .................................. 57
SPECIFICATIONS 58
Blood Pressure Monitor ......................................... 60
List of supported printers ...................................... 61
PC specification ...................................................... 61
Expected Service Life ............................................ 61
WARRANTY 61

SELVAS Healthcare User Manual BC380
4_Introduction
INTRODUCTION
You are kindly requested to be familiar with these directions before using this device and always keep it
together with the device. In case you are not sure about any directions or problems arising while using the
device, please contact with SELVAS Healthcare or its local distributor where is purchased. We will provide
you with detailed instructions.
INTENDED USE
This device measures impedance by bioelectrical impedance analysis method and provides lots of
information using measured impedance and inputted personal data (height, age, gender, weight).
It shows body composition of MBF, LBM, SLM, SMM, TBW, protein mass, mineral mass, etc. and information
regarding BMI, PBF, BMR, abdominal analysis, Target to control, segmental analysis, Body composition
change, etc.
Intended application location is professional healthcare facility environments, not home healthcare
environment.
CONTRAINDICATIONS
The device should not be used in following people:
・ Anyone who implants metallic materials like a pacemaker, defibrillator, stent, and metal suture in the
heart and great vessel etc.
・ Anyone who is equipped with the devices injecting electric current such as artificial heart, and heart
lung machine.
・ Anyone who is connected to liquid-filled catheter and other electronic devices of good conductivity.
・ Anyone who use following device can go into danger or cannot reach objective result due to
interference with other electric signals.
Using electronic stimulator for various purposes.
・ Devices injecting electric current or connecting and operating: ECG, EMG, and EEG.
・ Any other person who is diagnosed by the doctor to be influenced even by imperceptible micro-current.
Please consult with doctor before using this device to:
・ Any woman with contraceptive devices.
・ Any woman in pregnancy.
INTENDED PATIENT POPULATION
・ Age: No limit
・ Gender: No limit
・ Weight: 10 ~ 250 Kg
・ Height: 50 ~ 220 cm
・ Race: No special requirements
・ Health: A person who can stand up and maintain the measurement posture required by the machine.
WORD DEFINITIONS
To ensure safe operation and long term performance stability, it is essential that you fully understand the
functions, operating and maintenance instructions by reading this manual before operating your unit.
Particular attention must be paid to all warnings, cautions and notes incorporated herein.
The following conventions are used throughout the manual to denote information of special emphasis.
Warning
“Warning” indicates important information about the presence of a hazard which may
cause severe personal injury, loss of substantial property, damage if the warning is
ignored.
Caution
“Caution” indicates important information about the presence of a hazard which may
cause minor personal injury or property damage if the caution is ignored.
Note
“Note” indicates important information in order to notify installation, operation or
maintenance of this device. "Note" is important but not hazard-related.
Hazard warnings are not included here.
Introduction_5
CLASSIFICATION AND COMPLIANCE
1) This device is classified as;
・ Class 1 type-BF against electric shock
・ Ordinary equipment without protection against ingress of water
・ Equipment not suitable for use in presence of a flammable anesthetic mixture by standard of IEC
60601-1:2005/A1:2012 (Basic safety and essential performance of Medical Electrical Equipment)
2) This device is complied with Class A for Noise-Emission, Level B for Noise-immunity, by standard of IEC
60601-1-2:2014(Electromagnetic Compatibility Requirements).
Introduction

SELVAS Healthcare User Manual BC380
6_Introduction Introduction_7
Introduction
SAFETY SYMBOLS AND INFORMATION
The International Electro-technical Commission (IEC) has established a set of symbols for medical electrical
equipment which classify a connection or warning of any potential hazard.
The classifications and symbols are shown below. Save these instructions for your safety.
Degree of protection against electric shock : TYPE BF
Please observe operating instructions
General warning sign
General prohibition sign
General mandatory action sign
Caution sign
"OFF" (only for a part of equipment)
"ON" (only for a part of equipment)
Waste Electrical and Electronic Equipment (WEEE)
The device could be sent back to the manufacturer for recycling or proper
disposal after their useful lives. Alternatively the device shall be disposed in
accordance with national laws after their useful lives.
SYMBOL INFORMATION
Direct current
Date of manufacture
Manufacturer
CE mark
Authorized representative in the European community.
Keep dry
Fragile
Do not use blades to open
Handle with care
Medical Device
This way up
Serial No.
SYMBOL INFORMATION
MD

SELVAS Healthcare User Manual BC380
8_Introduction
Introduction
Introduction_9
SAFETY PRECAUTIONS
This device is designed and manufactured with consideration of the safety of the operator and subject and
also the reliability of the unit.
The following warnings, precautions and notes must be observed for safety;
Warning
During measurement of the body composition, a microcurrent of 180μA flows through
the body. Individuals who have any kind of implanted active medical devices, such
as pacemakers, should not use this equipment because the microcurrent can cause
malfunction in the implanted device.
Warning
To prevent fire hazard, use only a correctly wired (100-240VAC) outlet, and do not use a
MSO(Multiple Socket Outlet) that is not in compliance with IEC 60601-1.
Warning
To reduce the risk of electric shock or device damage, never plug-in or plug-out with
wet hands.
Warning
Physically disabled persons should not attempt to take measurements alone, but instead
should have their caretakers assist them in using the device.
Caution
The unit must be operated only by, or under supervision of a qualified person with our
company or our distributors.
Caution
If you have experienced any trouble with the unit, switch it off immediately, and contact
our company or its authorized dealer for assistance.
Caution
If you plan to connect any device from other manufacturers electrically or mechanically to
the unit, contact our company or its authorized dealer for instructions before doing so.
When you connect computer or other system to the unit (RS-232C), the attached systems
should be those certified by IEC 950 or equivalent standards for data processing equipment.
Configurations shall comply with the system standard IEC 60601-1:2005/A1:2012.
Everybody who connects additional equipment to the signal input part or signal output
part configures a medical system by standard IEC 60601-1:2005/A1:2012.
If in doubt, consult the A/S department of local distributor.
Caution
Avoid the following environments for storage;
・ Where the ambient temperature falls below -25℃ or exceeds 70℃.
・ Where the atmospheric pressure falls below 70kPa (700mbar) or exceeds 106kPa (1060mbar).
・ Where the humidity is over 93% non-condensing.
・ Where the unit is exposed to spray or splashing water.
・ Where the unit is exposed to dust.
・ Where the unit is exposed to water vapor.
・ Where the unit is exposed to salty atmosphere.
・ Where the unit is exposed to explosive gas.
・ Where the unit is exposed to excessive shocks or vibrations.
・ Where the angle of inclination of mounting surface exceeds 10 degrees.
・ Where the unit is exposed to direct sunlight.
Caution
This device needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided in the ACCOMPANYING
DOCUMENTS.
Caution
Cross contamination is possible because this equipment is used with bare hands and
feet. Refer to the cleaning and disinfecting methods in this manual.
Caution
Measurements may be impaired if this device is used near televisions, microwave
ovens, X-ray equipment or other devices with strong electrical fields. To prevent such
interference, use the meter at a sufficient distance from such devices or turn them off.
Prohibition
Do not disassemble or alter the device under any circumstances, as this could result in
electric shock or injury as well as adversely affect the precision of measurements. This
device is specified as Class 1 type BF unit under the standard IEC 60601-1:2005/
A1:2012 (Basic safety and essential performance of Medical Electrical Equipment).
Therefore, patients must not touch or handle inner side of the system at any time.
Prohibition
Do not to touch signal input, signal output or other connectors, and the patient
simultaneously.

SELVAS Healthcare User Manual BC380
10_Introduction
Introduction
Introduction_11
Prohibition
Never pour any liquid directly on the scale platform, as it may leak and cause internal
damage.
Prohibition
The unit has previously been adjusted in the factory for optimum performance.
Do not attempt to adjust switches or any other things except those specified in this
manual for operation.
Prohibition
Never jump on the Weighing Platform, there may be a risk of stumbling and malfunction of
the equipment.
Note
This equipment has been tested and found to comply with the limits for medical
devices according to IEC 60601-1-2:2014. These limits are designed to provide
reasonable protection against harmful interference in a typical medical installation.
This equipment generates uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference
to other devices in the vicinity. However, there is no guarantee that interference will
not occur in a particular installation. If this equipment does cause harmful interference
to other devices, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the following
measures:
・ Reorient or relocate the receiving device.
・ Increase the separation between the equipment.
・ Connect the equipment into an outlet on a circuit different from that to which the
other device(s) are connected.
・ Consult the manufacturer or field service technician for help.
Note
Place the Weighing Platform on a level and stable surface. If the equipment is used
when the Weighing Platform is unstable because not all feet are on the surface, there
may be a risk of stumbling or inaccurate measurement.
Note
Note that portable and mobile RF communications equipment can affect MEDICAL
ELECTRICAL EQUIPMENT.
Note
Consult a physician or a trained health professional for interpretation of measurement
results.
Note
In case of patients who have certain diseases, the estimates might be different
Note
Incorrect operation or failure of user to maintain the unit spares the manufacturer or his
agent of the responsibility for system’s non-compliance with specifications or responsibility
for any damage or injury.
This manual is made for informational purposes and this manual and device are not meant
to be a substitute for the advice provided by your own physician or other medical expert.
You should not use the information contained in the device for diagnosis or treatment of
health problems or prescription of medication by yourself. If you have or suspect that you
have a medical problem, consult with your physician promptly.
Defective units or accessories must be packed in the replacement cartons to be shipped
off from you to our company.
Shipping and insurance costs for return of defective unit must be prepaid by the users.
Warning
Do not modify this equipment without authorization of the manufacturer.
Warning
Connect the earth placed on the backside of this device to terminal plate to prevent any
electric shock from leakage current or a potential difference.
Warning
To avoid the risk of electric shock, this equipment must only be connected to supply
mains with protective earth.
Caution
Do not put anything other than the main unit and SELVAS’s Blood Pressure Monitor within
1.5 m from the patient.
Caution
Do not touch any other devices other than those specified by the manufacturer.
Caution
Any serious incident that has occurred in relation to the device should be reported to the
manufacturer and the competent authority of the Member State in which the user and/or
patient is established.

SELVAS Healthcare User Manual BC380
12_Introduction
Guidance for Electromagnetic compatibility (EMC)
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result
in improper operation. If such use is necessary, this equipment and the other equipment should be observed
to verify that they are operating normally.
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of
this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of
this equipment and result in improper operation.
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30cm (12inches) to any part of this equipment, including cables
specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Details about the electromagnetic compatibility (EMC) of the ACCUNIQ BC380 are given below.
Before using the ACCUNIQ BC380, be sure to read and understand the following information.
1) Guidance and manufacturer’s declaration - electromagnetic emissions
The ACCUNIQ BC380 is intended for use in the electromagnetic environment specified
IEC 60601-1-2:2014 (Fourth Edition).
2) Guidance and manufacturer’s declaration - electromagnetic immunity
The ACCUNIQ BC380 is intended for use in the electromagnetic environment specified
IEC 60601-1-2:2014 (Fourth Edition).
3) Guidance and manufacturer’s declaration - electromagnetic immunity 2
The ACCUNIQ BC380 is intended for use in the electromagnetic environment specified
IEC 60601-1-2:2014 (Fourth Edition).
4) Recommended separation distances between portable and mobile RF communications equipment and the
ACCUNIQ BC380
The ACCUNIQ BC380 is intended for use in the electromagnetic environment specified
IEC 60601-1-2:2014 (Fourth Edition).
What is the body composition?
The body composition of human body consists of moisture, protein, fat, and minerals in a certain ratio.
Total Body Water exists inside and outside the cell at the boundary of the cell membrane, and is regulated
and maintained within a certain range. Protein is a component that mainly forms the muscle layer with water
and is distributed throughout the whole body, including the extremities and intestines.
Lean fat mass is the sum of the body's non-fat components, that is, Soft Lean mass and bone mineral mass.
Fat Mass is stored in the subcutaneous and abdominal organs. It is decomposed and used as energy for
normal physiological functions. However, excess Fat mass can accumulate in the body and cause obesity.
A healthy person maintains a certain ratio of these body components, but conversely, an unhealthy person
may have an imbalance among body components, causing obesity, nutritional deficiencies, and osteoporosis.
What is obesity?
Several methods are used to diagnose obesity, the key to determining the degree of obesity is the amount
of Fat Mass. In general, obesity can be divided into ‘apparent obesity', which is overweight compared to
height, and 'hidden obesity', which has a high amount or ratio of Fat Mass. Therefore, although simple
obesity evaluation is possible through Body mass index (BMI), but percentage of body fat measured using an
impedance body fat analyzer can help for a more accurate diagnosis of obesity.
Necessity for body composition analysis
We can maintain a healthy body when human body components are harmonious and balanced, and the
imbalance in the amount and distribution of those components may be related to various diseases.
Impedance body fat analyzer is a device used for the purpose of analyzing the body composition constituting
the human body to understand the basic health condition and to prepare in advance against harmful factors
on health, which is one of the basic screening tools applied to the prevention and treatment of diseases.
Indicator for obesity judgment
• Percentage of Body Fat (%): It is the ratio of body fat to body weight. Finds the invisible obesity.
• Body Mass Index (kg/m2): Result of dividing weight by the square of height Has a strong relationship with
disease morbidity.
• Obesity Degree (%): It is a value indicating the degree of obesity of the current weight compared to the
standard weight.
Abdominal obesity
It is important not only to evaluate how much fat is accumulated in the body, such as percentage of body fat
or fat mass, but also to evaluate where in the body it is accumulated. Fat Mass is divided into subcutaneous
fat (stored under the skin) and visceral fat (stored between organs) according to the location where it is
accumulated.
WHR is the ratio of waist circumference to hip circumference, so it determines the body shape.
Segmental analysis
This device measures and evaluates body fat mass and soft lean mass in 5 segments of the right arm, left
arm, right leg, left leg, and trunk. Because the human body does not always maintain the correct symmetry,
rathern than questioning the slight difference, use it to evaluate how efficiently and positively you achieve
your goal by exercise rehabilitation treatment for increasing muscle strength or muscle mass based on the
initial analysis value.
ABOUT BODY COMPOSITION
About Body Composition_13
About Body Composition

SELVAS Healthcare User Manual BC380
14_About Body Composition
Body type
About 20 types of body shapes can be derived from body weight, skeletal muscle mass, and body fat mass,
and it determines the body shape closest to the user.
Biological age
It is the biological age considered according to the gender, actual age, and body composition results.
Even if the body composition of all measured people is included in the normal range, it cannot be assumed
that their conditions are all the same. Therefore, the biological age is determined by further subdividing the
criteria for the body composition, which can be used as a reference for physical health level.
Safety Management_15
Safety Management
SAFETY MANAGEMENT
This device is manufactured according to international standards, and user safety and device reliability can
be guaranteed only when the following precautions are observed.
1) The inside of this device can only be operated by a specialist who has been trained on the function and
structure of the device and has been certified by the manufacturer. Therefore, the user must not arbitrarily
manipulate the inside of the device under any circumstances, and the company shall not be responsible
for damage to the device, electric shock, fire, Etc.
2) This device was set to operate in an optimal state.
You shall not attempt to adjust any controls or switches other than those specified in the user manual.
If you want to make a change, please request a service from us or a designated vendor.
3) If any problem occurs with the device, turn off the power immediately and contact us or a designated
vendor for assistance.
4) Do not use or store this device in the following environments:
・Storage in a place where the ambient temperature is out of -25℃~70℃
・ Use in a place where the ambient temperature is out of 5℃~40℃
・ A place where atmospheric pressure is below 70 kPa (700 mbar) or above 106 kPa (1060 mbar)
・ Use in a place where humidity is less than 15% or more than 93%, and store in a place where humidity
is more than 93%
・ A place where the device is exposed to water vapor or water
・ A place where the device is exposed to dust
・ A place where the device is exposed to high concentrations of oil vapors
・ A place where the device is exposed to a salty atmosphere
・ A place where the device is exposed to an explosion hazard of gas or dust
・ A place where the device is exposed to excessive shock or vibration
・ A place where the surface inclination angle is more than 10°
・ A place where the device is exposed to direct sunlight
5) Be sure to use this device in a grounded state. When grounding, be careful not to come into contact with
anything that may cause electrical problems, such as gas, water pipes, lightning rods, and telephone lines.
6) Power must be supplied using the adapter provided by our company, and power supply accessories
such as wires and plugs must be free from defects and meet the specifications.
7) Be sure to turn off the main power switch before unplugging the power cord.
8) Do not operate the device with wet hands.
9) When using the main body in connection with options or peripherals, first turn off the power of the main
device. If you connect a wire while the power is on, an electric shock will occur to cause a breakdown.
10) If there are other power devices around, there is a risk of electric shock due to leakage current or
potential difference on the surface of conductors for users installing the device such as doctors, nurses,
assistants, etc. and customers measuring with this device, especially those who uses implantable current
devices in the body or other medical or electronic devices. Connect the equipotential terminal on the
back of the device to the grounding wire in the room for use.

SELVAS Healthcare User Manual BC380
16_Safety Management
11) Be sure to get inspection service before using the device that has foreign substances or has been
exposed to a special environment.
12) When reusing a device that has not been used for a long time, check whether there is no abnormality in
the appearance or function of the device before using it.
13) Observe other precautions for using general electric devices and the instructions specified in the user
manual.
Note
1. This device meets the requirements of IEC 60601-1 (Common Standards for Electrical
and Mechanical Safety for Medical Devices) and IEC 60601-1-2 (Electromagnetic
Compatibility for Medical Devices).
2. When disposing of this device, it must be disposed of in accordance with the medical
device regulations and the regulations of each country, so please follow the applicable
procedures.
Cautions
1. Be sure to use the dedicated cable provided by the manufacturer for power supply to
this device.
2. If this device is used near a television, microwave-based device, X-ray or other
devices that generate a strong electric field, measurement failure may occur or reliability
of the analyzed value may decrease.
So, keep a sufficient distance or turn off other devices.
Name and Function of Each Part_17
Name and Function of Each Part
NAME AND FUNCTION OF EACH PART
Main Components
The basic configuration of this device is as follows.
※ Model or specification of accessories can be changed according to market supply and demand.
※ To read the user manual in the USB memory, install the Adobe Acrobat Reader program on the latest
operating system. (https://get.adobe.com/reader/)
① Main body ② User manual ③ Power cable and adapter
④ USB cable ⑤ M6 bolts ⑥ 4mm hexagon wrench
⑦ Body hinge cover ⑧ A4 result sheet
① Main body ② User manual
③ Power cable and adapter
④ USB cable
⑤ M6 bolts
⑥ 4mm hexagon
wrench
⑦ Body hinge cover
⑧ A4 result sheet
OR

SELVAS Healthcare User Manual BC380
18_Name and Function of Each Part Name and Function of Each Part_19
Name and Function of Each Part
Options
1. Data Management Program
It is a program that allows you to manage your body composition
in an easy and systematic way. It evaluates and interprets body
fat measurement results in deep and easy-to-understand way,
and it is highly practical as it selects only the most concise and
essential contents to control body composition.
2. Result Sheet for Children
The result sheet for children is a sheet that easily displays
weight, height, and growth status above BMI percentile through
the child growth curve.
3. Automatic Blood Pressure Monitor
If Automatic Blood Pressure Monitor manufactured by SELVAS
Healthcare for hospital is connected to this device, the user
can easily check his/her blood pressure.
Detecting the patient with hypertension may help better manage
his/her blood pressure through weight control.
4. Thermal printer
The thermal printer enables quick and easy printing.
5. Thermal paper
It displays the output results in a simple and easy-to-understand
way.
6. Bluetooth Dongle
It allows wireless communication with a PC.
7. Ultrasonic Height Meter SM104
This is an instrument to measure subject’s height more accurately
and quickly.
It employs standoff determination method using ultrasonic sensor.
※ The model and specifications of the supplied accessories may be changed upon market supply and demand.
Cautions
When using commercially available thermal paper other than the thermal paper provided
by us, check the specifications of 55g/m2 before use. We shall not be responsible for
any malfunctions caused by the use of thermal paper that does not conform to the
specifications.
Note
The height measured in the morning is larger than the height measured in the evening, so
adults may show an average of 1-2 cm difference. This is because the vertebral cartilage
is maximally pressed in the evening by gravity and the person’s own weight.

SELVAS Healthcare User Manual BC380
20_Name and Function of Each Part Name and Function of Each Part_21
Name and Function of Each Part
・ Color LCD (Color TFT LCD monitor): A screen showing the input data, measurement process and results,
and at the same time, it has a built-in touch screen function, where you can input data.
・ Hand electrode: It is the part where the current input for measuring the user's resistance, the
measuring device and the user's hand come into contact.
Device appearance
Front part
Color LCD
Hand electrode
・ Scale with Foot electrode: A scale with Foot electrode attached.
・ Foot electrode : This electrode contacts to the user's foot to be measured.
It measures the body impedance.
Base part
Foot electrode
Foot electrode
Scale with Foot
electrode
・ Power switch:Power on/off switch
・ Adapter port:A terminal for connecting the adapter.
・ PC Connection port: A terminal for connecting the user’s PC to device.
・ Height Meter port: A port to connect Height Meter manufactured by SELVAS Healthcare, Inc.
・ WiFi port: A terminal for connecting a Wi-Fi printer.
・ USB memory port: A terminal for connecting an external memory stick
・ Bluetooth port: A terminal for connecting Bluetooth
・ Printer port: A terminal for connecting a printer to the device
・ Blood Pressure Monitor:A port to connect Blood Pressure Monitor manufactured by SELVAS Healthcare, Inc.
Rear part
Caution
Do not touch the patient while operating the rear panel.
Caution
Be careful for a patient not to touch the rear side during the measurement.
When using a USB memory, you must use a USB memory that uses the FAT32 file system.

SELVAS Healthcare User Manual BC380
22_Installation Installation_23
Installation
INSTALLATION
Power connection
1) Connect the adapter provided by our company to the Adapter port installed on the rear part of this device.
2) After connecting the wires, turn on the power switch.
Caution
1. You must turn the power off before connecting this device to peripherals.
(An electric shock may be applied to the device, which may affect its performance.)
2. When this device and peripheral devices are connected, the power supply must start
with this device to maintain the performance and safety of the device.
3. Be sure to use the adapter and cable provided by our company for power supply.
4. When turning on the device's power switch, be careful not to touch the base part
where there is a scale. If weight is applied to the base part, an error occurs in the '0'
point of the scale by that weight, making the weight measurement inaccurate.
5. Do not install the equipment where power can not be disconnected.
Caution
1. Connect only items that have been specified as part of the device or that have been
specified as being compatible with the device
2. In the event of a cybersecurity breach, stop using the device and contact the manufacturer.
3. When using Bluetooth communication, it may be affected or may be affected by
devices using other wireless systems nearby.
External connection
Connect the USB (B) port on the back of the device to the computer USB port using a USB cable.
PC connection - Wired (USB cable)
Caution
1. The PC that connects to the device must comply with IEC60950-1.
2. The PC must be kept up to date latest operating system for cybersecurity.
Therefore, make sure all the latest updates and security patches.
Note
1. If use USB port, the cable should be connected to the computer port.
2. When use computer port, USB driver should be installed at first.
For more information, please refer to the software user manual in the supplied USB
memory. It is recommended to use a USB cable shorter than 3m (121 inches).
Note
1. In order to save, search and retrieve the users’ data, the user should connect the
analyzer to the computer installed data management software offered free.
Printing is done through computer in this case.
2. The professional consulting software optionally provides various printouts.
In case of using the software, the pre-printed result sheet is not used.
3. Please refer to the software user manual in the supplied USB memory.

SELVAS Healthcare User Manual BC380
24_Installation Installation_25
Installation
Connect the Bluetooth Dongle provided by the company as an option to the Bluetooth terminal on the rear
side of the device. Please refer to the Bluetooth Setup section of the ACCUNIQ Manager manual.
PC connection - wireless (Bluetooth)
1) How to directly connect the device-printer
Connect the A4 printer to the USB (A) port on the back of this device with a printer cable.
How to connect printer
2) Connecting the device to computer-printer.
If you want to view the measurement results by connecting the A4 printer to the computer, connect it to
the computer using the USB cable to the USB (B) port on the back of the device, and then connect the A4
printer provided by our company with the printer cable. Then, you can print the analysis results.
Note
1. When you want to save or search customer data using the Data Management Program
provided as standard, you must connect the computer where this program is installed,
and in this case, you can print through the computer.
2. If you select professional consulting software as an option, various print contents are
provided, and special paper is not used in this case.
3. For information on how to install and use Windows programs, refer to the user manual
built in each program.
This device can be connected to the autonomous Blood Pressure Monitor as an option.
Connect the Blood Pressure Monitor (RS-232C) terminal on the rear part of this device and the Blood
Pressure Monitor with the provided Blood Pressure Monitor cable.
Blood Pressure Monitor connection
Note
For the RS-232C communication cable connecting the Blood Pressure Monitor, be sure
to use the dedicated cable (within 1.5 meters) provided.

SELVAS Healthcare User Manual BC380
26_Installation Installation_27
Installation
[Thermal printer FEED/CUT function]
・ FEED function
If you enter the password ‘5555’ and press ‘OK’
on the user setup screen, thermal paper will be
printed out.
・ CUT function
If you enter the password ‘8888’ and press ‘OK’
on the user setup screen, thermal paper will be
cut.
Note
Do not pull the paper during thermal paper printing. When printing is complete, the paper
is automatically cut.
Operation of the device during printing may cause devicefailure.
You have to replace thermal paper while the power is on.
1) Open the printer cover by turning the thermal printer fixing screw counterclockwise as shown in the picture.
2) Remove the protective thermal paper inserted into the thermal printer by pulling it slowly in the direction of
the arrow.
3) If you gently push the end of the thermal paper toward the printer inlet as shown in the picture,
the thermal paper will be printed and the paper will be cut automatically.
4) Close the printer cover and fix the printer cover by turning the fixing screw clockwise.
How to replace thermal paper in the thermal printer
1
3
4
2

SELVAS Healthcare User Manual BC380
28_System Setup System Setup_29
System Setup
The setup menu is made up of icons. The functions of each icon is as follows:
Type
4
8
3
7
2
6
1 5
9
SYSTEM SETUP
Enter the ‘System Setup’ screen
Press the “ ” icon at the bottom left of the initial
screen and enter the 0-0-0-0 password to enter
the ‘System Setup’ screen.
You can change the initially set password in Basic
settings-Password.
‘User Setup’ is built-in to control the operating environment of the device, so the user can set the device
environment.
Note
The contents of the device’s system setup are subject to change to improve performance.
Menu Setup item
1 Basic settings Date / Time
Date format
Select measurement unit
Volume
Measuring Music
Language
Password
Non-member use
2 Data management Check data / print / delete data
Copy data to Excel file
Data backup / restore / merge
3 Printer / Printing Printer setup
Set automatic printing preferences and the number of
automatic prints
Adjust print position
Search / Connect
4 Results ACCUNIQ exclusive paper
Change logo
Adult / Child outcome settings
Abdominal obesity analysis result settings
5 Communication Internet
Program
6 Weight Weight measurement / Input selection
Weight correction
Height correction
7 Management options Use optional equipment
Standard range
Cloud Setting
8 Display Setting Result Display For Summary
Lock Screen
9 Customer Support Problem solving
Product information

SELVAS Healthcare User Manual BC380
30_System Setup
Setup
・ Date/Time: Use the ∧ and ∨ buttons to set the current date and time.
・ Date format: Use the < and > buttons to select the desired date format from YYYY-MM-DD, MM-DD-YYYY
or DD-MM-YYYY.
・ Select measurement unit: The unit of weight and height measured or entered in the Body Composition
Analyzer is displayed in currently selected kg/cm.
・ Volume: Set the volume of the guidance voice.
・ Measuring Music: Set the background music that plays when measure body composition.
・ Language: Set the displayed language.
・ Password: Set the password to enter when entering user setup. The initially set password is ‘0000’ when
released. Please be careful that, if you forget your password after changing it, there is no way to find it again.
・ Non-member use: Enable/disable the member function.
Basic settings
Note
1. If you press the ‘Save’ button in the middle of setting, the setup you have created to
that point are input to the device and press the ‘ ’ button to return to the previous
screen.
To cancel the settings, press the ‘ ’ button to cancel the entered information.
The settings that were originally saved are maintained.
2. When managing data using software, the ‘Date/Time’ set in the PC is automatically
entered.
System Setup_31
System Setup
・ Check data / print / delete data: You can confirm, print, or delete data saved in the device.
・ Copy data to Excel file: All or part of the saved data is saved as an Excel file.
・ Data backup / restore / merge: Load data files saved in USB memory to the device or save data saved in
the device to USB memory.
Data Management
・ Printer setup: Select the printer type to be connected.
・ Set automatic printing preferences and the number of automatic prints: Select whether to print
automatically/manually and the number of prints.
・ Adjust print position: Adjust the print position when printing on special paper.
・ Search / Connect: Search the connected printer to select and connect.
Printer / Printing

SELVAS Healthcare User Manual BC380
32_System Setup
・ ACCUNIQ exclusive paper: Select paper for printing the result from special paper/A4 paper.
・ Change logo: Change the logo displayed on the top right of the result sheet.
・ Adult / Child outcome settings: Select a result sheet for adult or child.
・ Abdominal obesity analysis result settings: Select whether to evaluate abdominal obesity for children/
adolescents under the age of 18.
Results
・ Internet: Select the internet connection method (WiFi or Ethernet) and connect to the internet to connect
your device to the ACCUNIQ cloud service.
・ Program: Select the program you want to use, ACCUNIQ Manager or ACCUNIQ Dashboard, and select the
connection method (Bluetooth or USB cable).
Communication
Note
1. To use ACCUNIQ Dashboard, the date management app for administrators, and
ACCUNIQ Connect, the app for measurers, you must connect your measurement
device to the ACCUNIQ Cloud in the internet settings.
2. Bluetooth, BLE modules are only available for models specified as below.
Other modules are not supported.
- For Android devices: Bluetooth module HC-05-USB Bluetooth (supplied by the
company) or Parani SD1000U by Sena Technology (separately sold)
- For Android or iOS devices: BLE module ChipSense RCPORT-TD420 (separately sold)
System Setup_33
System Setup
・ Weight measurement / Input selection: Select whether to measure or input weight.
・ Weight correction: Correct the weight value.
・ Height correction: Correct the height value.
Weight
・ Use optional equipment: Select a connected optional device such as a Height Meter or Blood Pressure Monitor
・ Standard range: Select the adult BMI range and abdominal circumference range.
・ Cloud Setting: Check the ACCUNIQ Connect server address and save the time zone as the Universal
Time Coordinated (UTC) time zone of the region where you are using the device.
Management options
[Internet Settings]
1) Connecting to the Internet via WiFi
・
Select [Connect to WiFi] from Communication > WiFi Connection.
・
When the available networks are displayed, select the network you want to use and then select the [WiFi
Connection] button.
・
After entering the password, press the [OK] button to connect to WiFi.
2) Connecting to the Internet via Ethernet
・
Select [Connect to Ethernet] from Communication > WiFi Connection.
・
Connect an Ethernet cable to the Ethernet port labeled [PC] on the back of the device.
・
Select [Obtain an IP address automatically (DHCP)] or [Set IP manually (Static)] to connect to the Internet.
・
If you select [IP Manually (Static)]', please enter the IPv4 address, subnet mask, and gateway DNS server
information correctly.

SELVAS Healthcare User Manual BC380
34_System Setup
・ Problem solving: You can check the cause of the problem and how to solve it.
・ Product information: Check the software version of the impedance body fat scale and the serial number of
the device.
Customer Support
・ Setting Result Display For Summary: Set summarized result display item Set the summarized result display
items on the result screen.
・ Lock Screen: Set the Lock Screen function.
Display
Measurement and Analysis_35
Measurement and Analysis
MEASUREMENT AND ANALYSIS
Precautions for measurement
The reliability of the results obtained by analyzing body composition can be judged based on the accuracy.
If the precision of the device, which means how much the body composition analysis value matches the actual
body of the user, is called 'accuracy', the safety of the device that analyzes the same value under the same
physical condition and analysis environment is called 'reproducibility'.
In order to ensure the accuracy of the analysis value, the user must comply with the following:
In order to maintain the balance and health of body composition by examining changes in body composition
through continuous body composition analysis, and comparing the results, the analysis is carried out by
maintainging constant physical and environmental conditions.
In this case, it is more important to maintain the same conditions for analysis than to maintain the same conditions
for the accuracy of analysis results. What is used in this case, is the reproducibility of the device. When the
amount of food, the elapsed time after a meal, and the activity conditions (exercise, temperature, sauna, intake of
large amounts of drinks, urination, etc.) that occur within the time affecting the analysis are kept constant, and the
reproducibility of the device is secured, it means that the analysis results indicate the amount of continuous body
composition change, so it can be used to evaluate the desired change in body composition.
1. Water content increases after a meal, so measure on an empty stomach as much as possible.
・ Measure 3 to 4 hours after a meal.
・ Avoid caffeine or beverages which causes diuretic action, such as coffee, within 4 hours of measurement.
・ Drink 2 glasses of water 2 hours before the measurement.
2. Keep it stable so that the distribution of body fluid is constant before measurement.
・ Measure 3 to 4 hours after bathing, sauna, or sweaty exercise or activity.
・ Or measure before exercise or sweaty activity.
3. Do not drink alcohol for 24 hours before measurement.
4. Wear light clothing for measurement.
5. If you suddenly change your posture from a sitting position to a standing position, body water will come
down to your lower body and affect the measurement.
Measure after holding the standing position for 5 minutes.
6. Clean the measuring area and the electrode, and make sure that there is no foreign object between them.
7. Changes in room temperature may also affect the measurement results.
At room temperature around 20℃, measure after the body's warmth has stabilized.
8. Changes in weight and body composition occur throughout the day, so measure at a fixed time.
For people who stand for a long time, measuring in the morning will give more accurate analysis results.
9. Before measuring, go to the bathroom to remove as much body remnants as possible.
10. For measurements, keep the correct measurement method and posture.
Note
If the time zone is set incorrectly in [Cloud Settings] > [Time Zone Settings)', it may not
synchronize with the server and the measurement date in Accuniq Manager and on the
device may not be correct.

SELVAS Healthcare User Manual BC380
36_Measurement and Analysis
Correct Measuring Posture
・ Check whether the foot electrode or the electrode on the handle is clean.
・ Remove socks or stockings when using a foot electrode.
・ Remove sweat or foreign substances from your hands and body parts that will come into contact with the
hand or foot electrode.
・ Stand so that your feet contact the foot electrode below.
In this case, be careful not to get your pants or clothes caught between the electrodes and your feet.
・ Remove sweat or foreign substances from hands.
・ Hold the hand electrode so that your palm and fingers touch the entire electrode.
How to stand on a scale with a foot electrode
How to hold the hand electrode
Note
If 8 electrodes are not perfectly touched during the measurement, the result is not reliable
or the device quits the measurement.
1. When the subject has hands or feet that are too small to cover all electrodes sufficiently
for measurement, please pay attention to touch all electrodes fairly.
How one touches electrodes will affect the reliability of the analyzed value.
2. During measurement the subject should not be touched by another person or by
conductive materials.
3. If 8 electrodes are not perfectly touched during measurement, measuring is quit or the
data is not reliable.
(O) Standing correctly (X) The foot is biased
towards one electrode
(X) Foreign substance is stuck
(O) The correct holding posture (X) The hand is biased towards
one electrode
(X) The hand is biased
towards one electrode
Measurement and Analysis_37
Measurement and Analysis
・ After stepping on the foot electrode well, hold the hand electrode correctly and keep the arm at a 30˚angle to
the body.
・ Do not move or speak until the measurement is complete.
・ If contact with the electrode is not maintained during the measurement, the measurement may be interrupted.
・ Do not bend or shake your arm until the measurement is complete, remaining stationary.
Correct measuring posture

SELVAS Healthcare User Manual BC380
38_Measurement and Analysis
Measurement
Basic analysis
If you want to measure, stand on the device's scale.
1) Measure Weight
・ When you step on the scale, the screen changes
with a chime bell, and the weight measurement starts.
At this time, do not move or speak until the weight
is completely measured, and be careful not to place
your hands on other parts or lift an object while
measuring on the scale.
2) When the weight measurement is completed, the
measurement result is displayed.
Select either member examination or non-member
examination.
3) Enter personal information
Data required for analysis is entered in the following
order: ID, height, age, and gender.
・ Enter ID (for member examination)
・ The ‘ID’ field blinks first.
・ Enter your ID on the touchpad.
You can enter up to 20 characters including
alphanumeric characters.
・ Enter height
・ Enter the height of the user using the keypad and
the numeric keypad of the touchpad.
・ Analysis is not possible if the height of the user is out
of the input range.
・ After entering it, press the ‘Next’ button.
Measurement and Analysis
Measurement and Analysis _39
・ Enter your date of birth
・ Enter the measurer's date of birth using the numeric
keypad of the touchpad.
・ Enter gender
・
Select the gender, male or female using the touchpad.
4) Start measurement - fix foot position
・ Stand with your feet in contact with the foot
electrode correctly.
5) Start measurement - contact with the hand electrode
・ Wrap and hold the hand electrode with your entire
hand so that all your fingers contact with the
electrode.
Table of contents
Popular Scale manuals by other brands

RADWAG
RADWAG C315 Series user manual

Giropes
Giropes Baxtran BAP manual

OHAUS
OHAUS SD Series user manual

Melchioni
Melchioni PLANA user manual

Integrated Visual Data Technology Inc.
Integrated Visual Data Technology Inc. SkidWeigh ED2E-AT Elite Installation & Calibration Manual

KERN
KERN MBB operating instructions