Turn on the unit by sliding the switch on the top right of the Fluorotek 200
to the left.
Press the Measure button.
Press the Ch 360 to select the specific wavelength for the E. coli test.
Add 2 drops of buffer solution to a cotton swab provided with the test kit.
Rub the surface to be tested with the cotton swab.
Place the swab into the pre-filledtest vial provided with the test kit.
Remove swab & add two drops of reagent topre-filled test vial.
Remove the plastic light shield from Fluorotek 200 sample chamber.
Place testvialinto Fluorotek 200 with the tab facing the AccuStrata logo.
Replace the plastic light shield on the Fluorotek 200 sample chamber.
Press the Blank button.
Press the Measure button.
Wait for the 10-minute timer to complete.
A measurement will automatically be taken after 10 minutes.
Compare measured results to the threshold limit provided with the E. coli
test kit.
If the value on the screen is higher than the threshold limit, then E. coli has
been detected. The threshold value will be provided with test kit.