Acoustic control systems A1214 EXPERT User manual

OKP 42 7612
ULTRASONIC FLAW DETECTOR
А1214 EXPERT
USERS GUIDE
APYAS.412231.012 RE
Acoustic Control Systems
Moscow 2010


Ultrasonic aw detector А1214 EXPERT
Users guide 19.08.2010 3
Table of Contents
1 APPARATUS DESCRIPTION AND WORK .......................................... 6
1.1 DEVICE APPOINTMENT....................................................................... 6
1.1.1 Appointmentoftheawdetector ........................................................... 6
1.1.2 Serviceconditions................................................................................. 6
1.2 TECHNICAL CHARACTERISTICS ....................................................... 6
1.2.1 Thekeyparametersofthedevice ........................................................ 6
1.2.2 Instrumentalcharacteristicsofthedevice.............................................. 7
1.2.3 Characteristicsoftheprobepulsegenerator......................................... 7
1.2.4 Characteristicsofareceptionpath ....................................................... 8
1.2.5 Powersupplyandpowerconsumptioncharacteristics......................... 8
1.2.6 Constructivecharacteristicsandreliabilityparameters.......................... 9
1.3 APPARATUS ARRANGEMENT AND WORK ...................................... 10
1.3.1 Flawdetector’sarrangement ............................................................... 10
1.3.2 Maininterfaceprinciples...................................................................... 15
1.3.3 Flawdetector’soperatingmodesandtheirinterrelation..................... 15
1.3.4 Informationrepresentationonthescreen........................................... 16
1.3.5 Flawdetector’skeyboard.................................................................... 17
1.3.6 Usingofpictographs............................................................................ 20
2 PROPER USE .................................................................................... 22
2.1 OPERATIONAL RESTRICTIONS ....................................................... 22
2.2 APPARATUS PREPARATION FOR USE ........................................... 22
2.2.1 Flawdetectorswitchingonandoff ...................................................... 22
2.2.2 Choosingoftheconverter................................................................... 22
2.2.3 Installationofparameters ................................................................... 24
2.2.4 Calibrationoftheinclinedpiezoelectricconverter(PEC)..................... 25
2.2.5 Adjustmentofsensitivity ...................................................................... 28
2.3 APPARATUS OPERATING MODES ................................................... 29
2.3.1 MENUmode ........................................................................................ 29
2.3.2 SEARCHmode.................................................................................... 33
2.3.3 REVIEWmode..................................................................................... 38
2.3.4 MAGNIFIERmode.............................................................................. 39
2.3.5 B-SCANmode .................................................................................... 40
2.3.6 Auxiliarymodesandfunctions............................................................ 42
2.4 MEASURING ....................................................................................... 51
2.4.1 Measuringoftimeandcoordinates...................................................... 52
2.4.2 Measuringofthesignalamplitudes ..................................................... 54
2.5 USE OF SETTINGS DEPOSITORY.................................................... 55
2.5.1 Readingofcongurationfromdepository............................................ 56
2.5.2 Creatingandsavingnewcongurations.............................................. 57

4 Users guide 19.08.2010
Ultrasonic aw detector А1214 EXPERT
3 MAINTENANCE .................................................................................. 58
3.1 UTILITY POWER AND ENERGY CONSUMPTION ............................ 58
3.1.1 Powersupplyenergycontrol ............................................................... 58
3.1.2 Accumulatorpreburning....................................................................... 58
3.2 PERIODIC MAINTENANCE ................................................................ 59
3.3 POTENTIAL PROBLEMS.................................................................... 59
4 STORAGE ........................................................................................... 60
5 TRANSPORTATION ............................................................................ 61
ATTACHMENT А ............................................................................................ 62

Ultrasonic aw detector А1214 EXPERT
Users guide 19.08.2010 5
The present users guide (further in the text - a manual) contains
technicalcharacteristics,descriptionofthedeviceanditsactionprinciple.
Italsoincludesthedatanecessaryforcorrectoperationoftheultrasonic
awdetectorА1214EXPERT(furtherinthetext–awdetectorordevice).
Beforethebeginningofdeviceoperationpleaseattentivelystudythe
presentmanual.
Manufacturer’spermanentjobinthesphereofperfectingthedevice
possibilities, increasing of reliability and convenience of operation can
sometimes lead to some unprincipled changes which have not been
reectedinthepresenteditionofamanualandnotworseningdevice’s
technicalcharacteristics.
Thedeviceisissuedbythemanufacturer:
«Acoustic Control Systems» Ltd (“ACS” Ltd)
Russia, 115598, Moscow, street Zagoryevskaya, hultrasonic 10, building 4
Phone/fax: (495) 984 7462 (multichannel)
E-mail: [email protected]
Website: www.acsys.ru

6Users guide 19.08.2010
Ultrasonic aw detector А1214 EXPERT
1 APPARATUS DESCRIPTION AND WORK
1.1 DEVICE APPOINTMENT
1.1.1 Appointmentoftheawdetector
The aw detector А1214 EXPERT is related to ultrasonic general-
purposeawdetectorsformanualcontrol.
Theawdetectorisdesignedforsearch,denitionofcoordinatesand
estimationofvariousinfringementssizesincontinuityanduniformityofa
materialinproductsmadefrommetalsandplastic.
The aw detector allows to form, register and keep in non-volatile
memorytimerealizationsofechosignals.Theelectroluminescentdisplay
oftheelectronicblockprovidesdisplayingofechosignalsintheformof
A-involute, and also the images of control objects sections in the form
ofB-scan.Manualandautomaticmeasurementoftimeintervals,signals
amplitudesandautomaticcalculationofdefectscoordinatesareprovided.
ThesoftwareAdvancedDataManager(ADM),includedintotheset
ofdevicedeliveryallowstotransferthedatafromthedevice’smemoryto
theexternalpersonalcomputer(personalcomputer)fortheirsubsequent
analysis and documenting. Also there is a possibility to make device
adjustmentthroughthecomputer,toobserveonthecomputerscreenecho
signalsinrealtimeandscale,toprocessthem(computerawdetector
mode).Herewithcommunicationwiththecomputeriscarriedoutthrough
USBport.
1.1.2 Serviceconditions
Thedeviceisdesignedforoperationunderthefollowingenvironment
conditions:
– temperaturefromminus30toplus45centigradedegree;
– relativeairhumidityupto98%withthemaximumtemperature
35centigradedegree.
1.2 TECHNICAL CHARACTERISTICS
1.2.1 Thekeyparametersofthedevice
Thekeyparametersofthedevicearedisplayedinthetable1.

Ultrasonic aw detector А1214 EXPERT
Users guide 19.08.2010 7
Table1
Parameter’sname Value
Nominalworkingultrasoundfrequencies,MHz 0,5-15,0(±10%)
Dynamicrange,dB,notless 100
Rangeoftimeintervalschanges,mcs 1 - 1200
Rangeofultrasoundspeedinmaterialtuning,m/s 1000 - 15000
1.2.2Instrumentalcharacteristicsofthedevice
Instrumentalcharacteristicsofthedevicearedisplayedinthetable2.
Table2
Characteristic’sname Value
Limitofassumptivevalueofbasicabsoluteerror
inthicknessXmeasurements,mm,nomore ±(0,01X+0,1)
Limitofassumptivevalueofbasicabsoluteerror
intimeintervalmeasurementst,mcs,nomore ±(0,01t+0,1)
Nonlinearityofdevelopmentonavertical,%,no
more ±1
Limitsof adjustmentofadelayin development,
mcs 0,0-99,9
1.2.3Characteristicsoftheprobepulsegenerator
Characteristicsoftheprobepulsegeneratoraredisplayedinthetable3.

8 Users guide 19.08.2010
Ultrasonic aw detector А1214 EXPERT
Table3
Characteristic’sname Value
Theprobepulseform Meander
Numberofperiods 0,5-5,0
Theprobepulseparameters:
Amplitude(halfofswing),V
Durationoffronts,nomore,ns
20;100;200(±5%)
20
Frequencyoftheprobepulserepetitions,Hz 10;25;50;100;200
1.2.4Characteristicsofareceptionpath
Characteristicsofareceptionpatharedisplayedinthetable4.
Table4
Characteristic’sname Value
Frequenciesstripofareceptionpath,MHz 0,14-21,00
Effective value of the own noise, resulted in
toinput,mkV,nomore 20
Reorganizationrangeoftheattenuator,dB 0 - 90
Absolute error of the attenuator steps, dB, no
more ±0,2
Dynamicrangeoftimevariablegain,dB,notless 30
1.2.5Powersupplyandpowerconsumptioncharacteristics
Powersupplyandpowerconsumptioncharacteristicsaredisplayed
inthetable5.

Ultrasonic aw detector А1214 EXPERT
Users guide 19.08.2010 9
Table5
Characteristic’sname Value
Thepowersource Demountable
accumulator
Nominal value of the aw detector’s source
voltage,V 12
The electrical current consumed by the aw
detector under nominal value of the source
voltage:
Minimumlevelofthescreenluminescence,
mA,nomore
Maximumlevelofthescreenluminescence,
mA,nomore
300
500
Durationofacontinuouswork:
Minimumlevelofthescreenluminescence,h
Maximumlevelofthescreenluminescence,h
12
7
1.2.6Constructivecharacteristicsandreliabilityparameters
Constructivecharacteristicsandreliabilityparametersaredisplayed
inthetable6.
Table6
Characteristic’sname Value
Overalldimensionsoftheelectronicblock,mm 250x160x80
Weightwithpowerelements,kg 1,9
Screenresolution 320x240
Thesizesofadisplayingeldofthescreen,mm 115x86
Admissible range of temperatures under the
storageandtransportation,ºC -40-+50

10 Users guide 19.08.2010
Ultrasonic aw detector А1214 EXPERT
Characteristic’sname Value
Workingatmosphericpressurerange,kPa 84,0-106,7
Averagetimebetweenfailures,h,notless 32000
Targetdateofservice,years 5
1.3 APPARATUS ARRANGEMENT AND WORK
1.3.1 Flawdetector’sarrangement
Theawdetectorincludestheelectronicblockwiththereplaceable
storagebattery,asetofcablesforultrasonicconvertersofthevarioustype
connections,thechargingdeviceandvariousaccessories.
1.3.1.1 Theawdetector’selectronicblock
Theawdetector’selectronicblockprovideselectricimpulsesformation
forexcitationofthepiezoelectricconverter,strengtheningreceivedfromthe
piezoelectricconvertersignals,theirprocessing,visualization,formation
andrepresentationofmeasurementsresults,preservationofthedatain
non-volatilememory,datatransmissiontotheexternalcomputer.
Appearance of the electronic block is displayed on the picture 1.
Flawdetectormanagementiscarriedoutbymeansofthelmkeyboard.
Indications of signals, measurements results, aw detector conditions,
etc.arecarriedoutonthedisplayandbymeansoflight-emittingdiode
indicatorsonthedevicecase.

Ultrasonic aw detector А1214 EXPERT
Users guide 19.08.2010 11
The button for
handle position
adjustment
ASD Indicators
Sockets for
converters
connection
Picture1
Piezoelectric converter connection is carried out by means of the
coaxialcableswhichpresentinthesetofdelivery,throughsocketsLEMO
whicharelocatedinanicheoftherightwallofthecase.
Device’s power supply is carried out with the help of replaceable
accumulator,orfromtheexternalpowersourcewhichincludesintheset
ofdelivery,connectedtoLEMOsocket.Thesocketislocatedonaback
wallofthedevice(picture2).
Picture2
1.3.1.2 Chargingdevice
For the aw detector’s maintenance from external energy sources
andchargingofthereplaceableaccumulatorestablishedintheelectronic
blockthechargingdeviceisused.Itconsistsofthenetworkadapterand
chargerVS121.
The network adapter provides transformation of the network main
alternatingcurrentvoltage(100-240volt)andfrequency50-60Hzinto
constantvoltage12VnecessaryforchargerVS121work.
Theadapterisexecutedundertheschemewithpulsetransformation
ofvoltage.Itallowsconnectionwithnetworksofvariousvoltagevaluesand

12 Users guide 19.08.2010
Ultrasonic aw detector А1214 EXPERT
alsoreducesvoltageinstabilityinuenceswhicharepresentinindustrial
andhouseholdnetworks.
ChargingdeviceVS121transformsthedirectcurrentvoltage of12
V into necessary for aw detector’s work voltage and also carries out
accumulatorchargingandtrainingthatincreasesitsservicelife.
Dependingontheawdetector’sbatterydischargetheaccumulator
chargingcanlasttill6hours.Intheaccumulatorchargingprocesstheaw
detectorcancarryoutitsfunctionsinfullvolume(exceptacommunication
withcomputermode).
Toavoiddevicedamageunderwrongconnection,itisrecommended
toconnectdevicesinthefollowingsequence:connectthechargingdevice
cable to the electronic block; connect the charging device cable to the
charger;connecttheoutputcordofthenetworkadaptertothecharging
device; connect the network cable to the network adapter; connect the
networkcablewiththenetwork(picture3).
The electronic block
Charging device cable
Network cable
The network
adapter
Output cord of the network adapter
Charging device VS121
Picture3

Ultrasonic aw detector А1214 EXPERT
Users guide 19.08.2010 13
ATTENTION! DURING THE CONNECTION OF THE CHARGER AND
THE DEVICE WITH THE CABLE IT IS NECESSARY TO OBSERVE THE
SOCKETS KEYS AS IT IS SHOWN ON PICTURE 4. ARROWS SPECIFY
THE SOCKETS KEYS.
Picture4
Foroperatingmodeindicationontheforwardpanelofchargingdevice
VS121light-emittingdiodeindicatorsofthreevariouscolorsareplaced:
red, yellow and green. The red light-emitting diode designates voltage
presenceonchargingdeviceVS121input.Theyellowlight-emittingdiode
displaysaconditionofchargingprocess.AftertheconnectionofVS121to
thedeviceandswitchingitonaccumulatortestingbegins,theyellowlight-
emittingdiodeherewithblinks.Iftheaccumulatorwasnearlydischarged
thenthisprocesscantakeupto30minutes.Afterthatchargingprocess
begins,theyellowlight-emittingdiodeconstantlylights.Thegreenlight-
emitting diode signals about the termination of accumulator charging
process.
Asthereplaceableaccumulatortheassemblageof10nickel-metal-
hydride elements is used. In assemblage for short circuit and heating
preventionprotectionelementsareestablished.
ATTENTION:ITISFORBIDDENTODEMOUNTANDREPAIRTHE
DEMOUNTABLE ACCUMULATOR ON ONES OWN ACCOUNT!
1.3.1.3 Converters
Theawdetectorismadeforworkingwithcombinedandseparately-
combinedpiezoelectricconvertersofworkingfrequenciesfrom0,5to15,0
MHz
Inthedevicethegeneratorofbipolarprobingimpulseswithregulated
durationandnumberofperiodsisused.Thisprovidescompatibilitywith
piezoelectric converters which have integral interfacial inductances, or
don’thavethem,i.e.theawdetectoriscompatiblewiththemajorityof
piezoelectricconvertertypespresentedinthemarket.

14 Users guide 19.08.2010
Ultrasonic aw detector А1214 EXPERT
Dependingontheconvertertypetwotypesofcablesandaccordingly,
twowaysofconnectionwiththepiezoelectricconverterareused:
– Thecombinedconverters(typeS)areconnectedbymeansof
unarycableLEMO-LEMOtoasocketindicatedasIN(picture5).
Picture5
– Separately-combined (РС) converters (type D) are connected
bymeansofadoublecableLEMO-LEMO.ThesocketindicatedasOUT
servesforconnectionofthetransferringpiezoelectricelement,indicated
asIN-forreceivingpiezoelectricelement.UsuallyonРСconvertersthe
socketofatransferringpiezoelectricelementismarkedbyaredorawhite
point(picture6).
Picture6
1.3.1.4 Cablesandsockets
Connectingcablesofthedeviceareproducedfromthehigh-quality
coaxialcablesandsocketsofleadingworldmanufacturers(«Raychem»,
«LEMO»,etc.).
Applicationofthecableproductsdescribedaboveprovidesnecessary
exibility of cables at low temperatures (up to -30 centigrade degree),
reliablexingofconnectionsatsmalldimensionsandconnectionplaces
sockets and cables protection against fractures. This nally minimizes
refusalsinconnectingelementsandfacilitatesworkwiththedevice.

Ultrasonic aw detector А1214 EXPERT
Users guide 19.08.2010 15
1.3.2 Maininterfaceprinciples
In А1214 EXPERT there is an intuitive interface principle.
Associative menu pictographs in various modes, explanatory pictures
near the parameters, names and designations of keys allow to master
workwiththeawdetectoreasilyandsimply.
Thedataoutputonthescreenisrealizedinsuchamannerthatthe
necessaryforoperativecontrolinformationisalwaysthere.
Theawdetectorworkisaccompaniedbysoundsignalswhichcan
bedeactivated.
Theworkwith awdetectorisconsiderably facilitatedbypresence
ofsettingslibraryinwhichupto99congurationsofthedevicesettings
canbestored.Foreachcongurationusercanappointanameinlength
nomorethan14symbols.Beforeacongurationnameitsserialnumber
isautomaticallydisplayed.Thusawdetectoradjustmentundervarious
situationsandobjectsofcontrolcanbemadeinadvance.Thenyouneed
simplytochooseinthemenuthenecessaryone.
All aw detector options remain in the device’s memory after it’s
switchingoff,devicestoragewithoutthebatteryandundertheaccumulator
discharge.
1.3.3 Flawdetector’soperatingmodesandtheirinterrelation
Therearesomebasicworkmodesintheawdetectorе:theSEARCH,
theREVIEW,theMAGNIFIERandtheB-scan.Inthismodesformationof
aprobingsignal,strengtheningofacceptedsignals,theirrepresentation
on the aw detector’s screen, measurement performance and results
preservationinthedevice’smemoryarecarriedout.
IntheMENUmodechangeoftheawdetectoradjustmentparameters
is made on modes ОК or piezoelectric converter. The installation of
necessaryauxiliaryparametersisalsocarriedoutthere.
InthemodesSEARCH,REVIEWandMAGNIFIERthedeviceworkis
providedinthemodeofclassicalawdetector,allowingtovisualizesignals
intheformofA-involuteandthusmanuallyorautomaticallymeasurethe
signalsamplitudes,timeintervals,andalsodistancestakingspeedinto
account.
IntheSEARCHmodetheA-scan of a signal is displayed with the
maximumscreendisplayingofthenumericalmeasuringinformation.The
modeSEARCHisconvenientduringtheworkwithinclinedconverters.In
theREVIEWmodetheA-scanofasignalisdisplayedwiththemaximum
sizeonthescreenwhilethenumericalmeasuringinformationisminimized.

16 Users guide 19.08.2010
Ultrasonic aw detector А1214 EXPERT
This is convenient during the analysis of the general ultrasonic signals
formandworkwithdirectconverters.
In the MAGNIFIER mode the maximum quantity of the operative
measuringinformationispresentedonthescreenandthesizesofsignals
displaying area are limited. Herewith two A-scans are simultaneously
displayed-thegeneralandstretchedtimeintervalcorrespondingtoarea
oftherststrobe.Thefeatureofthegivenmodeusingismeasurement
withtheraisedaccuracy.
TheB-scanmodeismadeforformationofonecross-sectionsection
ОК(B-scans).
Besidesthemainthereareauxiliarymodes:STOP,COMMUNICATION,
CALIBRATIONOFTHEINPUTPIEZOELECTRICCONVERTERCORNER,
ADJUSTMENTOFTHESWEPTGAIN,DGS-DIAGRAMADJUSTMENT.
The STOP mode realizes “freezing” possibility which means
preservationandviewingofmeasurementsresultsandsinglerealizations
ofthe signalsreceivedin modesSEARCH,REVIEWandMAGNIFIER.
Thesavedlescanbetransferredtothepersonalcomputerforstorage
andusingforultrasoniccontrolprotocolsmaking.
In the COMMUNICATION mode data transmission to the personal
computeriscarriedout.Itlaunchesautomaticallyafterconnectiontheaw
detectorwithpersonalcomputer.Herewithmanagementofthedeviceis
completelytransferredtothecomputer.Duringtheconnectiontopersonal
computerthedevicecannotbeswitchedoff.
The CALIBRATION OF THE INPUT PIEZOELECTRIC CONVERTER
CORNER mode is used for semi-automatic procedure of input corner
measurementanddelayofinclinedpiezoelectricconverter.
ADJUSTMENTSWEPTGAINmodeisused for corrections of
signalsabatementwithdistanceeffect.
ADJUSTMENT DGS-DIAGRAM mode is used for adjustment of
sensitivityanddenitionoftheequivalentreectorsarea.
1.3.4 Informationrepresentationonthescreen
In the aw detector the monochrome graphic electroluminescent
displaywiththescreenresolutionof320х240pointsisusedasindicator.
Theworkingspaceofthescreenineachmodeisdividedintosome
functional areas. For example on picture 7 the screen is seen in the
REVIEWmode.

Ultrasonic aw detector А1214 EXPERT
Users guide 19.08.2010 17
3 1 2
4
5 6
7
Picture7
Thetoplineindicatesthenameofcurrentoperatingmodeandthe
serviceinformation.Intheeldoftheserviceinformationthereisacurrent
date and time , temperature inside of a device , equivalent time of
continuousdeviceworkduringaccumulator(timer)powersupplyfromthe
lastaccumulatorchargingmoment,powersupplysourcecondition.
Intherightpartofthetoplineinfunctionalmodesmeasuringinformation
canbedisplayed.
Inthebottompartofthescreentheareaofpictographsislocated.
Withtheirhelpthefollowingoperationsaremade:choiceofanoperating
mode,adjustmentofasignalsrepresentationway,adjustmentofstrobes
andthresholddevicesandsoforth.
The average part of the screen is used for results measurements
displaying,owingparametersandawdetectorsettings.
1.3.5 Flawdetector’skeyboard
Theawdetector’skeyboardisdisplayedonpicture8.

18 Users guide 19.08.2010
Ultrasonic aw detector А1214 EXPERT
CANCEL key ENTER key
The condition indicator
Device
ON/OFF
Display
brightness
settings
Block of control keys
SETTINGS key FREEZING key
Pictograph choosing keys
Monitor
Picture8
There are: tuning the power on/off key, 17 multipurpose keys and
threelight-emittingdiodeindicators.
The light-emitting diode indicator over a key shows aw
detector’s condition. Green color means that the device is charging
from the accumulator, the blinking green species in a low charge of
theaccumulator.Yellowcolor-adevicechargingiscarriedoutfromthe
networkadapter.Tworedindicatorsunderakey showhowoperation
monitorworksfortherstandsecondstrobesaccordingly.
The symbolical designation of their basic functions is put on keys.
The English-speaking designation of keys is chosen for unication of
device’sdesignandawdetector’soperationaldocumentationwhileusing
invariousregions.
Thekeys , andalsokeysofpictographschoosingforstrobes
managementandworkwithsweptgain/DSG-diagramrealizethefollowing
principle:«TheFirstpressing-Enter,repeated-Exit».Otherkeysofa
pictographschoicecarryoutswitchingofmodesonunitarypressing.
Regulatingkeys , , Inuenceonthe
currentactiveobject.Theiractionsaresimilarforvariousoperatingmodes
of the device.They are intuitively understood by the operator, i.e. their
symbolscorrespondtocharacteroftheiraction.
Forsomekeysthereisamodeofautorepeatwithaccelerationwhen
pressingmorethan1second.

Ultrasonic aw detector А1214 EXPERT
Users guide 19.08.2010 19
Whenthesoundisonpressingthekeys are accompanied by a
shortsignal.
The short description of keys appointments for the basic device
operatingmodesisresultedintable7.
Table7
Key
Keyfunctionsindifferentmodes
Search Rewiew Loupe B-scan STOP MENU
ApparatusswitchingONandOFF.
Youneedtoholdthekeynotlessthan0,5s
Time-baserangeswitching
Moving
thecursor
inB-scan,
rewiew
of the
A-scans
written
Choosing
ofactive
pictograph
Choosing
ofsettings
menu
mode
Metriccursormoving
Active
parameter
changing
Attenuatormanagement Threshold
shift -----
Choosing
of the
active
string
in the
settings
menu
Cross-overtotheMENUmode.
Whenholdingmorethan3secondsthewindowwithcurrent
congurationparametersisopened
MENU
modeexit
Cross-overtotheSTOPmode
Exitthe
STOP
mode
-----

20 Users guide 19.08.2010
Ultrasonic aw detector А1214 EXPERT
Key
Keyfunctionsindifferentmodes
Search Rewiew Loupe B-scan STOP MENU
Displaybrightnesssetting.
Eachpressingincreasesthebrightness.
Afterthemaximunbrightnessisachievedifpressagainthenitwill
becomeminimumagain.
Pectographschoosingbuttons
BasiclevelswitchingON/OFF
Recordof
thenext
section
Record
into
memory
Function
activation
Informationalwindowwithcurrent
congurationopening
Section
clearing ----- -----
Managementoftheseparatedevicefunctionsiscarriedoutbymeans
ofakeycombination(whilethe1stkeyispresseditisnecessarytopress
2ndkey).Thekeycombinationsusedandtheirappointmentaredisplayed
intable8.
Table8
Key 1 Key 2 Appointment
Apparatusswitchingonwithdefaultvalueofall
settingsandparameters
Directchoiceofadevicecongurationoptions
fromthesettingslibraryinactivemodes;
SEARCH,VIEW,LOUPEandB-SCAN
1.3.6 Usingofpictographs
Thefeature of the device’s interface is presence of associative
pictograph menu which are located in six rectangular windows in the
bottompartofthescreen.Pictographsaresymbolicalimages,associated
with object or operation they execute. Each mode has its own set of
pictographs.