Advantage PE-21 User manual

Required Equipment and Solutions
 • pHmeter
 • KnownvalueofsystemwaterpH
 • pHBuffer7.0
 • pHBuffer10.0
 • Washbottlelledwithdistilledorde-ionizedwater
 • Cleanbeakers
Preparation of the Electrode for Initial Use
1. Removetheprotectivebottleorcoverfromtheelectrodeand
thoroughlyrinsetheelectrodewithdistilledwater.Wipecarefullywith
acleanlabwipe.
2. Duringshipment,airbubblesmayhavemigratedintotheelectrode
sensingbulb.Holdtheelectrodeuptothelightandinspectthe
sensingbulbforairbubbles.Ifairisseen,carefullyshakethe
electrodedownward(likeathermometer)todispelanyairbubble
fromthesensingbulbatthetipoftheelectrode.
In-Line Standardizing the Electrode with Controller
1. ConnectyourpHelectrodetotheinputconnectoronthepH
controller.EnsurethattheelectrodeBNCandsolutionreference
connectionsaresecure.
2. Installtheelectrodeinthesamplelineandturnowon.Thetipof
theelectrodemustbepointeddownnomorethan40°fromvertical.
Wait2-3minutes.
3. Whilethereadingisstabilizingnavigatetothecontroller’spH
calibrationmenuandresettofactorydefaultsifavailable.
4. IfthepHreadingisstableandwithin1-2pHofyourtestedknown
valueforthesolutionperforma1pointcalibrationtomatchcontroller
andelectrodetotestedvalue.
Standardizing the Electrode to Solutions
1. ConnectyourpHelectrodetotheinputconnectoronthepH
controller.EnsurethattheelectrodeBNCissecure.
2. PlacetheelectrodeintoabeakercontainingpH7.00bufferandstir.
Whilereadingisstabilizingnavigatetothe2PointCalibrationmenu
ofcontroller,ifavailable.
3. Whenthereadingisstable,adjustthecontrollertoreadthevalueof
thepH7.00bufferforthelowcalibrationpoint.
4. Removetheelectrodefromthebuffer7.Rinsewithdistilledwater
andblotwithalabwipe.
5. Placetheelectrodeinabeakercontainingthebuffer10andrepeat
step3calibratingthehighcalibrationpoint.
Storing the Electrode
Short Term
Betweenmeasurements,storethepHelectrodeinabeakercontaining
pH4.0buffer.
Long Term
Whenstoringforlongerperiods,storethepHelectrodeinthestorage
bottleortheprotectivebootwhichcamewiththeelectrode.Ensurethat
thefoaminthestoragebottleorthecottonballintheprotectivebootis
totallywettedwithDlwatertokeepamoistenvironmentaroundthepH
bulbandjunction.Maintainthemoistenvironmentinthestoragebottleor
intheprotectivebootduringstorage.
Storingelectrodesformorethan2-3months,dry,oratextreme
temperatureswillshortenexpectedoperationallife.
Electrode Cleaning
Donotusestrongsolvents(e.g.acetone,carbontetrachloride,etc.)
tocleanthepHelectrode.Besuretorecalibratetheelectrodeafter
cleaning.
1. Iftheelectrodehasbecomecoatedwithoilorgrease,carefully
washtheelectrodeunderwarmtapwaterusingdish-washing
detergent.Rinsethoroughlywithfreshtapwaterfollowedbyarinse
withdistilledwater.SoaktheelectrodeinpHelectrodestorage
solutionfor30minutesafterthiscleaningprocedure.Recalibratethe
electrodebeforeuse.
2. Iftheelectrodehasbeenexposedtoproteinorsimilarmaterials,
soakinacidicpepsinfor5minutes.Rinsethoroughlywithdistilled
water.Recalibratebeforeuse.
3. Ifthepreviouscleaningproceduresfailtorestoreresponse,soakthe
electrodein0.1NHCIfor30minutes.Rinsethoroughlywithdistilled
water.Recalibratebeforeuse.
4. Ifelectroderesponseisnotrestored,replacetheelectrode.
Electrode Expected Service Life
1. Electrodeshaveanexpectedservicelifeincleanwatersolutions
of12-18months.Annualreplacementisrecommendedforbest
controllingandmonitoringresults.
2. Harshanddirtywaterwillshortenthelifeexpectancyofthe
electrode.
Warnings
1. Theelectrodetipmustbekeptwetatalltimes.Ifitisallowedtodry
outdamagecanoccurrequiringreplacement.
2. Iftheglasssurfaceiscrackedordamagedtheelectrodewillnotwork
properly.
PE-21 pH Probe
Instructions
www.AdvantageControls.com
918-686-6211
pH/ORP Electrode Products Warranty
Thisproductdesignedandsoldforuseincirculatingwaterapplications
iswarrantedtobefreefromdefectsinmaterialsandworkmanship
foraperiodofsix(6)months,providedthattheproductisusedin
accordancewiththeinstructionsprovidedandthattheproducthas
notbeensubjectedtobreakage,alteration,misuse,abuseorusedin
anapplicationnotnormallyintendedfortheproduct.Intheeventofa
warrantedfailurewithinthewarrantyperiod,contactyourrepresentative
duringregularbusinesshours,andaskforcustomerservice.
Pleasebepreparedtodiscussthedetailsofthedifculty.Ifnecessary,a
ReturnMaterialsAuthorization(RMA)numberwillbeissued.Materialsor
goodsreturnedwithoutanRMAwillnotbeaccepted.Returntheproduct
freightprepaid.
Thewarrantydescribedaboveisexclusiveandinlieuofallother
warrantieswhetherstatutory,expressorimpliedincluding,butnotlimited
to,anyimpliedwarrantyofmerchantabilityortnessforaparticular
purposeandallwarrantiesarisingfromthecourseofdealingorusageof
trade.Thebuyer’ssoleandexclusiveremedyisforrepairorreplacement
ofthenon-conformingproductorpartthereof,butinnoeventshall
dealersoragentsofanytierbeliabletothebuyeroranypersonforany
special,indirect,incidentalorconsequentialdamageswhethertheclaims
arebasedincontractintort(includingnegligence)orotherwisewith
respecttoorarisingoutoftheproductfurnishedhereunder.
Nootherrepresentationsofwarrantymadebyanyperson,including
dealers,employeesoragentsshallbebinding.
5/2014

Popular Measuring Instrument manuals by other brands

Pro OCC-CWDM-301-18B Operation guide

Pro

Pro OCC-CWDM-301-18B Operation guide

PCB Piezotronics IMI Sensors Y626B03 Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI Sensors Y626B03 Installation and operating manual

Quidel Sofia user manual

Quidel

Quidel Sofia user manual

Bosch gll 3-50 professional Original instructions

Bosch

Bosch gll 3-50 professional Original instructions

MOBILTEX RMU1 Installation and configuration guide

MOBILTEX

MOBILTEX RMU1 Installation and configuration guide

Stabila LD 400 operating instructions

Stabila

Stabila LD 400 operating instructions

PEWA AMPROBE DM-4 manual

PEWA

PEWA AMPROBE DM-4 manual

HunterLab LabScan XE Supplemental manual

HunterLab

HunterLab LabScan XE Supplemental manual

Pittsburgh 63656 owner's manual

Pittsburgh

Pittsburgh 63656 owner's manual

Rlide R2 Series manual

Rlide

Rlide R2 Series manual

AdValue Photonics AP-ML1 user guide

AdValue Photonics

AdValue Photonics AP-ML1 user guide

Dinel CLM-36 instruction manual

Dinel

Dinel CLM-36 instruction manual

Hach POCKET COLORIMETER II instruction manual

Hach

Hach POCKET COLORIMETER II instruction manual

jenway PFP7 instruction manual

jenway

jenway PFP7 instruction manual

Nokeval 2012 user manual

Nokeval

Nokeval 2012 user manual

Hanna Instruments HI 8915 instruction manual

Hanna Instruments

Hanna Instruments HI 8915 instruction manual

Cameron 351 installation manual

Cameron

Cameron 351 installation manual

Eldat RTS16 quick start guide

Eldat

Eldat RTS16 quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.