USER MANUAL
AD13
WaterproofORPTester
DearCustomer,
Thank you for choosing an Adwa product.
Please read carefully this manual before
starting operations.
This instrument is in compliance with the
EMCDirective89/336/EEC andLowVoltage
Directive73/23/EECforelectricalequipments.
For additional technical information, please
e-mailusat [email protected].
INTRODUCTION
AD13 is a waterproof ORP and temperature
tester. The housing has been completely
sealed against humidity.
Temperature values can be displayed in
°C or °F units.
The meter is factory calibrated both for
ORP (mV) and temperature ranges.
Measurements are highly accurate with
a unique stability indicator right on the
LCD. The model is also provided with a
low battery symbol which warns the user
when the batteries need to be replaced.
TheAD13PORPelectrode,suppliedwith
the meter, is interchangeable and can be
easily replaced by the user.
The encapsulated temperature sensor
allows fast and accurate temperature
measurement.
The meter is supplied complete with:
•AD13P ORP electrode
• 4 x 1.5V batteries, button type
• User manual
FRONT PANEL & DISPLAY
1. Dual line LCD
2. ON/OFF/ MODE button
3. ORP electrode & temperature sensor
4. Electrode body
5. Battery compartment (inside)
6. SET/HOLD button
7. Clip holder
1. Stability indicator (hourglass symbol)
2. Low battery warning indicator
3. Secondary LCD level
4. Primary LCD level
5. Measurement unit for primary level
TECHNICAL DATA
Range ±1000 mV
-5.0 to 60.0°C / 23.0 to 140.0°F
Resolution 1 mV
0.1°C / 0.1°F
Accuracy (@20°C/68°F)
±2 mV / ±0.5°C / ±1°F
Calibration Factory calibrated
Electrode
AD13P ORP electrode (included)
Battery Type 4 x 1.5V button type
Battery Life Approx. 300 h of use
Auto-off After 8 minutes of non-use
Environment
-5 to 50°C (23 to122°F); RH 100%
Dimensions 175.5 x 39 x 23 mm
Weight 100 g
ISTAD13R2 07/09
Turn the meter on
• Press and hold the ON/OFF/MODE
button until the LCD lights up.All the
used segments will be visible for one
second (or as long as the button is
pressed).
Freeze the display
• Whileinmeasurementmode,pressthe
SET/HOLD button. The reading will
be frozen on the LCD. Press any button
to return to normal mode.
Turn the meter off
• Whileinmeasurementmode,pressthe
ON/OFF/MODE button. OFF will
appear on the secondary display.
Release the button.
Note:If measurements are taken in
different samples successively,
rinse the probe thoroughly to
eliminate cross-contamination.
After cleaning, rinse the probe
with some of the sample to be
measured.
OPERATIONAL GUIDE MEASUREMENT&CALIBRATION
• Submergethe electrode in the solution
to be tested while stirring it gently.
• Measurements should be taken when
the stability indicator (hourglass)
disappears.
• The ORP (mV) value is shown on the
primaryLCDlevelwhilethesecondary
one shows the sample temperature.
Note: The ORP (mV) range is factory
calibrated.
SETUP
Setup mode allows the selection of
temperature unit.
To enter the setup mode, press the ON/
OFF/MODE button until CAL on the
secondary display is replaced by TEMP
and the current temperature unit (e.g.
TEMP °C).
Then use the SET/HOLD button to
select the desired option.
After the temperature unit has been
selected, press the ON/OFF/MODE
button to return to the normal measuring
mode.
ELECTRODE MAINTENANCE
When not in use, rinse the electrode
with water and store it with a few drops
of A7071S electrolyte solution in the
protective cap.
NEVER STORE THE ELECTRODE
IN DISTILLED OR DEIONIZED
WATER!
• If the electrode has been left dry, soak
thetipinelectrolytesolutionforat least
one hour to reactivate it.
• To prolong the electrode life, it is
recommended to clean it monthly by
immersing the tip in deionized water
for 30 minutes. Afterwards, rinse it
thoroughly with tap water and
recalibrate the meter.
• The electrode can be easily replaced
by unscrewing the body as shown
below.
When the batteries become weak, the
battery symbol on the LCD lights up to
indicatealowbatterycondition.Batteries
should be replaced soon.
To change the batteries, unscrew and
release the electrode body. Take out the
battery compartment and carefully
replace all four batteries while paying
attention to their polarity.
Reattach and tighten the electrode body
properly to ensure a watertight seal.
BATTERY REPLACEMENT

Other ADWA Test Equipment manuals

ADWA AD14 User manual

ADWA

ADWA AD14 User manual

ADWA AD11 User manual

ADWA

ADWA AD11 User manual

ADWA AD14 User manual

ADWA

ADWA AD14 User manual

ADWA AD12 User manual

ADWA

ADWA AD12 User manual

Popular Test Equipment manuals by other brands

Agilent Technologies U1604A quick start guide

Agilent Technologies

Agilent Technologies U1604A quick start guide

HP HP8648A instruction manual

HP

HP HP8648A instruction manual

WIKA CEP1000 operating instructions

WIKA

WIKA CEP1000 operating instructions

SmartSafe iSmartTool Series user manual

SmartSafe

SmartSafe iSmartTool Series user manual

National Instruments PXIe-5170 Getting started guide

National Instruments

National Instruments PXIe-5170 Getting started guide

Extech Instruments CT70 user manual

Extech Instruments

Extech Instruments CT70 user manual

Snap-On SVTS2000 manual

Snap-On

Snap-On SVTS2000 manual

barfield 2311FA User instruction manual

barfield

barfield 2311FA User instruction manual

Quick Charge C1600 plus operating instructions

Quick Charge

Quick Charge C1600 plus operating instructions

Atlona AT-DIS7-PROHD Q&A

Atlona

Atlona AT-DIS7-PROHD Q&A

PMK PHV Series instruction manual

PMK

PMK PHV Series instruction manual

Anritsu MT1000A Network Master Pro Operation manual

Anritsu

Anritsu MT1000A Network Master Pro Operation manual

ROD-L M25 quick start guide

ROD-L

ROD-L M25 quick start guide

Dräger Alcotest 3820 Instructions for use

Dräger

Dräger Alcotest 3820 Instructions for use

Tektronix 2430A user manual

Tektronix

Tektronix 2430A user manual

SecoRut Wireless ControlCase user manual

SecoRut

SecoRut Wireless ControlCase user manual

HYDAC FILTER SYSTEMS CTU 1 3 Series Operating and maintenance instructions

HYDAC FILTER SYSTEMS

HYDAC FILTER SYSTEMS CTU 1 3 Series Operating and maintenance instructions

Viking PB-100 Application note

Viking

Viking PB-100 Application note

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.