Agilent Technologies 1290 infinity User manual

Agilent Technologies
Agilent 1290 Infinity
Thermostat
User Manual

1290 Infinity Thermostat User Manual
Notices
© Agilent Technologies, Inc. 2011-2014,
2015
No part of this manual may be reproduced
in any form or by any means (including elec-
tronic storage and retrieval or translation
into a foreign language) without prior agree-
ment and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
G1330-90014 Rev. B
Edition
05/2015
Printed in Germany
Agilent Technologies
Hewlett-Packard-Strasse 8
76337 Waldbronn
Warranty
The material contained in this docu-
ment is provided “as is,” and is sub-
ject to being changed, without notice,
in future editions. Further, to the max-
imum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a par-
ticular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or perfor-
mance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the sep-
arate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accor-
dance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcon-
tract, Software is delivered and licensed as
“Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer soft-
ware” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation
or contract clause. Use, duplication or dis-
closure of Software is subject to Agilent
Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
receive no greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly per-
formed or adhered to, could
result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION
notice until the indicated condi-
tions are fully understood and
met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice,
or the like that, if not correctly
performed or adhered to, could
result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indi-
cated conditions are fully under-
stood and met.

1290 Infinity Thermostat User Manual 3
Contents
Contents
1 Introduction 5
Introduction to the Thermostat 6
Thermostat Operation 8
2 Site Requirements and Specifications 11
Site Requirements 12
Physical Specifications 15
Performance Specifications 16
3 Installing the Thermostat 17
Damaged Packaging 18
Optimizing the Stack Configuration 19
Installing the Thermostat 21
Transporting the Thermostatted Autosampler or Fraction Collector 32
4 Troubleshooting and Diagnostics 33
Agilent Lab Advisor Software 34
Overview of the Thermostat’s Indicators and Test Functions 35
Status Indicators 36
5 Error Information 39
What Are Error Messages 40
General Error Messages 41
Thermostat Error Messages 48
6 Maintenance 53
Warnings and Cautions 54
Introduction to Maintenance 56
Cleaning the Module 57
Exchanging the Power Supply Fuses 58

4 1290 Infinity Thermostat User Manual
Contents
7 Parts for Maintenance 61
Main Assemblies (External Parts) 62
Accessory Kit 63
HPLC System Tool Kit 64
Plastic Parts 65
8 Cable Identification 67
Cable Overview 68
Analog Cables 70
Remote Cables 72
BCD Cables 75
Auxiliary Cable 77
CAN/LAN Cables 78
External Contact Cable 79
RS-232 Cable Kit 80
9 Hardware Information 81
Electrical Connections 82
Early Maintenance Feedback 85
10 Appendix 87
General Safety Information 88
The Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC) 91
Radio Interference 92
Sound Emission 93
Agilent Technologies on Internet 94

5
1290 Infinity Thermostat User Manual
Agilent Technologies
1
Introduction
Introduction to the Thermostat 6
Thermostat Operation 8

6 1290 Infinity Thermostat User Manual
1Introduction
Introduction to the Thermostat
Introduction to the Thermostat
The Agilent 1290 Infinity Thermostat is designed for use with other
modules of the Agilent 1200 Infinity Series or with other LC systems if
adequate remote control inputs and outputs are available. The thermostat
is controlled from the Agilent 1200 Infinity Series Instant Pilot or from
your Agilent control software for LC systems.
The specially- designed thermostattable sample trays holds either
100 × 1.8 mL vials or two wellplates and 10 × 1.8 mL vials.
The thermostat contains Peltier- controlled heat exchangers. A fan draws
air from the area above the sample vial tray of the autosampler and then
blows it through the fins of the cooling/heating module. There it is cooled
or heated according to the temperature setting. The thermostatted air
enters the autosampler through a recess underneath the specially- designed
sample tray. The air is then distributed evenly through the sample tray
ensuring effective temperature control, regardless of how many vials are in
the tray.
In cooling mode condensation is generated on the cooled side of the
Peltier elements. This condensed water is safely guided into the leak
system.

1290 Infinity Thermostat User Manual 7
Introduction 1
Introduction to the Thermostat
Figure 1 Overview of the Thermostat
=ZVih^c`[Vch
8dcYZchVi^dcaZV`YgV^c
=ZViZmX]Vc\Zg
=ZVih^c`
EdlZghjeean
I86WdVgY
=ZViZmX]Vc\Zg[Vc

8 1290 Infinity Thermostat User Manual
1Introduction
Thermostat Operation
Thermostat Operation
Figure 2 Thermostat Principle
EZai^ZgZaZbZcih
VjidhVbeaZghVbea^c\jc^i
HeZX^VahVbeaZigVn
6^gdjiaZi
=ZVi^c\$Xdda^c\bdYjaZ
6bW^ZciV^g^c
I]ZgbdhiViiZYV^g
=ZViZmX]Vc\Zg
6^gX]VccZa^c
=ZViZYV^gdji

1290 Infinity Thermostat User Manual 9
Introduction 1
Thermostat Operation
The thermostat is equipped with a cooling/heating module which uses
Peltier elements for efficient air cooling. When turned on the front side of
the Peltier elements is heated/cooled according to the temperature setting.
A fan draws air from the sample tray area and blows it through the
channels of the heating/cooling module. The fan speed is determined
according to the environmental conditions (e.g., ambient temperature,
humidity). In the heating/cooling module the air reaches the temperature
of the Peltier elements and this thermostatted air is blown underneath the
special sample tray where it is evenly distributed and streams back into
the sample tray area. From there it is again drawn into the thermostat.
This “recycle” mode assures a very efficient cooling/heating of the sample
vials.
In cooling mode the opposite side of the Peltier element will become very
hot and to maintain the performance of the elements they have to be
cooled down. This is done with large heat exchangers in the back of the
thermostat. Four fans blow air from left to right through the instrument to
remove the heated air. The fan speed is controlled according to the
temperature of the Peltier elements.
During cooling condensation will appear in the heating/cooling module.
The condensed water will be guided out of the thermostat.

10 1290 Infinity Thermostat User Manual
1Introduction
Thermostat Operation

11
1290 Infinity Thermostat User Manual
Agilent Technologies
2
Site Requirements and Specifications
Site Requirements 12
Power Consideration 12
Power Cords 13
Bench Space 14
Environment 14
Physical Specifications 15
Performance Specifications 16
This chapter provides information on environmental requirements, physical and
performance specifications.

12 1290 Infinity Thermostat User Manual
2Site Requirements and Specifications
Site Requirements
Site Requirements
A suitable site environment is important to ensure optimum performance
of the instrument.
Power Consideration
The autosamplers comprises two modules, the autosampler module
(G1329A/B, G1367A- D, G1367E, G1377A, G2260A, G4226A, G5667A,
G4303A) or fraction collector (G1364A- C, G1364D, G5664A) and the
thermostat module (G1330B). Both modules have a separate power supply
and a power plug for the line connections. The two modules are connected
by a control cable and both are turned on by the autosampler or fraction
collector.
The autosampler or fraction collector power supplies have automatic
voltage selectors. Consequently there are no voltage selectors in the rear
of the two autosampler or fraction collector modules. The autosampler or
fraction collector has no externally accessible fuses, because automatic
electronic fuses are implemented in its power supply. The thermostat
power supply has two externally accessible fuses.
WARNING
Hazard of electrical shock or damage of your instrumentation
can result, if the devices are connected to a line voltage higher than specified.
➔Connect your instrument to the specified line voltage only.
CAUTION
Inaccessible power plug.
In case of emergency it must be possible to disconnect the instrument from the power
line at any time.
➔Make sure the power connector of the instrument can be easily reached and
unplugged.
➔Provide sufficient space behind the power socket of the instrument to unplug the
cable.

1290 Infinity Thermostat User Manual 13
Site Requirements and Specifications 2
Site Requirements
Power Cords
Different power cords are offered as options with the module. The female
end of all power cords is identical. It plugs into the power- input socket at
the rear. The male end of each power cord is different and designed to
match the wall socket of a particular country or region.
WARNING
Absence of ground connection or use of unspecified power cord
The absence of ground connection or the use of unspecified power cord can lead to
electric shock or short circuit.
➔Never operate your instrumentation from a power outlet that has no ground
connection.
➔Never use a power cord other than the Agilent Technologies power cord designed
for your region.
WARNING
Use of unsupplied cables
Using cables not supplied by Agilent Technologies can lead to damage of the
electronic components or personal injury.
➔Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
WARNING
Unintended use of supplied power cords
Using power cords for unintended purposes can lead to personal injury or damage of
electronic equipment.
➔Never use the power cords that Agilent Technologies supplies with this instrument
for any other equipment.

14 1290 Infinity Thermostat User Manual
2Site Requirements and Specifications
Site Requirements
Bench Space
The module dimensions and weight (see Table 1 on page 15) allow you to
place the module on almost any desk or laboratory bench. It needs an
additional 25 cm (10 inches) of space on either side and approximately
8 cm (3.1 inches) in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the
bench is designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Environment
Your modules will work at ambient temperatures and relative humidity as
described in Table 1 on page 15.
CAUTION
Condensation within the module
Condensation will damage the system electronics.
➔Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
➔If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.

1290 Infinity Thermostat User Manual 15
Site Requirements and Specifications 2
Physical Specifications
Physical Specifications
Table 1 Physical Specifications
Type Specification Comments
Weight 20.7 kg (46 lbs)
Dimensions (height ×
width × depth)
140 × 345 × 435 mm (5.5 × 13.5 × 17
inches)
Line voltage 100 – 240 VAC, ± 10 % Wide-ranging
capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 260 VA / 210 W / 717 BTU Maximum
Ambient operating
temperature
4 – 40 °C (39 – 104 °F) See warning “” on
page 15
Ambient non-operating
temperature
-40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 2000 m (6562 ft)
Non-operating altitude Up to 4600 m (15091 ft) For storing the module
Safety standards: IEC, CSA,
UL
Installation category II, Pollution degree 2 For indoor use only.
WARNING
Hot rear panel
Using the module at high environmental temperatures may cause the rear panel to
become hot.
➔Do not use the module at environmental temperatures higher than 50 °C (122 °F)

16 1290 Infinity Thermostat User Manual
2Site Requirements and Specifications
Performance Specifications
Performance Specifications
*Measurement conditions:
Table 2 Performance specifications Agilent 1290 Infinity Thermostat
Type Specification
Temperature range Settable from 4 °C to 40 °C in 1 ° increments
Temperature accuracy at ambient
temperatures < 25 °C and humidity < 50 %
3°Cto8°Catasetpointof4°C
*
Temperature accuracy at ambient
temperatures <30 °C and humidity <60 %
3°Cto9°Catasetpointof4°C
*
G1329B:
with 100-Vial Tray in vial location 2,10,92 and 100 vials filled with water
G1367A/G1367B/G1367C/G1377A/G1367E/G5667A/G2258A/G4226A:
For vials: Using the Thermostattable Tray (G1329-60011) or 100 Micro-Vial
(G4226-60021), both loaded with 100 vials. Temperature is measured in vial locations
1,10,23,25,45,75,91 and 100 (filled with 1 mL of water)
For well plates: Standard Tray (G2258-60011) for two Well Plates loaded with two
Agilent 96 Well Plate (5042-1386)

17
1290 Infinity Thermostat User Manual
Agilent Technologies
3
Installing the Thermostat
Damaged Packaging 18
Delivery Checklist 18
Optimizing the Stack Configuration 19
Installing the Thermostat 21
Stage 1: Preparing the Thermostat and Autosampler/ Fraction
Collector 22
Stage 2: Power Cable and Interface Cable Connection 25
Stage 3: Flow Connections 27
Stage 4: Installing the Sample Tray 29
Stage 5: Installing Tray Cover and Front Cover 31
Stage 6: Turning on the Thermostatted Autosampler/ Fraction
collector 31
Transporting the Thermostatted Autosampler or Fraction Collector 32

18 1290 Infinity Thermostat User Manual
3Installing the Thermostat
Damaged Packaging
Damaged Packaging
If the delivery packaging shows signs of external damage, please call your
Agilent Technologies sales and service office immediately. Inform your
service representative that the instrument may have been damaged during
shipment.
Delivery Checklist
Unpack the module. Ensure all parts and materials have been delivered.
The delivery checklist is shown in Table 3 on page 18. Please report
missing or damaged parts to your local Agilent Technologies sales and
service office.
CAUTION
"Defective on arrival" problems
If there are signs of damage, please do not attempt to install the module. Inspection by
Agilent is required to evaluate if the instrument is in good condition or damaged.
➔Notify your Agilent sales and service office about the damage.
➔An Agilent service representative will inspect the instrument at your site and
initiate appropriate actions.
Table 3 Thermostat Checklist
Description Quantity Part Number
Thermostat 1
Power cable 1 as ordered
Accessory kit (see “Accessory Kit” on
page 63 for more information)
1 G1330-68755
HPLC System Tool Kit, optional (see
“HPLC System Tool Kit” on page 64 for
more information)
1 G4203-68708

1290 Infinity Thermostat User Manual 19
Installing the Thermostat 3
Optimizing the Stack Configuration
Optimizing the Stack Configuration
If your autosampler/fraction collector is part of a system, you can ensure
optimum performance by installing the autosampler/fraction collector in
the stack in the position shown in Figure 3 on page 19 and Figure 4 on
page 20. This configuration optimizes the system flow path, ensuring
minimum delay volume. As the autosampler thermostat is very heavy, it
should always be installed at the bottom of a stack.
Figure 3 Recommended Stack Configuration (Front View)
>chiVciE^adi
9ZiZXidg
8dajbcXdbeVgibZci
HdakZciXVW^cZi
9Z\VhhZgdei^dcVa
Ejbe
I]ZgbdhiVi[dgi]Z6AH$;gVXi^dcXdaaZXidgdei^dcVa
6jidhVbeaZg6AH$;gVXi^dcXdaaZXidg

20 1290 Infinity Thermostat User Manual
3Installing the Thermostat
Optimizing the Stack Configuration
Figure 4 Recommended Stack Configuration (Rear View)
LAN to control software
CAN Bus cable
(to Instant Pilot)
CAN Bus cable
Remote cable
AC Power
Autosampler/ Fraction
Collector Cable
Other manuals for 1290 infinity
12
Table of contents
Other Agilent Technologies Thermostat manuals

Agilent Technologies
Agilent Technologies 1260 Infinity II User manual

Agilent Technologies
Agilent Technologies 1260 Infinity II User manual

Agilent Technologies
Agilent Technologies 1290 infinity User manual

Agilent Technologies
Agilent Technologies InfinityLab LC Series User manual

Agilent Technologies
Agilent Technologies G1330B User manual

Agilent Technologies
Agilent Technologies 1100 Series User manual