5 6
1.7 If Timed Test is selected
Press start/pause the display shows ‘Enter Run Time’
Press arrows to adjust time:-
Press start/pause to move between (Hrs) 00 (Mins) 00 (Sec) 00.
Press start/pause the pump runs and the timer starts. When the time has elapsed
the display shows ‘Test Completed - Check Flow’.
Press stop display shows ‘Test completed - Check Flow’.
Press stop display shows ‘Connect ow meter’ (now reconnect your ow meter to
check the ow rate).
Press start/pause display shows ‘Flow Check’ (any increment changes made when
setting the ow rate (see 1.4) will be shown.
NOTE: Wait 45 seconds for ow rate to stabilise.
Press stop display shows ‘Flow Check Complete’
Press stop to shutdown.
NOTE: If the pump switches itself o during either Free Running or Timed Test (i.e if the battery dies or
the sampling head gets blocked) on restart it will either show:-
1. ‘Test Incomplete, Restart Test?’, in which case just recommence test.
2. ‘Test Completed Check Flow’, in which case continue as instructions above.
3. ‘Low battery’, in which case either replace battery (within 10 minutes of disconnecting old one) or
charge pump directly from the mains for a short time to complete the ow check.
1.6 If Free Running is selected
Press start/pause the pump runs and the timer starts counting, (you can pause
during the test by pressing start/pause)
Press stop when the test is complete and the display shows ‘Test Stop? Are you
sure?’ If you are not sure, press start/pause to continue test, if you are sure:-
Press stop display shows ‘Test completed - Check Flow’.
Press stop display shows ‘Connect ow meter’ (now reconnect your ow meter to
check the ow rate).
Press start/pause, display shows ‘Flow Check’ (any increment changes made when
setting the ow rate (see 1.4) will be shown).
NOTE: Wait 45 seconds for ow rate to stabilise.
Press stop display shows ‘Flow Check Complete’.
Press stop to shutdown.
1.4 Switching on and setting the ow rate
NOTE: The two models are slightly dierent, please make sure you read the
correct model instructions.
VARIFLOW
Before switching on the VARIFLOW connect a ow meter to the air sampling head.
Press start/pause to switch on the unit. The display will show, in this order, the serial number,
the current rmware version and the battery status.
Press start/pause the display shows ‘Flow not set’.
Press either arrow to open Calibration Mode and to Set Flow.
Press arrows to select Low or High. (Low is approx 8 l/min, High is approx 16 l/min).
Press start/pause to start the pump. NOTE: Run for 45 seconds to allow to stabilise.
Press up/down arrows to adjust ow rate. (A single press advances the display by one
increment, holding while pressing advances the display by 10 increments).
Check the ow rate against the ow meter and adjust acording to ow requirements. (Any
increment changes will be saved for the post test ow check, see 1.6/1.7).
To conrm and set the ow rate press stop.
IMPORTANT: If pump is shutdown between setting the ow and starting the test,
ow will have to be reset by repeating the steps above.
VARIFLOW 8
Before switching on the VARIFLOW 8connect a ow meter to the air sampling head.
Press start/pause to switch on the unit. The display will show, in this order, the serial
number, the current rmware version and the battery status.
Press start/pause the display shows ‘Flow not set’.
Press either arrow to open and start Calibration Mode and to Set Flow.
The pump is set at aprox 8 l/min as a default, to ne tune this:-
Press start/pause to start the pump. NOTE: Run for 45 seconds to allow to stabilise.
Press up/down arrows to adjust ow rate. (A single press advances the display by
one increment, holding while pressing advances the display by 10 increments).
Check the ow rate against the ow meter and adjust acording to ow requirements.
(Any increment changes will be saved for the post test ow check, see 1.6/1.7).
To conrm and set the ow rate press stop.
IMPORTANT: If pump is shutdown between setting the ow and starting the test,
ow will have to be reset by repeating the steps above.
1.5 Setting test mode
Press start/pause the display will show ‘Set test mode’.
Press arrows to select ‘Free Running’ or ‘Timed test’.