Airel Quetin Pacific Instruction manual

Manual for Use

2
IMP-PI500-EN_ind J
12/12/2019

3
1. INTRODUCTION 7
1.1 General Information ...............................................................................7
1.1.1 Purpose of the manual ...................................................................... 7
1.1.2 À About PACIFIC equipment............................................................... 7
1.2 Classification/Regulations ......................................................................7
1.3 Safety instructions ................................................................................8
1.4 Guarantee.............................................................................................8
1.5 IGN®-Calbénium® protocol......................................................................9
1.6 Logotypes and icons...............................................................................9
2. DESCRIPTION 10
2.1 General description ..............................................................................10
2.1.1 PACIFIC Ambidextrous - NEO or EVO ................................................ 10
2.1.2 PACIFIC Fixed Unit - NEO or EVO...................................................... 11
2.1.3 PACIFIC CHIR / ORTHO ................................................................... 12
2.2 Detailed description .............................................................................13
2.2.1 PACIFIC unit and chair ................................................................... 13
2.2.2 NEO and EVO instrument tray ......................................................... 13
Instruments on whip arms 14
Calbépuls®14
2.2.3 Assistant station............................................................................ 14
2.2.4 Keyboard or control screen.............................................................. 15
Assistant station NEO / EVO touch keypad 15
CHIR / ORTHO touch keypad 15
NEO touchscreen - Smartcard 15
EVO.2 touch screen - USB stick 16
2.2.5 Foot pedal...................................................................................... 16
2.3 IGN®-CALBENIUM® system..................................................................17
2.4 Optional equipment ..............................................................................17
2.4.1 Basin............................................................................................. 17
2.4.2 Surgical lighting.............................................................................. 18
2.4.3 Amalgam separator ........................................................................ 18
2.4.4 Video surveillance camera................................................................ 18
3. USE 18
3.1 Power on/off.......................................................................................18
SUMMARY

4
3.2 Adjusting the headrest .........................................................................19
3.2.1 Mechanical locking.......................................................................... 19
3.2.2 Pneumatic assistance...................................................................... 19
3.3 Ambidextrous positioning......................................................................19
3.4 Foot pedal...........................................................................................20
3.4.1 Pneumatic or electric progressive pedal ............................................ 20
3.4.2 Wireless progressive pedal .............................................................. 21
3.5 NEO/EVO assistant keyboard controls ..................................................22
3.6 CHIR/ORTHO touch keypad controls .....................................................23
3.7 Additional controls under CHIR/ORTHO satellite ....................................25
3.8 "NEO" monochrome touchscreen controls ..............................................26
3.8.1 Inserting the user card.................................................................... 26
3.8.2 Creating a user card ....................................................................... 26
3.8.3 Using icons on NEO screen .............................................................. 27
3.8.4 Functions on the home screen.......................................................... 27
3.8.5 Settings menu ................................................................................ 29
Navigation in the settings menu 29
Clock and timer 29
Keyboard beep 29
Surgical lighting 30
Chair and basin positions 30
Glass-filling and bowl rinsing spray 31
Calbépuls 31
3.8.6 Turbine menu.................................................................................. 32
Selecting the turbine variation mode 33
Choice with or without alternospray 33
3.8.7 Motor menu ................................................................................... 33
Motor direction reversal selection using the pedal 34
3.8.8 Plaque remover menu...................................................................... 35
Plaque remover variation selection 35
3.8.9 MX micromotor menu...................................................................... 35
Motor torque as a % 37
Auto-forward time setting 37
3.9 “Evo.2” on-screen controls....................................................................38
3.9.1 User Authentication ........................................................................ 38
3.9.2 User management........................................................................... 39
Creating a user profile 39
Exporting a user profile 40
Importing a user profile 41

5
Changing the Password/Changing the language 42
Deleting a user 42
3.9.3 Functions on the home screen.......................................................... 43
3.9.4 Settings menu ................................................................................ 45
Navigation in the settings menu 45
Chair and basin positions 46
Clock and timer 46
Surgical lighting 47
Glass-filling and bowl rinsing spray 47
Calbépuls 48
Intercom picture 48
Electronic chart diagnostics 48
3.9.5 Work with user programmes............................................................ 49
Turbine menu 49
Selecting the turbine variation mode 50
Choice with or without alternospray 51
Selecting spray on chip 51
3.9.6 Plaque remover menu ..................................................................... 52
Choice with or without alternospray 53
Scaler variation selection 54
3.9.7 MX micromotor menu...................................................................... 54
Speed multiplier coefficient programming 56
Motor direction reversal selection using your foot 57
Auto-forward time setting 57
4. HYGIENE AND MAINTENANCE 58
4.1 Introduction ........................................................................................58
4.2 CALBENIUM® refill ..............................................................................58
4.2.1 Storage ......................................................................................... 58
4.2.2 Precautions for use......................................................................... 58
4.2.3 Refill on AMBIDEXTROUS PACIFIC NEO/EVO .................................... 59
4.2.4 Refill on FIXED PACIFIC UNIT .......................................................... 59
4.2.5 Refill on PACIFIC CHIR / ORTHO....................................................... 60
4.3 Maintenance guide ...............................................................................61
4.4 Maintenance procedures ......................................................................62
4.4.1 Purge the instrument cords............................................................. 62
4.4.2 Cleaning the bowl of the basin.......................................................... 62
4.4.3 Using Calbépuls .............................................................................. 62
4.4.4 Using Aspisept ............................................................................... 63

6
4.4.5 Cleaning the suction filters .............................................................. 63
4.4.6 Cleaning the amalgam separator filter .............................................. 64
4.4.7 Replacement of the amalgam separator cassette............................... 65
4.4.8 Disinfection of the suction cords ...................................................... 66
Daily disinfection - 1 L 66
Decontamination of instrument cords (beginning/end of the day) 67
Weekly disinfection - 2 L 68
4.4.9 Maintenance of external surfaces..................................................... 69
4.4.10 Fitting the cords onto the whip arms ................................................ 70
4.4.11 Intensive disinfection of the instrument cords.................................... 71
On PACIFIC NEO 71
On PACIFIC EVO 71
4.4.12 Cleaning the instrument holders and handles..................................... 72
4.4.13 Replacement of the filter cartridge................................................... 72
4.4.14 Checking air and water pressures - Water filter................................. 73
4.4.15 Cleaning and lubrication of micromotors, turbines, handheld parts ...... 74
5. PREVENTIVE MAINTENANCE 74
6. SIMPLE FAULTS & TROUBLESHOOTING 75
6.1 Chair Safety - Emergency ascent ..........................................................75
6.2 FAQ - Symptoms and Remedies .............................................................76
6.3 Foaming of sprays................................................................................77
6.4 Replacing a fuse ..................................................................................78
7. TECHNICAL CHARACTERISTICS 79
7.1 Dimensions of PACIFIC ambidextrous chair .............................................79
7.2 Dimensions of PACIFIC fixed column chair...............................................80
7.3 Dimensions of PACIFIC CHIR chair .........................................................81
7.4 Dimensions of PACIFIC ORTHO chair ......................................................82
7.5 Technical characteristics ......................................................................83
7.6 Wireless pedal.....................................................................................83
8. LIST OF COMPATIBLE INSTRUMENTS 84
9. WIRING DIAGRAMS 84
10. RECYCLING 84

7
1. INTRODUCTION
1.1 General Information
1.1.1 Purpose of the manual
This manual describes the use of AIREL-QUETIN PACIFIC dentistry equipment in its different
versions: NEO, EVO, CHIR, ORTHO.
• Read this manual carefully before using the PACIFIC equipment. It has been written
to allow you to use your equipment in the best conditions.
• The numerical values given in this manual and displayed as an example should not
be considered as recommended values, except when specified in the text.
• The figures are intended to illustrate the different functions of the PACIFIC. They are
not binding. They do not claim to represent all the equipment variants.
1.1.2 À About PACIFIC equipment
The PACIFIC equipment is a dental chair allowing the operator to detect and treat dental
problems and/or carry out preventive actions.
It must be used by dental professionals. The operator (practitioner) must have a degree in dental
surgery.
Patient population: from 4 years old.
1.2 Classification/Regulations
The AIREL-QUETIN PACIFIC dental equipment complies with the following standards:
EEC Directive 93/42 Class IIa medical device
(2195 - EC marking)
Standard IEC 60950-1 / 60417-
5019 Class I Electrical Protection. Functional insulation
with an earth terminal.
Standard IEC 60601-1 Type B
Standard IEC 60529 Protection against water spray IPX1
Standard EN 60601-1-2 The product does not cause electromagnetic
interference which may affect other devices. It
demonstrates appropriate immunity in the presence
of electromagnetic interference.
Standard IEC 80601-2-60 Special requirements for basic safety and essential
performance of dental equipment

8
1.3 Safety instructions
To avoid all risks of electrical shock, it is imperative that the device is connected to an electrical
supply network equipped with an earth.
Any electrical work on the device must be carried out by our after-sales department or by
qualified personnel.
AIREL-QUETIN provides this staff with any technical information needed for troubleshooting.
IMPORTANT RECOMMENDATIONS
- Do not use in the presence of an inflammable anaesthetic mixture.
- The connection of the fluids must comply with the regulations in force.
ISO 7494-2 Standard for dentistry - Dental units - Part 2: Water and air supply.
- Intermittent service.
- Turn off the PE8 unit every night after use.
- Applied parts: Saddlery, casing (protective cover)
- This device is intended for use by professionals only.
- Any modification of the EM device is prohibited.
- According to the requirements of standard NF EN 60601-1, the device may cause radio
interference and / or disturb the operation of a device located nearby. The use of accessories,
transducers and cables other than those specified, with the exception of equipment sold by the
manufacturer as a replacement for internal components, may result in increased emissions or
decreased immunity. of the device.
- The device must not be used alongside other devices or stacked with them.
If this condition can not be met, test the correct operation in this configuration.
- Portable and mobile RF communications devices must not be used within 30cm of the device.
Otherwise, degradation of the performance of the equipment could be felt.
WARNING ON NOISEY FROM SUCTION TUBE
The existing magnet inside of the suction flexible may interfere with pacemakers and ICD.
Therefore the practitioner shall ensure that the suction tube remains at opposite side of the
area of the device to avoid any risk during curing.
Down below are all the precautions of use :
• Ensure that the magnetic bit of the suction tube remains away from the pacemaker
• In case of audible signal coming from the pacemaker, please move promptly the
magnetic bit of the suction tube until the audible signal stops. The potential risk is so avoided.
1.4 Guarantee
All our devices are manufactured with the utmost care and are tested according to the ISO
7494-1 standard before being delivered to our customers.
Any new PACIFIC appliance is guaranteed against any defective parts for 2 years from the day of
delivery.
The relative information and specific to each device are listed on the serial label next to the
ignition button.
The guarantee includes the exchange or repair of the defective part, according to the
manufacturer's instructions.
The instruments (turbine, micromotor, plaque remover, scalpel, curing light, etc...) are subject
to special guarantee conditions (guarantee period and conditions of use) which are those of the
respective manufacturers.
Operation is only guaranteed for instruments provided by AIREL-QUETIN or for compatible
instruments.

9
The guarantee does not extend to light bulbs, glassware, coals, heat cartridges, capacitors,
fuses, customer-supplied parts, or any part subject to normal wear during the guarantee period.
The guarantee is withdrawn and AIREL-QUETIN is released from any responsibility in the
following cases:
• When modifications have been made to the device, outside the specifications provided by the
manufacturer, or when original parts have been replaced by parts from other sources.
• When the damage is due to faulty maintenance, misuse, fault of the operator or agents not
authorised by AIREL-QUETIN.
- Under no circumstances can the guarantee defined above cover any damage caused to
people and property due to the operation of the device.
- In case of dispute, by the express agreement of the parties, jurisdiction is given to the
Commercial Court of Créteil.
1.5 IGN®-Calbénium® protocol
The PACIFIC dental unit incorporates an IGN® hygiene system for spray water treatment
according to EEC directive 93/42 - Class IIb medical equipment.
An IGN®protocol ensuring the traceability and the conformity of operation according to standard
NF EN 1717 has been implemented in order to reach the "Zero permanent germ" objective.
The operations of this protocol are detailed in chap 4, page 58 with a maintenance guide.
IGN® and CALBENIUM® are a fundamental pairing for the hygiene of your PACIFIC unit:
- No product other than CALBENIUM® should be used in the IGN® system. A study has been
carried out on CALBENIUM® to ensure that it is compatible with all dental units and does not
cause any damage to internal components.
- The use of inappropriate products may result in oxidation of the components and
devulcanization of the conducting membranes. If such defects are noted, AIREL-QUETIN
reserves the right to remove the registration of use related to the IGN® protocol.
1.6 Logotypes and icons
Instruction manual
Dangerous voltage
Protective earthing
Type B
Warning
Manufacturing date
Manufacturer

10
2. DESCRIPTION
2.1 General description
The views below are generic, they do not represent all the equipment variants.
2.1.1 PACIFIC Ambidextrous - NEO or EVO
PACIFIC AMBI NEO / EVO
1
3
5
4
2
7
6
1Instrument tray (NEO or EVO screen)
2Unit and unit arm (column)
3PACIFIC patient chair
4Basin
5Assistant station
6Foot pedal
7PACIFIC chair foot

11
2.1.2 PACIFIC Fixed Unit - NEO or EVO
PACIFIC FIXED UNIT NEO / EVO
1
3
5
4
2
7
6
1Instrument tray (NEO or EVO screen)
2Unit arm
3Fixed unit (*)
4Assistant station
5PACIFIC chair foot
6PACIFIC patient chair
7Foot pedal
(*) This part consists of a column fixed to the ground.

12
2.1.3 PACIFIC CHIR / ORTHO
PACIFIC CHIR
1
2
3
4
5
6
1Instrument Satellite (NEO or EVO screen, or polycarbonate touch keypad)
2PACIFIC patient chair
3Assistant station
4Equipment arm
5PACIFIC chair foot
6Cervical tablet
PACIFIC ORTHO
1
2
45
6
There is no assistant station on the ORTHO model

13
2.2 Detailed description
2.2.1 PACIFIC unit and chair
14
2
3
1On-off button
2Headrest with mechanical locking or pneumatic assistance
3Imitation leather upholstery with memory foam
4Door access to the unit (ambidextrous version)
2.2.2 NEO and EVO instrument tray
The PACIFIC console is available in 2 versions, with instruments the right way up or upside down
depending on the practitioner's preference.
Calbépuls®
Handle
Transthoracic cervical
tablet
Instruments
Control screen
(2 models)
Automatically-
activated cords
The handle and the instrument stands can be removed and autoclaved.

14
2.2.2.1 Instruments on whip arms
Instruments upside down (up to 6 instruments).
Clearance and articulation of whip arms:
Maximum flexibility, combined with a long cord length.
2.2.2.2 Calbépuls®
There is no Calbépuls®on the PACIFIC CHIR and ORTHO models
Calbépuls® is used for the external decontamination of instruments
using a pressurised air jet and CALBENIUM® solution
It is used to clean instruments during a consultation, as well as
after each patient for a pre-cleaning of the instruments before they
are sterilised.
The decontamination and drying times are adjustable.
For more details, see the "Settings" menus of the EVO and NEO
keyboards.
2.2.3 Assistant station
Articulated arm
Aspisept®
Suction
filter
Touch control keypad
Automatically selected
suction cannulas
Smooth cords
CHIR model
The ergonomics of the assistant station with its flexible suction arm enable 2-handed or 4-handed
treatments.
Aspisept® is a system for the internal decontamination of suction cords using Calbénium®
solution. For its use, refer to § 4.4.4, page 63.

15
2.2.4 Keyboard or control screen
For the PACIFIC equipment there are 4 models of keyboard or control screen: two touch keypads
and two models of touchscreen.
These 4 models are waterproof, which gives them a long life.
2.2.4.1 Assistant station NEO / EVO touch keypad
The polycarbonate touch keypad opposite equips the assistant
station on the NEO and EVO models.
This keyboard is used to control the positions of the chair, the basin,
the glass-filling spray, and one auxiliary function.
Detailed explanations of the commands are on § 3.5, page 22.
2.2.4.2 CHIR / ORTHO touch keypad
The polycarbonate touch keypad opposite equips the practitioner
and/or assistant stations on the CHIR and ORTHO models.
This keyboard is used to control the positions of the chair, the basin,
the glass-filling spray, and one auxiliary function.
Detailed explanations of the commands are on § 3.5, page 22.
2.2.4.3 NEO touchscreen - Smartcard
The image opposite corresponds to the NEO home screen.
It is displayed when no instrument is selected.
As soon as an instrument is selected, the screen changes
automatically and displays the instrument specific commands.
The different menus are detailed in § 3.8, page 26.
This screen requires the insertion of a smart card to operate.
The PACIFIC comes with two "user" cards including default settings
and factory settings.

16
2.2.4.4 EVO.2 touch screen - USB stick
The image opposite corresponds to the EVO.2 home screen.
It is displayed when no instrument is selected.
As soon as an instrument is selected, the screen changes
automatically and displays the instrument specific commands.
The different menus are detailed in § 3.9, page 39.
The Pacific EVO.2 is delivered with factory settings stored in the unit
processor.
A USB stick is also supplied.
This allows you to save your customised settings so that you can
switch from one Pacific device to another and take your working
parameters with you (see § 3.9.2.3, page 42).
2.2.5 Foot pedal
The PACIFIC pedal comes in 6 versions with a "tread pedal" option for the progressive pedals.
The pedal controls for these different models are detailed in § 3.5, page 20.
Wireless electric progressive
foot control
Wireless electric progressive
crush foot control
Electric progressive foot control
Electric progressive crush foot
control
Pneumatic progressive foot
control
Pneumatic foot control

17
2.3 IGN®-CALBENIUM® system
Filtering cartridge
Water from the
network
Decontaminated
water
Calpépuls®
Purging
instruments
Aspisept®
Bottle of
Calbénium®
IGN®
The IGN® integrated in the PACIFIC unit allows the practitioner to work with a spray water treated
with CALBENIUM®.
The IGN®-CALBENIUM® system supplies instrument sprays, Calbépuls®and Aspisept®.
IGN® automatically dilutes the CALBENIUM® in the spray water.
Depending on your version of PACIFIC, you have the IGN® MAG or EVO.
The IGN® system is delivered ready for use with a full bottle of CALBENIUM®.
To change of bottle refer to § 4.2, page 58.
2.4 Optional equipment
To find out the options compatible with your PACIFIC model, please contact the Sales
department.
2.4.1 Basin
Removable bowl
Glass-filling spray
Bowl spray
The bowl-rinsing and glass-filling spray can be controlled by the keypad or the pedal.
The glass bowl is removable for easy maintenance.

18
2.4.2 Surgical lighting
Optional on the PACIFIC, the surgical lighting can be mounted on the
unit, suspended from the ceiling or on a floor column.
Available in halogen or LED version.
The surgical lighting is controlled by the touch screen (except cell
lighting).
2.4.3 Amalgam separator
Due to amalgam separator standards, a place is provided in the foot of the chair to enable the
basin to be connected to an amalgam separator.
For any operation concerning the amalgam separator, refer to the manufacturer's instructions.
2.4.4 Video surveillance camera
The camera is usually installed near the dental practice's intercom
system.
The Wi-Fi surveillance camera is compatible with the EVO.2 model.
The image is viewed directly on the console screen.
3. USE
The PACIFIC unit is delivered with factory settings that were used to check that it is working
properly.
The practitioner will have to modify these pre-settings in order to adapt the chair and the unit
to their setting and working methods.
3.1 Power on/off
On-off button
The console screen turns on automatically when the unit is switched on (*).
(*) The initialisation time is approximately 1 minute.

19
- Never switch the PACIFIC unit on/off using your foot. Press the button using your hand.
- Switch off the PACIFIC unit every night after use by turning the button to Off.
3.2 Adjusting the headrest
The tilt, rotation and position settings of the headrest are carried out manually. The PACIFIC
headrest comes in two versions: mechanically locked or pneumatically assisted.
3.2.1 Mechanical locking
Steps Action Illustration
1 Loosen the locking knob.
2 Manually adjust the triple articulation of the headrest.
3 Tighten the knob.
4 The bar can slide without being unlocked.
3.2.2 Pneumatic assistance
Steps Action Illustration
1 Press the setting button.
2 Manually adjust the triple articulation of the headrest, sliding in
the axis and orientation.
3 Release the button while holding the headrest for 1 to 2 seconds.
3.3 Ambidextrous positioning
The PACIFIC equipment is perfectly symmetrical regardless of the position of the practitioner
and the assistant. The equipment arms are mounted on independent pivots. Their independence
offers a maximum range of rotation.
Positioning both on the left and on the right takes just a few seconds.

20
3.4 Foot pedal
3.4.1 Pneumatic or electric progressive pedal
Progressive with tread pedal Progressive
4
3
2
1
3
4
2
1
8
75
6
9
Rep. Function Details
1Instruments control Progressive control of the selected instrument:
motor
variation, plaque remover power or turbine speed
2
Basin position call No instrument selected
Chip blower control On selected instrument
On CHIR AD: Spray selection With lever (1) selected
3
Bowl rinse control No instrument selected
On NEO and EVO.2:
Chip blower selection
On selected instrument
Loop by single presses on the button:
Chip air / Chip spray / No chip
On NEO and EVO.2:
"Motor direction reversal selection
using your foot"
Possibility of selecting "Reversal of the motor direction
at the foot" on this button
(see motor settings p 34 and p 57)
4
Glass-filling spray control No instrument selected
On NEO and EVO.2:
Spray selection
On selected instrument
Loop by single presses on the button:
Spray / alternospray / no spray
5Chair ascent Press and hold down
Chair program call 2 Single press
6Descent of the chair Press and hold down
Chair program call 1 Single press
7Tilting the backrest backwards Press and hold down
Chair program call 4 Single press
8Tilting the backrest forwards Press and hold down
Chair program call 3 Single press
9Chair safety
Pressing on the hoop of the pedal causes all movement
of the chair to stop, and moves the chair up by about
1 cm.
Table of contents