akrometrix DIC User manual

DIC
User Manual

DIC User Manual
Page 2 of 19 40-5000-01500 Rev. A
Akrometrix Proprietary
Information
No reproduction, adaptation,
or translation without prior
written approval.
Warranty
The information contained in
this document is subject to
change without notice.
Akrometrix makes no warranty
of any kind with respect to this
information.
Akrometrix shall not be liable
for any direct, indirect,
incidental, consequential, or
other damage alleged in
connection with the furnishing
or use of this information.
Trademark Credits
Microsoft®, Windows®,
Excel®, and Notepad® are
registered trademarks of
Microsoft Corporation.
TherMoiré® is a registered
trademark of Akrometrix, LLC.

DIC User Manual
40-5000-01500 Rev. A Page 3 of 19
DIC User Manual
Contents
1Introduction.................................................................................................................4
1.1 Overview................................................................................................................4
1.2 Warnings and Precautions.....................................................................................4
1.3 System Configuration.............................................................................................6
1.4 Technical Support..................................................................................................6
2Basic System Set-up..................................................................................................8
2.1 System Set-up.......................................................................................................8
2.2 System Alignment................................................................................................11
2.3 Calibration............................................................................................................13
2.4 System Removal and Shutdown..........................................................................13
3Troubleshooting .......................................................................................................15
3.1 Failure to Initialize the Cameras ..........................................................................15
Appendix A - Miscellaneous Information..................................................................19
A.1 Equipment Ratings...............................................................................................19
A.2 Transport .............................................................................................................19
A.3 Maintenance........................................................................................................19

DIC User Manual
Page 4 of 19 40-5000-01500 Rev. A
1Introduction
1.1 Overview
The DIC system is an add-on module to complement the Akrometrix TherMoiré AXP
or Studio PS400. It is designed for measuring both in-plane and out-of-plane
displacements of samples as a function of temperature. The module uses a full-field,
non-contact optical technique called digital image correlation (DIC). Samples are
measured inside a computer-controlled oven chamber with a temperature range from
ambient to 300°C.
This manual describes the system hardware, image acquisition software (DIC), and
data analysis software (Vic-3D). The DIC software is specific to the AXP and requires its
system hardware to operate. The Vic-3D software may reside on the AXP computer or
operate as a stand-alone application on a remote computer.
Before operating the system, the user should carefully read Section 1.2, which
describes safety and health hazards associated with the system. The TherMoiré AXP or
Studio PS400 system contains a high temperature oven, motorized sample stages, and
electrical systems, all with the potential to injure if used or maintained improperly.
Section 1.3 identifies system components and controls. Section 2 describes basic
procedures for system setup and shutdown.
Appendix A describes the DIC system assembly, individual components and
optical alignment of those components. It has recommended procedures for
maintenance, storage, and transport of the DIC system. This section also contains
equipment ratings, technical support material, and a description of software file types.
1.2 Warnings and Precautions
1.2.1 Health and Safety Hazards
The DIC module is designed to work with the AXP or Studio PS400 system. Do
not operate the DIC system separately. Doing so can cause potential damage of
the DIC components.
When installing or removing the DIC assembly, place the Akrometrix-supplied
oven shield on top of the oven glass lid to protect the glass from any object that
could be dropped from the DIC assembly.
Always keep the six bolts on the triangle supports while installing or removing the
DIC assembly.
During installation or removal of the DIC unit, the user is required to lean inside
the AXP or Studio PS400. Overhead obstacles such as the light source can
cause slight head injury. During installation, be careful and aware of any
overhead obstacles.
Turn off both light sources prior to disconnecting the optical fiber bundle.
Operation with the light on may cause severe discomfort due to the operator
being exposed to a very bright light with the optical fiber bundle disconnected.

DIC User Manual
40-5000-01500 Rev. A Page 5 of 19
1.2.2 AXP/Studio PS400 System Operating Precautions
Do not install additional software on the system computer, change computer
system settings, or upgrade system software or hardware without consulting
Akrometrix. Unauthorized modifications can cause problems in normal system
operation.
Handle the glass grating with care. Make sure there is adequate clearance
between sample and grating before closing the oven lid. If the grating is
discolored by smoke or other contamination, clean the grating using
recommended procedures before performing another heating cycle.
Do not touch the quartz heating elements in the IR heater bank. Do not operate
quartz heating elements except in the horizontal position. These actions can
reduce element life.
1.2.3 Warnings and Notes in this Manual
Warnings and Notes are marked throughout the manual with these icons:
Figure 1.1 Warning Icon
Figure 1.2 Note Icon
Warnings are specific health hazards for the operator or potential sources of system
damage. Notes highlight system limitations or automatic responses that may require
corrective action by the operator for successful operation.

DIC User Manual
Page 6 of 19 40-5000-01500 Rev. A
1.3 System Configuration
Figure 1.3 DIC System Diagram
Figure 1.3 shows the layout of the DIC system in the AXP or Studio PS400.
Essentially, the system consists of two 1.4 megapixel CCD cameras mounted above the
system’s computer controlled infrared oven. Unlike Shadow Moiré, DIC does not require
an optical grating. In fact, this grating must be removed in order for the technique to
work. The double pane glass window remains to insulate the oven and provide a
viewport into the thermal chamber. The system light source remains the same as well
and provides diffuse illumination for the sample under test.
1.4 Technical Support
For technical support, contact Akrometrix:
Akrometrix
2700 NE Expressway
Building B, Suite 500
Atlanta, GA 30345
404-486-0880
404-486-0890 (fax)
http://www.akrometrix.com
When contacting Akrometrix, please provide the system serial number, the version
numbers of the Akrometrix software being used, a description of the problem or
question, and contact information for reply. If the question concerns a particular
measurement or analysis, please provide electronic copies of the phase images,
Light Source
CCD Cameras
Monitor
Computer
Sample
Oven
Sample
Support
Servo
Motor +
Stage

DIC User Manual
40-5000-01500 Rev. A Page 7 of 19
reference images, and final results and a description of data acquisition and/or analysis
conditions. If the problem concerns changes or failure in general system operation,
please describe any events or system modifications that occurred immediately before
the problem arose.

DIC User Manual
Page 8 of 19 40-5000-01500 Rev. A
2Basic System Set-up
2.1 System Set-up
2.1.1 Loading the DIC System
Warning: The user should carefully read Section 1.2
(Warnings and Precautions) before installing the DIC
module into the system.
1. Ensure that any other modular systems, such as the MP10 Fringe Projection
System, have been completely removed per its removal instructions. The
instructions to follow assume the system is in its basic shadow moiré state.
2. Ensure that the TherMoiré system has been shut down.
3. Remove the grating from the oven lid before loading the DIC system. The DIC
system cannot be operated with the grating in place.
4. Place the oven shield on top of the oven lid. Then, carefully push the oven
shield back until it touches the air knife as shown in Figure 2.1.
Warning: The DIC Camera Sub-Assembly is heavy.
Accidentally dropping the whole assembly or
components from it could damage the components
themselves or the glass lid. The use of the Oven Shield
will help prevent both types of damage.
Figure 2.1 Oven Shield over the TherMoiré Oven Lid
5. Select the appropriate Camera Support Plate for the desired field of view
(FOV). The widest plate is used for the largest FOV and vice versa (Figure
2.2). When aimed at the same spot these plates should give the camera system
a stereo angle of ~15°.
Air Knife
Oven Shield
Oven

DIC User Manual
40-5000-01500 Rev. A Page 9 of 19
Figure 2.2 Large, Medium, and Small FOV Camera Adapter Plates
6. Attach the cameras to the plate using two ¼-20 button head screws (Figure
2.3). Only the widest mount has a front and back side. The back side has
counter-bored holes so that it can sit flush against the Common Crossbar
Mount Support. Their alignment is not critical at this point as they can be
adjusted to view the same point once the assembly is installed.
Figure 2.3 Camera Sub-assembly, Backside
7. Place the DIC assembly support (Common Crossbar Mount Support with two
Carriage Assembly Support Triangles) onto the two glide rails located on two
sides of the oven lid (Figure 2.1). Align the whole assembly such that the
cameras will be viewing the sample location. Attach them with three ¼-20
aluminum thumbscrews each. Make sure that the triangles are seated over the
glide rails properly before tightening the thumbscrews.
~88 x 66 mm
~66 x 48 mm
~45 x 34 mm

DIC User Manual
Page 10 of 19 40-5000-01500 Rev. A
Figure 2.4 DIC AXP Carriage Support Assembly
8. Slide the Camera Sub-assembly over the two pins located at the center of the
Common Crossbar Mount Support. There are three possible camera positions.
Choose the position which corresponds with the desired FOV. Some rocking
may be required to fully seat or unseat this assembly as the bushings and pins
are a precise fit.
9. Once the Camera Sub-Assembly is located on the pins and is flush with the
Common Crossbar Mount Support, place two aluminum thumbscrews into the
appropriate holes (marked by red circles in Figure 2.5) and tighten by hand.
10. Attach a trigger box cable and firewire cable to each camera. Secure them to
the top of the Common Crossbar Mount Support using the supplied Velcro strap
as shown in Figure 2.5.
Warning: Plugging or unplugging the 1394b firewire
cables while power is on can potentially damage 1394b
firewire devices.
Glide Rails (2)
Aluminum
Thumbscrews
(3) on Each Side
Common Crossbar
Mount Support
Support Triangles (2)

DIC User Manual
40-5000-01500 Rev. A Page 11 of 19
Figure 2.5 Fully Installed DIC system
11. Unplug the firewire cable from the AXP or PS400 camera.
Note: If the AXP system camera is connected at
startup, the software may produce an error and not
start.
2.1.2 System Start-Up
1. Please see the relevant base TherMoiré system manual for its start-up
procedures.
2. Turn on the Light Source Power switches. Make sure the Light Source
Adjustment knob is placed at approximately 90% of the maximum level.
3. Run the DIC Surface Measurement program by clicking its icon on the Studio
Manager bar. A start-up splash screen appears while the software initializes.
For operation of this program, refer to the Surface Measurement User Manual.
4. Check if there is any dust or other kinds of contaminations on the oven lid glass.
If so, clean the glass according to Appendix A.3.
2.2 System Alignment
1. Choose the largest calibration target that will fit within the cameras’ field of view
(FOV). Place the calibration target on the sample support rails such that it is
roughly in the center of the oven. The target should be oriented with the long
dimension running left to right.
2. Close the oven lid, being careful that there is enough of a gap above the
calibration target so that there is clearance when the lid closes.

DIC User Manual
Page 12 of 19 40-5000-01500 Rev. A
3. Press the Adjust Height button and lower the TherMoiré sample support so that
the level of the calibration target is the same as sample to be measured.
Typically this will be such that the bottom of the calibration target is within 5 mm
of the oven heater bulbs.
4. Adjust the right camera so that it is roughly centered on the calibration target
dot field.
5. Move the calibration target left or right and up or down to make its dot field fit
within the right camera’s FOV.
6. Align the left camera such that it matches the right camera’s FOV. There is a
hole in the Common Crossbar Mount support behind where the camera
attaches for the largest FOV. This allows for adjusting the left camera’s angle
while the camera mount plate is securely attached using the thumbscrews.
Note: Only the FOV that can be seen in both cameras
will be useful for analysis. When connecting the
cameras to their support bracket, it is important to line
them up so that they image the same FOV.
Figure 2.6 Two Images of Calibration Target in DIC Surface Measurement
7. Adjust aperture and focus so that the cal target is adequately bright and in
focus.
8. To get best focus, open the iris fully on one camera lens and reduce the light
source intensity to compensate.

DIC User Manual
40-5000-01500 Rev. A Page 13 of 19
9. Expand the images in the camera window by right clicking on either camera
image and unchecking Shrink To Fit. The camera window can also then be
expanded by dragging its edges to resize.
10. Unlock the focus ring on one lens and adjust its focus to maximum sharpness.
11. Lock the focus ring and close the lens iris.
12. Repeat steps 8 through 11 on the second camera lens.
13. Raise the light source intensity to original levels.
2.3 Calibration
Once the Cameras are properly aligned, a calibration needs to be performed before
measurement. Please see the Surface Measurement User Manual for instructions on
how to calibrate DIC system. After the calibration is done, no individual component
should be moved without recalibration.
2.4 System Removal and Shutdown
1. Close the DIC Surface Measurement application.
2. Press the LID RELEASE button, open the oven, and remove the sample from
the oven chamber.
Warning: The temperature control chamber cools
much more slowly than it heats. The sample supports,
sample and other components inside may retain
significant heat long after the thermal cycle is complete.
Use extreme caution in unloading the sample or other
components. The Process Temperatures for each loop
of the temperature controller display the sample
temperature, which may not be an accurate indicator
of the temperature of other oven components.
3. Close the oven lid.
4. Find the Akrometrix-supplied standby plate for the DIC system and place it on a
flat surface (Figure 2.7).
Figure 2.7 DIC Standby Plate
5. Place the Oven Shield on top of the oven lid as shown in Figure 2.1.

DIC User Manual
Page 14 of 19 40-5000-01500 Rev. A
6. Place lens caps on each of the DIC camera lenses.
7. Disconnect the two synchronization cables and two firewire cables from the top
of each camera. Carefully place them to the back or right side of the oven. Keep
them a safe distance from the oven body or any moving components.
Warning: The DIC Camera Sub-Assembly is heavy.
Accidentally dropping the whole assembly or
components from it could damage the components
themselves or the glass lid. The use of the Oven Shield
will help prevent both types of damage.
8. Loosen the six aluminum thumbscrews on the two triangles and take off the DIC
assembly support (Common Crossbar Mount Support with two Carriage
Assembly Support Triangles) from the oven lid. Attach the whole assembly onto
the Akrometrix-supplied standby plate with six aluminum thumbscrews (Figure
2.8).
Figure 2.8 DIC System Installed on Standby Plate
9. Cover the DIC system with a piece of cloth and/or place it in a storage room to
protect the optical components from dust and accidental damage.
10. Install the grating in the oven lid if the system needs to be operated in
TherMoiré mode.

DIC User Manual
40-5000-01500 Rev. A Page 15 of 19
3Troubleshooting
3.1 Failure to Initialize the Cameras
Symptom:
When opening the DIC Surface Measurement program, the cameras fail to be
initialized. A window may pop up as shown in Figure 3.1.
Figure 3.1 Camera Initialization Error
Cause:
One or more of the camera cables is not connected properly.
When a camera is first installed or replaced by another one, the driver is not
installed or updated.
When a camera cable is unplugged, the system may lose track of the camera. In
this case, the software cannot recognize the camera even if its cable is plugged
back in.
A trigger signal was not found.
Solution:
1. Make sure the firewire cables are connected to both cameras and to the firewire
hub. Also ensure the firewire hub is properly connected to the computer and is
powered up. If problem is not resolved, go to 2.
2. Make sure that the trigger box cables are connected to both cameras and the
box itself. Also ensure that the trigger box is properly powered on. A red
blinking light indicates power is available. If problem is not resolved, go to 3.
3. Check the cameras with the 1394camera utility program:
a. Run the utility program from StartAll Programs1394Camera Demo.
b. In the 1394 Camera Demo application window, select CameraCheck Link
(Figure 3.2).
c. A dialog shows up and states “Found 2 Cameras”. Press OK to continue
(Figure 3.3). If “Found 0 Cameras”, go to step 3.
d. Select either of the connected cameras: CameraSelect Camera (Figure
3.4).

DIC User Manual
Page 16 of 19 40-5000-01500 Rev. A
Figure 3.2
Figure 3.3
Figure 3.4
Figure 3.5
e. Initialize the camera: CameraInit Camera (Figure 3.5).
Warning: Selecting Yes will reset the camera
configuration (e.g. brightness, shutter speed, gain, etc.)
back to the defaults, which are not favorable conditions
for the measurement system. Akrometrix has set these
values from the factory. If they are accidentally reset
please contact support to receive instructions on how to
set these values.
f. Press No when asked to reset to powerup defaults (Figure 3.6).

DIC User Manual
40-5000-01500 Rev. A Page 17 of 19
Figure 3.6 Don’t Reset to Defaults
Figure 3.7 Show Camera
g. Display the camera output: CameraShow Camera (Figure 3.7).
h. A live view of the camera will be shown in the camera output window. Adjust
the light source intensity and/or lens iris as necessary.
i. If the camera is found and a live view can be seen in the utility program,
repeat Step 2 for the second camera. If both camera images can be seen,
close the utility program and reopen the DIC program. Otherwise, go to step
3.
4. Disconnect the non-functioning camera’s cable and reconnect it. In the device
wizard that appears:
a. Do NOT connect to Windows Update to search for software (Figure 3.8).
Figure 3.8
Figure 3.9
b. Install from a specific location (Figure 3.9).
c. Don’t search; choose the driver to install (Figure 3.10).

DIC User Manual
Page 18 of 19 40-5000-01500 Rev. A
Figure 3.10
Figure 3.11 Select Hardware Type
d. Select Imaging Devices as the Hardware Type (Figure 3.11).
e. Press Have Disk… and browse to “C:\Program
Files\CMU\1394Camera\Driver\1394Camera.inf” (Figure 3.12).
f. Continue Anyway when warned that the driver is not digitally signed (Figure
3.13).
g. Go to step 3.
Figure 3.12 Install From Disk Window
Figure 3.13 Windows Logo Compatibility Bypass

DIC User Manual
40-5000-01500 Rev. A Page 19 of 19
Appendix A - Miscellaneous Information
A.1 Equipment Ratings
Electrical*
Supply Voltage
220VAC, 50/60 Hz, 1Φ
Amp Rating
~1 A
*All electrical requirements supplied by host AXP system, no additional
facilities setup needed
Mechanical
Footprint
629 (w) x 355 (d) x 340 (h) mm
Weight
Camera Sub-Assembly: 0.6kg
Camera Sub-Assembly with Common Crossbar: 3.2kg
A.2 Transport
When transporting the DIC system to a new location, Akrometrix recommends that
certain components be removed from the system and packed and moved separately.
These include:
Cameras and lenses
Grating and grating frame
A.3 Maintenance
A.3.1 Optical Lenses
The camera lenses are exposed directly to the air and may collect dust over time.
Dust or dirt on the camera lenses will not be seen directly but may darken the images.
To clean dust on the lenses, use a non-flammable compressed air duster (canned
air), such as would be used for cleaning computer components. Blow away dust without
touching the lenses.
Warning: Do not use the tissue to rub the lens surface
back and forth. This may leave scratches on the glass
surface and damage the lens.
The lenses may also be contaminated with residues from dust or fingerprints. If this
takes place, they should be removed from the cameras and cleaned with lint-free lens
tissue (such as Kimwipes or other disposable absorbent papers designed for delicate
use and high-absorbency) and lens cleaning fluid. When using lens tissue to clean the
lenses, gently wipe the surface in one direction once and change to a new piece of
tissue to clean continuously until the lens is clean.
Table of contents
Popular Control Unit manuals by other brands

Paradox
Paradox PCS265 Installation and programming guide

Heath Zenith
Heath Zenith Wireless Outdoor Power Control 6022 owner's manual

Sys Tec Electronic
Sys Tec Electronic ECUcore-1021 Hardware manual

SIGMA TEK
SIGMA TEK S-DIAS DM 821 instruction manual

Abus
Abus PLAC90230 quick guide

Aero-motive
Aero-motive 19105 installation instructions