manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Akse
  6. •
  7. Measuring Instrument
  8. •
  9. Akse X3M Instruction Manual

Akse X3M Instruction Manual

X3M, X3M-H, Flash N, Flash N-H
Installation Manual
This instructions manual is common to both the instruments type X3M-D
and Flash-D.
COPYRIGHT
Akse S.r.l. All rights are reserved.
It is forbidden to duplicate, adapt, transcript this document without Akse written
authorization, except when regulated accordingly by the Copyright Laws.
Copyright© 2003-2004
WARRANTY
This product is covered by a warranty against material and manufacturing defects for a
period of 36 months period from the manufacturing date
The warranty does not cover the defects that are due to:
• Negligent and improper use
• Failures caused by atmospheric hazards
• Acts of vandalism
• Wear out of materials
Akse reserves the right, at its discretion, to repair or substitute the faulty products
The warranty is not applicable to the products that will result defective in consequence
of a negligent and improper use or an operating procedure not contemplated in this
manual.
RETURN AND REPAIR FORMALITIES
Akse accepts the return of instruments for repair only when authorized in advance.
For instrument purchased directly, the repair authorization must be requested to
Electrex directly by using the enclosed RMA form.
RE-SHIPPING OF REPAIRED PRODUCT
The terms for re-shipment of repaired products are ex-works, i.e. the transport costs
are at customer charge.
Products returned as detective but found to be perfectly working by our laboratories,
will be charged a fixed fee (40.00 Euro + VAT where applicable) to account for
checking and testing time irrespective of the warranty terms.
Safety
This instrument was manufactured and tested in compliance with IEC 61010 class 2
standards for operating voltages up to 250 VAC rms phase to neutral.
In order to maintain this condition and to ensure safe operation, the user must comply
with the indications and markings contained in the following instructions:
•When the instrument is received, before starting its installation, check that
it is intact and no damage occurred during transport.
•Before mounting, ensure that the instrument operating voltages and the
mains voltage are compatible then proceed with the installation.
•The instrument power supply needs no earth connection.
•The instrument is not equipped with a power supply fuse; a suitable
external protection fuse must be foreseen by the contractor.
•Maintenance and/or repair must be carried out only by qualified, authorized
personnel
•If there is ever the suspicion that safe operation is no longer possible, the
instrument must be taken out of service and precautions taken against its
accidental use.
•Operation is no longer safe when:
1) There is clearly visible damage.
2) The instrument no longer functions.
3) After lengthy storage in unfavorable conditions.
4) After serious damage occurred during transport
The instruments X3M and Flash-N must be installed in respect of all the local
regulations.
Operator safety
Warning:Failure to observe the following instructions may lead to a serious danger of
death.
-During normal operation dangerous voltages can occur on instrument terminals
and on voltage and current transformers. Energized voltage and current
transformers may generate lethal voltages. Follow carefully the standard safety
precautions while carrying out any installation or service operation.
-The terminals of the instrument must not be accessible by the user after the
installation. The user should only be allowed to access the instrument front panel
where the display is located.
-Do not use the digital outputs for protection functions nor for power limitation
functions. The instrument is suitable only for secondary protection functions.
-The instrument must be protected by a breaking device capable of interrupting
both the power supply and the measurement terminals. It must be easily reachable
by the operator and well identified as instrument cut-off device.
-The instrument and its connections must be carefully protected against short-
circuit.
Precautions: Failure to respect the following instructions may irreversibly damage to
the instrument.
-The instrument is equipped with PTC current limiting device but a suitable external
protection fuse should be foreseen by the contractor.
-The outputs and the options operate at low voltage level; they cannot be powered by
any unspecified external voltage.
-The application of currents not compatible with the current inputs levels will damage
to the instrument.
Mounting
Instruments size (mm)
Fixing and blocking
The instrument (as well as the optional modules) are fixed to the DIN rail by
means of the spring clip located on the rear side of the unit
Wiring diagrams
Power supply
The instrument is fitted with a separate power
supply with extended operating range. The power
supply terminals are numbered (10) and (11). Use
cables with max cross-section of 4 mm2.
Voltage connection
Connessioni voltmetriche
Use cables with max cross-section of 4 mm2and connect them to the
terminals marked VOLTAGE INPUT on the instrument according to the
applicable diagrams that follow.
Current connection
It is necessary to use external CTs with a primary rating adequate to the load
to be metered and with a 5A secondary rating. The number of CTs to be
used (1, 2 or 3) depends upon the type of network.
Connect the CT output(s) to the terminals marked CURRENT INPUT of the
instrument according to the applicable diagrams that follow.
Use cables with cross-section adequate to the VA rating of the CT and to the
distance to be covered. The max cross-section for the terminals is 4 mm2.
N.B. The CT secondary must always be in short circuit when not connected
to the instrument in order to avoid damages and risks for the operator.
Warning: THE PHASE RELATIONSHIP AMONG VOLTAGE AND
CURRENT SIGNALS MUST BE CAREFULLY RESPECTED. ALL
DISREGARD OF THIS RULE OR OF THE WIRING DIAGRAM LEADS
TO SEVERE MEASUREMENT ERRORS.
4W Star connection (4 wire)
Low voltage 3 CTs High voltage 3 PTs 3 CTs
Configuration 3Ph/4W Configuration 3Ph/4W
3W Delta connection (3 wire)
Connection with 3 CTs Connection with 1 CT
Low Voltage 3 CTs (unbalanced load) Low Voltage 1 CT (symmetrical and
balanced load)
Configuration 3Ph/3W Configuration 3Ph/3W-Bal
Connection with 2 CTs on L1 and L3
Low Voltage 3 CTs (unbalanced load) Low Voltage 1 CT (symmetrical and
balanced load)
Configuration 3Ph/3W Configuration 3Ph/3W-Bal
Outputs connection
Instrument set up The set up procedure allows to program the instrument
operating parameters. Entry in the programming procedure
is obtained by pressing the PROGRAM button that is
located on the upper right side of the instrument.
The key allows to scroll the various entry fields within a
set up page as well as to pass to the next page upon
scrolling all the fields of one page. The and keys
allow the modification of the flashing field being currently
selected.
The content of a field can be either numeric or a parameter controlling the device
behavior. The key advances to the next page, the key returns to the
previous page.
By pressing the PROGRAM button (while in any configuration page) the menu is
exited and the configuration entries so far performed are saved.
Configuration procedure
Communication characteristics configuration
This menu appear only upon connection to the instrument of an RS-485 or an
RS-232 optional module. The setting of the RS485 communication
characteristics requires to scroll the programming pages with two keys; The
key advances to the next page, the key returns to the previous
page The first page enables MODBUS communication protocol to
be set: the Slave Address, Transmission delay, RTU or
ASC (ASCII) mode. Transmission delay stands for the time
delay the instrument will wait prior to reply to a data query.
It is expressed in milliseconds, the
default value is 100 msec and a 0
setting is also possible.
All these data are correlated depending upon the stop bit
value.
Additional parameters regarding the the setting of
communication speed, number of data bits, parity, stop
bits are available in the next page.
Readings
The selection of the readings and of the reading pages is made by means of
the following keys: Voltage and frequency key, Currents key,
Powers key, Power factor key, Energies key, Life time,
indicator key, Move the selection up and down in the readings pages,
This key is not used in the readings pages, it displays instruments
information (loader version, application version, serial number, available
options).
A broader instructions manual including the Modbus protocol mapping
may be downloaded from our web page www.electrex.it.
DECLARATION OF CONFORMITY
Akse hereby declares that its range of products complies with the following
directives EMC 89/336/EEC 73/23CE 93/68 CE and complies with the
following product’s standard CEI EN 61326 – IEC 61326 CEI EN 61010 –
IEC 1010
The product has been tested in the typical wiring configuration and with
peripherals conforming to the EMC directive and the LV directive.
The Technical file report is available at:
Edition 3 January 2007 Subject to modification without notice

This manual suits for next models

3

Other Akse Measuring Instrument manuals

Akse Electrex FEMTO D4 70A PFA6431-02 User manual

Akse

Akse Electrex FEMTO D4 70A PFA6431-02 User manual

Akse Electrex D6 RS485 User manual

Akse

Akse Electrex D6 RS485 User manual

Akse Electrex 96 RS485 User manual

Akse

Akse Electrex 96 RS485 User manual

Popular Measuring Instrument manuals by other brands

Hioki FT3432 instruction manual

Hioki

Hioki FT3432 instruction manual

ColorVision Spyder2 Plus quick start guide

ColorVision

ColorVision Spyder2 Plus quick start guide

Xylem YSI Alyza IQ NH4 Operation manual

Xylem

Xylem YSI Alyza IQ NH4 Operation manual

Bushnell 1K owner's guide

Bushnell

Bushnell 1K owner's guide

Pulsar MDot instruction manual

Pulsar

Pulsar MDot instruction manual

3nh YG Series Operation manual

3nh

3nh YG Series Operation manual

SRM ORIGIN PowerMeter user guide

SRM

SRM ORIGIN PowerMeter user guide

Endress+Hauser Liquicap M FMI52 PFM operating instructions

Endress+Hauser

Endress+Hauser Liquicap M FMI52 PFM operating instructions

Jolly logic ALTIMETERTWO quick start

Jolly logic

Jolly logic ALTIMETERTWO quick start

Amprobe VPC-10 instruction manual

Amprobe

Amprobe VPC-10 instruction manual

Endress+Hauser Proline Promass E TB2 HART operating instructions

Endress+Hauser

Endress+Hauser Proline Promass E TB2 HART operating instructions

LINOVISION IOT-S300AP product manual

LINOVISION

LINOVISION IOT-S300AP product manual

Bosch GLM 50 Professional Original instructions

Bosch

Bosch GLM 50 Professional Original instructions

Apogee 2349 DLTM4 instructions

Apogee

Apogee 2349 DLTM4 instructions

Palstar PM2000AM Technical manual

Palstar

Palstar PM2000AM Technical manual

TSI Incorporated velocicalc 9545 Operation and service manual

TSI Incorporated

TSI Incorporated velocicalc 9545 Operation and service manual

Dent ELITEpro XC Operator's guide

Dent

Dent ELITEpro XC Operator's guide

Megger MPQ1000 Getting started

Megger

Megger MPQ1000 Getting started

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.