1.Characteristics ・ Applications
zCapable of measuring moisture contents of timbers・wooden products by inserting (touching) a probe to
measuring objects.
zMC (Moisture content) mode makes it possible to measure all kinds of objects.
Switching of the mode of a body is capable of comparing and classifying states of dryness・wetness of
measuring objects with a displayed numerical value of 1 to 100(without unit) which leads to quick and easy
checking
Woods
(H,S)
Wooden products like Timbers, Constructing materials, Packing materials, Plywood,
Compiled woods, Particle board, Furniture, Flouring woods, Textile materials,
Household materials, etc.
Moisture content
(MC-6)
Judging measurement results by comparing displayed numerical values of moisture
content of textiles, foods, chemical synthesis, ceramics and so on, the methods of which
were considered difficult to perform.
2. Specifications
Model AQUA SEARCH SERIESTG-101
Method DCelectricresistancetype Dryweightedstandardmoisture%
Range 6.0 to 50.0% , 1 to 100(MC mode)
Resolution 0.1%, 1(MCmode)
Accuracy ±3% (read-out value)
Mode Broad leaf tree (Hard), Conifer (Soft), Moisture content comparison
Average Value Display Max. 20 points of average value data (Switch off to delete)
Display LCDindicationwithHOLDfunction
In addition to a measuring value, measuring mode, HOLD, TEMP,
Upper limit value setting and Battery residue are displayed
Upper limit value setting Setting of optional numerical value of upper limit
6.5 to 49.5% (0.5% step), MC: 2 to 99 (1 step)
Probe Small stabbing(Inserting)type 2 needle electrode
Temperature compensating Automatic temperature compensating function (ON-OFF function available)
Power Alkali dry battery LR03 (1.5V) × 4, Non-stop running 100 hrs (approx.)
Auto-power OFF
Operating temperature 0 to 40℃(Non-condensing)
Dimensions 80(W) × 35(H) × 150(D) mm, 320g
Accessories Carrying case, spare needles, Hexagon wrench (2 mm)
Option SB metal net electrode, Conducting rubber electrode, Roller electrode,
Moisture reading checker
* Specifications and appearances subject to change due to improvement without prior notice.
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

3. Names
3-1 Main body
① Power SW (POWER)
Switch ON or OFF. Switch ON after a probe has been hooked up to a main body.
A display turns automatically to Error [E01] and Power-OFF because of Self Analysis Function when any
abnormality is detected in the inner circuitry of a body.
② Hold・Average Key (HOLD/AVE)
It changes ON or OFF alternately whenever pushed.
A push of long periods (approx. 1 sec.) measures Average Statistical Mode (Max. measuring point 20).
An average value can automatically be displayed after 20 points have been measured.
Repress the key to stop the measurement for getting an average value at less than 20 points.
A push of long periods (approx. 2 sec.) releases the mode and turn to a normal measurement.
And POWER OFF also releases the mode.
③ Mode Change Key (G)
It turns to H (Broad leaf tree:HARD), S (Conifer:SOFT) or MC-6 (Moisture content) mode for each moisture
measurement.
④ Numerical value setting key (▲ , ▼)
This is a key to set to an upper limit value.
The「▲」,「▼」key moves up and down by a 0.5% notch (a 1 notch for MC mode) and keep the key pushed
for a fast forward winding (rewinding).
The「▼」Key makes it possible to delete a point only of a latest measuring value at Average Statistical Mode
in a measuring process.
Switch the Automatic temperature compensating function to ON or OFF. Refer for details to 「7. About
temperature compensating」.
Probe connector receptacle
Probe connector
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

⑤ Lock key (LOCK)
A push of a Lock key locks all the keys except POWER SW ①which prevents erroneous operations.
Switch OFF to release the lock.
⑥ Upper Limit Value Key (H)
Switch on Upper Limit Value Mode.
Set to an optional numerical value with a numerical value setting key ④.
⑦ LCD Display
Measuring Modes, measuring values, upper limit setting, temperature compensating, battery residue and so on
are to be indicated on the LCD.
⑧ Probe connector receptacle
This is a connector to be hooked up to a probe.
⑨ Hand strap
To protect this instrument from dropping, make sure that the hand strap be hung through over an operator’s
wrist.
⑩ Battery Case (back Lower part of a main body)
Batteries are to be placed.
3-2 Probe
Small type 2 needles probe
⑪ Probe Connector
⑫ 2 needle electrode
⑬ Needle fixing block
⑭ Hexagon socket set screw for needle fixing
⑮ Handle
⑯ Probe Cord
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

4. How to use
4-1 Connection or disconnection of probe
( ensure that Power Source of a main body is OFF).
・ Hook up the Probe connector ⑪to a main body to meet the
Arrow Mark of the Probe Connector ⑪with the Mark of Probe
Connector Receptacle ⑧and insert it carefully up to the end of a
stroke.
・ Switch Power OFF to disconnect Probe Connector from the
receptacle.
*Unnatural force in inserting and pulling off may cause damages to connectors in a receptacle and a cord.
4-2 POWER ON・OFF
・Press Power SW ①to emit a“ Peep” sound ,and “LLL” and Measuring
Mode ”H” Broad Leaf Tree mark are displayed ⑦.
・ Press again to switch OFF, emitting a “Peep” sound to turn off the display.
・ Availability of Auto-Power-OFF preventing batteries from being
consumed when Power is forgotten to switch off. An interval of 3 minutes without operation of measuring
will automatically switch off.
*Press Power SW ①to ON or OFF at an interval of 3 to 5 seconds to prevent operating errors.
4-3 Changes of Measuring Mode
・When Power is ON、「H」Broad Leaf Tree Mode is displayed.
・ For changes of Measuring Mode, press a Mode-Change「G」key ③to change the following turns by each
press.
When Power is OnLCD Display
Hard (broad leaf tree) H
↓ ↓
Press 「G」key. Soft (conifer)S
↓↓
Press 「G」key. MC ModeMC‑6
Press 「G」key
zReturning again to ON from OFF, the mode automatically returns to an initial measuring modes. Confirm a
measuring mode when Power is ON.
z「TEMP」is not displayed with the MC mode of Moisture Content because there is no temperature
compensating with the mode.
Probe connector receptacle
Probe connector
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

4-4 Changes of Hold Mode
A press of Hold・Average Key ②turns to the Hold Mode with a「peep」sound.
Each press of the key changes the Hold Mode alternately to ON or OFF and
「Hold」is displayed on LC Display ⑦when ON is in process. The displayed
measuring value is on hold until a next measuring value has been input.
*A press of long periods (approx. 1 sec.) turns to Average Value Mode
(temporal memory). To release it, press again and hold the Hold ・Average Key
②for long periods (approx. 2 sec.).
4-5 Setting of Upper Limit Values
・Press the Upper Limit Value Key「H」⑥ when Power ON is in process.
・A buzzer emits「peep」sounds and the「H」,「●」marks of an upper limit
value indications and「50.0%」of the range of the upper limit are displayed.
・Press and hold a「▼」key ④until it has reached to a desired upper limit value.
Keeping the「▼」key ④pressed changes to a fast winding of numerical values.
・Release your finger of the「▼」key ④when the values has reached to a
desired value.
・When the value has reached too small beyond the desired one, press the「▲」
Key ④to adjust the value.
・Set an upper limit value by pressing the「H」⑥ key again. ex) set upper limit value to 20%
・A buzzer emits a “ Peep” sound and the upper limit setting mark「●」is
indicated on the edge of the left side of LCD display ⑦.
*An interval between setting values is a 0.5% notch.
*“ TEMP”, “HOLD” is not displayed during in process of setting.
*Possible to activate the same operations even at the MC mode.
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

・At a real time measuring, it sounds a buzzer “ Peep, Peep, Peep “ when a measuring value exceeds a setting
value, and the upper limit value and the measuring value alternately blinks on the display while being probed to
contact. Releasing the probe turns to「LLL」on display.
・At a Hold Mode measuring, when it exceeds the setting value and even after releasing the probe off the object,
the upper limit value and the measured value are alternately kept blinking.
・Press a「H」key ⑥again and repeat the same procedures as before to change upper limit values.
Possible to operate the same procedures during measuring in process (Hold mode or Non-Hold mode).
*When Upper Limit Value is unnecessary for an initial operation, press again the 「H」key after pressing first
the「H」key. In this case the Upper Limit Value is set at 50%, Maximum.
*To delete an Upper Limit Value, set the Upper Limit Value at 50 % or switch Power to OFF.
And the Upper Limit Value will also be deleted when Power switches OFF with Auto-Power-Off.
4-6 Changes of Key -Lock Mode
A press of a Lock Key ⑤locks all the keys except Power SW ①and
prevents erroneous operations. Switch off Power to release.
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

5. Measurements
5-1 Real time measurement
1. Turn on Power SW ①and confirm an indication of LCD.
Display of a measuring mode 「H」Broad Leaf Tree
and 「LLL」
zTurning on Power while inserting the probe to the object indicates 「LLL」once on the display and then
displays a measuring value in 2 to 3 seconds.
2. Select a mode suitable for the measuring object by pressing a Mode Change key「G」③.
(refer to 4-3 changes of measuring modes page at 5)
3. Wait for reading the display until the display has become stable by inserting Electrode into the measuring
object.
4. Press a Hold・Average key ②to display 「0.0 %」with a sound of “Peep”.
When probing an object, it emits “peep, peep” sounds and when releasing it,
a “ peep” sound again with a measured valued indicated on LCD ⑦.
* Do not probe other object until the above sounds of 3 times has finished
otherwise that may cause measurement errors.
An indicated value does not change when a measured value indicates not
more than the measurement range.
When a measured value indicates not less than the measurement range,
「HHH」blinks with sounds of “ peep, peep, peep”.
Releasing a probe displays「0.0%」.
5. Press a Hold・Average key ②again for releasing the Hold state.
Note) When stubbing a probe into a measuring object:
Keep Hold Mode released until stubbing has been completed. Make sure of changing to Hold Mode to
read a measured value after finishing the stubbing.
(Stubbing during a Hold mode in process may cause measurement errors because it displays data while
in process of stubbing.)
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

5-2 Measurement of Average Value Indication (Temporary Memory)
1. A long press (approx. 1 sec.) of a Hold・Average Key ②performs temporary
Memory Measurement to display Average Value Indications.
( A max. measurement point capable for Temporary Memory is 20)
2. This Temporary Memory can be deleted when Power is OFF
(and Auto-Power OFF is as well)
3. Average Value Measurement is performed in a measurement of Hold Mode.
Measured data memory is capable of memorizing temporarily「AV E 1」,
「AV E 2」,・・・・ and 「AV E 2 0 」at the Maximum.
4. A press of the Hold・Average ②key during in process of the measurement
turns to Average Value mode and on LCD ⑦「AV E _ _ 」is indicated and shows
the average value measured up to the points.
Then by operating numerical value setting keys「▲」,「▼」④, measured
values can be recalled and confirmed.
On the place of〔AV E 〕on display, recalled measured numbers (1 to 20) are
indicated and each measured value can also be confirmed respectively.
And then〔HOLD〕or〔AV E 〕is not indicated.
*During this mode, Auto Power OFF can not be operated.
*An Average Value Mode can not return to a measuring operation.
When returning, refer again to 7. below and try to set again.
5. When reading erroneous data during a measurement of Average Value Indications, a long press and hold
(approx. 1 sec.) of a Numerical Value Setting Key 「▼」 ④is capable of deleting the only last data stored
( *Only possible to operate the latest measured data.)
6. When measured numbers has reached 20 items, it automatically turns to
Average Value Indication mode in a second (the data at the 20th can not be
deleted by operating any keys)
Release the mode to return to a measuring operation.
↓1 second after
7. A long press and hold (approx. 2 sec.) of the Hold・Average key ②make it
possible to release the mode.
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

zPower SW key ①and Hold・Average key ②can only be used during a measurement in process of
Average Value Indication.
zDuring a measurement of Average Value Indication, a setting function of Upper Limit Values turns ineffective.
When actuating Average Value Indication measurement during a use of a setting function of Upper Limit
Value, the setting function of an Upper Limit Value turn ineffective.
But, even in this case Upper Limit Value is being in process of storing unless Power is switched OFF, and it
returns to the Upper Value setting when a measurement of Average Value Indications has finished.
6. Measurement of MC(Moisture Content) mode
Select Moisture Content Mode by pressing a Mode Change key ③after
turning Power to ON.
This MC mode is capable of using broad objects relating to Resistance.
Changes of electric resistance depending on dryness and moisture indicates numerical values
「1 to 100」(No Unit). Checking relatively comparative values makes it easy to classify and compare objects.
This MC measurement mode as well as when measuring Moisture can use all the functions although each of them
does not display〔%〕.
7. About Temperature Compensating
・Automatic Temperature Compensating Function is set to ON at the initial
setting of this meter and the LCD ⑦indicates〔TEMP〕.
(There is no Temperature Compensating with the MC mode measurement.)
・Auto Temperature Compensation function is performed at a basis of an
indicated value of 20 C.
・ Turn Auto Temperature Compensation function to OFF for measurements
when measuring objects are heated up because of a baking (dry) heat treatment.
・ Differences of temperatures between a main body and a probe cause
measurement errors.
・ Press 「▲」 and 「▼」 of Numerical Value setting keys ④at the same
time to change Auto Temperature Compensation to ON or OFF.
And then that make it possible to understand the changes whether〔TEMP〕is indicated or not on
the LCD ⑦, depending on the key operations.
・ A setting of ON or OFF of Auto Temperature Compensating is kept alive even if the mode changes.
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

●Refer to the table below for compensate the temperature and kind of wood of electric resistance type wood
moisture meter.
※Extracted from the「Electric Moisture Meters」issued in 1960 by THE NIKKAN KOGYO SHINBUN, Ltd
Moisture(%)
Temperature(℃)
+2.0 +2.5 +3.0
+2.0 +2.5 +3.0
+2.0 +2.0 +3.0
+1.5 +2.0 +2.5
+1.5 +1.5 +2.0
+1.0 +1.5 +2.0
+1.0 +1.0 +1.5
+1.0 +1.0 +1.0
+0.5 +0.5 +1.0
+0.5 +0.5 +0.5
0 +0.5 +0.5
000
0 -0.5 -0.5
-0.5 -0.5 -0.5
-0.5 -0.5 -1.0
-1.0 -1.0 -1.0
-1.0 -1.0 -1.5
-1.0 -1.5 -2.0
-1.5 -1.5 -2.0
-1.5 -2.0 -2.5
-2.0 -2.0 -3.0
-2.0 -2.5 -3.0
0 to 1
2 to 3
4 to 5
6 to 7
8 to 9
10 to 11
12 to 13
14 to 15
26 to 27
28 to 29
30 to 31
16 to 17
18 to 19
20 to 21
22 to 23
above or 21
above or 40
below 0
4 to 11 12 to 20
32 to 33
34 to 35
36 to 37
38 to 39
24 to 25
Moisture(%)
Kind of wood
Ash , Elm -0.5 -1.0 -1.0 -1.0
Japanese oak , Linden 0 0 +0.5 +0.5
Beech 0 0 0 +0.5
Cherry tree , Chestnut tree 0 0 +0.5 +0.5
Katsura wood , Birch +0.5 +0.5 +1.0 +1.0
Oak , Zelkova 0000
Apitong -1.0 -1.0 -2.0 -2.5
Castor aralia , Lauan +0.5 +1.0 +2.0 +4.0
Japan cedar +0.5 +0.5 +1.0 +2.0
Japanese cypress , Pine +0.5 +1.0 +2.0 +3.0
Fir 0 0 +1.0 +2.0
Magnolia obovata +1.0 +1.5 +3.0 +4.0
4 to 7 8 to 11 12 to 20 21 to 30
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

8. Remarks on measurements
Measuring values may disperse according to measuring circumstances.
Electric Resistor Type Moisture meters use the applications of sensitively electric characteristics of measuring
objects against the moisture contents and the corresponding relations never be a perfectly correlative one- to- one.
In case of measuring wood barks, grains, faces, knots ,albeit the same material , or woods with changes in quality,
salty, contamination, and furthermore with special processing, chemical treatments, it may be necessary to avert
measuring such unfavorable objects, and compensate or average measured values. As for compensating, it is, in
general, based on moisture measurement methods in accordance with a drying process.
(refer to previous page for compensate the kind of wood)
9. Dry batteries
9-1 Indication of Voltage-drop
Batteries run short when a battery mark indicating the state of charge of the batteries turns to 1 on the left lower
side of LCD ⑦. By sliding open the lid of the battery case ⑩placed on the backside of a main body, replace all
the old batteries with new ones (watch battery life). Alkali dry battery(LR03) × 4
In case batteries are kept used with a battery mark 1, the mark turns to a battery frame only (no charge of the
batteries) and in Approx. 5 seconds Power will switch OFF, ceasing further uses.
◆Battery residue indication
wellremained lessremained leastremained
Power will switch off in 5 seconds.
Replace them with new ones.
9-2 Batteries while not in use.
・ Batteries drain a minimum even while not in use.
・ In case of keeping a meter unused for more than a month, remove the batteries from the case for storing.
10. Maintenance ・ Preservation
* Clean and wipe off dirt with soft clothes. Pay attention to connector part, electrodes and grooves of electrodes
of a probe and keep them clean and dry.
* Prevent a meter from impulse, direct sunlight, and high humidity.
* Choose a clean and dry place for preservation.
Remove batteries from the case for a long-term preservation.
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

11. Others
11-1 How to use moisture reading checker (option)
Touch the electrode to the checker illustrated below when Power is ON.
It is a good condition if indicated values are as per listed on the checker.
If indicated values deviate, contact a store or our sales office to get it corrected. (subject to pay)
* Turn a Temperature Compensating Function to OFF to use the checker.
Refer to「7 About Temperature compensating」at page 10 in details.
* Moisture Reading Checker has 2 different checking points in a front side and a back side.
Use the checker with confirmation of the object to be checked and a measuring mode without fail.
11-2 Needle electrode cover (Attaching and detaching of Needle electrode cover)
As illustrated below, insert the needle electrode cover to the electrode by pinching the both side of the cover with
fingers. If it is tight, pick the cover by pushing it against the needle electrode.
Note: Please align the cover in line with the electrodes to push together for the insertion.
For detaching just remove the cover from the electrode by pinching it with fingers.
Unnecessary, unnatural force may cause damages. Attach or detach the cover with full care.
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925

11-3 Replacement of needle electrodes
The needle electrode can be replaced loosing a Hexagon socket set screw for needle fixing with a furnished
accessory Hexagon wrench.
Make sure that the hexagon socket set screw is set and screwed to strike a notch face of the needle electrode,
otherwise it may cause the electrode to loose and drop off the body.
Storage:
This moisture meter incorporates a Liquid Crystal Display (LCD).
If the display is heated above 50℃(120°F) it may be damaged.
This can happen if the moisture meter is left in a car parked in strong sunlight.
Needle electrode
Hexagon wrench
Hexagon socket set screw
Notch
Hexagon socket set screw for needle fixing
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting) Fax: 021-8690 6777
Mobile: +62 816 1740 8925
Table of contents
Other ALAT UJI Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Clavis
Clavis Type 7 user manual

YOKOGAWA
YOKOGAWA CW500 user manual

FORENSICS DETECTORS
FORENSICS DETECTORS FD-311 user manual

S+S Regeltechnik
S+S Regeltechnik THERMASGARD RGTM1 Series Operating Instructions, Mounting & Installation

BEHA
BEHA Unitest SF 100 Instuction manual

Eaton
Eaton Power Xpert PXM 4000 User and installation manual