
10
DO
9.5 DISSOLVED OXYGEN CALIBRATION
9.5.1 The DO sensor comes empty without electrolyte. Please remove
the Membrane Cap and 70% fill with electrolyte. Replace
Membrane Cap fully closed. Cap should not be loosened and
re-tighten again as this will affect its proper operation.
9.5.2 This meter employs a galvanic Dissolved Oxygen sensor. The
sensor is required to be polarized for 90 minutes upon first time
connected to the meter with batteries powering it. After which,
there is no need for polarization. This polarization sequence is
required every time whenever probe is disconnected from the
meter or when batteries are removed
9.5.3 If E03 display during polarization, it is normal. Once sensor is fully
polarized, display will return to normal.
9.5.4 Polarization sequence is also required every time the electrolyte
or membrane is changed.
9.5.5 While the DO sensor is an electro-chemistry sensor, it consists of a
semi-permeable membrane and a chamber filled with
electrolyte. This makes it volatile and requires calibration every
time before measurements.
9.5.6 Calibration cannot be performed whenever the low battery sign
. appear or while E03 appear during calibration. Please
change batteries before proceeding.
9.5.7 Calibration can be easily performed in air. Make sure the
environment has good air flow and not crowded and only at
temperature of within 10°C of the measured temperature in the
liquid. Example: If measure liquid is 15°C, air temperature must
be between 5 to 25°C.
9.5.8 Switch on unit and remove the Sensor Cover.
9.5.9 Press key to set display to select icon flashing and
press to display reading % in display panel. Calibration
cannot be performed in ppm mode.
9.5.10 Hold the probe in air with sensor pointing downward, wait for the
DO reading to stabilize fully for 10 seconds. Meaning reading
remain the same for 10 seconds without changes in a digit.
9.5.11 Hold down key to enter calibration mode.
9.5.12 Percentage reading on the Dissolved Oxygen display panel will
flash while Temperature Display Panel will alternate displaying
and the measured temperature reading.
9.5.13 Wait another 10 seconds and press to confirm. Display will
momentarily show 100% and return to normal measuring mode.
9.5.14 Calibration is completed. On subsequent measurement,
recalibrate as needed if display does not show desire % in air.