alerta sensoflo-hybrid ALT-HYB2/05 User manual

Instruction
manual
UK Version
ALT-HYB2/05
Sensaflo-Hybrid Mattress c/w
Evacuation & SensaCare Pump
www.alertamedical.com
PAGE 02
Alerta Medical is a world leading designer and manufacturer of innovative, cutting-edge
alternating pressure relieving mattress systems for the healthcare market, delivering
solutions for care providers to proactively maximise patient safety and comfort.
Operating Principles
• The Alerta Sensaflo-Hybrid can be used just as
a high specification foam pressure redistribution
mattress, and also functions as a balanced-air
self-adjusting system with foam-filled air cells, and
the option to add a choice of pumps to provide
alternating pressure relief, and a built in Alerta
Evacuation System of handles and straps.
• Use as static or dynamic system, no requirement
to remove patient from bed to change from foam
system to alternating air system. Alerta Sensaflo-
Hybrid also boasts a unique adaptable heel zone
with gel-foam pads, individually removable to
reduce to zero pressure on the heel as required for
unrivalled pressure redistribution and
improved circulation.
• Alternating pressure at a sufficient rhythm avoids
prolonged vascular compression which is likely to
cause tissue hypoxia.
• The patient is an intended operator. The patient can
use the device and its accessories according to this
manual.
Guidelines For Use
• The traditionally accepted method of pressure sore
control is to turn the client at regular intervals and
to observe the areas of risk regularly.
• Alternating pressure mattress is intended to
compliment this treatment, Observation of the
client and turning when necessary should still be
considered as part of the treatment.
• Always consult a physician or health professional
before using this device.
Control Panel Features
• Pressure adjustable knob easy operation for setup
of patient weight (max patient weight setting
250kgs).
• Auto pressure control through air sensor and
microcomputer control.
• 10mins, 15mins, 20mins, 25mins cycle time options.
• Static, alternate, max firm and lock modes.
• Low pressure, power failure and alternate failure
warning lights and alarm.
Installation/Use
1. Unpack the mattress and place on bed
with the feet logo at the foot end.
2. Hang the compressor (pump) on the
bed using the hooks.
3. Connect 2-pipe connector to end of mattress
4. tubing (supplied in pump box)
5. Connect the mattress feed pipes to
the compressor (pump) connector by
matching the appropriate outlets.
6. Connect the electrical power supply cable to
the compressor (pump).
7. Guide the electrical power supply cable from
the compressor (pump) along the bed to the
plug socket and plug in.
8. Press switch to turn on the compressor (pump).
9. The pump will enter static mode initially until
pressure reaches 15mmHg (about 30 minutes),
then it will change to alternate mode and
keep inflating the mattress to the configured
pressure. Once configured pressure is reached,
the pump will start functioning normally.
10. Select the pressure corresponding to the
patient’s morphology.
11. For nursing and caring convenience, press the
max firm button to automatically inflate the
mattresses to the maximum level. The pressure
will return to alternate mode after 20 minutes.
12. Max firm inflates whole mattress to max level.
Static is different, static ceases alternate
function but allows pressure to still be selected.
Lock button
1. Auto: Control panel will automatically
lock in 5 minutes without operation.
2. Manual: press lock button for 2 seconds to lock
the panel, press again for 2 seconds to unlock the
panel.
SensaCare Truflo Technology™ – constant auto pressure sensing through an air sensor and
microcomputer regulator to provide ultra-low interface pressure and promote an even ambience
for natural sleep patterns and stable blood circulation with Ultra-Silent™ pump operation.
Making cutting-edge medical equipment affordable PAGE 03
Memory Function
After Power Failure
If the power supply is stopped and the switch is
ON position, when power resumes, the pump will
resume to the configured mode before power
failure. However if the pump is in Static or Max
Firm mode, the pump will be reverted to Alternate
Mode after re-switching on the pump.
Emergency CPR Deflation
In the event of an emergency, remove the
mattress pipe connector block from the
pump and turn off the compressor (pump).
The air will discharge from the mattress with
the patient’s own weight. To resume normal
operation follow the above steps.
Safety Measures
• DO TAKE CARE when operating
this electrical equipment.
• DO NOT leave connected to
electrical supply when not in use.
• DO NOT expose the equipment
to excessive humidity.
• DO NOT use corrosive cleaning products such
as industrial degreasers or acetic solvents.
• DO NOT attempt to relocate this equipment whilst
connected to the electrical supply.
• DO NOT position the equipment to make it
difficult to operate the power plug used to
isolate the equipment circuits electrically from
the mains supply.
• WARNING To avoid the risk of electric shock, this
equipment must only be connected to a mains
supply with protective earth.
Cleaning/Disinfection
Air mattress system must be cleaned thoroughly
between patients to avoid cross contamination.
• DO TAKE CARE to disconnect from the
electrical supply before attempting to clean.
• DO TAKE CARE to use a cloth impregnated
with a mild detergent solution.
• DO TAKE CARE to follow manufacturers
solution concentrations
• DO TAKE CARE not to exceed manufacturer’s
recommendations before wiping off cleaning
solution.
• Clean using a single use wipe or damp cloth
and a solution of mild surface detergent /
disinfectant with CE mark and neutral water.
Rinse thoroughly with clean water. If the cover
is heavily soiled a 0.1% (1,000ppm) chlorine
solution and cold water can be used. Where
required, covers can be laundered in the
washing machine as follows; main wash 65°C
for not less than 10 minutes or 95°C for not less
than 3 minutes.
• Drying Covers can be tumble dried on low heat
setting. Drying temperature must not exceed
60°C.
Service & Warranty
This device is not self-serviceable.
Maintenance must ONLY be performed by
an authorised Technician. Failure to use genuine
Alerta parts will void the warranty of this product.
We warrant the product to be free from
manufacturing defects for the period of
1 year from date of purchase.
We will make available upon request; component
part lists, descriptions, or other information that will
assist service personnel to repair those parts of ME
equipment that are designated by the manufacturer
as repairable by service personnel.
Environmental Protection
The device and accessories are to be disposed of
according to local regulations after their service lives.
Alternatively, they can be returned to the dealer or
the manufacturer for recycling or proper disposal.
Mattress Characteristics
Specification:
Max. User Weight: 250 Kgs
Mattress Dimensions: 200x90x14cm
uninflated / 200x90x16cm inflated
Materials: TPU cell, nylon PU
multi-way stretch cover
Pillow function
Compressor (pump)
Characteristics
Specification:
Cycle Time: 10/15/20/25 mins
Air Output: 8L/min
Electrical Power Supply:
220-240V - 50-60Hz
Warranty: 12 months unless
otherwise specified
Making cutting-edge
medical equipment
affordable
www.alertamedical.com
Head Offi ce
Alerta Medical
4 Symington Place
Riverside Business Park
Irvine
KA11 5DE
United Kingdom
EU 2017/745
Rev: 2021.10
CS Lifesciences Europe Ltd
The Black Church
St. Mary’s Place
Dublin 7
D07 P4AX
Ireland
eurep@cslifesciences.com
EC REP

Other alerta Indoor Furnishing manuals

alerta bariatric 2 User manual

alerta

alerta bariatric 2 User manual

alerta sensaflex 3000 User manual

alerta

alerta sensaflex 3000 User manual

Popular Indoor Furnishing manuals by other brands

Hooker Furniture 366-75-008 Assembly instruction

Hooker Furniture

Hooker Furniture 366-75-008 Assembly instruction

Northern Pal user manual

Northern

Northern Pal user manual

Habitat Parker Extending Table Assembly instructions

Habitat

Habitat Parker Extending Table Assembly instructions

Dorel Asia DA6126FQ Instruction booklet

Dorel Asia

Dorel Asia DA6126FQ Instruction booklet

Furniture of America CM7182D Assembly instructions

Furniture of America

Furniture of America CM7182D Assembly instructions

Powell 16D8216CSBLU Assembly instructions

Powell

Powell 16D8216CSBLU Assembly instructions

Ticaa Tolly Instructions for use

Ticaa

Ticaa Tolly Instructions for use

Room essentials 417087 Assembly instructions

Room essentials

Room essentials 417087 Assembly instructions

World Market 551917 Assembly instructions

World Market

World Market 551917 Assembly instructions

Regency ONRTSP4224 instructions

Regency

Regency ONRTSP4224 instructions

Bassett Inspired CHAPMAN BP-CHTX operating instructions

Bassett

Bassett Inspired CHAPMAN BP-CHTX operating instructions

deVRIES Willington 40465L Assembly instructions

deVRIES

deVRIES Willington 40465L Assembly instructions

Greentouch 1495KDI-28-247 Assembly / care & use instructions

Greentouch

Greentouch 1495KDI-28-247 Assembly / care & use instructions

Hodedah HI29-A Assembly instructions

Hodedah

Hodedah HI29-A Assembly instructions

Unfinished Furniture of Wilmington HAMPTON OT-2013X Assembly instructions

Unfinished Furniture of Wilmington

Unfinished Furniture of Wilmington HAMPTON OT-2013X Assembly instructions

Noble House Home Furnishings Chair Assembly instructions

Noble House Home Furnishings

Noble House Home Furnishings Chair Assembly instructions

Furinno 13192 Assembly instruction

Furinno

Furinno 13192 Assembly instruction

Santalucia Mobili QUADRA California Closet manual