Model 1000+ Printer – User Manual
1. ABOUT THE 1000+P AUDIOMETER
The AMBCO Model 1000+P is a microprocessor controlled pure tone air conduction
audiometer with automated screening and automated threshold test feature.
2. FEATURES AND SPECIFICATIONS OF THE MODEL 1000+P
Meets performance requirements of ANSI S3.6-1996
OPERATING TEMPERATURE: 15° to 30°C (59°F to 86°F)
PRODUCT TYPE: Pure Tone, Type 4 Audiometer
TEST FREQUENCIES
AUTOMATIC THRESHOLD MODE (HZ): 500, 1000, 2000, 3000, 4000, 6000, 8000 (with
validity check at 1000 Hz)
AUTOMATIC SCREENING MODE (HZ):
TEST 1: 500, 1000, 2000, 4000
TEST 2: 1000, 2000, 3000, 4000
MANUAL MODE: 250, 500, 1000, 2000, 3000, 4000, 6000, 8000
FREQUENCY ACCURACY: Less than 3% – Typically better than 2%
TOTAL HARMONIC DISTORTION: Less than 2.5%, typical 1%
HEARING LEVEL RANGE: 0 to 90 dB in 5 dB increments
POWER SUPPLY: 107-127 VAC 60 Hz, 10 Watts, UL approved (Optional 220
VAC 50/60 Hz available)
AC Adapter Output: 12V DC nominal at 500mA
HEARING LEVEL ACCURACY /Attenuator linearity: ± 1 dB
STIMULUS: Continuous, pulse or warble
Rise/fall time: 20 - 50 ms
DATA OUTPUT: RS-232 9600 Baud, 8 bits, no parity, DTR handshake
EARPHONES: TDH-39 or DD45, 10 OHM
CASE: Injection molded, impact resistant A.B.S. plastic
DIMENSIONS: Width 10 inches, Length 8 inches, Height 4.2 inches
TOTAL WEIGHT: 4.0 Lbs.
WARRANTY: 5 Years on base system
PRINT TIME: Typically less than 20 seconds (printer dependent)
3. SETTING UP THE AMBCO MODEL 1000+P with the Printer
3.1. Connect the printer cable securely to the audiometer and to the printer by
completely tightening all four thumbscrews, two thumbscrews on each end of cable.
Connect the printer AC adapter to the printer and the audiometer AC adapter to the
audiometer recheck to be sure that both adapter plugs are fully inserted into each
jack.
3.2. Locate the red & blue plugs on the cord connected to the headset and the black plug
on the cord connected to the patient response switch. The three plugs need to be
10-1004, Rev. A DCO 17–008, 11–13–17 Page 3 of 9