Installation Guide
Publication Nº 805384
Issue 14
Language: English
09 January 2017

Health and Safety Information
Read all of the instructions in this booklet - including all the WARNINGS and CAUTIONS
- before using this product. If there is any instruction which you do not understand, DO
NOT USE THE PRODUCT.
Safety Signs
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or
personal injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in minor or
moderate injury to the user or users, or result in damage to the product or to property.
NOTE
Indicates a potentially hazardous situation which, if not avoided, could result in damage or the loss of
data.
Signs and Symbols used on equipment and Documentation
Caution, risk of electric shock.
Caution, attention to possibility of risk of damage to the product, process or surroundings. Refer
to instruction manual.
Caution, hot surface.
Protective Conductor Terminal.
Observe precautions for handling electrostatic discharge sensitive devices.
Equipment Operation
Use of this instrument in a manner not specied by Land Instruments International may be hazardous.
Read and understand the user documentation supplied before installing and operating the equipment.
The safety of any system incorporating this equipment is the responsibility of the assembler.
Protective Clothing, Face and Eye Protection
It is possible that this equipment is to be installed on, or near to, machinery or equipment operating at
high temperatures and high pressures. Suitable protective clothing, along with face and eye protection
must be worn. Refer to the health and safety guidelines for the machinery/equipment before installing
this product. If in doubt, contact Land Instruments International.
Electrical Power Supply
Before working on the electrical connections, all of the electrical power lines to the equipment must
be isolated. All the electrical cables and signal cables must be connected exactly as indicated in these
operating instructions. If in doubt, contact Land Instruments International.
Storage
The instrument should be stored in its packaging, in a dry sheltered area.
Unpacking
Check all packages for external signs of damage. Check the contents against the packing note.
Lifting Instructions
Where items are too heavy to be lifted manually, use suitably rated lifting equipment. Refer to the
Technical Specication for weights. All lifting should be done as stated in local regulations.
IMPORTANT INFORMATION - PLEASE READ

Health and Safety Information
Read all of the instructions in this booklet - including all the WARNINGS and CAUTIONS
- before using this product. If there is any instruction which you do not understand, DO
NOT USE THE PRODUCT.
Safety Signs
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or
personal injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in minor or
moderate injury to the user or users, or result in damage to the product or to property.
NOTE
Indicates a potentially hazardous situation which, if not avoided, could result in damage or the loss of
data.
Signs and Symbols used on equipment and Documentation
Caution, risk of electric shock.
Caution, attention to possibility of risk of damage to the product, process or surroundings. Refer
to instruction manual.
Caution, hot surface.
Protective Conductor Terminal.
Observe precautions for handling electrostatic discharge sensitive devices.
Equipment Operation
Use of this instrument in a manner not specied by Land Instruments International may be hazardous.
Read and understand the user documentation supplied before installing and operating the equipment.
The safety of any system incorporating this equipment is the responsibility of the assembler.
Protective Clothing, Face and Eye Protection
It is possible that this equipment is to be installed on, or near to, machinery or equipment operating at
high temperatures and high pressures. Suitable protective clothing, along with face and eye protection
must be worn. Refer to the health and safety guidelines for the machinery/equipment before installing
this product. If in doubt, contact Land Instruments International.
Electrical Power Supply
Before working on the electrical connections, all of the electrical power lines to the equipment must
be isolated. All the electrical cables and signal cables must be connected exactly as indicated in these
operating instructions. If in doubt, contact Land Instruments International.
Storage
The instrument should be stored in its packaging, in a dry sheltered area.
Unpacking
Check all packages for external signs of damage. Check the contents against the packing note.
Lifting Instructions
Where items are too heavy to be lifted manually, use suitably rated lifting equipment. Refer to the
Technical Specication for weights. All lifting should be done as stated in local regulations.
IMPORTANT INFORMATION - PLEASE READ
Contact Us
UK - Droneld
Land Instruments International
Tel: +44 (0) 1246 417691
Email: [email protected]
Web: www.landinst.com
USA - Pittsburgh
AMETEK Land, Inc.
Tel: +1 412 826 4444
Web: www.ametek-land.com
China
AMETEK Land China Service
Tel: +86 21 5868 5111 ext 122
Email: [email protected]
Web: www.landinst.com
India
AMETEK Land India Service
Tel: +91 - 80 67823240
Email: [email protected]
Web: www.landinst.com
For further details on all AMETEK Land oces, distributors and representatives, please visit
our websites.
Return of Damaged Goods
IMPORTANT If any item has been damaged in transit, this should be reported to the carrier and to the
supplier immediately. Damage caused in transit is the responsibility of the carrier not the supplier.
DO NOT RETURN a damaged instrument to the sender as the carrier will not then consider a claim. Save
the packing with the damaged article for inspection by the carrier.
Return of Goods for Repair
If you need to return goods for repair please contact our Customer Service Department. They will be able
to advise you on the correct returns procedure.
Any item returned to Land Instruments International should be adequately packaged to prevent damage
during transit.
You must include a written report of the problem together with your own name and contact information,
address, telephone number, email address etc.
Design and Manufacturing Standards
The Quality Management System of Land Instruments International is approved to BS EN ISO 9001 for
the design, manufacture and on-site servicing of combustion, environmental monitoring and non-contact
temperature measuring instrumentation.
Approvals apply in the USA
Operation of radio transmitters, telephones or other electrical/electronic devices in close proximity
to the equipment while the enclosure doors of the instrument or its peripherals are open, may cause
interference and possible failure where the radiated emissions exceed the EMC directive.
The protection provided by this product may be invalidated if alterations or additions are made to the
structural, electrical, mechanical or pneumatic parts of this system. Such changes may also invalidate the
standard terms of warranty.
Copyright
This manual is provided as an aid to owners of Land Instruments International’s products and contains
information proprietary to Land Instruments International. This manual may not, in whole or part, be
copied, or reproduced without the expressed written consent of Land Instruments International Ltd.
Copyright © 2016 Land Instruments International.
MARCOM0311, Issue 7, 01 November 2016
Approvals apply in India

Blank

Preface
Installation GuideLSP-HD Mountings & Accessories
Blank

Page 1
Installation Guide LSP-HD Mountings & Accessories
1 Introduction
This document provides installation details for the range of mountings and
accessories for Landscan LSP-HD Scanners.
Note
The equipment must be used, maintained and serviced by properly
trained personnel, capable of following the procedures given in these
instructions. Safety may be impaired if the system is not used as
indicated.
It is important to check the equipment with which you have been supplied and
then read these instructions thoroughly before proceeding with operation.

Page 2
Installation GuideLSP-HD Mountings & Accessories
2 System Overview
It is important to check the equipment with which you have been supplied and
then read these instructions thoroughly before proceeding with operation.
The Landscan mounting accessories comprise:
• Basic Mounting Plate (Land Part Nº 031.520)
• Right Angle Mounting Bracket (Land Part Nº 031.438)
• Water Cooled Air Purge Baseplate Kit (Land Part Nº 031.496)
• Plate Mounted Blower Unit (Land Part Nº 099.471)
• LSP Purge / Cooling System (Land Part Nº 800095)
• Light Floatline Housing (Land Part Nº 802663)
• Floatline Housing System (Land Part Nº 092.732)
• Light Steel Housing (Land Part Nº 802664)
• Heavy Steel Housing (Land Part Nº 802665)
The Lanscan LSP-HD power supply accessories comprise a choice of:
•
IEEE802.3at Power over Ethernet Desktop Power Supply (Land Part Nº 805024)
•
IEEE802.3at Power over Ethernet Power Supply for Cabinet Installation (Land
Part Nº 807105)
• Landscan LSP-HD Power Supply Unit (PSU) (Land Part Nº 805037)
• Landscan LSP-HD Floatline Service Panel (Land Part Nº 805140)
The LSP-HD cables (between the scanner head and the power supply accessory
comprise a choice of:
• 15m High Temperature Cable
(Land Part Nº 804977)
• 15m Cat 5e High Quality Ethernet Cable
(Land Part Nº 805038)
• 50m Cat 5e High Quality Ethernet Cable
(Land Part Nº 805039)
• Floatline Housing Cable Assembly
Power Supply
Accessory
(Optional)
Plate Mounted
Blower unit
Mounting Accessory
LSP-HD
Scanner
LSP-HD
Cables
Floatline
services cable
(Optional)
PSU I/O
Processor
Process Control
PC (WCA)
Landscan
Configuration
Software
Customer Connections

Page 3
Installation Guide LSP-HD Mountings & Accessories
The LSP-HD I/O Processors comprise a choice of:
• LSP-HD I/O Processor 1 (Land Part Nº 805773) – 4 digital in, 4 digital out,
2 analogue in, 4 analogue out
• LSP-HD I/O Processor 2 (Land Part Nº 805774) - 8 digital in, 12 digital out,
4 analogue in, 12 analogue out
• LSP-HD I/O Processor 3 (Land Part Nº 805775) – Customisable I/O
• LSP-HD I/O Processor for Single Floatline Measurement (Land Part Nº
805778)
• LSP-HD I/O Processor for Four Floatline Measurement (Land Part Nº
805779)
• Power Supply Unit for LSP-HD I/O Processor (Land Part Nº 805953)
The LSP-HD software accessories comprise a choice of:
• LandscanProcessorCongurationSoftware(fromversion3.51)
• Landscan WCA Software (from version 5.7)
2.1 LSP-HD Cable assembly
Note
When mounting an LSP-HD scanner in either the Floatline Housing, Light
Steel Mounting or Heavy Steel Mounting, you must disassemble and
remove the L-shaped connector from the LSP-HD cable, thread the cable
through the cable gland on the mounting and re-assemble the connector.
There are two types of connector used with LSP-HD:
• Phoenix Connector (See Section 2.1.1)
• Binder Connector (See Section 2.1.2)
2.1.1 Phoenix Connector disassembly
1) Unscrew the bezel that grips the outer sheath of
the cable.
2) Slide the bezel back along the length of the ca-
ble a little.

Page 4
Installation GuideLSP-HD Mountings & Accessories
3) Unscrew the innermost knurled ring
from the main body of the connector.
4) Slide the main body of the connector
down the length of the cable a little.
5) Note down the wiring schedule of the
four conductors, then using a small
atobjectsuchasascrewdriver,
prise up the securing mechanism
to release the conductors from the
connector.
6) Remove the connector from the cable
including both the connector main
body and the bezel that grips the
outer sheath.
7) Thread enough cable through the
relevant cable gland in the LSP-HD
mounting hardware and re-assemble
the connector on the other side of
the cable gland by working through
steps 1-6 in reverse order.

Page 5
Installation Guide LSP-HD Mountings & Accessories
2.1.2 Binder Connector disassembly
1) Unscrew the threaded clamp ring
which holds the cable and terminal
block in the main body of the
connector.
2) Unscrew the inner metal clamp ring
and slide the plastic clamp and inner
rubber washer along the cable.
Remove the lid of the connector
shell. Make sure that you retain the
thinrubbersealforretting.
2) The four wires in the end terminal block are in two twisted pairs: A Black/
White pair and a Black/Red pair.
Unscrew the terminals to disconnect the wires, making sure that you retain
them in the correct twisted pair.
3) Remove the connector shell, clamp rings, washers etc. from the cable and
pass the free end of the cable through the cable gland on your mounting
accessory.
4)
White/Black twisted pair
Red/Black twisted pair
View on front of connector
Correct orientation
of notch
1
2
3
4
Re-connect the wires to the terminal
block according to the cable schedule
below:
Terminal Nº Colour
1 Black 1
2 White 1
3 Black 2
4 Red 2
1 - Twisted pair 1
2 - Twisted pair 2

Page 6
Installation GuideLSP-HD Mountings & Accessories
5)
Orientation of notch on connector
View inside rear of connector
When re-assembling the connector,
make sure that the terminals are
orientated so that the locating notch
in the connector is in the position
shown in the illustrations.
6) Reposition the lid of the connector shell, ensuring the thin rubber seal is
seated correctly on the lid.
7) Tighten both of the metal threaded metal clamp rings onto the connector
body to complete the re-assembly.
ThecablecannowbettedtotheLandscanLSP-HDThermometer.

Page 7
Installation Guide LSP-HD Mountings & Accessories
3 Summary of Specifications
Blower (Land part Nº 099.471)
Operating Temp Range: -40 to 50°C / -40 to 120°F
Sealing: IP65withcovertted
Weight: 28.5 kg / 63lb
Refer to the Blower Unit User Guide for full details
Basic Mounting Plate (Land Part Nº 031.520)
Operating Temp Range: 5 to 60°C
Water Cooled Air Purge Baseplate Kit (Land Part Nº 031.496)
Operating Temp Range: 5 to 100°C
LSP Purge / Cooling System (Land Part Nº 800095)
Operating Temp Range: 5 to 120°C
Light Floatline Housing (Land Part Nº 802663)
Operating Temp Range: 5 to 150°C
Floatline Housing System (Land Part Nº 092.732)
Operating Temp Range: 5 to 150°C
Light Steel Housing (Land Part Nº 802664)
Operating Temp Range: 5 to 150°C
Heavy Steel Housing (Land Part Nº 802665)
Operating Temp Range: 5 to 150°C
Desktop Power Supply (Land Part Nº 805024)
Operating Temp Range: 0 to 45°C
Input voltage range: 100 to 240VAC
AC Input Frequency: 50 to 60Hz
Refer to the manufacturer's datasheet for full details
Power Supply for Cabinet Installation (Land Part Nº 807105)
Operating Temp Range: 0 to 60°C
Input voltage range: 100 to 240VAC
AC Input Frequency: 50 to 60Hz
Refer to the manufacturer's datasheet for full details

Page 8
Installation GuideLSP-HD Mountings & Accessories
Landscan LSP-HD Power Supply Unit (PSU) (Land Part Nº 805037)
Operating Temp Range: 5 to 50°C
Voltage range: 100 to 240V
Frequency range: 50 to 60Hz
Sealing: IP65
Environment: Indoor use only
Warning
The PSU is designed to be mounted vertically on a wall or other
appropriate structure. Mounting the unit in any other orientation
mayincurariskofpersonalinjury.
Landscan LSP-HD Floatline Service Panel (Land Part Nº 805140)
Operating Temp Range: 5 to 50°C
Maximum Power: 1.4kW (including blower)
Voltage Range* 100-240V AC
Frequency range* 50-60Hz
2.4A max excluding blower
Sealing: IP65
Environment: Indoor use only
Blower Connection (optional): 110, 220, 240VAC
Warning
The safe operation of the service panel is only guaranteed when
Land-speciedblowersareconnectedtotheblowerterminals.
*Note -withblowertted,thesupplyvoltageandfrequencyis
restricted to that of the blower.
The Service Panel is designed to be mounted vertically on a wall
or other appropriate structure. Mounting the unit in any other
orientationmayincurariskofpersonalinjury.
Landscan LSP-HD I/O Processor
Input voltage: 24VDC
Incoming Data Connection: RJ45 Ethernet
Alarms: System Alarm, Remote Client Alarm
Analogue Outputs: Max. 48 on 12 modules
Digital Outputs: Max. 16 on 4 modules
Analogue Inputs: Max. 16 on 8 modules
Digital Inputs: Max. 16 on 4 modules
Default IP Address: 10.1.10.110
Subnet: 255.255.00

Page 9
Installation Guide LSP-HD Mountings & Accessories
4 Maintenance
Warning
The safe operation of the service panel is only guaranteed when
Land-speciedblowersareconnectedtotheblowerterminals.
The body of the accessories and mountings can be cleaned by wiping with a
clean cloth moistened with either water or a mild, neutral detergent.
DO NOT use alcohol, chemicals or abrasive cleaning products.
Hosesandpipesshouldalsobecheckedforleaks,cracks,changandother
wear and tear.

Page 10
Installation GuideLSP-HD Mountings & Accessories
Blank

Page 11
Installation Guide LSP-HD Mountings & Accessories
5 Installation Diagrams
This section of the Installation Guide contains a series of installation drawings,
covering the installation of the LSP-HD and its mountings, accessories and
services.
5.1 System Components
740-069 Light Floatline System
Component Land Part Nº
Power Supply Unit 805037
• PSU (Mechanical Installation Diagram) 740.066 (Sheet 1)
• PSU (Electrical Installation Diagram) 740.066 (Sheet 2)
Blower 099.471 / 099.472 / 099.473
Light Floatline Housing 802.663
740-064 Full Floatline System
Component Land Part Nº
Service Panel 805140
• Service Panel (Mechanical Installation Dia-
gram)
740.065 (Sheet 1)
• Service Panel (Electrical Installation Dia-
gram)
740.065 (Sheet 2)
Blower 099.471 / 099.472 / 099.473
Floatline Housing 092.732
740-067 Heavy Steel System
Component Land Part Nº
Power Supply Unit 805037
• PSU (Mechanical Installation Diagram) 740.066 (Sheet 1)
• PSU (Electrical Installation Diagram) 740.066 (Sheet 2)
Blower 099.471 / 099.472 / 099.473
Heavy Steel Housing 802.665
740-068 Light Steel System
Component Land Part Nº
Power Supply Unit 805037
• PSU (Mechanical Installation Diagram) 740.066 (Sheet 1)
• PSU (Electrical Installation Diagram) 740.066 (Sheet 2)
Blower 099.471 / 099.472 / 099.473
Light Steel Housing 802.664

152.0mm [5.984"]
240.0mm [9.449"]
176.0mm [6.930"]
276.0mm [10.866"]
HOLE Ø8.5mm [0.335"]
TO ACCEPT
M8 SCREWS
22.5mm
[0.886"]
52.5mm
[2.066"]
80.0mm [3.150"]
20.0mm
[0.787"]
71.0mm
[2.795"]
61.0mm
[2.401"]

740.033
740.033
A
INSTALLATION DRAWING RIGHT ANGLED MOUNTING BRACKET
N.T.S.
LAE 16/11/04
2004
#1132 LSP ACCESSORIES
3 OF 4
0.1
1°
0.4
1.6
0.25
0.50
A
B
C
D
12 3 45 6
THIS DRAWING MUST NOT BE COPIED OR USED OTHER THAN FOR
THE PURPOSE FOR WHICH IT WAS ISSUED WITHOUT THE WRITTEN
PERMISSION OF:-
DRG. TITLE
MATERIAL
FINISH
DRAWN
CHECKED
MODIFICATION DATE ISSUE
DATE
DRG.No.
PROJECTION
TOLERANCES U.O.S.
DIMNS. ±
ANGLES ±
CASTING ±
SURFACE FINISH
ALL SHARP CORNERS REMOVED
INTL. CORNERS R
EXTL. CORNERS R
APPROVED DATE
ORIGINAL
SCALE
DRG.No.
DRG.TO
BS 308
DIMENSIONS
ARE IN
MILLIMETRES
LAND INSTRUMENTS INTERNATIONAL LTD
DRONFIELD
SHEFFIELD
S18 1DJ
C
DO NOT SCALE
SHEET
ORIGINALLY USED ON PRODUCT TYPE
Revision
222.0mm [8.74"]
52.0mm
[2.047"]
24.0mm
[0.945"]
4 HOLES Ø8.5mm [0.335"]
TO ACCEPT M8 SCREWS
TO SECURE TO PLANT
100.0mm
[3.937"]
184.0mm[7.244"]
240.0mm [9.449"]
152.0mm[5.984"]
174.0mm [6.85"]
286.0mm [11.26"]
20.0mm
[0.787"]
4 HOLES Ø8.5mm [0.335"]
TO ACCEPT M8 SCREWS
FOR MOUNTING SCANNER