Amos ATP200 User manual

ATP200
CAROUSEL TISSUE PROCESSOR
Operation Manual
www.amos-scientific.com
AMOS SCIENTIFIC PTY.LTD. ABN37 159 778 140
ATP200 CASOUSEL TISSUE PROCESSOR
----1--
Ver:AMOSATP200OP20151127
Table of content:
1- Product description
2- Technical specifications of the product
3- Product parts
4- Installation instructions and product use
5- Cleaning
6- Safety and warnings
7- Maintenance and possible problems and solutions
8- Warranty
9- Symbols
Congratulations!
This product is the result of many years of research and development so that you have a product
with excellent quality.
We are glad for your appreciation of AMOS that throughout the years has been gradually
dedicating itself to deliver trustworthy products and which care for the expectations of our
Customers.
It is indispensable to read this manual before switching on any product operation.
NOTE: Every procedure performed by the automatic tissue processor must be
accompanied by a qualified operator. AMOS is not responsible for damage to the
processed material caused by not monitoring or no observation to the comments of this
manual.
1-) Product description:
CAROUSEL TISSUE PROCESSOR –MODEL ATP200 was developed for use in Pathological
Anatomy Laboratory. It is indispensable for the automatic preparation of histological tissue
sample, teeth, bones and other materials.
The product provides:
•Procedure Manual
•Warranty certified
•Power cable
•Spare fuse
•03 stainless steel baskets with capacity for 85 cassettes each
•10 beakers (glassware) for chemical reagents and 01 spare
•02 paraffin mugs
•01 no-break
•500 cassettes
2-) Techniques specifications of the product:
•Voltage: 100-240 V/50-60 Hz (Optional - without selector switch. The voltage choice
must be made in the act of the purchase).
•Input Maximum Current: 127 V = 2,97A e 220 V = 1,80 A
•Potency 415 W
•Fuse: 20 mm, normal, corrente 10A (127 V e 220 V).
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----2--
•Thermal Protector: Disarm to high temperature when it reaches 212 °F (100 °C).
General shutdown of the equipment. Automatic rearm below 203 °F (95 °C).
•Operation Mode: Continuous
•Protection Against Incoming Water Harmful: IPX0
•Capacity: 2 liter of reagent or paraffin in each Becker and up to 85 cassettes for basket.
•Temperature control: Environment temperature up to 90 °C with a range of up to + -
2 °C through Digital Thermostat.
•Hat diameter: 810 mm
•External diameter: 810 mm
•Base depth: 650 mm
•Base width: 460 mm
•Equipment total height: 70 mm
•Net weight: 68 kg
•Gross weight: 110 kg
•Packing External Dimensions: 910 mm X 910 mm x 970 mm
•Product in agreement with IEC 61010, Class I
•Product in agreement with IEC 61010-2010 and 98/37 Regulations.
3-) Product parts:
4-) Product use instructions:
Read the instructions for using the equipment in the best possible way for
greater durability of the equipment. Pay attention to the precautions that
need to be taken while handling the equipment.
Read the Warranty Terms.
NOTE: This product does not require any technical intervention for their installation. Just
follow the instructions below:
1. Before connecting the product to power source, make sure if the network voltage is the
same of product.
2. Clean this product as determined by this manual.
3. Connect no-break into the power outlet. It is a bivolt no-break and it can be powered on
in 110 V or 220 V.
4. Connect the equipment to one of the no-break outputs.
Front Panel
Paraffin Mugs
Conectors
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----3--
5. Turn on the back key on the equipment.
1- In the front panel display, the following message will appear.
1- TO SET THE CLOCK AND CALENDAR:
*If the date and time aren't correct according to current date, follow the following step.
In the Start Screen, press the [Presets] key
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----4--
In the Screen below, select [Data/Hora]
In the Screen below, click on [+] or [-] and set the Day, Month, Year, Hour and Minute
Click on the BACK ARROW to come back and then this screen appears
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----5--
Click on the BACK ARROW to come back and then this screen appears
2- PARAFFIN WORK TEMPERATURES PROGRAMMING:
*Calibrate the Temperature of Paraffin Mugs in agreement with the Paraffin Temperature to be
used and in agreement with the environmental temperature, so that Paraffin turns into fluid and
without "butterfat" on the surface.
In Start Screen, press the [Presets] key
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----6--
In the Screen below please enter the factory password 1234 and press [Enter] key
In the Screen below select [Prt Gerais]
In the Screen below select [+] or [-] and setting the Desired Temperature of T1 and T2 Mugs.
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----7--
Click on the BACK ARROW to come back and then this screen appears
Click on the BACK ARROW to come back and then this screen appears
3- TO EFFECT A WORK PROGRAMMING:
The equipment is delivered from the factory with the following programmings:
*Program 1: 60-minute bath changes
*Program 2: 30-minute bath changes
*Program 3: 15-minute bath changes
*Program 3: 1-minute bath changes
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----8--
In Start Screen, press the [Presets] key
In the Screen below select [Prt Pgm 1] or any other that you want to modify or make a new
Programming
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----9--
In the Screen below press [+] or [-] and set the Time of each Immersion that will be in
accordance to the used reagents.
Press the Arrow and go to the screen below
Press the Arrow and go to the screen below
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----10--
Press the Arrow and go to the screen below
Click on the BACK ARROW to come back as the following screen appears
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----11--
4- HOW TO LOAD THE BASKETS:
In the Start Screen, press the [Manual] key
In the Screen below press [Sobe], [Desce] or [Gira], according to the desired movement.
[Sobe]: This option will raise the Cover ("Hat") to charge the basket for Processing
[Desce]: This option will descend the Cover ("Hat") to dip the basket and if needed, to start the
Processing
[Gira]: This option will turn the Cover ("Hat") until the basket gets to the Desired Place
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----12--
Click on the BACK ARROW to come back and then this screen appears
5- HOW TO START AN IMMEDIATE PROGRAMMING:
In the Start Screen, press the [Presets] key
In the Screen below select the Program that will be used
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----13--
In the Screen below select if you want to start the Program immediately or with delay
Immediate start
Click on [INÍCIO IMEDIATO] (immediate start) and the Selected Program will be started and
the screen below appears
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----14--
Starting with delay
Touch [INÍCIO RETARDO] (start with delay) as the screen below is shown. Select the day,
month, hour and minute that you want the Selected Program to start.
Click on [OK] Screen and the Screen below will appear indicating which the selected Program
is and when it will start
6- HOW TO RESET THE MANUFACTURES PARAMETERS
In the Start Screen, press the [Presets] key
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----15--
In the Screen below type the password 99 and then press the [Enter] key
In the Screen below choose between yes [SIM] or no [NÃO]
If you chose yes, the next Screen will appear
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----16--
Verified through [SIM] key, the screen below will appear
7- HOW TO CANCEL A CYCLE IN PROGRESS
In the Start Screen, press the [Inicia Ciclo] key
In the Screen below select the Program that will be used
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----17--
In the Screen below select if you want to start the Program immediately or with delay
In the Screen below will appear the following Screen
If you choose to cancel the Cycle, press [ESC] and the following Screen will appear
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----18--
Confirm by pressing [SIM] and the Cycle will be canceled, or press [NÃO] and the cycle will
continue.
5-) Cleaning:
Product surfaces are easy to clean.
To clean the product you need to unplug it from the power and wait for the inner
container cool. For cleaning product and paraffin removal that is deposited on the surface,
use a soft cloth soaked in kerosene.
After cleaning with kerosene, you can apply with soft, moist cloth, and a neutral detergent.
Solvent as Xylene and its derivatives does not have to be used for
cleaning the equipment. This will cause irrecoverable damages to the
Product!
6-) Cautions and Warnings:
Read the product handling instructions carefully and instruct with the technical details before
use.
- Make sure that the product and the laboratory electric current match, and then connect the
product in the power outlet.
- The product tank can get very hot during the process so, beware of burning.
- Change the paraffin always when it is necessary.
The product was designed for using exclusively to process the material!
Risk of fire!
Flammable fuels or substances should not be handled close to the product!
- The Product cannot be used without Paraffin in the Mugs.
- The Paraffin Mugs need to be connected with each other, and in the correct positions.
- For any repair or maintenance, call the Authorized Service Center.
7-) Maintenance and troubleshooting:
- Every 15 days ascend the equipment hat with the panel manual system using the function [1]
[SOBE], drip 3 lube oil drops that came with the Equipment in the felt washer of the central
pipe of ascent, and return the hat to the initial position selecting the function [3] [DESCE].
- For any repair or maintenance, the product must be turned off and unplugged from the power
outlet.
- In the chart below there are some problems which may occur and their causes and solutions.
- If the problem does not fit in these descriptions, please contact immediately our Service Center
and see more details at the end of the manual.
ATP200 CASOUSEL TISSUE PROCESSOR
ATP200 CASOUSEL TISSUE PROCESSOR
Ver:AMOSATP200OP20151127
----19--
Problem
Cause
Solution
Nothing happens when
the key power is on
(ON).
Is the input power cable
connected correctly to the power
outlet?
Check if the input power cable is connected
to the power outlet and to the equipment.
Put the key at ON position.
Check for power into the outlet.
Check if the power ON/OFF key
is on (ON)
Fuse
Remove the fuse from the fuse holder and
check if it is not broken or with the filament
interrupted. If it is necessary, replace it for
similar.
The paraffin either heats or does
not, but does not melt
The wire of mug is broken or with poor
contact.
Sensor with malfunction
Paraffin mug resistor with malfunction
Call the Authorized Service Center.
Error in the movement engine
Burns the Fuse of up/down motor or the
Spinning Motor
Contact our Service Center.
If the power is reestablished the Program
will continue starting from where it was,
with a 3-hour delay, which is enough for
the Paraffin to melt and stabilize.
Power failure during the Program
Execution

Popular Computer Hardware manuals by other brands

EMC2 VNX Series Hardware Information Guide

EMC2

EMC2 VNX Series Hardware Information Guide

Quanmax KEDB-1000 Quick installation guide

Quanmax

Quanmax KEDB-1000 Quick installation guide

Adaptec 7210S Replacement procedure

Adaptec

Adaptec 7210S Replacement procedure

HP MCS-200 user guide

HP

HP MCS-200 user guide

Thermalright Macho 120 manual

Thermalright

Thermalright Macho 120 manual

Sundrax DimGate DIN user manual

Sundrax

Sundrax DimGate DIN user manual

ekwb EK-FC1080 GTX TF6 Backplate INSTALLATION AND MOUNTING MANUAL

ekwb

ekwb EK-FC1080 GTX TF6 Backplate INSTALLATION AND MOUNTING MANUAL

TYAN TS65-B7120 Service engineer's manual

TYAN

TYAN TS65-B7120 Service engineer's manual

Fujitsu Semiconductor Controller MB89950/950A Hardware manual

Fujitsu

Fujitsu Semiconductor Controller MB89950/950A Hardware manual

mikroElektronika 4x4 RGB click quick start guide

mikroElektronika

mikroElektronika 4x4 RGB click quick start guide

Simplex 4001-9813 installation instructions

Simplex

Simplex 4001-9813 installation instructions

Spirent Vertex Application note

Spirent

Spirent Vertex Application note

AMX AXU-SPL4 Schematic diagram

AMX

AMX AXU-SPL4 Schematic diagram

NEC IECUBE2 user manual

NEC

NEC IECUBE2 user manual

Hitachi H8/3048 Series Hardware manual

Hitachi

Hitachi H8/3048 Series Hardware manual

Renesas UM-WI-038 user manual

Renesas

Renesas UM-WI-038 user manual

Nvidia GeForce2 Ti SP6600Ti manual

Nvidia

Nvidia GeForce2 Ti SP6600Ti manual

Pickering 40-575A user manual

Pickering

Pickering 40-575A user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.