SG021/19BM
Press the [P] button to select flow rate setting. Use the (+) or (-) button to
set the desired rate
Press the [P] button to confirm value and select the Alarm sound setting.
Press [-] button to turn the sound off or the [+] button to turn it on.
Press the [P] button to confirm value and select the Infusion Break setting.
Press [-] button to reduce the number or the [+] button to increase.
Press the [P] button to confirm value and select the partial volume setting.
Press [-] button to reduce the volume or the [+] button to increase.
Press the [P] button to confirm value and the pump will take a short time to
position the plunger at the programmed position.
Starting infusion and setting infusion time
Start the infusion by pressing the [+/ON] button.
You are given the option to prime the infusion set [Pr]. Press and hold the
[P] button to deliver the required prime up to a maximum of 1.5mL.
Press the [+] button again to start the infusion at any time.
Resetting plunger position
At the end of an infusion the plunger will automatically return to its
programmed start position.
To return this earlier, switch the pump into [StoP] mode by pressing the [+]
and [-] keys together. Then press the [P] and [-] buttons together until [END]
appears on the screen.
The syringe rubber plunger will not be withdrawn with this procedure.
Programming the pump
Setting procedure
With the keyboard unlocked, and the pump displaying [OFF] pressing the [P]
button will allow you to step through a menu to set the following parameters
in turn.
Set Flow rate 0.2 mL/hr to 300 mL/hr
End of infusion alarm On/Off
Infusion Break Points IP.1 to IP.5 (Normally set to IP.1)
Syringe Partial Volume 1mL steps