MultiSure GK Link Blood Glucose and Ketone Monitoring System
2
Intended Use
Your MultiSure GK Link Blood Glucose and Ketone Meter, MultiSure GK blood
glucose test strip, and MultiSure GK blood ketone test strip are in vitro diagnostic
medical devices to quantitatively measure blood glucose in fresh capillary whole
blood drawn from
ngertips, palm, or forearm, and blood ketone in venous whole blood or capillary
whole blood from ngertips. Alternative site testing for glucose test should be
performed only during steady-state (when glucose is not changing rapidly). They
are intended for self-testing outside the body (in vitro diagnostic use) by people
with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes
control and should only be used by a single patient and it should not be shared. They
are not intended for diagnosis or screening of diabetes and DKA or for neonatal use.
Do not alter your treatment on the basis of test results of this meter without previously
consulting your doctor or healthcare professional.
IMPORTANT:
• Dehydration – Severe dehydration may lead to inaccurate blood glucose test result.
If you suspect you are severely dehydrated, contact your healthcare professional
immediately.
• Hematocrit range – A hematocrit range that is higher than 70% or lower than 10%
can cause inaccurate blood glucose test results; A hematocrit range that is higher
than 60% or lower than 20% can cause inaccurate blood ketone test results.