Arjo 2000 Rhapsody User manual

System 2000
08.AR.09_1EN • 10/2020 • Page 1/2
QUICK REFERENCE GUIDE – CLEANING AND DISINFECTION
Caregiver obligations
WARNING
To avoid eye and skin damage, always use protective glasses and gloves. If contact occurs rinse with plenty of water.
If eyes or skin becomes irritated, contact a physician. Always read the material safety data sheet of the disinfectant.
To prevent cross-contamination, always follow the disinfection instructions in the Instructions For Use.
To avoid eye or skin irritation, never disinfect in the presence of a patient.
Clean and disinfect with disinfection unit (Still Bath and Hydrosound System)
NOTE
For disinfection of the Hydromassage system start on page 2.
Remove visual residue
1. Raise the bathtub to convenient working position.
2. Make sure all functions are turned off.
3. Open the drain.
4. Detach and place all accessories at the bottom of the bathtub.
5. Shower and rinse the bathtub and the accessories with water to remove
visible residue. Turn the accessories and rinse again.
Clean
6. Put on protective gloves and eye wear.
7. Lift out the disinfection spray handle.
8. Press the disinfection button.
9. Spray disinfectant into the bathtub until the ow meter shows correct mixing
value according to the Instructions For Use. Adjust if needed.
10. Spray disinfectant in the drain. Scrub the drain with a brush.
11. Close the drain.
12. Spray disinfectant on all internal bathtub surfaces and on all sides of the
accessories.
13. Spray into the surface overow outlet.
14. Press the disinfection button to turn it off.
15. Scrub the surfaces of the bathtub, loose parts and accessories thoroughly
with a soft bristle brush.
16. Open the drain.
17. Rinse all parts clean with plenty of water with a temperature of
25 °C (77 °F).
18. Soak a cloth in disinfectant and wipe all other contact areas:
the outside of the bathtub, metal sensory knobs and hand control.
19. Wet a new cloth with water and remove the disinfectant.
Occasionally rinse the cloth in running water.
Disinfect
20. Spray disinfectant in the drain.
21. Close the drain.
22. Spray disinfectant on all internal bathtub surfaces and on all sides of the
accessories.
23. Allow appropriate contact time according to the instructions of the
disinfectant.
24. Push the disinfection button to end the disinfection cycle.
Automatic Disinfection only:
If the preset disinfection time has not elapsed other functions are disabled
for at least 2 minutes.
25. Put back the disinfection spray handle.
26. Open the drain.
27. Push the button for shower.
28. Rinse all parts clean with water having a temperature of 25 °C (77 °F).
29. Soak a cloth in disinfectant and wipe all other contact areas:
the outside of the bathtub, metal sensory knobs and hand control.
30. Allow appropriate contact time according to the instructions of the
disinfectant.
31. Wet a new cloth with water and remove the disinfectant.
Occasionally rinse the cloth in running water.
32. Leave to air dry.
Between patients
Clean and disinfect System 2000 bath before every use.
Rhapsody, Primo, Harmonie
08.AR.09_1EN • 10/2020 • Page 2/2
Clean and disinfect the Hydromassage System
With automatic disinfection
1. Raise the bathtub to convenient working position.
2. Close the drain.
3. Unscrew the screws and remove the suction inlet screen.
Clean the suction inlet screen separately.
4. Press the disinfection button to start the disinfection cycle.
Disinfectant comes out from the Hydromassage system nozzle.
The disinfection cycle ends with the Hydromassage system being rinsed
with water.
5. Spray disinfectant into the surface overow outlet.
6. Continue with the remaining disinfection instructions according to
Clean and disinfect with disinfection unit (Still bath and Hydrosound
System), on page 1.
7. Allow appropriate contact time according to the instructions of the
disinfectant.
8. Press the Shower button and rinse the surface overow outlet.
9. After completed disinfection, attach the suction inlet screen with the screws
and tighten them gently.
Without automatic disinfection
1. Raise the bathtub to convenient working position.
2. Open the drain.
3. Unscrew the screws and remove the suction inlet screen.
Clean the suction inlet screen separately.
4. Pull out the turnable Hydromassage system nozzle if a treatment hose is
going to be connected.
5. Insert the treatment hose, if any, in the Hydromassage system outlet.
6. Spray disinfectant into the suction inlet until disinfectant comes out of the
Hydromassage system nozzle/treatment hose.
7. Spray disinfectant into the surface overow outlet.
8. Allow appropriate contact time according to the instructions of the
disinfectant.
9. Press the disinfection button to turn it off.
10. Press the shower button and connect the shower handle to the suction inlet.
11. Turn the Hydromassage system knob to its highest intensity.
12. Press the trigger and rinse until the water comes out clear from the
Hydromassage system nozzle/treatment hose.
13. Remove the treatment hose, if any.
14. Put back the nozzle into the Hydromassage system outlet.
15. Rinse the surface overow outlet with water.
16. Continue with the remaining disinfection instructions according to
Clean and disinfect with disinfection unit (Still bath and Hydrosound
System), on page 1.
17. After completed disinfection, attach the suction inlet screen with the screws
and tighten them gently.
Clean and disinfect with a spray bottle
Remove visual residue
1. Raise the bathtub to convenient working position.
2. Make sure all functions are turned off.
3. Open the drain.
4. Detach and place all accessories at the bottom of the bathtub.
5. Shower and rinse the bathtub and the accessories with water to remove
visible residue. Turn the accessories and rinse again.
Clean
6. Put on protective gloves and eye wear.
7. Spray disinfectant in the drain. Scrub the drain with a brush.
8. Close the drain.
9. Spray disinfectant on all internal bathtub surfaces and on all sides of the
accessories.
10. Spray into the surface overow outlet.
11. Scrub the surfaces of the bathtub, loose parts and accessories thoroughly
with a soft bristle brush.
12. Open the drain.
13. Rinse all parts with plenty of water with a temperature of 25 °C (77 °F).
14. Soak a cloth in disinfectant and wipe all other contact areas:
the outside of the bathtub, metal sensory knobs and hand control.
15. Wet a new cloth with water and remove the disinfectant.
Occasionally rinse the cloth in running water.
Disinfect
16. Use a spray bottle and spray disinfectant in the drain.
17. Close the drain.
18. Spray disinfectant on all internal bathtub surfaces and on all sides of the
accessories.
19. Allow appropriate contact time according to the instructions of the
disinfectant.
20. Open the drain.
21. Push the button for shower.
22. Rinse all parts clean with water having a temperature of 25 °C (77 °F).
23. Soak a cloth in disinfectant and wipe all other contact areas:
the outside of the bathtub, metal sensory knobs and hand control.
24. Allow appropriate contact time according to the instructions of the
disinfectant.
25. Wet a new cloth with water and remove the disinfectant.
Occasionally rinse the cloth in running water.
26. Leave to air dry.
ArjoHuntleigh AB
Hans Michelsensgatan 10
211 20 Malmö, Sweden
www.arjo.com

Other manuals for 2000 Rhapsody

1

This manual suits for next models

2

Other Arjo Bathroom Aid manuals

Arjo Malibu User manual

Arjo

Arjo Malibu User manual

Arjo Miranti User manual

Arjo

Arjo Miranti User manual

Arjo MIRANTI CEB 8X0X Installation guide

Arjo

Arjo MIRANTI CEB 8X0X Installation guide

Arjo MAA4031-M User manual

Arjo

Arjo MAA4031-M User manual

Popular Bathroom Aid manuals by other brands

Pressalit MATRIX R4820 Mounting instruction

Pressalit

Pressalit MATRIX R4820 Mounting instruction

Roth Mobeli 1400222H2S instruction manual

Roth

Roth Mobeli 1400222H2S instruction manual

Sensea S08-WH Assembly-Use-Maintenence Manual

Sensea

Sensea S08-WH Assembly-Use-Maintenence Manual

HOESCH ZERO Instructions for use

HOESCH

HOESCH ZERO Instructions for use

Moretti Mopedia RS926 instruction manual

Moretti

Moretti Mopedia RS926 instruction manual

Invacare Aquatec 900 user manual

Invacare

Invacare Aquatec 900 user manual

Guardian G1-200MX1 User instructions and warranty

Guardian

Guardian G1-200MX1 User instructions and warranty

lopital Lotus 5900 8100 Instructions for use

lopital

lopital Lotus 5900 8100 Instructions for use

identités 812025 instruction manual

identités

identités 812025 instruction manual

Invacare H112-5 user manual

Invacare

Invacare H112-5 user manual

Rebotec 363.00.20 user manual

Rebotec

Rebotec 363.00.20 user manual

Pressalit VALUE IV R1620 Mounting instruction

Pressalit

Pressalit VALUE IV R1620 Mounting instruction

Patterson Medical 09 107 8393 quick start guide

Patterson Medical

Patterson Medical 09 107 8393 quick start guide

Pressalit R1192 Mounting instruction

Pressalit

Pressalit R1192 Mounting instruction

NRS Healthcare P05218 User instructions

NRS Healthcare

NRS Healthcare P05218 User instructions

lopital Tango XL 5100 5705 Instructions for use

lopital

lopital Tango XL 5100 5705 Instructions for use

pba 4CN.NYD.0002 user manual

pba

pba 4CN.NYD.0002 user manual

NRS N05370 User instructions

NRS

NRS N05370 User instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.