Normal operating range
Low
Accompanied with a short beep when a button is pressed
Critical
Icons alternate with two short beeps every 10 seconds
< 5%
25%
100%
75%
50%
Display symbols
08
ABC
Battery Charge Level
Enhanced Charging System (ECS) Indicator
Displayed when charging units with ECS.
Maintenance Required (Wrench symbol with a letter character)
Shows for 7 seconds with 2 beeps when the lift goes from sleep mode and flashes each 5
seconds to draw user's attention.
Contact your Arjo service centre for maintenance.
Over Duty Cycle
Flashes when the lift is used for a long period without allowing it to cool down.
The overheat protection is activated when a function exceeds its duty cycle, is becomes disabled
while all other functions remain active.
The red light stays on during the cooling period and a single beep is heard.
The symbol disapears when the function has cooled down and it is ready for use again.
General Malfunction (Warning symbol with a numeric character)
Flashes when the lift encounters a malfunction.
The lift needs to be inspected or repaired by a qualified service technician.
The red light stays on and a single beep is heard.
Overweight
Flashes when a transfer is attempted with a load exceeding the lift's capacity.
The red light stays on with 4 rapid beeps until the patient is lowered.
The function is disabled while all other functions remain active.
001-31325-EN Rev 0 • 05/2020 • Page 2/4