ASTEC AD-3100 User manual

AD-3100
Operation Manual
Drawer Type Incubator
ATTENTION
In order for you to use the product properly, please read this
instruction manual thoroughly before use and retain it for future
reference. Also, when using the product, keep this manual where
you can retrieve it easily.
Please note that the warranty we provide only covers the product
itself. We cannot compensate for any damage and/or loss to the
contents incurred by the malfunction of the product.

AD-3100 Drawer Type Incubator Operation Manual Rev. 4
ASTEC CO., LTD.
2
Index of Contents………………………………………………………………………...2
1. Safety Precautions............................................................................................. 3
2. Warranty and Liability ...................................................................................... 7
3. Important Notice.............................................................................................. 8
4. Warning Labels/Bio Hazard/Copy Rights............................................................ 9
5. Specifications/ Accessories.............................................................................. 10
6. Dimensions ....................................................................................................11
7. Parts Description............................................................................................ 12
8. Description of Control Panel............................................................................ 13
9. Prior to Installation......................................................................................... 14
10. Initial Cleaning and Disinfecting .................................................................... 15
11. Humidity for the Incubation Chamber............................................................. 16
12. Connecting Gas............................................................................................ 17
13. Main Power................................................................................................. 18
14. Gas Stop Mode............................................................................................. 19
15. Daily Checkup (Gas Concentration)................................................................ 20
16. Gas Circulation / Automatic Calibration.......................................................... 21
17. Operation Procedure ..................................................................................... 22
18. Adjustment and Calibration Menu .................................................................. 23
19. Flow Chart of Adjustment and Calibration....................................................... 25
20. Temperature Calibration................................................................................ 26
21. CO2and O2Adjustment/Calibration Procedure................................................. 27
22. Power Monitoring Relay Unit ........................................................................ 28
23. Alarms........................................................................................................ 29
24. Troubleshootings.......................................................................................... 30
25. Gas Pressure Regulator ................................................................................. 33
26. Precaution for set up ..................................................................................... 34
27. Cleaning and Disinfectant.............................................................................. 35
28. Precaution to keep Incubator clean.................................................................. 36
29. Wiring Diagram ........................................................................................... 37
30. Maintenance ................................................................................................ 39
31. Gas Flow Diagram........................................................................................ 43
Factory Settings……………………………………………………………………………..44
Index of Contents
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts
Index of Contexts

AD-3100 Drawer Type Incubator Operation Manual Rev. 4
ASTEC CO., LTD.
3
Thank you very much for purchasing ASTEC Drawer type incubator AD-3100. In order for you to use
this instrument properly and safely, please read this instruction manual thoroughly prior to use and retain
it for your future reference. The specifications and design of this instrument, as well as the contents of
this instruction manual, are subject to change for improvement without prior notice.
1. Safety Precautions
This section covers important safety issues. These cautionary instructions below are provided to prevent
any injuries or damages to you and other people that may be caused by unsafe and improper handling of
this instrument. The instructions are classified into 3 categories: DANGER, WARNING, and CAUTION,
in order to clearly show the degree of urgency and the extent of hazard and damage which may be
incurred if the said instructions are not followed. Please be sure to read them carefully, follow them
strictly and use this product as directed. The following symbols are used in this manual.
There is an imminent likelihood that failure to follow instructions
could result in death or serious personal injury.
Improper handling of this instrument could result in death or
serious personal injury.
Improper handling of this instrument could result in personal injury and/or
property damage.
This symbol represents Danger, Warning and Caution
This symbol represents prohibition statement.
This symbol represents directive statement.
Warnings
Connect the instrument to a line plug according to local or county standards and to the specifications.
Do not use the instrument for different purposes from those shown by the manufacturer. (See Chapter 1
for General Descriptions)
Do not switch on the instrument without having waited at least 20 seconds after switching off.
Do not open the instrument before reading the servicing specific conditions described in the manual.
Follow instruction contained in the manual in case of changing any parts
To move the instrument, at least two people are required. We suggest contacting the distributor or the
manufacturer directly for servicing.
DANGER
WARNING
CAUTION

AD-3100 Drawer Type Incubator Operation Manual Rev. 4
ASTEC CO., LTD.
4
Do not use this instrument outdoors. It could cause an electrical short or a lethal shock.
Ask the manufacturer or your sales agent for the setting up. This will prevent negative
events such as gas leaks, water leaks or any other unexpected problems caused by
mishandling.
The location where the instrument will be placed must be steady, firm and leveled.
Do not leave this instrument in humid or wet conditions. It could cause an electrical short.
Single application of electric power source for this incubator according to rated voltage
described in label on the incubator is required. Parallel usage of the power source with the
other devices could cause abnormal heating/ignition on plug.
The power plug should be inserted all the way into the power outlet. Make sure that the
power plug is clean and dust-free when used. Do not plug/unplug the power cable with
soaked hands. Improper plugging causes ignition/fire due to heat generation.
Make sure that the instrument is securely earth-grounded.
Do not earth the instrument to the gas pipe, water pipe and lighting rod. This will cause
an electrical shock.
Connect only to a suitable power source specifically assigned for the incubator. Using the
correct power source specifically assigned for the incubator is extremely required.
Make sure the type of gas and appropriate gas pressure are used with the instrument. All
the gas connections with gas tube(s) must be firmly connected and tightened with
cable-tie.
Ventilate the room in which you use carbon dioxide with the instrument. It can harm your
body without adequate air ventilation. Do not breath carbon dioxide directly.
Do not plug any wire or pin into gas ports and/or sampling port. It could cause electrical
shock.
Do not place flammable objects in the incubator, such as alcohol, benzene, ethanol or
propane gas. Especially when you use oxygen, do not use matches or lighters or
cigarettes in the laboratory. Never use the flammable objects with the instrument.
WARNING

AD-3100 Drawer Type Incubator Operation Manual Rev. 4
ASTEC CO., LTD.
5
Toxic, pathogenic and radioactive substances should be kept away from the instrument.
If this instrument is to be used with toxic, pathogenic, radioactive, and/or other harmful
specimens, be sure to conduct such procedures in an isolated facility in compliance with
the safety regulations. Failure to do so will harmfully affect human body and surrounding
environment. Also, in order to protect the service technician, please issue a Safety
Declaration Letter.
During maintenance and cleaning, make sure to power off the instrument and unplug the
power cord. Switch off and unplug, or shut off the power supply when servicing this instrument.
Failure to do so could cause an electric shock and/or personal injury.
In case of malfunctions, turn off the power switch in the back of the instrument, and then,
unplug the power cord.
When the incubator is disposed, please be concerned for environment safety.
When connecting the plug of the power cord into the outlet, check the polarity of the plug of
the power cord and the outlet. Connect the outlet of L(Live) pole and L(Live) pole of the
power cord correctly. Incorrect connection can cause electric shock.
In case the instrument will be unused for a long period of time, remove the water in the
water pan and dry out the chamber completely. Make sure chambers are dry when turning
on the incubator. It causes sensors failure.
Do not inhale any chemicals, organic compounds, dust or moisture when cleaning.
Do not splash the water to the outside of the instrument especially on the top of it. Do not
operate the incubator with soaked hands. It causes electric shock and ignition.
Do not disassemble, service and remodel unless instructed by manufacturer. It causes
ignition or injury due to abnormal action of incubator.
Room temperature where the incubator being used must be maintained
between 24 –28°C.
When double-stack the incubators, use the optional brackets. The top incubator may fall by
earthquake or other unexpected cause or accidents.
Do not take apart or modify the instrument without first-hand supervising.