12021/09 | P000402.P
The AtriCure ASB3 Switch Matrix is a reusable accessory interface module that allows
simultaneous connection of AtriCure® Isolator™ Transpolar™ ablative handpiece and
pen devices to the ASU RF generator. Devices will be operational only when the ASB3
selector switch is set to the device specic position. Please refer to the applicable
handpiece and pen device Instructions For Use. Please refer to the ASU User’s Manual
for descriptions of RF energy delivery, modes of operation, output power, operating
frequency, and approved temporary cardiac pacing and stimulating accessories.
The ASB3 has been tested and found to comply with the limits for medical devices in
IEC 60601-1-2:2001. These limits are designed to provide reasonable protection against
harmful interference in a typical medical installation. The ASB3 generates and can
radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity. However,
there is no guarantee that interference will not occur in a particular installation. If the
ASB3 does cause harmful interference to other devices, which can be determined by
turning the ASB3 o and on, the user is encouraged to try to correct the interference by
one or more of the following measures; (1) reorient or relocate the receiving device, (2)
increase the separation between the ASB3 and the other devices, (3) connect the ASB3
into an outlet on a circuit dierent from that to which the other device(s) are connected,
and (4) contact the AtriCure service representative for assistance.
Inspect the following interface cables for any signs of physical damage to the cable and
connectors. If physical damage is found or the interface cables do not perform within
specication, notify AtriCure. All returns must have approval from AtriCure. The cables
include: (1) A000442-ASB/ASU RF Interface Cable, (2) C001237-ASB/ASU Footswitch
Interface Cable, and (3) A000479-ASB3 PSS Interface Cable.
Caution: Federal (US) law restricts this device to sale by or on the order of a
physician
European Representative:
AtriCure Europe B.V.
De Entree 260
1101 EE Amsterdam
The Netherlands
+31 20 7005560
ear@atricure.com
Manufacturer:
AtriCure, Inc.
7555 Innovation Way,
Mason, Ohio 45040 USA
1.866.349.2342
Dangerous Voltage Manufacturer
Type CF Debrilla-
tion-Proof Applied Part
Authorized represen-
tative in the European
Community
Alternating Current
Product complies with the
requirements of directive
93/42/EEC
Equipotential Refer to Instruction
Manual
Footswitch Fuses
Power ON
Power OFF
Caution Non-Sterile
Federal (USA) law restricts
this device to sale by or on the
order of a physician.
Waste Electrical and
Electronic Equipment
Humidity Storage Limit Temperature Storage
Limit
Environmental Conditions: (1) operational temperature +10°C to +40°C, (2) storage
temperature -35°C to +54°C, and (3) humidity 15% to 90% relative humidity.
Perform annual preventative maintenance procedures to ensure all ASB3 components
are functioning properly, including but not limited to: (1) electrical power cords for
fraying, damage, and proper grounding, (2) AC power switch, (3) indicator damage
(Power On, Footswitch Present, or Handpiece Select), (4) handpiece connector damage,
cracking or inability to insert and latch handpiece plug, (5) carrying handle damage,
inability to latch or rotate, (6) rubber feet damage, cracking or inability for the ASB3 to
remain stabile on a at surface, (7) interface cable fraying or damage, and (8) footswitch
connector damage cracking or inability to insert and latch footswitch plug.
The ASB3 does not have any serviceable parts. For servicing issues, contact AtriCure, Inc.
The ASB3 Accessory Cleaning and Disinfection Instructions; (1) use a mild detergent
(prepared to its specications) and a damp cloth to clean the exterior of the Switch
Matrix, (2) do not submerge or allow uids to enter the chassis, (3) do not use caustic,
corrosive, or abrasive cleaning materials, and (4) the ASB3 and auxiliary device cable
cannot be sterilized.
Types and rating of fuses: (1) 100-120V, 220-240V, ~50/60 Hz, and (2) 1.25A/250V,
T-lag, 5 x 20mm, UL Recognized, IEC Approved. Replace fuses as marked.
Follow local governing ordinances and recycling plans regarding disposal or recycling of
device components.
Disclaimer - Under no circumstances will Atricure, Inc. be responsible for any incidental,
special or consequential loss, damage, or expense, which is the result of the deliberate
misuse of this product, including any loss, damage, or expense which is related to
personal injury or damage to property.
AtriCure Switch Matrix ASB3
Instructions for Use
P000402.R
2021/12