Beijing Sanhe Beauty S & T Co,.Ltd PINXEL User manual

1
Operation Manual
BEIJING SANHE BEAUTY reserves the right to make changes to its products or specifications to improve
performance, reliability, or manufacturability. Information furnished by BEIJING SANHE BEAUTY is believed to
be accurate and reliable. However, BEIJING SANHE BEAUTY assumes no responsibility for its use. In any of
case, specifications are subject to change without notice.
CONVENTIONS Used in this Manual
The following conventions in the form of notes, cautions and warnings are used in this manual:
Note
The content of this Note offers general information that is important to keep in mind.
Caution
ACaution alerts the user to the possibility of a problem with the device associated with its use or
misuse. Such problems include system malfunction, failure, and damage to the system or other

2
property. The caution statement includes a precautionary action that should be taken to avoid the
hazard.
Warning
AWarning alerts the user to the possibility of injury, death, or serious adverse reactions associated
with the use or misuse of the system.
Table
Table o
of
fContents
Contents
1. Before Use .............................................................................................................
A. Notice …………………………………………………………………………………….
B. Symbols (Label) ………………………………………………………………………….
2. System Features ..................................................................................................
3. System Composition ...........................................................................................
4. Description of Product .......................................................................................
A. Main Unit
B. Main Unit (Rear)
C. Hand-piece
5. Operating Instruction ..........................................................................................
A. Preparations
B. Safety checking before operation
C. Operation procedure
D. Error message
6. Specification ......................................................................................................
A. System classification
B. Electrical specification
C. Mechanical specification
7. General Cautions ..............................................................................................

3
1. Before Use
A. Notice
Before the usage of this equipment, safety precautions and operating procedures should be
acknowledged thoroughly.
In general the devices might cause side-effects and accidents in case of improper use.
Please be careful about all potential risks of danger and conduct all necessary precautionary
actions according to this manual.
Please also be noted that this equipment should be considered for all repairing works,
upgrading and other necessary technical works by our instruction.
B. Symbols (Label)
Warning Alert Label for potential risks in usage of laser devices, will be adhered on
Main body.
2. System Features
Pinxel (Fracitonal RF Technology allows effectively treatment of deeper dermis, epidermis,
and upper dermis. RF energy is emitted at the end of fractional RF micro-needle; this will treat
target area - deeper dermis - without damage of epidermis.
The fractional RF is useful for minimal invasive epidermal rejuvenation.
RF energy does not target pigment-making cells in the skin. Pinxel cures and improves
wrinkle, scar, skin rejuvenation, and skin tightening safely and efficiently with minimized PIH,
using RF energy.

4
Special features of Pinxel ;
a) Minimizing downtime
b) Available to treat both epidermis and dermis layers
c) It is suitable for all skin types
d) Lesser risk of PIH and side effects
e) RF Fractional and RF micro-needle technologies are in one platform.
3. System Composition
*Front view
2
4
1
2
3
4
5
6
7
8
9

5
Description
1LED Screen
2Key Switch
3Emergency off Button
4Main unit
5RF fractional hand piece
6Earth grounding bar
7RF micro-needle hand piece
8Holder for hand pieces
9Holder screw.
* Back view
1
2

6
Description
1Holder neck
2Ventilation windows
3RF Micro needle hand piece connecter
4Grounding bar connecter.
5RF fractional hand piece connecter
6Power code connecter.
7Foot switch connector
3
4
5
6
7

7
System Composition
[ RF fractional hand piece ] [ RF micro-needle hand piece ]
Grounding bar Foot switch Power cord
4. Control Panel
1
2
3

8
Description
1Displaying operation status
2Key switch: Main power switch. Used for on/off of equipment.
3Emergency stop switch : Switch for operation halt at emergency
5. Operating Instruction
A. Preparation
1) Connect power cable into the appropriate socket, Identify safety.
2) Make sure air cooling tube is securely connected on main unit.
3) Check the using voltage and connect the power cord.
4) Check whether the exposure part is polluted and foreign materials exist.
Check each mode works properly before you operate it on patients!!
B. Safety check before operation
1Renew the patient’s treatment record.
2Check if all operation materials including micro-needle kit are sterilized in normal condition.
3To start operation procedure, we recommend that you check if each modes works properly
before you apply it on patient in order to prevent unexpected abnormalities.
4Establish proper parameter considering patient’s condition assessment and clinic treatment
guideline.
5Make patient’s skin ready for treatment, such as anesthesia cream, if necessary.
6The equipment should stay with STAND BY mode except using period for surgical operation.
7All operation procedure status should be monitored on displaying screen.

9
We recommend starting with low level first then increasing gradually!!
C. Operation procedure
1) Make sure power cord is securely connected, and fuse is equipped in right position on main
body.
2) Turn power ‘on’position, then power switch on front side of main unit will be activated.
3) Make sure grounding bar and foot switch are properly connected.
4) Initial screen will be displayed as following upon key switch on, and stand-by screen will be
shown after 2 ~3 seconds.
< Initial Screen >
Micro Needl RF Mode
5) Select ‘micro needle RF mode’, and then you can find screen like below picture.

10
6) Set each parameters in prior to operation in accordance to the following description.
No. Items Descriptions
1intensity To determine output level (1∼10 level)
2Polarset
To select Mono-polar or Bio-polar
3Interval
To set the time between first shot and next shot
Single/0.5/1.0/1.5/2.0/2.5/3.0 sec.)
4Active pulse To set continuance of output time per each shot (10~300 msec.)
5Depth To set the depth of penetration of micro needle (0.5∼3.0mm)
6Shot count To count out put on each mode.
7Back To return to initial screen
8Standby
(Ready) To select standby and ready mode.
Fractional RF Mode
No. Items Descriptions
1Intensity To determine output level (1∼10 level)

11
2Interval
To set the time between first shot and next shot
Single/0.5/1.0/1.5/2.0/2.5/3.0 sec.)
3 Active pulse To set continuance of output time per each shot (10~300 msec.)
4Shot count To count out put on each mode.
5Back To return to initial screen
6
Standby
(Ready) To select standby and ready mode.
When you stop operation please return to initial screen and off the system.
Please be noted that it is not completely power ‘off’ during stand-by status. It will
be completely power off by pressing main power switch in rear side of main unit.
6. Specification
A. System Classification
1) Electric shock protection: Class I equipment,
2) Electrical device class by IEC: Class IIa
3) Protection against ingress of liquids: Ordinary equipment
4) Mode of operation: Stand-by and Continuous in timer
B. Electrical Specification
1) Input rating : 100 ~ 240VAC ±10% , 50/60Hz
2) Power consumption : 300 VA in max.
3) System
1Main frequency :2 MHz ±10%

12
2Output (micro needle mode) : max. 30 watt, 250 V p-p,200Ω
3Output (Fractional RF mode) : max. 20 watt, 300 V p-p,200Ω
4Mode : Micro needle RF (mono or bi polar) / Fractional RF
5Grounding bar : Ø20 x 110mm, SUS-303
4) Display : LED Screen
C. Mechanical Specification
1) Dimension : W280mm x L505mm x H1,100mm
2) Weight : approximately 20kg
3) Standard acc’y: 2 handpieces / ground bar / foot switch / power cord
7. General Cautions
A. Environment and Restrictions for use
1) Environment: Do not use where humidity is high (example: bathroom)
[1] Operating Temperature : 10°C ~ 40°C (With enclosure)
[2] Operating Humidity : 10% ~ 80%, Non-condensing
[3] Storage Temperature : 10°C ~ 60°C (With enclosure)
[4] Storage Humidity : 10 - 90%, Non-condensing
[5] Pressure : 500 ~1060 hPa
B. Cautions before use
1) Check operation procedure preliminary.
2) Check buttons to make sure they work normally.
3) Do not modify without our consent.
4) Check all codes connections to make sure they are not separated.
5) Check the others except the previous articles when using the equipment after a long halt in

13
the operation. Check whether the other instruments are normal and work safely.
6) A bunch of cords can be a cause of overheat. Do not use cables in a state of that.
7) Do not step on plugs or give strong impact on them.
8) Do not use inflammables or fire.
C. Cautions during use
1) Stop using the equipment and power off immediately when its malfunction is found.
2) Power off immediately when power is failed. And return operation handles, switches etc. to
the original position.
3) This equipment should not be used in contiguity of patient monitor equipments.
4) Only persons who had special training are allowed to use.
5) The equipment should be used after consulting with a medical specialist.
6) Pay attention to following cautions during use.
1Check whether there are abnormal conditions in the equipment or patient.
2If any abnormal condition in the equipment or patient is found, take care of patient first.
3Then, stop using the equipment. Start reusing after checking patient and the equipment.
4Not qualified persons, including patient, are not allowed to contact the equipment.
5Do not use the equipment where chemicals are kept or gas is generated.
D. Cautions after use
1) Power off after returning operation handles, switches etc. to the original position according to
the procedure when operation is finished.
2) Don’t do unreasonable actions (example: holding cord) to cord connection part when
removing cord.
E. Maintenance
1) Do not repair at discretion when out of order. Ask the store or maker.
2) Do not open the case of the equipment or controller.

14
8,Packing List
Component Quantity
Main Body 1
RF fractional hand piece 1
RF Needle hand piece 1
Ground Bar 1
Key 2
Fuse 2
Foot switch 1
Protective goggle 1
Power cord 1
Holder for hand piece 1
Holder screw 1
RF fractional treatment tips 10
RF Needle treatment tips 20
9,Treatment Procedure
A, Cautions before operating patients
•Cleanse always the medical treatment area before the operation to remove the foreign
substance like cosmetics.
•The patient must wear the protective goggles.
•Begin the energy for treatment from low power to prevent any side effect, and lower the
patient's treatment expectancy level, and it will be better to perform the operation with
frequent treatments.
B, Treatment procedure
1, wear anesthesia cream on the treatment area (Anesthesia is suitable for about 45-55 minute)
After clean up anesthesia cream from face.
2, Perform an operation with the Pinxel (Refer to basic protocol for the energy setting)
3, The treatment as the sequence of the followings:
1Forehead
2Around eyes
3Nose

15
4Under eyes
5cheek
6Around lips
7Chin and under chin
8Jaw lines
4, After operation, strong cooling is essential for about 20 minutes.
5, Then perform the cooling and moisturizing care.
C, Cautions when and after performing operations
•Every treatment should be done as one passing of full face according to basic protocol.
•And you can do 2~3 passing additionally according to the lesions.
•Strong cooling after performing an operation is essential, and cooling and moisturizing
care are needed to enhance the treatment effect and to prevent any side effect.
•To enhance the cooling effect, after you treated the half of face, please cool down the
parts with roller which you already treated while you treat the other half of face also it is
very important that all the patients should cooling the face completely after treatment.
(We recommend you to use cooling up to 40 to 60 minute continuously by roller or Ice
pack...etc)
•Cooling is very important to prevent any side effect and when patient face still feel hot
after cooling , then please cooling their face by using fan or air conditioner blow cooling
air to the face.
•After treatment, there appear Edema and Erythema. Edema goes for 1-2 days and
Erythema goes for1-2 weeks.
D, Basic protocol
Indication Level Needle Tip
(mm)
ON time Passsessions Delay weeks
Large pore 3-5 1.0-1.5mm 100-150ms 2-3 3-4 4-5
Acne/Acne scar 3-5 1.0-1.5mm 150-200ms 2-3 3-4 4-5
Skin rejuvenation 3-5 0.5-1.0mm 50-100ms 1-2 3-4 4-5
Skin resurfacing 3-5 0.5-1.0mm 50-100ms 2-3 3-4 4-5
Skin tightening 3-5 1-1.5mm 50-100ms 1-2 3-4 4-5
Body tightening
(Abdominal)
5-8 2.0-3.0mm 100-150ms 2-3 3-4 4-5

16
Stretch marks 5-8 2.0-3.0mm 100-200ms 2-3 3-4 4-5
Neck 3-5 0.5-1.0mm 50-100ms 1-2 3-4 4-5
Wrinkle/Fine line 3-5 1-1.5mm 150-200ms 2-3 3-4 4-5
10, Question and Answer
1, What is the difference of Fractional RF and Fractional Co2 laser?
Fraction RF is better than Fractional Co2 laser, Since Fractional laser is single wavelength, the
indications is small range, and only have limitation on skin color. The aborbed energy is different
with different pigment color. But the result of Fractional RF is not effected by the pigment and
suitable for all kind of skin color. Obviously result can be seen after each treatment.
2, How about the safety of Fractional RF?
Fractional RF treatment has passed the FDA and CE certifcation, the safety has been proved
according to clinical report.
3, What kind of skin problem that the Fractional RF can treat?
Fractional RF could treat wrinkles, skin loose, Acne scar, large pores. In some treatment, the
pigment and vascular has been treated in some degree. This is a multiple treatment and it can
improve the skin property all round.
4, Is it suitable for sensitive skin?
The sensitive skin is suitable for the Fractional RF treatment, it could change the sensitivity in
some degree.
5, Is it suitable for the dark color skin?
It is suitable for the dark color skin, since there is not light power only RF energy. The dark skin
could use the treatment.
6, Does the skin become thinner after the treatment?
The Fractional RF treatment will not make the skin thinner. Oppositely, it could active the growth
of collagen in skin to make the skin healthier.
7, What will happen on skin after the treatment?
At the treatment, the redness and slight swelling will happen, but the customer could totally
accept. After 24 hours, the treatment area will show some needle shape escharotics. The
customer could use make up to cover the treatment area, and not effect normal life and work.
8, Is it necessary for long time treatments? How long time to do another treatment?
Obviously result will happen even after one treatment. But as the time pass by, the skin will
become loose. So we suggest do another treatment after 6 weeks interval. Each course could do
3-5 treatment.

17
9, How long time the result can be kept?
According to different skin property, single treatment could keep 2-6 month. 3-10 years can kept
after one whole course.
9,Contact Details
Company:Beijing Sanhe Beauty S & T Co,.Ltd
Adds: 3rd floor, #6E Building, Liando U Valley, Tongzhou District, Beijing,
China
Tel: +8610 81597649 / 81596921
Web: http://www.sanhebeauty.com
Email: Gxl@sanhebeauty.com
Table of contents
Popular Personal Care Product manuals by other brands

Conair
Conair NE150SBC Instruction booklet

Orliman
Orliman Thermomed Smart 4102 Use and maintenance instructions

Oral-B
Oral-B iO 3768 instruction manual

Profi Care
Profi Care PC-GRB 3081 instruction manual

Pacific Supply
Pacific Supply Ortop AFO manual

Abbott
Abbott Agilis HisPro DS3H010-38 Instructions for use