Beurer medical GL 44 User manual

GL 44
EN
Lancet needles / Lanzetten / lancettes / lancetas / lancette
GL 44
LD 03
Instructions for use
Blood glucose monitor
Step by step


Beurer GL44 3
Contents
1 Getting to know your device ...................................................................................................4
1.1 Scope of delivery and accessories ............................................................................................5
1.2 Replacements ..........................................................................................................................6
1.3 Functions of the device .............................................................................................................6
1.4 Signs and symbols ...................................................................................................................6
2 Warnings and safety notes .....................................................................................................8
3 Description of device and accessories ..............................................................................11
3.1 Blood glucose monitor ............................................................................................................11
3.2 Display symbols .....................................................................................................................12
3.3 Test strips ..............................................................................................................................12
4 Initial use and basic settings ................................................................................................14
4.1 Removing the battery insulation strip, replacing the batteries ...................................................14
4.2 Making and changing basic settings ........................................................................................14
5 Taking the blood glucose measurement ............................................................................15
5.1 Preparing to take a blood sample ............................................................................................15
5.2 Taking a blood sample ............................................................................................................16
5.3 Reading the result and highlighting measurements ..................................................................17
5.4. Post-processing and disposal ................................................................................................17
5.5 Evaluating measured blood glucose values .............................................................................18
5.6 Functional check using control solution ...................................................................................19
6 Measurement memory ...........................................................................................................22
6.1 Displaying individual values ....................................................................................................22
6.2 Displaying average blood glucose values ................................................................................23
6.3 Displaying average blood glucose values for labelled values ....................................................24
6.4 Reset to basic settings ...........................................................................................................25
6.5 Transferring measurements to a PC .........................................................................................26
7 Storing, maintaining and disinfecting the device .............................................................27
7.1 Cleaning ................................................................................................................................27
7.2 Disinfection ............................................................................................................................27
8 What if there are problems? .................................................................................................28
9 Technical specifications ........................................................................................................29
10 Comparison of measured values with laboratory values ................................................31
11 Usage limits for specialist personnel from the healthcare sector ................................33
12 Instructions for use for the lancing device LD 03 .............................................................35
12.1 Proper use ............................................................................................................................35
12.2 Warnings and safety notes ....................................................................................................36
12.3 Device description ................................................................................................................36
12.4 Initial use ..............................................................................................................................36
12.5 Usage...................................................................................................................................36
12.6 Post-processing and disposal................................................................................................38
12.7 Cleaning and maintenance ....................................................................................................39
12.8 Disposal ...............................................................................................................................39
13 Warranty/service .....................................................................................................................39

4 Beurer GL44
1 GettinG to know your device
Dear customer,
Thank you for choosing a product from our range. Our name stands for high-quality, thoroughly tested
products for applications in the areas of heat, weight, blood pressure, blood glucose, body temper-
ature, pulse, gentle therapy, massage and air.
Please read these instructions for use carefully and keep them for later use. Be sure to make them
accessible to other users and observe the information they contain.
With kind regards,
Your Beurer team.
Getting to know your device
The GL44 blood glucose monitor is intended for fast and simple blood glucose measurement of
fresh capillary blood samples, either for self-testing or in a clinical environment by trained personnel.
It enables you to measure your blood glucose quickly and easily, store the measured values and dis-
play the average of all measured values, thereby providing optimum assistance for monitoring your
diabetes. The test is performed exclusively externally (IVD).
The large backlit display shows measured values clearly. The user-friendly design with handy test
strips and the simple controls with just a small number of buttons guarantee simple, yet reliable
measurements.
The device can be connected directly to a PC using the USB cable provided. You can evaluate the
measured values on your PC using blood glucose diary software and use the results to monitor your
blood glucose values.
Blood glucose diary software is available to download for free at www.beurer.com.

Beurer GL44 5
1.1 Scope of delivery and accessories
Check that the set packaging has not been tampered with and make sure that all contents are present.
Before use, ensure that there is no visible damage to the device or accessories and that all packaging
material has been removed. If you have any doubts, do not use the device and contact your retailer
or the specified Customer Service address.
A1 blood glucose monitor
B1 lancing device
C10 test strips
D10 sterile lancet needles (5 x 28G blue, 5 x 33G purple)
E2 button cell batteries, 3 V CR2032 (already inserted)
F1 practical case
G1 USB cable
These instructions for use, additional information
•
If the packaging has sustained considerable damage or the contents are incomplete, please return
the system to your retailer.
•
The blood glucose monitor, test strips and additionally available control solutions have been spe-
cially designed to complement each other. For this reason, use only test strips and control solutions
that have been approved for this monitor.
Note
•
Use original manufacturer accessories only.

6 Beurer GL44
1.2 Replacements
You can also obtain test strips, control solution and lancets without a prescription.
Item REF
50 test strips REF 464.15
50 test strips, individually film packed REF 464.17
100 test strips REF 464.13
Control solution LEVEL 3 and 4 REF 464.16
100 soft touch lancets 33G REF 457.24
100 lancet needles 28G REF 457.01
100 safety lancets REF 457.41
200 safety lancets REF 457.42
1.3 Functions of the device
This device is intended for measuring the blood glucose content in human blood. It is also suitable
for self-testing at home.
The monitor enables you to quickly and simply:
•
Measure your blood glucose level
•
Display, label and save measured values
•
Display the average measured blood glucose value from the last 7, 14, 30 and 90 days
•
Display the average of the labelled measured blood glucose values from the last 7, 14, 30 and 90
days
•
Set the time and date
•
Transfer stored measured values to a PC for evaluation (additional accessories required)
The monitor also includes the following monitoring functions:
•
Warning in the event of unsuitable temperatures
•
Battery replacement display for low batteries
•
Warning that test strip is insuciently filled
Warning
•
Do not use the device to diagnose diabetes; it is intended for regular monitoring only.
•
Consult your GP with regard to insulin doses.
1.4 Signs and symbols
The symbols on the packaging and type plate of the monitor and accessories represent the following:
In vitro diagnostic device Manufacturer
Serial number Observe the instructions for use

Beurer GL44 7
Temperature limit
+2°C to +30°C Not for re-use/for single use
only
Use by Green dot (Der Grüne Punkt):
German dual waste collection
system
18 M
Maximum shelf life after initial
opening in months Contents sucient for <n> tests
LOT
Batch designation
/
Art.-Nr. Order number
Sterilised by radiation (lancets) Unit of measurement for blood
glucose value
Warning, see accompanying
documents Biohazard, risk of infection
Medical device (MDR Symbol) CE labelling
This product satisfies the
requirements of the applicable
European and national
directives.
Read the instructions
In the instructions for use, the symbols represent the following:
Warning
Warning instruction indicating a risk of injury or damage to your health/your patient's health.
Important
Safety note indicating possible damage to the device/accessory.
Note
Note on important information.

8 Beurer GL44
2 warninGs and safety notes
Risk of infection
All components of the monitor and its accessories may come into contact with human
blood and are therefore a possible source of infection.
Warning
•
Blood glucose values are displayed in mg/dL
or mmol/L.
You risk damaging your health if you meas-
ure your blood glucose value using a unit of
measurement with which you are not famil-
iar, misinterpret the values and subsequently
take incorrect measures. Therefore, please
ensure that this monitor displays a unit of
measurement with which you are familiar.
The unit of measurement accompanies each
blood glucose value.
Please contact Customer Services if the device displays the incorrect unit of measure-
ment.
•
When using the monitor for various persons, observe the generally applicable regulations regarding
disinfection, safety and contamination.
•
Medical carers and others who use this system on several patients must be aware that all prod-
ucts or objects that come into contact with human blood must be handled, even after cleaning, as
though they could transfer pathogens.
•
The lancing device is suitable for self-testing. Never share the lancing device or lancet needles with
others or amongst various patients (risk of infection!).
•
Use a new sterile lancet needle for each blood sample (for single use only).
General notes
Warning
Do not use the device in the vicinity of strong electromagnetic fields and keep it away from radio
systems or mobile telephones.
Measuring the blood glucose content
Warning
•
The measurements taken by you are for your information only – they are no substitute for a medi-
cal examination! Consult your GP regularly regarding your measured values. Never independently
alter the procedures prescribed by your GP.
•
Despite the simple usage of the Beurer GL44 monitor for self-monitoring of blood glucose levels,
you may possibly need to obtain instructions for using the system from your healthcare profes
-
Example

Beurer GL44 9
sional (for example, your GP, chemist or diabetes consultant). Only proper use will guarantee exact
measurements.
•
This device may be used by people with reduced mental capabilities provided that they are super-
vised or have been instructed on how to use the device safely and are fully aware of the consequent
risks of use.
•
A lack of water, high fluid loss, for example perspiration, frequent passing of water, severe hypo-
tension (low blood pressure), shock or hyperosmolar hyperglycaemic non-ketotic coma may lead
to incorrect measurements.
•
A haematocrit value between 20% and 60% has no significant influence on the measurements.
•
A very high or low haematocrit value (proportion of red blood cells) may lead to incorrect meas-
urements. In the event of a very high haematocrit value (above 60%), the displayed blood glucose
value may be too low; in the event of a very low haematocrit value (below 20%), it may be too high.
Consult your GP if you do not know your haematocrit value.
•
Do not use the test strips to measure blood glucose values on newborns.
•
Do not use NaF or potassium oxalate anticoagulants to prepare for venous blood samples.
•
Do not test any severely ill patients using this device.
•
Use fresh whole blood only. Do not use serum or plasma.
•
Use capillary blood without squeezing the penetration area. Squeezing the area causes the blood
to be diluted with tissue fluid and this may lead to an incorrect measurement.
•
Do not use the test strips above an altitude of 7010 m.
•
Very high levels of humidity may influence the test results. Relative humidity of more than 90% may
lead to inaccurate results.
Note
•
The Beurer GL44 measuring system is intended for measuring capillary and venous whole blood.
Storage and maintenance
Warning
•
Store the monitor and its accessories out of the reach of small children and pets. Small parts, such
as lancet needles, parts of the lancing device, batteries or test strips may be life-threatening when
swallowed. If swallowed, seek medical attention immediately.
•
The test strip box contains desiccant, which may irritate the skin or eyes when inhaled or swal-
lowed. Keep the box out of the reach of children.
The monitor is made from precision and electronic components. The accuracy of the measured val-
ues and service life of the device depend on its careful handling:
•
Protect the device and its accessories from impacts, humidity, dirt, marked temperature fluctua-
tions and direct sunlight. Do not store the device, test strips and control solution in your vehicle,
in the bathroom or in a cooling appliance!
•
Do not drop the device

10 Beurer GL44
Batteries/Saving measured values
Notes on handling batteries
•
If your skin or eyes come into contact with battery fluid, rinse the aected area with water and seek
medical assistance.
•
Choking hazard! Small children may swallow and choke on batteries. Store batteries out of
the reach of small children.
•
Observe the plus (+) and minus (-) polarity signs.
•
If a battery has leaked, put on protective gloves and clean the battery compartment with a dry cloth.
•
Protect batteries from excessive heat.
•
Risk of explosion! Do not throw batteries into a fire.
•
Do not charge or short-circuit batteries.
•
If the device is not to be used for a relatively long period, take the batteries out of the battery com-
partment.
•
Use identical or equivalent battery types only.
•
Always replace all batteries at the same time.
•
Do not use rechargeable batteries.
•
Do not disassemble, open or crush the batteries.
Note
•
The stored blood glucose values are retained when the batteries are replaced. If applicable, the
date and time must be reset after replacing the batteries.
•
Use lithium-ion batteries only.
Repairs
Note
•
Do not open the device. Failure to comply with this instruction will void the warranty.
•
Do not repair the device yourself. Proper operation can no longer be guaranteed in this case.
•
Do not dismantle the lancing device into individual parts, except in the steps described in these
instructions.
•
Please contact Customer Services for repairs.
Disposal
Warning
•
It is essential to comply with the generally applicable safety precautions for handling blood when
disposing of materials from the monitor. Dispose of all blood samples and materials with which
you or your patients come into contact correctly in order to prevent injury and infection of other
persons.
•
After use, dispose of test strips and lancets in a puncture-proof container.
Note
Empty, completely flat batteries must be disposed of through specially designated collection boxes,
recycling points or electronics retailers. You are legally required to dispose of the batteries.

Beurer GL44 11
The codes below are printed on batteries containing harmful substances:
Pb = Battery contains lead
Cd = Battery contains cadmium
Hg = Battery contains mercury.
For environmental reasons, do not dispose of the device in the household waste at the end
of its service life. Dispose of the device at a suitable local collection or recycling point in your
country. Dispose of the device in accordance with EC Directive – WEEE (Waste Electrical and
Electronic Equipment). If you have any questions, please contact the local authorities responsible
for waste disposal.
3 description of device and accessories
3.1 Blood glucose monitor
Front
1
Receptacle for test strips, illuminated
2 Display
3 Highlight button
4 On/O button
5 PC connection
6 “+ –” rocker switch
Rear
1 Battery compartment
(bottom side)
2 Reset button
1
4
2
3
5
6
reset
CR2032
CR2032
1
2

12 Beurer GL44
3.2 Display symbols
1 Date
2 Symbols for highlighting measurements
3 Measured value display, HI, LO display, average
blood glucose value, Err
4 Blood droplet symbol
5 Speaker symbol
6 Hand symbol
7 Test strip symbol
8 Blood glucose unit mmol/L
9 Blood glucose unit mg/dL
10 Battery replacement symbol
11 Memory symbol
12 Temperature symbol
13 Time
Note
The monitor is supplied with the following basic settings:
•
Acoustic signal o
•
Backlighting o
Warning
Ensure that you are using the device with the correct blood glucose unit (either mg/dL or mmol/L)
setting for you.
If in doubt, consult your GP.
3.3 Test strips
Front Rear
3
2
1
1 Gap for blood input
2 Grip area
3 Contacts
You can identify the rear by
the contact tracks.
1
2
3
4
5
68
9
10
11
12
13
7

Beurer GL44 13
Insert the test strip into the de-
vice so that the contacts are
pointing inside the slot.
Make sure that the front of the
test strip is facing you.
Note
Read carefully the following information on handling and storing your test strips. The test strips will
only provide accurate measurements if all information is followed.
Warning
Use each test strip only once and for one patient only!
Handling test strips
Note
•
Securely close the test strip box immediately after taking out a test strip.
•
Do not use the test strips if they have expired. The use of expired test strips may lead to incorrect
measurements. The expiry date is located next to the hourglass symbol on the box or on the re-
spective film packaging of the individual test strips.
•
Test strips expire 18 months after the box is opened. Make a note of the expiry date (opening date
+ 18 months
18 M
) on the label. The shelf life is limited to the expiry date (see date next to the hour-
glass symbol ). This does not apply for individual test strips, which are to be used immediately
after opening.
•
Discontinue use of the test strips if one of the two expiry dates ( /
18 M
) has passed.
•
You can touch any part of the test strip with clean, dry hands.
•
Use the test strip for measurement immediately after removing it from the box/film packaging.
•
Do not bend, cut or otherwise modify the test strips.
•
Do not use test strips that have come into contact with fluids.
Storing test strips
Note
•
Keep the test strips in a cool, dry place above +2°C and below +30°C. Do not expose the test strips
to direct sunlight or heat. Do not store in your vehicle, in the bathroom or in a cooling appliance.
•
Permitted relative humidity: below 90%.
•
The test strips must be stored in the original box/unopened film packaging – never use other con-
tainers.

14 Beurer GL44
4 initial use and basic settinGs
4.1 Removing the battery insulation strip, replacing the batteries
Note
•
Two batteries are included in delivery of the
blood glucose monitor. These have already
been inserted into the battery compartment.
•
Remove the insulation strip before initial
use.
1
Remove the battery compartment lid on the underside of the device.
2
When replacing the batteries, remove all batteries. The device retains the date and time as
long as one battery is still inserted. If necessary, reset the date and time (see “Making and
changing basic settings”, page 14).
3
Insert two new CR 2032 3 V batteries. Make sure that the batteries are inserted the correct
way round in accordance with the markings. Observe the graphic in the battery compartment.
4
Close the battery compartment lid again carefully.
Note
•
The batteries are almost empty if the battery replacement symbol appears. Replace both
batteries as soon as possible.
•
If “ ” appears on the display, the battery power level is so low that no more measurements are
possible.
4.2 Making and changing basic settings
1
Remove the batteries and reinsert them. Alternatively, press the “+” button
and the On/O button for a minimum of 5 seconds.
An acoustic signal sounds.
The year display flashes.
2
Setting the date and time
Note
•
You must set the date and time. Otherwise, you will not be able to save your measured values
correctly with a date and time and access them again later.
•
The time is displayed in the 24-hour format.
reset
CR2032
CR2032

Beurer GL44 15
2
Set the year (calendar to 2099) by pressing the “+” or “–” button. Confirm by pressing the On/
O button [4].
The day display flashes.
Proceed as described above for the day, month, hour and minute.
“” and “ ” are displayed.
3
Switching backlighting on/o
To switch the backlighting on/o, press the “+” or “−” button. “ ” and “ ” are displayed
for switched on and “ ” and “ ” are displayed for switched o. The background of the
display is simultaneously illuminated for a few seconds. Confirm by pressing the On/O button
[4]. “ ” and “ ” are displayed.
4
Switching the acoustic signal on/o
To switch the acoustic signal on/o, press the “+” or “–” button. “ ”, “ ” and the speaker
symbol are displayed for switched on, and “ ” and “ ” are displayed for switched o.
Confirm by pressing the On/O button [4].
5
The monitor is now ready for use.
5 takinG the blood Glucose measurement
Warning
•
If the protective disc on a lancet needle has already been removed, do not use the lancet needle.
•
If you drop the lancing device with an inserted lancet needle, carefully pick it up and dispose of the
lancet.
Important
•
Use the lancing device only with lancet needles from the same manufacturer. Using other lancet
needles may prevent the lancing device from working properly.
•
If you are using a third-party lancing device, please read the respective instructions for use.
5.1 Preparing to take a blood sample
1
Using the lancing device, you can take a blood sample from the fingertip. To make the pro-
cedure as painless as possible, do not take samples directly from the centre of the fingertip,
but slightly to either side.
Warning
•
In the event of suspected hypoglycaemia: take blood from the fingertip only.
Reason: changes to blood glucose levels can be detected quickly in blood samples taken
from the fingertip.
2
Prepare the following items: monitor, test strip box or test strips in film packaging, lancing
device, sterile lancet needles.

16 Beurer GL44
3
Wash your hands with soap and warm water before taking a blood sample. This not only en-
sures optimal hygiene but also encourages good blood circulation at the puncture area on the
finger. Dry your hands carefully.
Warning
If you have used alcohol for cleaning, ensure that the area has fully dried prior to measuring.
5.2 Taking a blood sample
Warning
•
Change the puncture area for each measurement, e.g. a dierent finger or the other hand. Repeat-
edly using the same area may cause inflammation, numbness or scarring.
•
Without the cap, there is a danger of injury from the exposed lancet.
•
Do not squeeze your finger to obtain a larger drop of blood. If squeezed, the blood is diluted with
tissue fluid and this may lead to an incorrect measurement.
•
Please note that insucient blood circulation at the puncture area, e.g. caused by cold temperatures
or illness, may lead to incorrect results.
Important
Do not apply any blood samples or control solutions to the test strip before inserting it in the monitor.
Please also note the following:
•
If the blood glucose test results do not match how you feel, carry out another test using blood from
your fingertip.
•
DO NOT change your treatment purely on the basis of a measurement that was carried out using
blood taken from an alternative area. Carry out another test with blood from your fingertip in order
to confirm the test result.
•
If you often fail to notice that you have a low blood glucose level, carry out a test using blood from
your fingertip.
1
If there is an insucient amount of blood, repeat steps 1 to 12 of the lancing device instructions
at the end of these instructions for use with a greater penetration depth.
2
Discard the first drop of blood. Always only take measurements with the second drop of blood.
3Hold the blood input gap (at the tip of the test strip) to
the drop of blood until the gap is completely filled and
the monitor in the display starts counting backwards. Do
not press the penetration area to the test strip. The blood
must not be spread. The blood is sucked into the gap.
Note
Error message “002” appears on the display if the gap was not correctly and suciently filled
with blood. Repeat the measurement using a new test strip and a greater penetration depth.
OK

Beurer GL44 17
Note
•
Do not apply blood to the sides of the test strips.
•
Do not add blood later if the device does not start the measurement. Remove the test strip and end
this test. Use a new test strip.
•
The device switches itself o if the test strip has already been inserted into the device but no blood
is added to the test strip within two minutes. Briefly remove the test strip and reinsert it so that the
device automatically switches itself back on.
•
Contact Customer Services if you are unable to fill the test strip with blood correctly.
•
If you are measuring in a dark environment, press the On/O button to switch the device on. The test
strip light switches on and makes it easier to insert the test strip. The backlight is also switched on in
the results display.
5.3 Reading the result and highlighting measurements
Reading the result
As soon as the gap is suciently filled with blood, the device performs the blood glucose measure-
ment. The monitor counts down for approx. five seconds.
The measurement is then shown on the display.
Read the measured value. For explanations and actions for the measured values, see chapter “5.5
Evaluating measured blood glucose values”. If an error message is displayed, read chapter “8. What
if there are problems?”
Highlighting measured values
You have the following options to label measured values.
Before meals.
After meals.
General highlighting (e.g. after exercising).
Highlighting measured values enables you, your GP or diabetes consultant to better monitor your
blood glucose values. For example, you can display the average values of all measurements taken
before meals.
The measured value can be labelled as soon as it is displayed. You cannot label it later.
For this purpose, briefly press the highlight button [3].
a) Pressing once labels the value with .
b) Pressing again labels the value with .
c) Pressing yet again labels the value with .
d) Pressing a final time removes the label.
The selected label is stored in the memory of the device when it is switched o.
5.4. Post-processing and disposal
Remove the test strip from the device and carefully dispose of it according to the currently applicable
regulations to avoid infecting others.

18 Beurer GL44
5.5 Evaluating measured blood glucose values
Your blood glucose monitor can process measured values between 20 and 630 mg/dL (1.1 and 35.0
mmol/L). The “ ” warning is displayed for measured values below 20 mg/dL (1.1 mmol/L). The “ ”
warning is displayed for measured values above 630 mg/dL (35.0 mmol/L).
Warning
•
If you suspect that the blood glucose results are incorrect, first repeat the test and, if applicable,
perform a functional test using control solution. Seek medical advice if dubious results persist.
•
Immediately seek medical attention if your symptoms are not in line with your measured blood glu-
cose values and you have observed all instructions for the Beurer GL44 blood glucose measuring
system.
•
Do not ignore symptoms of too high/low blood glucose levels. Consult your doctor.
Blood glucose values
The following tables list blood glucose values based on the STANDARDS OF MEDICAL CARE IN
DIABETES2016 from the ADA (American Diabetes Association).
Time of the blood glucose
measurement Normal blood
glucose values Increased risk of
diabetes (prediabetes)* Diabetes
On an empty stomach (fast-
ing plasma glucose) Below 100mg/dL
Below 5.6 mmol/L 100–125 mg/dL
5.6 – 6.9 mmol/L ≥ 126 mg/dL
≥ 7.0 mmol/L
Two hours after an oral
glucose tolerance test (con-
sumption of 75g) Below 140 mg/dL
Below 7.8 mmol/L 140–199 mg/dL
7.8 – 11.0 mmol/L ≥ 200 mg/dL
≥ 11.1 mmol/L
*The risk increases continually, beginning with values below the lower limit of the range and increasing disproportionately towards
the upper limit of the range.
Overview of glycaemic recommendations for non-pregnant adults with diabetes
A1C < 7.0%*
< 53 mmol/mol*
Preprandial capillary plasma glucose 80–130 mg/dL*
4.4–7.2 mmol/L*
Peak value of postprandial capillary plasma
glucose** < 180 mg/dL*
10.0 mmol/L*
*For individual patients, more or less strict glycaemic targets may be appropriate. The target values should be adjusted depending
on the length of time the person has had diabetes, age/life expectancy, accompanying diseases, known cardiovascular diseases or
advanced microvascular complications, hypoglycaemia unawareness, as well as individual patient considerations.
**The postprandial glucose value can serve as a target value if the A1C values are not met despite the preprandial glucose targets
having been reached. Postprandial blood glucose measurements should be taken one to two hours after the start of a meal, as this
is when diabetics' values are generally at the highest.

Beurer GL44 19
Evaluating critical measured values
Display Blood glucose Action
Very low blood glucose
level
under 20 mg/dL (under
1.1mmol/L)
Seek medical attention
immediately.
Low blood glucose level
under 70 mg/dL (under
3.9mmol/L)
Have a suitable snack.
Follow your GP's instructions.
High blood glucose level
on an empty stom-
ach, above 100 mg/dL
(5.6mmol/L)
2 hours after a meal
over 140 mg/dL
(7.8mmol/L)
If this high value persists 2 hours
after your last meal, this may in-
dicate hyperglycaemia (high blood
glucose). Seek medical attention
to coordinate any measures, if ap-
plicable.
High blood glucose level,
possibly ketones
over 240 mg/dL
(13.3mmol/L)
Perform a ketone test. For this pur-
pose, seek medical attention.
Very high blood glucose
level
over 630 mg/dL
(35.0mmol/L)
Take another measurement using a
new test strip. If the display is the
same as before: seek medical as-
sistance immediately.
5.6 Functional check using control solution
The control solution is used to test the entire blood glucose monitoring system. This helps to deter-
mine whether the monitor and the test strips are working optimally together and whether the test is
being performed correctly.
Perform the control solution test if you suspect that the monitor and/or the test strips could be faulty
or if you have repeatedly measured unusual blood glucose values. Also test the monitor if it has been
dropped or is damaged. The control solution is available separately. For the control solution test,
please observe the additional notes in the instructions for use for the control solution.
Important
•
Do not use control solutions by a dierent manufacturer. Correct functioning of your monitor can
only be tested using Beurer control solutions (LEVEL3 + LEVEL4).

20 Beurer GL44
•
Control solution measurements: When using the device, specialist personnel must follow statutory
guidelines.
•
Do not apply any blood samples or control solutions to the test strip before inserting it in the mon-
itor.
Performing a functional test using control solution
Warning
To obtain correct results, the monitor, the test strip and the control solution must be the same tem-
perature. For the “Functional test using the control solution”, the temperature is to be between 20°C
and 26°C.
The check performed at room temperature is used as a general functional check. The operating range
specified in the technical specifications is valid without restriction.
1
Hold the monitor so that the display is facing you.
2
Insert a test strip into the slot on the monitor with the contacts first. Ensure the front of the test
strip is facing you (see “Test strip”, page 12).
3
The device automatically switches on and briefly shows the initial display. The device is ready
for measurement as soon as the hand and the symbol flash.
IMPORTANT: Control solutions and blood react to temperature influences in dierent ways.
It is therefore of vital importance that control solution measurement is always performed in
control solution mode. If this mode is not used, results may be obtained that are outside the
target range.
Note
Press the rocker switch “+” or “–”, to change to the control mode. “ ” is shown on the dis-
play. This means that the result value is not stored in the memory, therefore not influencing
your measured value statistics. Pressing “+” or “–” again causes “ ” to disappear from the
display and the value is stored normally in the memory.
4
A clean surface is required to correctly perform a functional test.
Shake the control solution well before use.
Undo the cap and squeeze two drops next to each other on the
clean surface without touching them. Use the second drop for the
measurement.
Note
Do not apply the drop directly to the test strip to avoid contaminating the remaining control
solution in the bottle by touching the test strip with the tip of the bottle.
5
Hold the input gap (at the tip of the test strip) to the drop of control solution until the gap is
completely filled and the display of the monitor starts counting backwards.
When the gap is
suciently filled with solution, the device performs a measurement. The device counts down
for approx. five seconds. The measurement is then shown on the display.
Table of contents
Other Beurer medical Blood Glucose Meter manuals
Popular Blood Glucose Meter manuals by other brands

Sejoy
Sejoy BG-205 owner's booklet

MHC Medical Products
MHC Medical Products Easy Touch user manual

pts Diagnostics
pts Diagnostics CardioCheck PA user guide

Lifescan
Lifescan SureStep Blood Glucose Monitor Operator's guide

Wellion
Wellion LEONARDO GLU/CHOL plus owner's guide

Microdot
Microdot Xtra user manual