TABLE OF CONTENTS
SYMBOLS
...................................................................................................................................................................
4
TECHNICAL DATA
.....................................................................................................................................................
5
SAFETY INSTRUCTIONS & WARNINGS
..................................................................................................................
6
GENERAL
................................................................................................................................................................
6
SHOCK PREVENTION
...........................................................................................................................................
8
FIRE AND EXPLOSION
..........................................................................................................................................
8
EQUIPMENT WARNING LABELS
..........................................................................................................................
8
ASSEMBLY INSTRUCTIONS
.....................................................................................................................................
9
Assembly Instructions for
Ministand
Lift
..................................................................................................................
9
ADJUSTING THE WIDTH OF THE LEGS
............................................................................................................
13
L
IFT ASSEMBLY EXPLODED VIEW
.......................................................................................................................
14
With shifter base
....................................................................................................................................................
14
With pedal base
.....................................................................................................................................................
15
With power base
....................................................................................................................................................
16
LIFT PARTS LIST
......................................................................................................................................................
17
SHIFTER BASE EXPLODED VIEW
..........................................................................................................................
18
#7
00.20061.09 (SHIFTER BASE KIT ASSEMBLY)
..............................................................................................
18
PEDAL BASE EXPLODED VIEW
.............................................................................................................................
19
#700.20060.09 (PEDAL BASE KIT ASSEMBLY)
..................................................................................................
19
POWER BASE EXPLODED VIEW
...........................................................................................................................
20
#700.20062.09 (POWER BASE KIT ASSEMBLY)
................................................................................................
20
CONTROL BOX DETAI
LS (with integrated charger)
.............................................................................................
21
BATTERY PACK DETAILS
......................................................................................................................................
22
CHARGER WIRING DETAILS
..................................................................................................................................
23
BATTERY & CONTROL BOX WIRING DETAILS (with integrated charger)
........................................................
24
BATTERY MAINTENANCE & SERVICE
..................................................................................................................
25
MAINTENANCE AND REPLA
CEMENT OF BATTERIES
.....................................................................................
25
MAINTENANCE & SERVICE
....................................................................................................................................
26
MAINTENANCE AND REPLACEMENT OF BATTERIES
.....................................................................................
26
MONTHLY INSPECTION DETAILS
..........................................................................................................................
28
DIAGRAM
..................................................................................................................................................................
29
LOGBOOK
................................................................................................................................................................
.
30
WARRANTY
..............................................................................................................................................................
33
COPYRIGHT 2003
WARNING!
All right
s reserved. CONFIDENTIAL.
The reproduction of this document or the transmittal in any form of any information, contained herein,
without the authority in writing of an
officer of
the manufacturer is prohibited.
TECHNICAL SPECIFICATIONS AND DESIGN SUBJECT TO CHANGE WITHOUT NOTICE.

4
WARNING: this symbol is
intended to alert the user
to hazards or unsafe
practices which could
result in serious bodily
harm.
NOTE:
this symbol
offers helpful information
concerning certain
operating procedures.
CAUTION: this symbol is
intended to alert the user
of the presence of
important operating and
maintenance instructions
which could prevent
product damage or
possible personal injury
.
SYMBOLS
DO NOT ATTEMPT TO USE THIS EQUIPMENT WITHOUT UNDERSTANDING
THIS MANUAL.
To ensure safe operation,
ca
refully
read the entire manual, especially the
section on Safety Instructions and Warnings , before installing, operating or
servicing this equipment.
If anything is not completely understood, please contact your supplier for more
details. Failure to comp
ly with warnings in this manual may result in injury.
Keep this manual with lift and refer to it as required. Contents of this manual
are subjected to change without prior notice to users.
CONSULT ACCOMPANYING
DOCUMENTS
BOOM "UP"
BOOM "DOWN"
BASE "CLOSED"
BASE "OPENED"
INDICATES CHARGING
MODE
INDICATES
CHARGER
POWER IS "ON"

5
TECHNICAL DATA
DIMENSIONS
WEIGHT DATA
Hoisting height a
t:
Description
kg
l
bs
Description
Min
Max.
Total lift
45.5
100
mm
in
mm
in
Shipping
55
121
A
Total lift height
1090
42.9
1580
62.2
Complete base
17.7
39
B
Hoisting height at connecting points
955
37.6
1540
60.6
Complete mast with controls &
actuator
10.86
24
C
Connecting points to base end
310
12.2
625
24.6
Complete boom
4.5
10
mm
in.
Foot plate
3.3
7
D
Knee pad to base end (Min.)
490
464.6
Knee pad
2.7
6
D
Knee pad to base end (Max.)
610
24.0
Batteries pack
4.5
10
E
Knee pad to
gound (Min.)
520
20.5
Shifter arm
2 4
E
Knee pad to ground (Max.)
610
24.0
TECHNICAL DATA
F
Base floor clearance
50
2.0
Description
kg l
bs
G
Base height
110
4.3
Maximum Permissible Load (Kg (lbs))
159.1
350
H
Base total length
1050
41.3
Maximum
hoisting capability
199.1
438
I
Foot plate to ground (Low)
110
4.3
IP rating control std. #700.160XX
IP21
I
Foot plate to ground (Hi)
IP rating control opt. #700.161XX
IP24
J
Total internal base width (legs closed)
565
22.2
IP rating hand remote
control
IP44
J
Total internal base width (legs open)
980
38.6
Control voltage output (Vdc)
27.0Vdc
K
Total external base width (legs closed)
700
27.6
Battery charger input (Vac)
90 to 240
K
Total external base width (legs open)
1100
43.3
Battery char
ger output (VA max)
27
L
Connecting points distance
440
17.3
Up & down current limiting (Amps.)
10A
Turning radius
830
32.7
Minimum door requirement
711
28.0
Fuse size (Amp.)
10A
Duty Cycle 16% (#700.06553)
10min./Hr.
Duty
Cycle 10% (#E6507)
6min./Hr.
Sound power level ( dBA )
40
ENVIRONMENTAL CONDITIONS:
Normal operation: 10
º C to 40º
C, 0% to 90% R.H., 700
-
1060hPa.
Equipment is not suitable for use in the presence of flammable anaesthetic mixture with
air
or with nit
rous oxide.
Equipment classification with respect to protection from electric shock: Internally powered.
Degree of Protection against Electric Shock: Type BF
.

6
SAFETY INSTRUCTIONS & WARNINGS
GENERAL
IMPORTANT
CAREFULLY READ THESE INSTRUCTIONS OR SERIOUS INJURY
MAY RESULT.
KEEP THESE INSTRUCTIONS WITH THE LIFT AT ALL TIMES.
READ OPERATING MAINTENANCE INSTRUCTIONS IN THIS MANUAL BEFORE
INSTALLING, OPERATING
OR SERVICING THIS EQUIPMENT.
YOUR LIFT is for transferring patients only. Do not use the lift for a
ny other purpose.
ALWAYS carry out the daily checklist before using the lift.
BHM medical patient lifts are designed specifically for use with slings and accessories
manufactured by BHM Medical. We
cannot ensure the safety of a transfer, and therefore
can
not be held responsible, for any incident that may occur as a result of the use of
slings or accessories produced by another manufacturer. In order to ensure the safety of
the patient and that of the caregiver use only products manufactured by BHM Medical
with this patient lift.
This mobile floor lift is intended to be used for patients within the specified weight limit
indicated for the lif
t. Do not attempt to lift more than the weight limit indicated.
BEFORE attempting to transfer, the patient must be assessed by a qualified
professional.
This mobile patient lift must be used by a caregiver with proper training to work with the
patient to be transferred. This patient lift should never be used by a patient on their
own.
ONLY trained and qualified caregivers should transfer a patient. DO NOT attempt to use
the lift if you have not been properly trained to do so.
ALWAYS be prepared before attempting to transfer a patient.
DO NOT use a sling that is not recommended for the lift.
NEVER use a damaged, torn or
frayed sling.
ALWAYS place the sling around patient according to the instructions enclosed.
FOLLOW lifting procedures outlined in this manual.
USE all controls and safety features only according to the rules specified in this manual.
Never attempt to forc
e a control or button on the lift.
DO NOT store the lift
in a shower, bath or other area with high humidity.
IMPORTANT: keep all components of the lift clean and dry, and have electrical and
mechanical safety checkpoints done as instructed in the Maintenance section of this
manual.
Replace any precautionary or instruction labels that cannot be easily read.
Do not attempt to manoeuvre the lift by pushing on the mast, motor shaft, boom or
patient.
ALWAYS manoeuvre the lift with the handles provided.
DO NOT push a loaded lift at speeds which exceed a slow walking pace (3 km/hour or
0.8 meter/second).

7
SAFETY INSTRUCTIONS & WARNINGS
DO NOT push the lift over uneven or rough ground, particularly if loaded
DO NOT attempt to push or pull a loaded lift over a floor obstruction which the casters
are unable to ride over easily.
DO NOT BUMP the lift down steps, loaded or unloaded.
DO NOT park a loaded lift on any sloping surface.
DO NOT use an electric lift in a shower
This mobile floor lift is not intended to be a transport device or to be used to transport a
patient over any long distance.
BASE should not be opened as much as possible for optimum safety.
DO NOT USE, in any way, the actuator as a handle to push or pull the lift. If the actuator
is used as a handle
, the caregiver and the patient safety is greatly compromised
.
Do not put fingers, hands or feet where space is limited (see diagram below *). This
could pinch, cut or seriously injure any parts exposed.
* All models

8
SAFETY INSTRUCTIONS & WARNINGS
SHOCK PREVENTION
AVOID violent shock during transportation.
DO NOT touch or use a lift with bare conductors or a damaged power cord. Electrically
live equipment can electrocute a patient. If the lift or charger has any exposed or
damaged wires, contact y
our local dealer immediately.
DO NOT splash or expose electric parts of the device to water or moisture.
CHECK nameplate for voltage and cycle requirements. These requirements differ by
country. Do not attempt to use the lift in an area that has a different voltage and cycle
requirements.
READ the battery and charger instructions thoroughly before using or storing them
.
FIRE AND EXPLOSION
Batteries may explode,
leak and cause personal injury i
f not disposed of properly.
Do not place or store the battery und
er direct sunlight or near a heat source.
Do not dispose of in fire.
Do
not
short the battery terminals.
Do not incinerate.
Flush with water if electrolyte (Acid) comes in contact with skin or eyes.
Batteries must be recycled or disposed of according to local law regulations. When
returning batteries, insulate their terminals with adhesive tape, etc. otherwise the residual
electricity in use batteries may cause fire or explosion.
IMPORTANT
:
BATTERIES NEED TO BE CHARGED FOR A MINIMUM OF 8 HOURS PRIOR
TO
USING THIS LIFT FOR THE FIRST TIME.
EQUIPMENT WARNING LABELS
Inspect all precautionary labels on the equipment. Order and replace all labels that
cannot be easily read.

9
ASSEMBLY INSTRUCTIONS
Assembly Instructions for
Ministand
Lift
There are three
different models of
Ministand
lift :
With legs activated by a shifter
(A).
With legs activated by a pedal
(B).
With electrical legs (activated by the control box)
(C).
The assembly
instructions remain
the same for
all
lifts
unless specified otherwise.
1.
Remove all parts from carton and inner boxes, make sure all components are
present (see checklist on the main box sticker
).
Check for any shipping
damage. Do not discard carton or packing material until you are completely
satisfied there is no damage
to the lift.
With shifter
models only

10
2.
Remove the securing bolts
on the base mast post.
Slide mast over the post
and secure with the two
bolts and lock washers.
3.
For power base shifter
only
:
Unpack the actuator
from the mast prior to
fixing it to the floor lift.
Actuato
r wire must be
oriented upward and its
label must be visible.
Make sure to fix both ends
properly to the mast
bracket and the power
base shifter. Both cotter
pins must be properly
installed and secured.
4.
Install boom assembly into
position at the top of the
mast and secure it with the
shoulder bolt, the two sling
sleaves and lock nut.
Tighten securely.

11
5.
Install
the top extremity of the actuator
and fix it to the boom bracket with the
clevis pin assembly.
6.
Insert the knee support union into th
e
mast bracket and adjust with the knob
place below the bracket. Insert the
knee pad on the knee support union
and adjust for proper height with the
knob placed behind the knee pad.
7.
Clip the footplate on the base.

12
8.
Insert the shifter arm onto the
shifter plate assembly (for
model
s
with shifter arm
only).
9.
Place batteries into the control
box.

13
ADJUSTING THE WIDTH OF THE LEGS
The base of the lift can be set with the legs opened or closed. A
djustments
of the base
varied according to the model:
Shifter base: The adjustments are made by changing the position of the shifter:
Disengage the shifter plate by pulling the shifter arm then push it to the left will
close
the legs.
Pushing the shifter arm to the right
will
open
the legs (Fig.1)
.
Pedal base : The adjustments are made by changing the position of the pedal
.
Press down the left side of the pedal will
close
the legs. Press down the right
side of the pedal
will
open the legs (Fig. 2).
Power base
: The adjustments are made by using the
control buttons OPEN and
CLOSE on the control box or on the hand control (Fig. 3).
Since the residents centre of gravity is positioned over the centre of the chassis, in most
situations the resident can be comfortably transported with the base closed. Ke
eping
the resident at the lowest point within the base, lowers the centre of gravity, increasing
resident security. The residents lateral movement is also reduced by the design of the
arms
.

14
LIFT ASSEMBLY EXPLODED VIEW
With
shifter base

15
LIFT ASSEMBLY EXPLODED VIEW
With
pedal
base

16
LIFT ASSEMBLY EXPLODED VIEW
With
power base

17
LIFT PARTS LIST

18
SHIFTER BASE EXPLODED VIEW
#700.20061.09 (SHIFTER BASE KIT ASSEMBLY)

19
PEDAL BASE EXPLODED VIEW
#700.20060
.
09
(PEDAL BASE KIT ASSEMBLY)

20
POWER BASE EXPLODED VIEW
#700.20062.09 (POWER BASE KIT ASSEMBLY)
This manual suits for next models
1
Table of contents
Other BHM Medical Medical Equipment manuals
Popular Medical Equipment manuals by other brands

smirthwaite
smirthwaite 130 manual

Stryker
Stryker SERFAS user manual

Johnson & Johnson
Johnson & Johnson DePuy Synthes Battery Power Line II Instructions for use

Joerns Healthcare
Joerns Healthcare EasyCare User & service manual

Dräger
Dräger Evita V800 Pocket guide

BeaconMedaes
BeaconMedaes Z MED Medical Air System Installation, operation and maintenance instructions