BodyTrace BT104 User manual

v1.0
BT104
USER MANUAL
BLOOD PRESSURE MONITOR
START
STOP
START
STOP

2
Components of the Device
START
STOP
CUFF
CUFF TUBE
START/STOP
BUTTON
DISPLAY
CUFF TUBE
CONNECTOR
BATTERY
COMPARTMENT
BATTERY
DOOR
SYSTOLIC
MEASUREMENT
DIASTOLIC
MEASUREMENT
PULSE
MEASUREMENT
IRREGULAR
HEARTBEAT
INDICATOR

3
If this is your rst time using the device:
1. Slide open the battery door on the back of the device.
2. Install the batteries provided with the device.
Follow the diagram inside the battery compartment
for correct polarity — the springs should align with
the negative sign on the batteries.
3. Slide the battery door closed.
Installing the Batteries

4
Securing the Cu
1. Plug the connector on cu tube into the device.
2. Expose your left upper arm by removing or adjusting
clothing and jewelry. Make sure blood ow is not
constricted by a rolled up sleeve.
3. Open the cu and loosen fully.
4. Orient the cu so that the tube exits towards
the hand.
5. Place your arm through the cu loop, with your palm
facing up.
6. Position the cu’s edge about an inch (2-3 cm) above
the elbow.
START
STOP

5
8. Tighten the cu evenly around your arm by pulling on
the end — make sure the Ф marker stays aligned with
the center of your arm.
9. Wrap the end of the cu over your arm to secure it in
place. Don’t make it too tight — allow a nger to t
between the cu and your arm.
7. Align the Ф marker (located to the right of tube
exit) with the center of your arm.
Ф
2 - 3 cm (1 inch)

6
Taking a Measurement
10. Lay your arm on a table with your palm facing up.
The cu should be at the same height as your heart.
Sit up straight and rest your feet at on the ground.
Make sure the tube is not kinked or pinched.
11. Press the START/STOP button
on the device to turn it on.
12. Wait for the device to power on, and perform the
measurement while inating the cu.
13. When the measurement concludes, results are
shown on the display and the cu automatically
deates.
14. Remove the cu after it deates.

7
15. The device automatically transmits measurement
results. During this process an indicator is shown in
the middle of the display.
16. When the measurement is successfully transmitted,
the device shows the following indication before
powering o
Transmission
in progress Transmission
successful
Notes:
• If you wish to abort the measurement while the cu is
inating, press and hold the START / STOP button
for 5 seconds.
• If the device detects an irregular heartbeat, the display
will show this indicator along with the
measurement results. While this may not be a reason
for immediate concern, contact your primary care
provider if you see this symbol frequently.

8
Precautions
• Only use Duracell alkaline AA batteries with this device as
these oer the longest battery life. Other brands may not be
able provide sucient power for cellular data transmission.
• This device is for adults. Do not use it on children
unless otherwise instructed by a medical professional.
• Do not use or store the device near infants or in an
area they can reach.
• Do not apply the cu on an arm that has:
• an unhealed wound,
• another medical device attached, or
• an intravenous drip or blood transfusion attached.
• Do not use this device if you have an implanted cardiac
device, such as a pacemaker or debrillator.
• Measurements performed by this device may not be
accurate during pregnancy.
• This device is intended for indoor, home use.
• If in doubt, contact your medical professional before use.
• Do not use the device where ammable gases are present,
or in environments with high oxygen concentration.
• Measurements may not be accurate if the device is used
in close to proximity devices generating strong
electromagnetic elds, such as microwave ovens,
wireless charging stations, X-ray machines, etc.
• Wireless communication devices, such as Wi-Fi access
points, mobile phones, cordless phones, may cause
interference that may aect the accuracy of
measurements. A minimum distance of 13 feet (4 meters)
should be kept from such devices during measurement.

9
Troubleshooting
Problem
Device
doesn’t power on
when START/STOP
button is pressed.
or
Cu does not
inate.
(display remains
o)
Cu does not
inate.
(display turns on)
Cu doesn’t form a
loop around arm.
Possible Reason
Batteries
completely drained.
Cu not plugged in
correctly.
Cu not assembled.
Batteries inserted
with incorrect
polarity.
Resolution
Replace batteries
with new
Duracell AA.
Make sure the
connector on cu
tube is plugged all
the way into the
device.
To form a loop:
pass the end of the
cu furthest away
from the tubing
through the long
metal D-ring.
The smooth side
(without the hook
and loop fasteners)
should be on the
inside of the cu
loop.
Remove batteries
and reinstall them
making sure the +
and - polarity
symbols on the
batteries match
those inside the bat-
tery compartment.

10
Problem
Readings
too high or low.
or
Readings
very signicantly
between
measurements.
Measurement
results dier
from those
measured at
doctor’s oce.
Error code
shown after
measurement.
Possible Reason
Cu not positioned
correctly.
Measurements
taken in a medical /
professional
environment
(white coat eect)
may aect your
blood pressure.
Measurement
accuracy may be
aected or
transmission
may have been
unsuccessful.
Movement during
measurement.
Resolution
Review instructions
on Securing the Cu,
and retry
measurement.
Measurements taken
in a more relaxed,
home environment
may avoid eects
experienced in
clinical settings.
Continue taking
readings as usual.
If in doubt, contact a
medical professional.
Consult the error
code table on page
11-12 for more
information and
further
instructions.
Place your arm on
a table with your
palm facing up. The
cu should be at the
same level as your
heart. Make sure you
stay still and don’t
talk during
measurement.

11
Error Codes
Code
Er 1
Er 7
Er 9
Er 2
Er 3
Er 4
Er 5
Er 6
Description
Battery low
Internal error
Provisioning
error
Movement
detected
Measurement
out of range
No cellular
coverage
Cellular connec-
tion error
Data transmis-
sion error
Resolution
Replace batteries with
four Duracell AAs.
Replace batteries with four Duracell
AAs. If the issue persists, contact
customer support.
Retake measurement.
If the issue persists, contact
customer support.
Avoid moving or talking while taking
a measurement. After securing the
cu and assuming the correct
posture, consider waiting for ve
minutes to settle down before
starting a measurement.
Conrm that the cu is secured
correctly on your exposed arm.
Remove all clothing and jewelry
from your arm.
If you feel unwell, contact a
medical professional.
Move to another
area, closer to a window.
Use the device at a
location where you
get trong cellular signal
with your mobile phone.

12
Code
Er 10
Er 15-
24
Er 11
Description
Subscription
suspended
System error
Subscription
terminated
Resolution
Contact customer support.
Contact customer support.

13
Storage and Maintenance
• Only use Duracell AA batteries with the device.
• Store your device and cu in a clean and dry place.
• Avoid exposure to direct sunlight, moisture, extreme
heat, cold or humidity.
• Avoid exposure to vibration, shaking or
other mechanical stresses.
• Make sure the rubber tube of the cu is not squeezed,
stretched or kinked during use and storage.
• Avoid twisting or folding the cu and tube too tightly as
this may cause damage over time.
• Clean both device and cu with a soft, dry cloth.
If necessary use a dampened cloth and natural
detergent. Do not use alcohol, benzene or
other harsh chemicals.
• Avoid washing or splashing the device or cu with water.
They are not water resistant.
• Do not try to repair or modify the device at home. If you
suspect incorrect operation, contact customer support
for further assistance.
• Do not use or store the device near infants or in an area
they can reach.

14
Symbols
Symbol Description
Type BF applied part
Relative humidity range
Date of manufacture
Temperature range
Serial number / IMEI
Cu circumference
Caution: consult instructions for use
Model number
Complies with FCC regulations
Do not dispose of this product as unsorted
municipal waste, take it for recycling.
Ingress Protection Eective against ngers
or small objects; dripping water
REF
SN
IP21

15
Model
Cu Size
Device Weight
Dimensions
Ingress Protection
Applied Part
Software Version
Measurement
Method
Rated Cu
Pressure
Accuracy
Operating
Conditions
Storage
Conditions
Measurement
Range
BT104
22 - 42 cm
274 g (excl. batteries)
102 x 143 x 73 mm
IP21
Type BF
A01
Oscillometric measurement
0 - 299 mmHg (0.0 - 39.9 kPa)
Pressure:
Pulse:
±3 mmHg (0.4 kPa)
±5 %
Temperature:
Relative humidity:
Pressure:
5 - 40 °C
15 - 90 %
700 - 1060 kPa
Temperature:
Relative humidity:
Pressure:
-20 - 60 °C
0 - 93 %
700 - 1060 kPa
Systolic pressure:
Diastolic pressure:
Pulse:
Technical Specication
60 - 230 mmHg
(8.0 - 30.7 kPa)
40 - 130 mmHg
(5.3 - 17.3 kPa)
40 - 199 beats/min
Power Supply 4 x 1.5V alkaline AA batteries

16
Intended Use
Compliance Statement
Contact Information
This digital blood pressure monitor was designed to
measure the blood pressure and heartbeat rate of adults
with an arm circumference of 22 - 42 cm (9 - 17 inches).
This device is intended for indoor, home use.
FCC ID: XMR201404UC15
This device complies with part 15 of the FCC Rules.
Operation is subject to the following two conditions:
1. This device may not cause harmful interference, and
2. this device must accept any interference received,
including interference that may cause
undesired operation.
Please contact your program’s customer support.