manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Brasch
  6. •
  7. Security Sensor
  8. •
  9. Brasch GSE2-CM-24 User manual

Brasch GSE2-CM-24 User manual

Field Calibration Instructions
Overview
Brasch Generation 2 sensor boards are calibrated at the factory and have an expected
calibration duration of approximately two years. Environmental conditions may shorten or
prolong this period of time, but once the sensor parameters drift outside of acceptable
tolerances, the sensor board must be either be recalibrated or replaced.
This document outlines the procedure for performing field calibration of all Brasch Generation
2 sensor boards including:
Carbon Monoxide Models Nitrogen Dioxide Models Combination Models
G E2-CM-24 G E2-ND-24 G E2-NCM-24
G E2-CM-120 G E2-ND-120 G E2-NCM-120
G E2-CM-Remote G E2-ND-Remote G E2-NCM-Remote
TRN 2-CM-Analog TRN 2-ND-Analog TRN 2-NCM-Analog
TRN 2-CM-Digital TRN 2-ND-Digital TRN 2-NCM-Digital
GDCP-CM-Remote GDCP-ND-Remote GDCP-NCM-Remote
Required:
•Voltmeter
•Potentiometer Adjustment Tool
•100 PPM Carbon Monoxide Test Gas (for CO sensors)
•5.0 PPM Nitrogen Dioxide Test Gas (for NO2 sensors)
•Gas Flow Regulator
•Tubing
•Calibration Hood
Note: All of the above required parts (except a voltmeter) are included with Brasch CO and
NO2 Test Gas Kits.
FG01
Rev 1.2 – January 10, 2022 1
Procedure:
1. Turn power OFF to the sensor board
2. Move the calibration jumpers (JP4) so that both pins are covered by the jumpers
3. Turn power ON to the sensor board
4. Connect the negative voltmeter lead to COM (TP3)
5. Connect the positive voltmeter lead to AN OUT (TP7)
6. et the voltmeter to measure DC voltage (VDC)
a) If possible, set the meter to display the average voltage to smooth out readings
7. Use the potentiometer adjustment tool on R2 until the meter reads approximately 5.0 V
a) This number will fluctuate but should remain stable at or near 5.0 V
8. Remove the top calibration jumper on JP4
9. Apply test gas directly to the sensor
a) The voltage at TP7 will begin to drop
10.Once the voltage stabilizes, remove the bottom calibration jumper on JP4
a) tabilization requires approximately 5 minutes from the application of gas
11. Remove test gas from the sensor
12.Verify that the voltage at TP7 returns to approximately the same value as was set in
step 7
a) tabilization may take approximately 5 minutes from the removal of gas
13.Cycle power to the sensor board to verify that calibration data has been saved
14.Repeat this procedure for any remaining sensor boards
FG01
Rev 1.2 – January 10, 2022 2
Troubleshooting:
•If any of the following errors are present, calibration data was corrupted and this
procedure needs to be repeated.
◦tandalone Detectors: 9997
◦BM Transmitters: 0x0008
◦Control Panel Remotes: ensor # Calibration Invalid
•If achieving stable voltages is not possible, it may be necessary to replace the sensor
and/or the sensor board.
◦Once the useful lifespan of the sensor is exceeded, the sensor can no longer be
recalibrated. Both carbon monoxide and nitrogen dioxide sensors have an expected
lifespan of up to 10 years; however, this period can be reduced with increased
exposure to the target gas or sensor poisons. Replacing the sensor should yield
more stable voltages.
◦Over time, control electronics can degrade or suffer damage. If this is the case,
replacing the sensor board should yield more stable voltages.
FG01
Rev 1.2 – January 10, 2022 3
COM (TP3)
Calibration Jumpers
[ZERO]
[ PAN]
AN OUT (TP7)
R2 R1
FG01
Rev 1.2 – January 10, 2022 4
Revision: 1.2
Issue Date: January 10, 2022
Document Number: FG01
© Brasch Environmental Technologies, LLC
All Rights Reserved
Technical Support Contact Information
Phone: 314-291-0440 ext. 1006
Fax: 314-291-0646
Email: [email protected]
Website: www.braschenvtech.com/support-for-existing-systems
General Contact Information
Phone: 314-291-0440
Fax: 314-291-0646
Email: [email protected]
Website: www.braschenvtech.com

This manual suits for next models

17

Popular Security Sensor manuals by other brands

Ecolink WST-611 installation manual

Ecolink

Ecolink WST-611 installation manual

D-tect Systems Mini Rad-D Product guide

D-tect Systems

D-tect Systems Mini Rad-D Product guide

no-flame 83197 instruction manual

no-flame

no-flame 83197 instruction manual

DSC WLS904PL-433 installation instructions

DSC

DSC WLS904PL-433 installation instructions

Rycom SnapTRACK HPTX quick start guide

Rycom

Rycom SnapTRACK HPTX quick start guide

Niko SERVODAN 41-771 user manual

Niko

Niko SERVODAN 41-771 user manual

Azbil AUD10C1000 user manual

Azbil

Azbil AUD10C1000 user manual

Deteck DT10C manual

Deteck

Deteck DT10C manual

hyfire ALTAIR HFI-PA-05 manual

hyfire

hyfire ALTAIR HFI-PA-05 manual

ProVero ID-0055 Instruction

ProVero

ProVero ID-0055 Instruction

koban KDP1 360FP manual

koban

koban KDP1 360FP manual

ACS A1214 Expert Operation manual

ACS

ACS A1214 Expert Operation manual

EVERSPRING EVR HSP02 manual

EVERSPRING

EVERSPRING EVR HSP02 manual

DAITEM SH101AX installation manual

DAITEM

DAITEM SH101AX installation manual

Master Vega BT quick start guide

Master

Master Vega BT quick start guide

THORLABS APD110 Series Operation manual

THORLABS

THORLABS APD110 Series Operation manual

SECUMAR GOLF 275 HV ALU quick start guide

SECUMAR

SECUMAR GOLF 275 HV ALU quick start guide

Sunx NA1-11 instruction manual

Sunx

Sunx NA1-11 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.