2.4 Basic System Calibration
The following procedure can be used for quick, accurate adjustment and calibration of
vibration measurement systems for a reference acceleration level of 10 ms–2 RMS/14.14 ms–2
peak for Type 4294 (3.16 ms–2 RMS/4.47 ms–2 peak for Type 4294-002). Calibration using
reference values of velocity or displacement is also possible, as mentioned in step 6).
1) Attach the system transducer to the vibration table (finger-tight is sufficient) of the
calibrator using either a 10–32 UNF Stud YQ-2962 of Mounting Disc DB-2996. See
section 2.3.
2) Connect the transducer to the vibration measurement system using a suitable low-noise
cable, for example, that supplied with the transducer. Where a preamplifier with adjustable
sensitivity conditioning is employed, the accelerometer sensitivity specified on the
calibration chart should be entered prior to setting the measuring range of the read-out
instrument.
3) Switch on the preamplifier and indicating instrument and select the acceleration mode and
the highest indicating range on the appropriate instruments.
4) Press the button on the side of the calibrator to vibrate the transducer. The transducer is
now subjected to a sinusoidal acceleration of 10 ms–2 RMS ±3% (4294-002: 3.16 ms–2
RMS ±3%) at a frequency of 159.15 Hz (that is, ω= 1000 rad s–1).
5) Switch the range controls of the preamplifier and/or indicating instrument to obtain an on-
scale meter indication without the overload lamps of the instruments lighting.
The indicating instrument should now indicate a reference acceleration level of 10 ms–2
RMS (4294-002: 3.16 ms–2 RMS). If not, adjust the sensitivity potentiometer of one of the
instruments until the correct reference level is indicated. Once correctly set, avoid further
adjustment of the sensitivity potentiometer, as the calibration will be lost.
Note: If an indicating instrument with peak indicating mode is employed, then it should be
adjusted to indicate a reference level of 14.14 ms–2 peak (4294-002: 4.47 ms–2 peak).
6) For velocity and displacement calibration of the vibration measurement system, the
preamplifier/measuring instrument should be equipped with integration networks. The
appropriate measurement mode should therefore be selected before switching the
calibrator on. The relevant calibration levels are as follows:
Velocity: 10 mms–1 RMS/14.14 mms–1 peak ±3%
(4294-002: 3.16 mms–2 RMS/4.47 mms–2 peak ±3%)
Displacement: 10 μm RMS/14.14 μm peak ±3%
(4294-002: 3.16 μs–2 RMS/4.47 μs–2 peak ±3%)
Fig.2.3
Example of calibration of a
vibration measurement
system
4294 or 4294-002
Calibration Exciter
2635
Charge Amplifier
Digital
Voltmeter