8
8) BUFR_1 (First buffer): selects the value of the first buffer for the pH electrode calibration.
The preset buffers are compensated for temperature. The ATC indication is displayed in the
lower line. The USER buffer, defined by the user, is not compensated for temperature: to
change its value, go to the "SET USER BUFR" step.
Please see the paragraph dedicated calibration on page.14.
9) BUFR_2 (Second buffer): selects the value of the second buffer for the pH electrode cali-
bration. No standard solution values close (less than 2pH) to the first buffer are proposed.
This is also true for the USER buffer, defined by the user at the "SET USER BUFR" step: if,
for example BUFR_1=6.860 and USER=5.000, the BUFR_2 USER does not appear among
the values because it is too close to BUFR_1. The selection NIL (=no buffer) disables the
current buffer. The USER buffer, defined by the user, is not compensated for temperature.
Please see the paragraph dedicated calibration on page14.
10) BUFR_3 (Third buffer): selects the value of the third buffer for the pH electrode calibra-
tion. No standard solution values close (less than 2pH) to the first buffer are proposed. This
is also true for the USER buffer, defined by the user at the "SET USER BUFR" step: The se-
lection NIL (=no buffer) disables the current buffer. The USER buffer is not compensated
for temperature.
Please see the paragraph dedicated calibration on page 14.
11) SET USER BUFR (Set user pH buffer): this is a buffer whose value can be defined by the
user using the arrows; all values from 0 to 14pH are available. This buffer is not compen-
sated for temperature, so the buffer value must be set at the actual solution temperature. As
an alternative, the correct value according to temperature can be set in the calibration phase.
12) RCD MODE (Record mode): the instrument captures a pH/mV and a temperature value
every second. If the RCD MODE parameter is set to "pH" (factory default), the maximum
and minimum values displayed using FUNC/ENTER refer to pH: the indicated temperature
is that measured at the maximum and minimum pH and is not the maximum and minimum
temperature.
If the RCD MODE parameter is set to "tp" (=temperature), the maximum and minimum
values displayed using FUNC/ENTER refer to temperature: the indicated pH is that meas-
ured at the maximum and minimum temperature and is not the maximum and minimum pH.
Finally, if the RCD MODE parameter is set to "Indep" (=independent), the maximum and
minimum values displayed using FUNC/ENTER are independent: the indicated pH and
temperature are the maximum and minimum measured values but are not necessarily re-
ferred to the same measurement moment.
13) Probe type: the message "PRBE_TYPE" is scrolled in the comment line. The main line in
the center of the display shows the type of temperature probe connected to the instrument.
following probes can be connected to the input:
•temperature probes Pt100 complete with SICRAM module
•4 wire Pt100 probes through module TP47
•2 wire Pt100 probes through module TP47
•2 wire Ni1000 probes through module TP47
Upon turning on the instrument automatically detects the probes fitted with SICRAM mod-
ule, the Delta Ohm Pt1000 and Ni1000: the Probe Type menu item is configured by the in-
strument and cannot be modified by the user.
The temperature probes direct 4 wire Pt100, the Pt1000 and the Ni1000 are manufactured by
Delta OHM. When turned on they display the message "NO_PRBE_SER_NUM". In this