BW Technologies GasAlertMicro5 User manual

iERP: 128900
D6258/1 [English]
© BW Technologies 2009. All rights reserved.
1, 2, 3, 4, and 5-Gas Detector
Quick Reference Guide

Limited Warranty and Limitation Liability
BW Technologies LP (BW) warrants the product to be free from defects in material and workmanship under normal use and service for a period of
two years, beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer.
BW’s warranty obligation is limited, at BW’s option, to refund of the purchase price, repair or replacement of a defective product that is returned to
a BW authorized service center within the warranty period. In no event shall BW’s liability hereunder exceed the purchase price actually paid by the
buyer for the Product.
This warranty does not include:
a) fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use;
b) any product which in BW’s opinion, has been misused, altered, neglected or damaged, by accident or abnormal conditions of operation,
handling or use;
c) any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation of unapproved
parts on the product; or
The obligations set forth in this warranty are conditional on:
a) proper storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other applicable
recommendations of BW;
b) the buyer promptly notifying BW of any defect and, if required, promptly making the product available for correction. No goods shall be
returned to BW until receipt by the buyer of shipping instructions from BW; and
c) the right of BW to require that the buyer provide proof of purchase such as the original invoice, bill of sale or packing slip to establish that the
product is within the warranty period.
THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER’SSOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR APARTICULAR PURPOSE. BW SHALL NOT BE LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL, OR BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential
damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or
unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
Contacting BW Technologies by Honeywell
Email us at: info@gasmonitors.com
Visit BW Technologies by Honeywell’s website at: www.gasmonitors.com
USA: 1-888-749-8878 Canada: 1-800-663-4164
Europe: +44(0) 1295 700300 Other countries: +1-403-248-9226

1
GasAlertMicro 5/PID/IR
Introduction
The quick reference guide provides basic information to operate
the GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro
5 IR gas detectors. For complete operating instructions, refer
to the GasAlertMicro 5/PID/IR User Manual provided on the
CD-ROM.
The GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro
5 IR detectors (“the detector”) are designed to warn of hazard-
ous gas levels above user-defined alarm setpoints.
Note
Unless reference is made to a specific detector
model, the GasAlertMicro 5, GasAlertMicro 5 PID,
and GasAlertMicro 5 IR detectors are referred to as
GasAlertMicro 5/PID/IR.
Note
The detector is shipped with English as the default
displayed language. Additional languages provided
are French, German, Spanish, and Portuguese. The
screens for the additional languages are displayed on
the detector and in the corresponding quick reference
guide.
Safety Information - Read First
Use the detector only as specified in this quick reference guide
and the user manual, otherwise the protection provided by the
detector may be impaired.
The detector is a personal safety device. It is your responsibility
to respond properly to the alarm.
Read the following Cautions before using the detector.

GasAlertMicro 5/PID/IR
Quick Reference Guide
2
aCautions
•Warning: Substitution of components may impair Intrinsic
Safety.
• Caution: For safety reasons, this equipment must be
operated and serviced by qualified personnel only. Read and
understand the user manual completely before operating or
servicing.
• Charge the detector before first-time use. BW recommends
the detector be charged after every workday.
• Charge the battery pack immediately when a low battery
alarm occurs.
• Read and adhere to the battery cautions provided in
Replacing Battery Cells and Packs on page 17.
• BW recommends that the combustible sensor be checked
with a known concentration of calibration gas after any
exposure to contaminants/poisons such as sulfur
compounds, silicon vapors, halogenated compounds, etc.
• BW recommends to bump test the sensors before each
day’s use to confirm their ability to respond to gas by
exposing the detector to a gas concentration that exceeds
the alarm setpoints. Manually verify that the audible and
visual alarms are activated. Calibrate if the readings are not
within the specified limits.
• Calibrate the detector before first-time use and then on a
regular schedule, depending on use and sensor exposure to
poisons and contaminants. The sensors must be calibrated
regularly and at least once every 180 days (6 months).
• Calibrate only in a safe area that is free of hazardous gas in
an atmosphere of 20.9% oxygen.
• Only the combustible gas detection portion of this
instrument has been assessed for performance by CSA
International.
• The combustible sensor is factory calibrated to 50% LEL
methane. If monitoring a different combustible gas in the
% LEL range, calibrate the sensor using the appropriate
gas.
• Caution: High off-scale LEL readings may indicate an
explosive concentration.
• Protect the combustible sensor from exposure to lead
compounds, silicones, and chlorinated hydrocarbons.
Although certain organic vapors (such as leaded gasoline
and halogenated hydrocarbons) may temporarily inhibit
sensor performance, in most cases, the sensor will recover
after calibration.
• For use only in potentially explosive atmospheres where
oxygen concentrations do not exceed 20.9% (v/v).
• Any rapid up scaling reading followed by a declining or
erratic reading may indicate a gas concentration beyond
upper scale limit, which can be hazardous.
• Extended exposure of the GasAlertMicro 5, GasAlertMicro 5
PID, or GasAlertMicro 5 IR to certain concentrations of
combustible gases and air may stress the detector element
that can seriously affect its performance. If an alarm occurs
due to a high concentration of combustible gases,
recalibration should be performed, or if needed, the sensor
replaced.
• The BW pump module (M5-PUMP) is certified for use with
the GasAlertMicro 5, GasAlertMicro 5 PID, and
GasAlertMicro 5 IR models only.

3
GasAlertMicro 5/PID/IR
Safety Information - Read First
• Protect the PID sensor from exposure to silicone vapors.
• When calibrating O3and ClO2sensors that are located in
the Toxic 2 position of the detector, a single gas calibration
cap must be used to ensure accurate calibration.
• Replace the CO2sensor only in a safe area that is free of
hazardous gas in an atmosphere of 20.9% oxygen.
•Warning: The battery pack (M5-BAT08) is equipped with a
lithium battery that may present a risk of fire or chemical
burn hazard if misused. Do not disassemble, heat above
212ºF (100ºC), or incinerate.
•Warning: Do not use any other lithium batteries that are not
designed for use with the GasAlertMicro 5, GasAlertMicro5
PID, or GasAlertMicro5 IR detectors. Use of any other cell
can cause fire and/or explosion. To order and replace the
M5-BAT07 or the M5IR-BAT08 lithium battery pack, contact
BW Technologies by Honeywell.
•Warning: Lithium polymer cells exposed to heat at 266°F
(130°C) for 10 minutes can cause fire and/or explosion.
• Dispose of used lithium cells immediately. Do not
disassemble and do not dispose of in fire. Do not mix with
the solid waste stream. Spent batteries must be disposed
of by a qualified recycler or hazardous materials handler.
• Keep lithium cells away from children.
• Before using common products around sensors, refer to
Sensor Poisons and Contaminants on page 4.
The detector contains either alkaline batteries or a lithium
polymer battery. Refer to the following warnings.
ce Warning
This instrument contains alkaline batteries. Do not mix
with the solid waste stream. Spent batteries must be
disposed of by a qualified recycler or hazardous materials
handler.
ce Warning
This instrument contains a lithium polymer battery.
Dispose of lithium cells immediately. Do not disassemble
and do not dispose of in fire. Do not mix with the solid
waste stream. Spent batteries should be disposed of by a
qualified recycler or hazardous materials handler.

GasAlertMicro 5/PID/IR
Quick Reference Guide
4
Sensor Poisons and Contaminants
Several cleaners, solvents, and lubricants can contaminate and
cause permanent damage to sensors. Before using cleaners,
solvents, and lubricants in close proximity to the detector sen-
sors, read and adhere to the following caution and table.
aCaution
Use only the following BW Technologies by Honeywell
recommended products and procedures:
• Use water based cleaners.
• Use non-alcohol based cleaners.
• Clean the exterior with a soft, damp cloth.
• Do not use soaps, polishes, or solvents.
The following table lists common products to avoid using
around sensors.
Cleaners and
Lubricants Silicones Aerosols
Brake cleaners Silicone cleaners
and protectants
Bug
repellents
and sprays
Lubricants Silicone based
adhesives, sealants,
and gels Lubricants
Rust inhibitors Hand/body and
medicinal creams
containing silicone
Rust
inhibitors
Window and glass
cleaners Tissues containing
silicone
Window
and glass
cleaners
Dishsoaps Mold releasing agents
Citrus based
cleaners Polishes
Alcohol based
cleaners
Hand sanitizers
Anionicdetergents
Methanol (fuels
and antifreezes)

5
GasAlertMicro 5/PID/IR
Parts of the GasAlertMicro 5/PID/IR
Parts of the GasAlertMicro 5/PID/IR
Item Description
1Liquid crystal display (LCD)
2Pushbuttons
3Audible alarms
4Toxic 2 sensor
5Toxic 1/PID sensor (Micro 5 PID)
or
Toxic 1/IR (CO2) sensor (Micro 5 IR)
6Visual alarm indicators (LEDs)
7LEL sensor
8Oxygen sensor
9Pushbuttons
10 Battery pack
11 Alligator clip

GasAlertMicro 5/PID/IR
Quick Reference Guide
6
Display Elements
Item Description
1Alarm condition
2Automatically span sensor
3Gas cylinder
4Gas identifier bars
5Battery life indicator
6Passcode lock
7Data transmission
8Clock
9Stealth mode
10 Optional pump indicator
11 Datalog card indicator (optional)
12 Alarm condition (low, high, TWA, STEL, or
multi) or view TWA, STEL and peak (MAX)
gas exposures
13 Automatically zero sensor

7
GasAlertMicro 5/PID/IR
Pushbutton
Pushbutton
Pushbutton Description
A• To activate the detector, press A.
• To deactivate the detector, press and hold Auntil the countdown is complete.
G
• To increment the displayed value or scroll up, press G.
• To enter the user option menu, press Gand Hsimultaneously and hold until the countdown is
complete.
• To clear the TWA, STEL, and peak (MAX) gas exposure readings, press Cand G
simultaneously and hold until the countdown is complete.
• To view the date and time, alarm setpoints (TWA, STEL, low, and high) of all sensors, and the
LEL correction factor (if applicable), press G.
H
• To decrement the displayed value or scroll down, press H.
• To initiate calibration and setting alarm setpoints, press Cand Hsimultaneously and hold until
the countdown is complete.
C• To view the TWA, STEL, and peak (MAX) readings, press C.
• To acknowledge latched alarms, press C.

GasAlertMicro 5/PID/IR
Quick Reference Guide
8
Connecting the Gas Cylinder to the Detector Single Gas Calibration Cap
aCaution
If an O3or ClO2sensor is located in the Toxic 2 sensor
position, a single gas calibration cap must be used to
ensure accurate calibration.
Only use the single gas calibration cap during the
calibration process.
To calibrate O3and ClO2sensors using the single gas calibra-
tion cap, refer to the following procedures and illustrations.
1. Insert the cap into the Toxic 2 sensor position on the
detector. Press firmly until the release tabs click.

9
GasAlertMicro 5/PID/IR
Connecting the Gas Cylinder to the Detector
2. Connect the calibration hose to the gas cylinder
and to the intake inlet on the cap.
Note
The arrow on the cap indicates the direction of gas
flow from intake to outtake.
Removing the Cap
Using your thumb, push forward against both the inlet and
outlet simultaneously to remove the cap from the detector.

GasAlertMicro 5/PID/IR
Quick Reference Guide
10
Confined Space Sampling
To measure hazardous gas in a confined space, refer to the
following illustration and steps 1-5.
aWarning
To measure hazardous gas in a confined space, the
sample probe must be used with the pump module.
aCaution
To prevent the teflon lining inside the Tygon tubing
from causing a blockage when connecting it to the
sample probe, the end of the tubing must be flared.
Refer to steps 2 and 3.
1. Attach the pump connector end of the teflon-lined
Tygon tubing to the pump module.
2. Gently insert needle nose pliers into the other end
of the teflon-lined Tygon tubing. Using a circular
motion, flare the end of the tubing.
3. Connect the flared end of the tubing to the sample
probe. Ensure the teflon-lining does not separate
from the Tygon tubing, as it will block the tube and
generate a pump alarm.
4. Activate the detector. Ensure all connections are
secure before sampling.
5. Insert the sample probe into the confined space.
aWarning
Depending upon the length of the tubing and the type of
gas in the confined space, allow a minimum of 3 seconds
per foot of tubing to ensure the readings stabilize before
entering the area.
Example: 50 ft. = 2.5 minutes

11
GasAlertMicro 5/PID/IR
Calibration
Calibration
Calibrate only in a safe area in an
atmosphere of 20.9% oxygen.
1. Activate the detector. To enter calibration,
press and hold Cand H
simultaneously. The detector beeps and
flashes to the corresponding countdown.
The LCD then displays Starting
calibration.
2. flashes while the detector
zeros all of thesensors (except CO2) and
calibrates the oxygen sensor. If a sensor
fails to auto zero, that sensor will bypass
the span.
If calibrating a Micro 5 or Micro 5 PID,
proceed to step #4.
Step #3 for Micro 5 IR Only
3. The Zero-CO2? screen displays. Press
Cto zero the CO2sensor, or press Ato
bypass.
If Cis pressed to zero the CO2sensor,
the following screens display:
- Apply CO2zero gas now (nitrogen
must be used to zero the CO2sensor).
- The Auto-Zero CO2screen displays,
and flashes.
4. The following three screens display:
- Apply span gas now to calibrate
- or press Cto select sensor(s)
- or press Ato skip calibration
If none of the buttons are pressed,
proceed to step #6.
If Cis pressed, proceed to step #5.
If Ais pressed, proceed to the end of
step #7.
5. Select which sensor to span. Press G
or Hto scroll to the required sensor
and then press Cto select. Sensors
must be spanned in the following order:
- Exotics (NH3, ClO2, O3, and CO2)
- Single gases
- Quad gases (H2S, CO, O2and LEL).
6. Attach the calibration cap and apply gas
at a flow rate of 500 ml/min.
Kflashes while the detector determines
which gas is being applied.
After 30 seconds, flashes and
a countdown displays while the detector
completes the span.

GasAlertMicro 5/PID/IR
Quick Reference Guide
12
For additional information about performing calibrations and
bump tests, refer to the GasAlertMicro 5/PID/IR User Manual.
7. When the span is complete, the following
three screens display:
- Calibration successful
- Press Gto apply a new cal gas
- Press Hto end span
Repeat steps #4-7 to calibrate the
remaining sensors.
The LCD displays the following options:
- Press Cto set the calibration due
dates or press Ato bypass.
8. Press Gor Hto change the calibration
due date. Press Cto accept the value
and proceed to the next due date. If a
sensor fails or does not span, the
calibration due date cannot be changed
for that sensor.
The LCD displays the following options:
- Press Cto change the calibration due
dates or press Ato bypass.
9. Press Gor Hto change the alarm
setpoint. Press Cto save the value and
proceed to the next setpoint.
Define the remaining setpoints. The
detector beeps twice when all of the
alarm setpoints have been defined or
bypassed.
10.When calibration is complete, Saving
calibration displays.
Note
Only use the calibration cap and
single gas calibration cap during
calibration and bump tests.

13
GasAlertMicro 5/PID/IR
Alarms
Alarms
Refer to the following table for information about alarms and corresponding screens. During an alarm condition, the backlight
activates and the LCD displays the ambient gas readings.
Alarm Screen Alarm Screen
Low Alarm
• Fast beep
• Slow flash
•Land target gas bar flash
• Vibrator alarm activates
TWA Alarm
• Fast beep
• Slow flash
•L and target gas bar flash
• Vibrator alarm activates
High Alarm
• Constant beep
• Fast flash
•L and target gas bar flash
• Vibrator alarm activates
STEL Alarm
• Constant beep
• Fast flash
•Land target gas bar flash
• Vibrator alarm activates

14
GasAlertMicro 5/PID/IR
Quick Reference Guide
Alarm Screen Alarm Screen
Multi Alarm
• Alternating low and high alarm beep
and flash
•Land target gas bars flash
• Vibrator alarm activates
Over Limit (OL) Alarm
• Fast beep and flash
•Land target gas bar flash
• Vibrator alarm activates
Sensor Alarm
• One beep every 15 seconds
•FAIL flashes above the failed sensor
Automatic Deactivation Alarm
• Eight beeps and flashes
• displays
• Vibrator alarm temporarily activates
•Battery depleted screen displays and
the detector deactivates
Low Battery Alarm
• One beep and two flashes every
25 seconds
• flashes
aCaution
Charge the battery immediately when
a low battery alarm occurs.
Normal Deactivation
• Three beeps and flashes

15
GasAlertMicro 5/PID/IR
Alarms
Note
If enabled, during an alarm condition the Latching Alarms option causes the low and high gas alarms (audible, visual, and
vibrator) to persist until the alarm is acknowledged and the gas concentration is below the low alarm setpoint. The alarm can
be temporarily deactivated by pressing C, but the LCD displays the high peak concentration until the alarm no longer exists.
Local regulations may require the Latching Alarms option be enabled.
Alarm Screen Alarm Screen
Confidence Beep
• One beep, one flash, and one
vibration every 10 seconds
Note
The Confidence Beep option is
automatically disabled during a
low battery alarm.
Pump Alarm
• Screen flashes:
- Pump flow change detected
- Check for blocked inlet
- or press Cto run a pump test
• Two fast beeps and alternating
flashes
• Vibrator alarm activates
•Land Jflash
MMC Fail Alarm
• One beep every 5 seconds
• flashes

GasAlertMicro 5/PID/IR
Quick Reference Guide
16
User Options Menu
To access the user options, press and hold Gand Hsimulta-
neously until the detector completes the countdown. To scroll
through the options press Gor H. Press Cto select the
option. The following options are available.
Exit: Exits the user options menu.
User Options:
•Backlght (backlight): Enables/disables the automatic back-
light in low-light conditions.
•Confibeep (confidence beep): If enabled, the confidence
beep provides continuous confirmation that the detector
is operating correctly (audible beep every 10 seconds).
•Due-lock (calibration user lockout): If enabled, upon
startup a passcode is required to operate a detector
that is overdue for calibration.
•Latch (latched alarm): Enable to ensure an alarm persists
until it is acknowledged (press Cto acknowledge).
•Passcode (passcode protection): Enable to prevent
unauthorized personnel from accessing the user options
menu, calibration function, and alarm setpoint adjust
function.
•Safe (safe mode): If enabled, Safe displays continuously on
the LCD unless an alarm condition occurs.
Sensors:
•Sens on (sensor enabled): Enables/disables a sensor
(the detector continues to operate if a sensor is disabled).
aWarning
Use extreme caution when disabling a sensor. The
disabled sensor cannot detect and alarm against the
applicable gas.
•Span gas: Define the span gas concentration for each sensor
(must match the gas concentration on the gas cylinder).
•Stel period (Short Term Exposure Limit): The short term expo-
sure limit (STEL) provides protection for workers from over
exposure to high concentrations of gas, and is based on user-
defined 5-15 minute intervals.
When the maximum STEL is reached, the detector
alarms to notify the worker. Set the STEL period from
5-15 minutes (applicable to toxic sensors only).
•TWA method (Time Weighted Average): This option is a safety
measure that calculates the accumulated averages of gases to
warn the worker when the maximum average is accumulated.
Select one of the following options:
- OSHA Method: 8 hour moving average - oldest value
(first hour) is replaced by the newest value (ninth hour).
- ACGIH Method: Infinite accumulated average to 8 hours -
total accumulation, whether it is 2 hours or 8 hours.
• Resolution: Defines the resolution of the gas measurement
as either regular or extra (if applicable).

17
GasAlertMicro 5/PID/IR
Maintenance
•%volCO
2:Enable to display the carbon dioxide (CO2) reading
as %vol.
•%volCH
4:Enable to display the LEL readings as %vol
assuming a methane environment.
•Correction Factor (%): Enter a compensation factor for
hydrocarbons other than methane. The factor can only
be applied if the LEL sensor has been calibrated with
methane (LEL only).
•Autocal (automatic oxygen calibration) Enable/disable
the detector to automatically calibrate the oxygen sensor
during startup.
Logger (Datalogging): Define how often the detector records a
datalog sample (once every 1 to 127 seconds).
Clock: Define the date and time for the detector.
Language: Displays the LCD screens in English, Français
(French), Deutsch (German), Español (Spanish), or Prtugês
(Portuguese).
Note
The detector is shipped displaying English as the
default language.
Maintenance
To maintain the detector in good operating condition, perform
the following basic maintenance as required.
• Calibrate, bump test, and inspect the detector on a regular
schedule.
• Maintain an operations log of all maintenance, bump tests,
calibrations, and alarm events.
• Clean the exterior with a soft damp cloth. Do not use
solvents, soaps, or polishes. Refer to Sensor Poisons and
Contaminants on page 4.
• Do not immerse the detector in liquids.
Replacing Battery Cells and Packs
aWarning
To avoid personal injury and/or property damage, adhere
to the following battery cautions:
• Replace the alkaline cells or rechargeable battery pack
immediately when the detector emits a low
battery alarm.
• Use only batteries that are recommended by BW
Technologies by Honeywell.
• Use only approved alkaline batteries that are properly
installed in the battery pack. Refer to Specifications.
• To order lithium battery packs (M5-BAT08/M5-BAT07),
contact BW Technologies by Honeywell.

GasAlertMicro 5/PID/IR
Quick Reference Guide
18
• Charge the batteries and battery packs using only a
recommended BW charger. Failure to adhere to this
caution can lead to fire and/or explosion.
• The detector must be deactivated to charge the
battery pack.
• Do not calibrate the detector immediately after
charging is complete.
• Both the lithium battery pack and the alkaline battery
pack are user-changeable in hazardous locations, but
the alkaline battery cells inside the pack can only be
replaced in a safe area that is free of hazardous gas.
•Warning: The M5-BAT08 and M5-BAT07 battery packs
are equipped with lithium batteries that can present a
risk of fire or chemical burn hazard if misused. Do not
recharge, disassemble, heat above 212°F (100°C), or
incinerate.
• Warning: Do not use any other lithium batteries with
the GasAlertMicro 5, GasAlertMicro 5 PID, and
GasAlertMicro 5 IR detectors. Use of any other cell
can cause fire and/or explosion.
• Warning: Lithium polymer cells exposed to heat at
266°F (130°C) for 10 minutes can cause fire and/or
explosion.
• Dispose of used lithium cells immediately. Do not
disassemble and do not dispose of in fire. Do not mix
with the solid waste stream. Spent batteries must be
disposed of by a qualifies recycler or hazardous
materials handler.
• Keep lithium cells away from children.
Replacing Alkaline Batteries
aWarning
Always deactivate the detector before removing the
battery pack. The battery packs are user-changeable in
hazardous locations, but the alkaline battery cells inside
the pack must be changed in a safe area that is free of
hazardous gas.
This manual suits for next models
2
Table of contents
Other BW Technologies Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

F&F
F&F LE-03MW user manual

Larson Electronics
Larson Electronics LZRWL-CRN-RED-200MW-120V quick start guide

Bushnell
Bushnell Elite 1600 Rangefinder owner's manual

Twinno
Twinno T6055 operating manual

Hanna Instruments
Hanna Instruments pH 210 instruction manual

Emerson
Emerson Energy Meter Installation and operation manual