Campbell NL115 User manual

NL115
Ethernet Interface and
CompactFlash®Module
Issued: 24.3.14
Copyright © 2006-2014 Campbell Scientific, Inc.
Printed under licence by Campbell Scientific Ltd.
CSL 629
USER MANUAL


Guarantee
This equipment is guaranteed against defects in materials and workmanship.
This guarantee applies for twelve months from date of delivery. We will
repair or replace products which prove to be defective during the guarantee
period provided they are returned to us prepaid. The guarantee will not apply
to:
•Equipment which has been modified or altered in any way without the
written permission of Campbell Scientific
•Batteries
•Any product which has been subjected to misuse, neglect, acts of God or
damage in transit.
Campbell Scientific will return guaranteed equipment by surface carrier
prepaid. Campbell Scientific will not reimburse the claimant for costs incurred
in removing and/or reinstalling equipment. This guarantee and the Company’s
obligation thereunder is in lieu of all other guarantees, expressed or implied,
including those of suitability and fitness for a particular purpose. Campbell
Scientific is not liable for consequential damage.
Please inform us before returning equipment and obtain a Repair Reference
Number whether the repair is under guarantee or not. Please state the faults as
clearly as possible, and if the product is out of the guarantee period it should
be accompanied by a purchase order. Quotations for repairs can be given on
request. It is the policy of Campbell Scientific to protect the health of its
employees and provide a safe working environment, in support of this policy a
“Declaration of Hazardous Material and Decontamination” form will be
issued for completion.
When returning equipment, the Repair Reference Number must be clearly
marked on the outside of the package. Complete the “Declaration of
Hazardous Material and Decontamination” form and ensure a completed copy
is returned with your goods. Please note your Repair may not be processed if
you do not include a copy of this form and Campbell Scientific Ltd reserves
the right to return goods at the customers’ expense.
Note that goods sent air freight are subject to Customs clearance fees which
Campbell Scientific will charge to customers. In many cases, these charges are
greater than the cost of the repair.
Campbell Scientific Ltd,
Campbell Park, 80 Hathern Road,
Shepshed, Loughborough, LE12 9GX, UK
Tel: +44 (0) 1509 601141
Fax: +44 (0) 1509 601091
www.campbellsci.co.uk


PLEASE READ FIRST
About this manual
Please note that this manual was originally produced by Campbell Scientific Inc. primarily for the
North American market. Some spellings, weights and measures may reflect this origin.
Some useful conversion factors:
Area: 1 in2(square inch) = 645 mm2
Length: 1 in. (inch) = 25.4 mm
1 ft (foot) = 304.8 mm
1 yard = 0.914 m
1 mile = 1.609 km
Mass: 1 oz. (ounce) = 28.35 g
1 lb (pound weight) = 0.454 kg
Pressure: 1 psi (lb/in2) = 68.95 mb
Volume: 1 UK pint = 568.3 ml
1 UK gallon = 4.546 litres
1 US gallon = 3.785 litres
In addition, while most of the information in the manual is correct for all countries, certain information
is specific to the North American market and so may not be applicable to European users.
Differences include the U.S standard external power supply details where some information (for
example the AC transformer input voltage) will not be applicable for British/European use. Please
note, however, that when a power supply adapter is ordered it will be suitable for use in your country.
Reference to some radio transmitters, digital cell phones and aerials may also not be applicable
according to your locality.
Some brackets, shields and enclosure options, including wiring, are not sold as standard items in the
European market; in some cases alternatives are offered. Details of the alternatives will be covered in
separate manuals.
Part numbers prefixed with a “#” symbol are special order parts for use with non-EU variants or for
special installations. Please quote the full part number with the # when ordering.
Recycling information
At the end of this product’s life it should not be put in commercial or domestic refuse
but sent for recycling. Any batteries contained within the product or used during the
products life should be removed from the product and also be sent to an appropriate
recycling facility.
Campbell Scientific Ltd can advise on the recycling of the equipment and in some cases
arrange collection and the correct disposal of it, although charges may apply for some
items or territories.
For further advice or support, please contact Campbell Scientific Ltd, or your local agent.
Campbell Scientific Ltd, Campbell Park, 80 Hathern Road, Shepshed, Loughborough, LE12 9GX, UK
Tel: +44 (0) 1509 601141 Fax: +44 (0) 1509 601091
www.campbellsci.co.uk


i
Contents
PDF viewers: These page numbers refer to the printed version of this document. Use the
PDF reader bookmarks tab for links to specific sections.
1. Introduction................................................................1
2. Cautionary Statements..............................................1
3. Initial Inspection ........................................................2
4. Quickstart...................................................................2
4.1 Physical Setup......................................................................................2
4.2 Communicating via Ethernet................................................................3
4.2.1 Step 1: Configure Datalogger........................................................3
4.2.2 Step 2: LoggerNet Setup...............................................................4
4.2.3 Step 3: Connect.............................................................................4
4.3 Programming the Datalogger to Send Data to the NL115....................5
4.4 CF Card Data Retrieval........................................................................5
5. Overview.....................................................................6
5.1 Status LEDs..........................................................................................6
5.2 Power ...................................................................................................6
5.3 CF Card Data Retention.......................................................................6
6. Specifications ............................................................7
7. Operation....................................................................8
7.1 TCP/IP Functionality ...........................................................................8
7.1.1 Communicating Over TCP/IP.......................................................8
7.1.1.1 Data Callback.....................................................................8
7.1.1.2 Datalogger-to-Datalogger Communication........................9
7.1.2 HTTP Web Server.......................................................................10
7.1.3 FTP..............................................................................................11
7.1.3.1 FTP Server .......................................................................11
7.1.3.2 FTP Client ........................................................................13
7.1.4 Telnet..........................................................................................14
7.1.5 Ping.............................................................................................14
7.1.6 Serial Server................................................................................14
7.1.6.1 Serial Input.......................................................................14
7.1.6.2 Serial Output ....................................................................14
7.1.7 TCP ModBus ..............................................................................15
7.1.8 DHCP..........................................................................................15
7.1.9 DNS ............................................................................................15
7.2 File Formats .......................................................................................15
7.2.1 Data Files ....................................................................................15
7.2.2 Program Files..............................................................................15
7.2.3 Power-up Files (Powerup.ini) .....................................................16
7.2.3.1 Creating and Editing Powerup.ini ....................................16
7.2.3.2 Applications .....................................................................17
7.2.3.3 Program Execution...........................................................17
7.2.3.4 Example Powerup.ini Files...............................................18
7.2.4 Camera Files ...............................................................................19
7.3 Programming......................................................................................19

ii
7.3.1 The CardOut() Instruction...........................................................19
7.3.2 Program Examples......................................................................19
7.3.2.1 Ring Mode........................................................................19
7.3.2.2 Fill-and-Stop Mode..........................................................20
7.3.2.3 Mixed Modes ...................................................................20
7.3.3 Table Size and Mode ..................................................................21
7.4 CF Card Data-Retrieval Details .........................................................21
7.4.1 Via a Communication Link.........................................................21
7.4.1.1 Fast Storage/Data-Collection Constraints ........................22
7.4.2 Transporting CF Card to Computer ............................................22
7.4.2.1 Converting File Formats...................................................22
7.4.2.2 Reinserting the Card.........................................................23
7.4.2.3 Card Swapping.................................................................23
Appendix
A. CF Card Maintenance ............................................ A-1
A.1 Formatting CF Card .........................................................................A-1
A.1.1 Windows Explorer ....................................................................A-1
A.1.2 CR1000KD ...............................................................................A-2
A.1.3 LoggerNet File Control.............................................................A-2
A.2 Checking CF Card Integrity.............................................................A-3
Figures
4-1. NL115 attached to a CR1000...............................................................2
4-2. DevConfig setup ..................................................................................3
4-3. LoggerNet setup...................................................................................4
6-1. NL115 Ethernet/CompactFlash Module ..............................................8
7-1. Datalogger home page........................................................................11
7-2. FTP root directory..............................................................................12
7-3. FTP CRD directory ............................................................................12
7-4. CardConvert.......................................................................................23
Table
7-1. Powerup.ini Commands.....................................................................17

1
NL115 Ethernet and CompactFlash®
Module
1. Introduction
Campbell Scientific’s NL115 Ethernet/CompactFlash®Module provides two
independent capabilities: (1) it enables 10Base-T Ethernet communications and
(2) stores data on a removable CompactFlash (CF) card. It allows the datalogger
to communicate over a local network or a dedicated Internet connection via
TCP/IP. It also expands on-site data storage and provides the user with a
convenient method of transporting data from the field back to the office. This
small, rugged communication device connects to the 40-pin peripheral port on a
CR1000 or CR3000 datalogger.
This manual describes how to use LoggerNet to connect to your datalogger with
an NL115 attached. You can also use other software packages such as PC400,
RTDAQ, or LoggerLink Mobile Apps for iOS and Android.
Campbell Scientific recommends that only FMJ CF cards be used with the NL115.
For more information on CF cards, see Campbell Scientific’s application note CF
Card Information (3SM-F).Note that it is not necessary to purchase the FMJ CF
cards directly from Campbell Scientific as long as the FMJ model number matches
those shown in the application note.
Before using the NL115, please study:
Section 2, Cautionary Statements
Section 3, Initial Inspection
Section 4, Quickstart
The Quickstart section explains how to quickly begin using an NL115 for
straightforward Ethernet communications and data-storage operations. The
remainder of the manual is a technical reference which describes in detail such
operations as: TCP/IP functionality, file formats, datalogger programming, and
data retrieval.
2. Cautionary Statements
The NL115 is rugged, but it should be handled as a precision scientific
instrument.
The #28033 surge suppressor and/or a shielded 10Base-T Ethernet cable
should be used for locations susceptible to power surges and for cable lengths
longer than 9 ft.
Always power down the datalogger before installing to or removing the
NL115 from the datalogger.
The first time an NL115 is attached to a datalogger, the datalogger’s memory
has to be reorganized to allow room in memory for the IP stack. To avoid the
loss of data, collect your data before attaching the NL115 to a datalogger.
An NL115 with a serial number less than 10297 requires a firmware update to
function properly when attached to a CR3000 datalogger with a serial number
greater than 6260. This update must be performed by Campbell Scientific. If
you require this firmware update, see the Assistance section at the front of this
manual for information on returning your NL115 to Campbell Scientific.

NL115 Ethernet and CompactFlash®Module
2
3. Initial Inspection
Upon receipt of the NL115, inspect the packaging and contents for damage. File
damage claims with the shipping company.
4. Quickstart
This section describes the basics of communicating via Ethernet and storing and
retrieving datalogger data. These operations are discussed in detail in Section 7,
Operation.
4.1 Physical Setup
Always power down the datalogger before installing to or
removing the NL115 from the datalogger.
After powering down the datalogger, plug the NL115 into the datalogger
peripheral port (see Figure 4-1). Attach Ethernet cable to the 10Base-T port. If
using the #28033 surge protector, connect the other end of the Ethernet cable to
the #28033 and connect another Ethernet cable to the other end of the #28033.
Restore power to the datalogger. Insert formatted CF card. (For instructions on
formatting a CF card, see Appendix A, CF Card Maintenance.)
A CF card does not need to be present in order to use the NL115’s
TCP/IP functionality.
Figure 4-1. NL115 attached to a CR1000
CAUTION
NOTE

User Manual
3
4.2 Communicating via Ethernet
4.2.1 Step 1: Configure Datalogger
a. Connect serial cable from PC COM port to the datalogger RS-232 port.
b. Open Campbell Scientific’s Device Configuration Utility. Select the device
type of the datalogger (CR1000 or CR3000), the appropriate serial port, and
baud rate. Connect to the datalogger.
c. Under the TCP/IP tab, input the IP Address, Subnet Mask, and IP gateway.
These values should be provided by your network administrator.
Figure 4-2. DevConfig setup
d. Press the Apply button to save the changes and then close the Device
Configuration Utility.
A temporary IP address may be obtained from a DHCP server. For
more information, see Section 7.1.8, DHCP.
The NL115 must be connected to the datalogger before configuring
the datalogger with the Device Configuration Utility. If it is not
connected, the TCP/IP settings will not be displayed.
NOTE
NOTE

NL115 Ethernet and CompactFlash®Module
4
4.2.2 Step 2: LoggerNet Setup
The next step is to run LoggerNet and configure it to connect to the datalogger via
the Ethernet port.
a. In LoggerNet’s Setup Screen, press Add Root and choose IPPort. Input the
datalogger’s IP address and port number. The IP address and port number
are input on the same line separated by a colon. (The datalogger’s default
port number is 6785. It can be changed using Device Configuration Utility
or by modifying its value in the Status Table.)
b. Add a PakBus port and set the desired baud rate.
c. Add the datalogger (CR1000 or CR3000). Input the PakBus address of the
datalogger.
Figure 4-3. LoggerNet setup
4.2.3 Step 3: Connect
You are now ready to connect to your datalogger using the LoggerNet Connect
Screen.
Datalogger program transfer, table-data display, and data collection are now
possible.

User Manual
5
4.3 Programming the Datalogger to Send Data to the NL115
The CardOut() instruction is used in the datalogger program to send data to the
CF card. CardOut() must be entered within each DataTable declaration that is to
store data on the CF card. The file is saved on the card with the name
stationname.tablename and a .DAT extension.
The CardOut() instruction has the following parameters:
StopRing: A constant is entered for the StopRing parameter to specify whether the
DataTable created should be a ring-mode table (0) or a fill-and-stop table (1).
Size: The Size parameter is the minimum number of records that will be included
in the DataTable. If –1000 is entered, the size of the file on the card will be the
same as the size of the internal table on the datalogger. If any other negative
number is entered, the memory that remains after creating any fixed-size tables on
the card will be allocated to this table. If multiple DataTables are set to a negative
number, the remaining memory will be divided among them. The datalogger
attempts to size the tables so that they will be full at the same time.
In the following example, the minimum batt_voltage and a sample of PTemp are
written to the card each time the data table is called. The StopRing parameter is 0
for ring mode. This means that once the data table is full, new data will begin
overwriting old data. The size parameter is –1, so all available space on the card
will be allocated to the table.
DataTable(Table1,1,-1)
CardOut(0 ,-1)
Minimum(1,batt_volt,FP2,0,False)
Sample(1,PTemp,IEEE4)
EndTable
To prevent losing data, collect data from the CF card before
sending the datalogger a new or modified program. When a
program is sent to the datalogger using the Send button in the
Connect Screen of LoggerNet or PC400, an attribute is sent
along with the program that commands the datalogger to
erase all data on the CF card from the currently running
program.
4.4 CF Card Data Retrieval
Data stored on cards can be retrieved through a communication link to the
datalogger or by removing the card and carrying it to a computer with a CF
adapter. With large files, transferring the CF card to a computer may be faster
than collecting the data over a communication link. CF card-data retrieval is
discussed in detail in Section 7.4, CF Card Data-Retrieval Details.
Removing a card while it is active can cause garbled data and
can actually damage the card. Always press the Initiate
Removal button and wait for a green light before removing
card.
CAUTION
CAUTION

NL115 Ethernet and CompactFlash®Module
6
LoggerNet’s File Control should not be used to retrieve data
from a CF card. Using File Control to retrieve the data can
result in a corrupted data file.
5. Overview
The NL115 Ethernet/CompactFlash Module enables 10Base-T Ethernet
communication with the datalogger. It also has a slot for a Type I or Type II
CompactFlash (CF) card (3.3 V, 75 mA).
The NL115/CF card combination can be used to expand the datalogger’s memory,
transport data/programs from the field site(s) to the office, upload power-up
functions, and store JPEG images from the CC640 camera. Data stored on cards
can be retrieved through a communications link to the datalogger or by removing
the card and carrying it to a computer. The computer can read the CF card either
with the CF1 adapter or #17752 reader/writer. The CF1 adapter allows the PC’s
PCMCIA card slot to read the CF card; the #17752 reader/writer allows the PC’s
USB port to read the CF card. User-supplied CF adapters may also be used.
LoggerNet’s File Control should not be used to retrieve data
from a CF card. Using File Control to retrieve the data can
result in a corrupted data file.
5.1 Status LEDs
There is one red-green-orange LED (light-emitting diode) and two buttons: Initiate
Removal and eject. The LED indicates the status of the module. The LED will
flash red when the CF card is being accessed, solid green when it is OK to remove
the card, solid orange to indicate an error, and flashing orange if the card has been
removed and has been out long enough that CPU memory has wrapped and data is
being overwritten without being stored to the card. The Initiate Removal button
must be pressed before removing a card to allow the datalogger to store any
buffered data to the card and then power off. The eject button is used to eject the
CF card. Note that if the eject button cannot be pressed, it may have been disabled
by bending it to the right. Straighten and press the eject button to eject the CF
card.
5.2 Power
The Ethernet/CompactFlash module is powered by 12 Vdc received from the
datalogger through the peripheral port.
Additional power consumption occurs if the Status LED is continuously lit. At
12 Vdc, an additional 1 mA is drawn by the red or green LED; an additional
2 mA is drawn by the orange LED.
5.3 CF Card Data Retention
The module accepts CompactFlash (CF) cards which do not require power to
retain data.
Typically, a CF card can be erased and rewritten a minimum of 100,000 times.
Industrial CF cards, graded for 2,000,000 write cycles, are recommended for most
applications.
CAUTION
CAUTION

User Manual
7
6. Specifications
Datalogger Compatibility: CR1000, CR3000
Storage Capacity: Depends on card size (up to 16 GB supported;
to use card size greater than 2 GB, the
datalogger operating system must be OS 25 or
greater)
Power Requirements: 12 V supplied through datalogger’s peripheral
port
Current Drain: 19 mA (CR1000 with NL115 attached, no
Ethernet cable attached, not actively
communicating over Ethernet nor accessing
the CF card)
20 mA (CR1000 with NL115, Ethernet cable
attached)
20 mA (CR1000 with NL115 communicating
over Ethernet)
43 mA (CR1000 with NL115 communicating
over Ethernet and accessing CF card)
2 mA (Ethernet port has been put to sleep
using the IPNET Power() CRBasic
instruction)
Add 1 mA to current drain if red or green
Status LED is continuously on.
Add 2 mA to current drain if orange Status
LED is continuously on.
Operating Temperature Range: –25
to +50
C Standard
–40
to +85
C Extended
EMI and ESD Protection: Meets requirements for a class A device
under European Standards
Application of Council Directive(s):
89/336/EEC as amended by 89/336/EEC and
93/68/EEC
Standards to which conformity is declared:
EN55022-1; 1995 and EN50082-1: 1992
Cable Requirements: Use a straight-through Ethernet cable when
the cable is run from the hub to the NL115.
Use a crossover Ethernet cable when the
cable is run directly from the computer to the
NL115. Use a shielded cable and/or the
#28033 surge suppressor when the cable
length is more than 9 ft.
Typical Access Speed: 200 to 400 kbits s–1
Memory Configuration: User-selectable for either ring style (default)
or fill-and-stop.

NL115 Ethernet and CompactFlash®Module
8
Software Requirements: LoggerNet 3.2 or later
PC400 1.3 or later
Dimensions: 10.2 x 8.9 x 6.4 cm (4.0 x 3.5 x 2.5 in)
Weight: 154 g (5.4 oz)
Figure 6-1. NL115 Ethernet/CompactFlash Module
7. Operation
7.1 TCP/IP Functionality
This section describes the main TCP/IP functionality of a datalogger with an
NL115 attached. Additional functionality may be added in the future. For more
information, refer to the Information Services section of the datalogger manual
and CRBasic Editor Help.
7.1.1 Communicating Over TCP/IP
Once the datalogger, the NL115, and LoggerNet have been set up as described in
Sections 4.1, Physical Set-up, and 0, Communicating via Ethernet, communication
is possible over TCP/IP. This includes program send and data collection. These
are straightforward operations and are accomplished through LoggerNet’s
Connect Screen. For more information, see the LoggerNet manual.
Data callback and datalogger-to-datalogger communications are also possible over
TCP/IP, as well as the creation of simple HTML pages to view datalogger
variables using a web browser.
7.1.1.1 Data Callback
The following program is an example of doing data callback over TCP/IP. It first
checks to see if a port to the LoggerNet Server already exists. (The LoggerNet

User Manual
9
Server is assumed to be at the default PakBus address, 4094.) If not, a socket to
LoggerNet is opened using the TCPOpen() instruction. The SendVariables()
instruction is then used to send data.
PROGRAM
' CR1000
' IP_Callback.cr1
' LoggerNet server Pak Bus Address assumed = 4094
' PC IP address assumed = 192.168.7.231
' LoggerNet IPPort "IP Port Used for Call-Back" = 6785
' LoggerNet IPPort "Call-Back Enabled" is checked
' LoggerNet CR1000 "Call-Back Enabled" is checked
' LoggerNet PakBusPort "PakBus Port Always Open" is checked
' IP Call-back using auto-discover (-1) neighbor in SendVariables
Public PanelTemperature, BatteryVoltage, Result1, dummy1
Dim Socket as LONG
DataTable (CLBK1,1,1000)
DataInterval (0,0,Sec,10)
Sample (1,PanelTemperature,FP2)
Sample (1,BatteryVoltage,FP2)
EndTable
BeginProg
Scan (5,Sec,6,0)
PanelTemp (PanelTemperature,250)
Battery (BatteryVoltage)
If not Route(4094) then Socket = TCPOpen ("192.168.7.231",6785,0)
SendVariables (Result1,Socket,-1,4094,0000,100,"Public","Callback",dummy1,1)
CallTable CLBK1
NextScan
EndProg
7.1.1.2 Datalogger-to-Datalogger Communication
Communication between dataloggers is possible over TCP/IP. In order to do this,
a socket must be opened between the two dataloggers using the TCPOpen()
instruction. The socket opened by this instruction is used by the instructions
performing datalogger-to-datalogger communication.
The example program below gets the battery voltage from a remote datalogger and
sends its panel temperature to the remote datalogger. The remote datalogger is at
IP address 192.168.7.125 and port 6785 is used for communication between the
dataloggers. The remote datalogger must have its battery voltage stored in a
public variable, BattVolt. It must also have a Public variable declared,
PTemp_Base. This will be used to store the panel temperature of the base
datalogger.

NL115 Ethernet and CompactFlash®Module
10
PROGRAM
'CR1000
'DL-to-Dl_Comms_1.cr1
'Send this program to CR1000 #1
'Remote CR1000 #2 has PBA = 2, IP addr = 192.168.7.125, and port 6785
Public BattVolt,, BattVolt_Remote
Public PTemp
Public Result1, Result2
Dim Socket as LONG
DataTable (Test,1,-1)
DataInterval (0,12,Sec,10)
Minimum (1,BattVolt,FP2,0,False)
EndTable
BeginProg
Scan(2,Sec,0,0)
Socket = TCPOpen("192.168.7.125",6785,0)
BatteryVoltage(BattVolt)
PanelTemp(PTemp,250)
GetVariables (Result1,Socket,-1,2,0000,50,"Public","BattVolt",BattVolt_Remote,1)
SendVariables (Result2,Socket,-1,2,0000,50,"Public","PTemp",PTemp_Base,1)
CallTable(Test)
NextScan
EndProg
7.1.2 HTTP Web Server
Typing the datalogger’s IP address into a web browser will bring up its home
page, as shown in Figure 7-1. This default home page provides links to the
current record in all tables, including data tables, the Status table, and the Public
table. Clicking on a Newest Record link will bring up the latest record for that
table. It will be automatically refreshed every 10 seconds. Links are also
provided to the last 24 records in each data table. Clicking on a Last 24 Records
link will bring up the last 24 records for that table. The Last 24 Records display
must be manually refreshed. In addition, links are provided to all HTML files, all
XML files, and all JPEG files in the datalogger.

User Manual
11
Figure 7-1. Datalogger home page
If there is a default.html file on the datalogger, this will automatically become the
user-configurable home page.
The WebPageBegin/WebPageEnd declarations and the HTTPOut() instruction
can be used in a datalogger program to create HTML or XML files that can be
viewed by the browser. For more information on using these instructions, see the
datalogger manual or CRBasic Editor Help.
FileOpen() and FileWrite() can be used to create HTML pages, but
this requires first writing the file to the datalogger’s USR drive. It is
less convenient, and the page will be only as current as it is written
to the file.
7.1.3 FTP
7.1.3.1 FTP Server
With an NL115 attached, the datalogger will automatically run an FTP server.
This allows Windows Explorer to access the datalogger’s file system via FTP. In
the FTP world, the “drives” on the datalogger are mapped into directories (or
folders). The “root directory” on the datalogger will include CPU and possibly
USR and/or CRD. The files will be contained in one of these directories. Files
can be pasted to and copied from the datalogger “drives” as if they were drives on
the PC. Files on the datalogger drives can also be deleted through FTP.
NOTE

NL115 Ethernet and CompactFlash®Module
12
Figure 7-2. FTP root directory
Figure 7-3. FTP CRD directory
In order to use FTP, the datalogger’s FTP User Name and FTP Password must be
set. This is done using Device Configuration Utility.
Other manuals for NL115
1
Table of contents
Other Campbell Network Hardware manuals

Campbell
Campbell NL115 User manual

Campbell
Campbell NL121 User manual

Campbell
Campbell NL200 User manual

Campbell
Campbell RTMS SYSTEMS User manual

Campbell
Campbell NL240 User manual

Campbell
Campbell 38485 User manual

Campbell
Campbell 4WPB100 User manual

Campbell
Campbell NL120 User manual

Campbell
Campbell SC932A User manual

Campbell
Campbell NL200 User manual