Candela GentleMax Pro User manual

Alexandrite Long Pulse Laser 755nm
GENTAL MAX PRO
YOU JUST CAN NOT COPYAN ORIGINAL.
COPYRIGHTS RESERVED.

Catalog
1.Appearance.............................................1
1.1 Appearance & Components...........
1.2 Packing list.....................................
2. Safety ............................................5
2.1 Electric safety.................................
2.2 Optical safety..................................
2.3 Safety labels...................................
3. Specifications........................................... 8
3.1 Laser specification..........................
3.2 Headpiece specification..................
3.3 Environment....................................
4. Assembling...............................................
11
4.1 Assembling .............................
4.2 Water-cooling system......................
5. Product Operations Guide........................ 15
5.1 System operation............................
5.2 Treatment parameters.......................
5.3 Operation procedure.......................
5.4 Clinical step.....................................
5.5 Switch off..........................................
6. Routine maintenance................................. 21
6.1 Preventive maintenance...................
6.2 Routine maintenance........................
6.3 Visual inspection...............................
6.4 Instrument Cleaning..........................
6.5 Emergency stop switch test..............
6.6 Replacement of insurance fuse........
6.7 Replace the cooling water.................
6.8 Move and transport the machine.......
6.9 Warranty............................................
6.10 Troubleshooting .................................
7. Fault handling..............................................
25
7.1 Fault handling guide...........................
8. Clinical.........................................................
28
8.1 Mechanism of action..........................
8.2 Suitable conditions.............................
8.3 Cycle-time-setting..............................
8.4 Discomfort feeling..............................
8.5 Attention.............................................
8.6 Operation note matters......................
8.7 Other items.........................................

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
1. Appearance & Components
1.1 Appearance & Components
6. Fiber
8. Headpiece
9. Emergency
4.Monitor
3. Right panel
7. Cooling tubes
2. Handrail
10. Key switch
5. Bracket
11. Front panel
1. Caster
12. Finger switch
(Noted :This photo is only considered as a reference; please reply with the real laser which may
has differences at some details from the photo showed above. )
1. Caster:
2. Handrail:
3. Right panel.
4. Monitor:
5. Bracket:
6. Fiber:
7. Cooling tubes:
8. Headpiece:
9. Emergency:
10. Key switch:
11. Front panel.
12. Finger switch.
to support and transport the equipment.
to move the equipment.
to operation by software.
to catch the headpiece.
to conduct laser energy.
to hold t the wires and cooling tubes.
headpiece for treatment.
to stop for emergency.
to switch on the laser.
1 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
13. Cooling liquid storage:to store the cooling liquid
14. Left panel.
15. Cooling device port: to connect a separate cooling device for skin.
16. Handrail.
17. Water tank: to store the cooling water for laser chamber.
18. Electric socket: to plug in power supply.
19. Foot-pedal: to connect a foot-pedal.
20. Cooling window.
21. Castor.

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
1.2 Packing list
Laser bodyX1
Headpiece with cold wind
cooling, spot-size 6-18mm
Electric cableX1
Foot switchX1
Water chanelX1
KeysX2
GogglesX1
Eye Glasses X1
3 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
2. Safety
This article shows the details of safety when operation. The operator and technician
must be certified after professional training of safety, know well the requirements of safety
and steps to operate the laser.
2.1 Electric safety
1.Make sure the voltage of laser requiring will be adapted by local electricity supply.
2. 10 minutes after switching off and plug off the cable, only the maintenance can be
started. No leave the laser without anyone taking care of by power on, or else injure will be
happened to human or laser itself.
3. Ensure that all the operators and their assistants know how to operate the laser
and know well how to switch off laser when emergency.
4. There is high voltage inside the laser, only qualified technicians or workers from
Perfectlaser can repair of maintain the laser.
5. The laser connects earth with a wire of yellow and green color, photo A. Fine earth
connection is very important.
A
(Different regions may have a different style electric plug, above photo only can be
reference.)
2.2 Optical safety
1. Laser headpiece is of high energy output, no look at directly without any
protection.
2. Ensure the operating room is no mirrors or any articles which can function as a
mirror to reflect the laser.
3. Ensure that all the operators and assistants know well the knowledge of optical
safety.
4. Ensure that all the operators and assistants and workers are protected well by
goggles, and the patients are wearing the eye-patch.
5. As the high laser peak power and density, operation error will damage the skin.
6. Laser 1064nm is invisible; the eyes will be blanked forever if wrongly shooing into.
7. Laser 755nm is visible red, and the eyes will be blanked forever if wrongly
5 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
shooting into. Also, staring for long while, discomfort will be caused.
8. Laser definitely cannot shoot at eyes even goggles and eye-patch are well
wearing.
9. The remarkable area should plant a label plate “ Caution of Laser”
2.3 Safety Labels
1. Product plate
Displays the product name, model no., and serial no., and voltage/frequency input
and the date of manufacture
2.
3. ETL LISTED label
Laser label.
4. Invisible laser radiation, avoid eye or skin exposure, and the U.S. patents the
product covered.
6 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
5. Water tank label
6. Hot water label
7. Caution, no remove the covers.
8. Caution for Class 4 laser.
7 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
8 / 32
9.
10.
11.
3 Specifications
3.1 Machine specification

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
3.2 Headpiece
Candela Gentlemax Pro
Items
Details
Control
Monitor
10 inches Digital Real Color LCD
Interface
Touch
Control system
Single chip
Headpiece
Headpiece/spot-size
6-18mm spotsize with cold air coolilng
Wave-length
1064nm&755nm
Cooling
Cooling system
Water cooling
water temperature
5-25 ℃auto control
Skin cooling
internal air cooling machine optional cold air cooler
Interface
Max Energy
100J @ 1064nm, 60J@755nm
Pulse width
10-100ms
Pulse amount
1-5T
Pulse interval
5-100ms
Pulse frequency
0.25, 0.5, 1Hz,2Hz
Power system
Power supply part
3680 watts with double IGBT
IGBT part
Made in Japan
Capacitance capacity
5000ufx8=40000uf
In charge voltage
110V 220V AC
Capacitance
10000ufx8
Laser board
Lamp
100x176x10mm Made by UK
Lamp amounts
2 pieces
Rod
Alexandrite Rod 755nm
LBO
PP-LN
Wavelength output
755nm
Fiber
Laser fiber
Made by Japan, MITSUBISHI
Fiber spec
1mm diameter with 2200mm length
Size & Weight
Size
400(W)x800(L)x1150mm(H)
Net weight
95KG
9 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
Headpiece: A
Headpiece B
Headpiece: C
A: Spot-size 6- 18mm adjustable; Skin cooling temperature -5 to 5℃;
Cold air skin cooling.
3.3 Environment
1. Air quality:
The system should work in a non-corrosive environment. Corrosive substances, such
as acids, etc., may damage the surface of the wires and electronic equipment. Dust in
the air should be kept to the minimum. Metal particles have a corrosive effect on the
electronic device.
2. Temperature and humidity
Working temperature:10-30 ℃
keep temperature:0-55 ℃
Working Relative Humidity:max 80%
Storage relative humidity:Max 90%
3. Atmospheric pressure:860hPA— 1060hPA
4. Environmental requirements: Avoid the sunshine
10 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
4 Assembling
4.1 Assembling
1. Fiber support arm installation
1. Find the fiber support arm.
Inset the arm.
2. Fix arm tightly with screws.
3. Adjust the angle and screw it.
4. Hang up.
5. Hang up.
6. Put the headpiece into the
socket.
11 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
2.Headpiece connection.
1. Take out the cap of fiber and
insert into the fiber port
2. Fix fiber tightly with screw.
3. Prepare the finger switch and
insert into the port.
7. Fix fiber tightly with screw.
8. Prepare the cooling tube and
insert into the port.
9. Insert into the port deeply.
3.Add water
1. Find the water injection and
open the overflow.
2.Add water by water injection.
3. Overflow has water flowing out
then close them.
4.Connect foot-pedal and electric plug.
12 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
1. Prepare the electric plug.
2.Insert into the plug.
3.Switch on.
4.2 Water-cooling system
Due to the generated huge heat of the xenon lamp when do treatment, and the
treatment handle cooling the skin produce a lot of heat, so the machine needs to add a
certain amount of pure water before work.
Water added standard:Distilled water or purified water(avoid to use tap water or
mineral water, otherwise it will seriously affect the working life of the machine and the
cooling effect.)
1. Water tank capacity:15L.
2. the shortest period of replace cooling water: three months.
3. Cooling water filling (see 4.1.3 step):
a. Water monitoring program.
b. water monitored. Machine built-in flow monitoring devices. The machine
starts to work, then flow monitoring module to work. If the water monitoring
module can not feel water, the machine will automatically lock the user
interface and prompt "Water is not circulating", accompanied by sound of the
buzzer.
c. Water temperature monitoring. Machine built-in temperature monitoring
sensor, monitoring the temperature of the cooling water into the water
circulation system. The upper temperature limit is set to 50℃. If the cooling
water temperature reaches the set limit, the user interface is automatically
locked, and prompt "temperature too high", and accompanied by sound of
the buzzer. Machine built-in radiator module, at room temperature (25℃),
and operating room ventilation is good, the machine can guarantee
continuous operation. If the room temperature is too high, or operation room
ventilated effect is bad, or water filling not enough,after the machine working
for some time, it may reach the upper temperature limit, then alarm and stop
working. At this approach: First, Wait for the machine cooled to room
temperature. Second, change the water with normal temperature. If the
summer temperature of some certain areas is too high (above 35℃), the
13 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
operating room should install air-conditioning equipment to ensure the
temperature about 25 ℃before using the machine; otherwise the water
temperature will be a sharp rise in a short time, which will not all shortens the
working time of the machine , but also affect the service life of the machine.
d. Water quality monitoring. Machine Built-in cooling water quality monitoring
module. if the water is impurities, and water quality is bad, the machine will
no longer monitor the water flow, suggesting that "water is not circulating." It
should replace the cooling water in this time.
e. water level monitoring. The machine has no built-level monitoring module or
device.
External water level displays the current flow in the water tank. The water line
should be 7 or more.
14 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
5 Monitor Operations Guide
This chapter describe the treatment handle of Alex. And Nd Yag laser system, the
proposed treatment parameters and operation method.
Please make sure to read in detail before using the product.
5.1 System operation
5.1.1 Start the machine。
1. Make sure the power cord is connected, insert the key, screw to the right,
"Cartagena" sound, the start device inside the machine will start the machine, the user
interface will light.
5.1.2 Monitor operation guide
2. The main display of the machine is light, enter the initialization interface. In the
same time variety of detection devices inside the machine run. The following welcome
page shows on the screen (Figure 5.1.2).
Figure 5.1.2
This is the first logo page. Click anywhere of this page to enter next page for wavelength
selection. (Figure 5.1.3)
3. At this page, please select the wavelength related for treatment. YAG is 1064nm, Alex is
755nm. Now we click “YAG” for example.
15 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
Figure 5.1.3
4. The below page shows the main operating page for parameters adjustment. ( Figure
5. 1.4).
Figure 5.1.4
16 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
.
press this button back to wavelength selection page.
b. Press this button for applications reference,
The page will show the page as below,
At this page, the YAG Applications can be selected by, Hair Removal, Diffuse Redness,
Vascular Lesions, and Facial Treatment . Or press this back to the main
page .
c. Energy selection, At this page, the energy can be selected from
10- 100 Joules (1J=1000mj).
17 / 32
a

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
d. Pulse width selection,
By this page, the pulse width can be selected from 10- 100ms.
e. Frequency selection,
By this page, the frequency can be selected from 0.5-2.0HZ.
f. By this page, the spot-size can be selected by
6mm(diameter), 8mm, 10mm, 12mm, 15mm, 18mm.
g. The volume for cold air can be adjusted by three grade.
h.
i.
j.
Switch on or off the cold air cooling for skin.
Press 5 seconds to enter into the engineer mode by password.
To check clients information installed.
k. Cold air blowing Pulse duration and Pulse interval selection.
The pulse duration and pulse interval can be selected by 0ms, 50ms, 100ms, 500ms,
1000ms. If Pulse duration =1000ms, and Pulse Interval=0ms, it means the cold air is
blowing the skin continually.
l. The cooling liquid need to be replaced.
m. Press red circle to simmer up laser, and then press foot-pedal
or finger switch, the laser starts to shoot. Press blue circle, the laser simmer down for
standby station.
Press to shift wavelength from YAG to Alex.
4. “Alex” page. The alex page shows following,
18 / 32

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
The difference from YAG, the Alex page, only the energy adjust scale is from
10-60Joules, the others keeps same as YAG. So here we do not explain again for
details.
5.2 Treatment parameter
Check the “Application” page.
5.3 Treatment progress
Take the755nm wavelength as an example. 1064nm wavelength is the same, which will
not be illustrated here again.
The steps are as following:
1. Make sure the machine is installed complete;
2. Make sure the water is full filled;
3. Position the machine correctly beside the treatment table; clean the treatment area;
4. Start the machine;
5. Enter the welcome page;
6. Enter the Alex page;
7. Set up the parameters as required;
8. Press “standby” key to ready for treatment;
9. Now the laser is ready to shoot.
10. Get the headpiece ready on the treatment area;
11. Step the pedal ( or finger switch) to get started; the light should be out on the skin and
the hair should be burned.
Note: to keep the continuous treatment, the pedal should be stepped all the time. Release
the pedal when the treatment needs to stop.
19 / 32
Other manuals for GentleMax Pro
1
Table of contents
Other Candela Hair Removal System manuals

Candela
Candela MGentleLASE Operating and maintenance manual

Candela
Candela MGentleLASE Manual

Candela
Candela New GentleYAG Manual

Candela
Candela New GentleYAG Manual

Candela
Candela New GentleYAG Use and care manual

Candela
Candela GentleLASE Manual

Candela
Candela MGentleLASE Use and care manual

Candela
Candela New GentleYAG Manual