caron 6045 Manual

MODEL 6045
INST-PM-6045

Revision Log
Revisions
Description
Letter
Date
A
09-05-2017
Initial release.
B
04-04-2018
New cover.

INST-PM-6045 Rev B 04/04/2018
Parts Required: (1) Air Filter, (2) Atomizers
Tools Required: Ladder, Flatblade Screwdriver, Phillips Screwdriver, (2) 8″
Crescent Wrenches, Flatblade Screwdriver (may be necessary)
Parts included in Preventative Maintenance Kit for Model 6045 are shown below.
Note: Photos may have a different appearance depending on chamber options.
1. Remove air filter cover by pulling outward on both sides near bottom as shown
below.
INSTRUCTIONS FOR REPLACING AIR FILTER AND ATOMIZERS IN
MODEL 6045
REPLACING AIR FILTER
Air Filter Cover
Air Filter
Atomizers

INST-PM-6045 Rev B 04/04/2018
2. Slide air filter out of air filter cover and discard as shown below.
3. Install new air filter with seam next to angled edge of air filter cover as shown
below. Note: Orient cardboard side of air filter towards chamber.
4. Install air filter cover with angled edge at top while aligning air filter with guide
tracks on both sides and snap into place as shown below. Note: Right side is
shown in photos.
Air Filter
Guide Track
Angled Edge on Air Filter Cover
Air Filter Seam

INST-PM-6045 Rev B 04/04/2018
Note: Photos may have a different appearance depending on chamber options.
1. Turn off chamber and unplug power cord.
2. Remove shelves from chamber.
3. Remove shelf channels located near top of chamber if necessary.
4. Remove false floor from chamber by pulling access port upwards while bending
edges downward as shown below.
5. Remove (4) screws securing intake diffuser using Phillips screwdriver bit as shown
below. Note: Retain screws.
6. Remove intake diffuser by pulling towards front of chamber.
REPLACING ATOMIZERS
Disconnect electrical power before removing interior parts.
Intake Diffuser Mounting Screws (as viewed from inside chamber under intake diffuser)
Intake Diffuser
False Floor Access Port

INST-PM-6045 Rev B 04/04/2018
7. Remove (2) screws securing drip pan using Phillips screwdriver bit as shown
below. Note: Retain screws.
8. Pull drip pan downward as shown below. Note: If necessary use flatblade
screwdriver to pry drip pan down as shown below.
9. Remove and discard old atomizers using (2) 8″crescent wrenches as shown below.
10.Install new atomizers.
11.Reinstall drip pan. Note: If necessary use flatblade screwdriver to install drip pan.
12.Secure drip pan with retained (2) screws using Phillips screwdriver bit.
13.Align tabs on intake diffuser with drip pan.
14.Secure intake diffuser with retained (4) screws using Phillips screwdriver bit.
15.Reinstall false floor.
Drip Pan Mounting Screws
Drip Pan
Drip Pan
Atomizer
Atomizers

INST-PM-6045 Rev B 04/04/2018
16.Reinstall shelf channels.
17.Reinstall shelves.
18.Continue to instructions for replacing additional accessory preventative
maintenance items if applicable or plug in power cord and turn on chamber.
This manual suits for next models
1
Table of contents
Other caron Climate Chamber manuals