Cavitron Plus G136 Manual

1
with
Détartreur ultrasonique avec
Escarificador ultrasónico con
Ultraschall-Zahnsteinentfernungsgerät mit
Scaler ultrasonico con
Ультразвуковая система для удаления зубного камня с Технологией
Tap-OnTM
-Technologie
Technologie Tap-OnTM
Tap-OnTM
tecnología
Tap-OnTM
tecnologia
Технология Tap-OnTM
Tap-OnTM
-Technologie
Technologie Tap-OnTM
Tap-OnTM
tecnología
Tap-OnTM
tecnologia
Технология Tap-OnTM
Tap-OnTM
-Technologie
Technologie Tap-OnTM
Tap-OnTM
tecnología
Tap-OnTM
tecnologia
Технология Tap-OnTM
Tap-OnTM
-Technologie
Technologie Tap-OnTM
Tap-OnTM
tecnología
Tap-OnTM
tecnologia
Технология Tap-OnTM
Tap-OnTM
-Technologie
Technologie Tap-OnTM
Tap-OnTM
tecnología
Tap-OnTM
tecnologia
Технология
Tap-OnTM
Cavitron®Plus
Ultrasonic Scaler
Please read carefully and completely before
operating unit.
Prière de lire attentivement et complètement
avant la première utilisation de l’appareil.
Por favor lea cuidadosamente y en su totalidad
antes de operar la unidad.
Bitte vor Inbetriebnahme der Einheit sorgfältig
und vollständig durchlesen.
Si prega di leggere attentamente e
completamente prima di utilizzare l’apparecchio.
Пожалуйста, внимательно и полностью
прочтите перед использованием устройства.
Directions For Use
Mode d’emploi
Instrucciones de uso
Gebrauchsanweisung
Istruzioni per l’uso
Указания по применению

2
Quick Start Installation Instructions
Set unit to OFF position.
Symbol: means o.
Connect POWER supply. Connect WATER supply by
pushing BLUE water hose into
water port.
Install BATTERIES in wireless
foot pedal unit.
Connect AIR supply (if equipped)
by pushing BLACK air hose
into air port.
Hold upright and depress foot
pedal until handpiece
FILLS WITH WATER.
TIP: adjust lavage control to higher
ow rate to increase water ow.
Tap-On™ Technology does not function
unless an insert is in the handpiece.
TIP: Handpiece is detachable and
autoclavable. Make sure the electrical
connections are dry before connecting.
Place ULTRASONIC INSERT
into handpiece.
Adjust LAVAGE CONTROL on
handpiece to desired setting.
Consult Directions For Use for
daily maintenance, usage
techniques and more.
30K Steri-Mate
30K Steri-Mate
30K Steri-Mate
1
2
3
4
Adjust POWER LEVEL.
Turbo
Purge
1 2 3
45Connect HANDPIECE to cable.
6
89
10 11 12
Water
Air
Set unit to ON position.
Symbol: means on.
The indicator light will be
green when unit is on.
Turbo
Purge
7
teri-Mate
1
2
3
4

3
To activate Tap-On™ Technology:
Tap wireless foot pedal one time quickly.
Relax foot as you scale or air polish. To deactivate Tap-On™ Technology:
Tap wireless foot pedal one time.
Turbo: When pressed,
increases power delivered
to the system by up to
25%; turbo power remains
on until the button is
pressed again.
Boost: Temporary, hands-
free activation by pressing
the foot pedal all the way to
the oor; enables quick
removal of tenacious calculus.
Tip: Quick release of pedal
from Boost will not deactivate
Tap-On™ Technology.
Purge: Remove insert
from Handpiece and
press Purge button; water
will purge through
system for two minutes
or until you press the foot
pedal or press the
Purge button again.
Finer Lavage Water Control:
Conveniently adjust lavage
water control directly on the
handpiece;
1 = lowest water ow
6 = highest water ow
> 6 = ush or purge
Prophy Mode Auto Cycles
(available on Cavitron® JET Plus
Systems): Automatically cycles
between air polish and rinse
without pumping the foot pedal.
Choose short, medium or long
prophy cycle times. Tap-On™
Technology will activate prophy
mode auto cycles.
User controls
Turbo Purge
Boost
30K Steri-Mate
1
2
3
4
Quick Start User Guide

4
Quick Reference Guide
Diagnostic Display
ON/OFF
Illuminates when the Main Power
On/Off Power switch is in the “ON”
(I) position.
TURBO
Offers the ability to increase power
to the system by up to 25% with
the push of a button. Purple arrows
illuminate when in use.
BOOST
Illuminates when the Boost Mode is
activated by the Tap-On TM Wireless
Foot Pedal. To activate, fully
depress Tap-OnTM Foot Pedal to the
second position (all the way to the
floor). To deactivate, release Tap-
OnTM Foot Pedal to the first position.
PURGE BUTTON
Illuminates when Purge function
is activated. To activate Purge,
remove insert from handpiece, turn
the Handpiece Lavage Control to
maximum water flow, press the
Purge Button on the Diagnostic
Display. Water will purge through
system lines for two minutes. To
deactivate during two minute cycle,
press Purge button again or press
Tap-OnTM Foot Pedal.
SERVICE
Illuminates when the system is not
functioning properly. This display
has three distinct modes:
• Slow blink (1 blink per
second) means the system is
not operating within factory
specifications.
• Fast blink (3 blinks per second)
indicates an improper set-up.
• Steady light indicates the system
is overheating.
LOW BATTERY
Illuminates when the Tap-OnTM Foot
Pedal battery power is approaching
end of life. Replace batteries as
instructed in
section 7.9.
Power Control
Power Level Control
Turn knob to select ultrasonic power
level for operation. Turning the knob
clockwise increases the distance the
insert tip moves (the stroke) without
changing frequency; turning knob
counter-clockwise decreases the distance
the insert tip moves (the stroke) without
changing the frequency.
RINSE
Rinse mode is used during an ultrasonic
scaling procedure when lavage is required
to flush the procedural area. To activate,
turn Power Level Control Knob fully
counter-clockwise until a “click” is heard.
BLUE ZONE
Provides an extended low-power range
for improved patient comfort when
subgingivally scaling.
RINSE
Rinse mode is used during an ultrasonic
scaling procedure when lavage is required

5
TABLE OF CONTENTS
QUICK START INSTALLATION
INSTRUCTIONS . . . . . . . . . . . . . .2
QUICK START USER GUIDE . . . . . . 3
QUICK REFERENCE GUIDE:
DIAGNOSTIC DISPLAY . . . . . . . 4
INTRODUCTION . . . . . . . . . . . . . 6
PRODUCT OVERVIEW. . . . . . . . . .6
TECHNICAL SUPPORT . . . . . . . . 6
SUPPLIES & REPLACEMENT PARTS . 6
INDICATIONS FOR USE . . . . . . . 7
CONTRAINDICATIONS . . . . . . . 7
WARNINGS . . . . . . . . . . . . . 7
PRECAUTIONS
4.1 System Precautions . . . . . . . . 7
4.2 Procedural Precautions . . . . . 7-8
ADVERSE REACTIONS . . . . . . . . 8
INFECTION CONTROL
6.1 General Infection Control . . . . . 8
6.2 Water Supply Recommendations . 8
INSTALLATION INSTRUCTIONS
7.1 Water Line Requirements . . . . . 8
7.2 Electrical Requirements . . . . . . 8
7.3 Unpacking the System. . . . . . .9
7.4 System Installation . . . . . . . . 9
7.5 Power Cord Connection. . . . . .9
7.6 Water Supply Line Connection . . 9
7.7 Tap-OnTM Foot Pedal Battery
Installation/ Replacement. . . . .10
7.8 Tap-OnTM Foot Pedal
Synchronization . . . . . . . . .10
CAVITRON®Plus Scaler DESCRIPTION
8.1 System Controls . . . . . . . . . 11
8.2 Diagnostic Display Indicators
and Controls . . . . . . . . . . 12
8.3 Handpiece/Cable . . . . . . . . 13
8.4 Cavitron® 30K™ Ultrasonic
Inserts . . . . . . . . . . . . . . 13
8.5 Tap-OnTM Wireless Foot Pedal
Operation . . . . . . . . . . . . 14
8.6 Accessories and User
Replaceable Parts . . . . . . . . 14
8.6.1 Accessories . . . . . . . 14
8.6.2 User Replaceable Part
Kits . . . . . . . . . . . 14
SYSTEM SETUP, OPERATION
AND TECHNIQUES FOR USE
9.1 Handpiece Setup . . . . . . 14-15
9.2 Turbo Mode. . . . . . . . . . . 15
9.3 Boost Mode . . . . . . . . . . . 15
9.4 Patient Positioning . . . . . . . 15
9.5 Performing Ultrasonic
Scaling Procedures . . . . . . . 15
9.6 Patient Comfort Considerations . 15
SYSTEM CARE
10.1 Daily Maintenance. . . . . . . 16
Start-up procedures
at the beginning of the day . . 16
Between patients . . . . . . . 16
Shut-down procedures at
the end of the day . . . . . . 16
10.2 Weekly Maintenance . . . . . 16
10.3 Monthly Maintenance . . . . . 17
Water Line Filter
Maintenance . . . . . . . . . 17
TROUBLESHOOTING
11.1 Troubleshooting Guide . . 17-18
11.2 Technical Support and Repairs . 18
WARRANTY PERIOD. . . . . . . . .18
SPECIFICATIONS. . . . . . . . . 18-19
CLASSIFICATIONS . . . . . . . . . . 19
DISPOSAL OF UNIT . . . . . . . . . 19
ELECTRO-MAGNETIC COMPATIBILITY
PRECAUTIONS . . . . . . . . . 20-22
QUICK REFRENCE GUIDE:
TROUBLESHOOTING . . . . . . . . 23
English

6
INTRODUCTION
Congratulations!
Your decision to add the Cavitron®Plus Ultrasonic Scaler
with Tap-On™ Technology to your practice represents a wise
investment in good dentistry.
For over four decades, dental professionals have preferred
the clinical benefits and labor-saving advantages inherent in
Cavitron ultrasonic scalers. Clinical studies and independent
research have confirmed the speed, efficiency and versatility
of ultrasonic scaling.*
DENTSPLY Professional is an ISO 13485 registered
company. All DENTSPLY Professional medical devices sold
in Europe are CE marked in conformance with Council
Directive 93/42/EEC.
Website: www.professional.dentsply.com
CAUTION: United States Federal Law restricts this
device to sale by or on the order of, a licensed dental
professional.
PRODUCT OVERVIEW
The Cavitron®Plus Ultrasonic Scaler is a precision
engineered and manufactured instrument. It contains
controls and components for ultrasonic scaling. The system
produces 30,000 strokes per second at the ultrasonic
insert’s working tip that when combined with the cavitational
effect of the coolant lavage creates a synergistic action
that is designed to “power away” even the heaviest calculus
deposits while maintaining operator and patient comfort.
The Cavitron Plus Ultrasonic Scaler is equipped with the
Sustained Performance System™ (SPS Technology), which
offers a constant balance between scaling efficiency and
patient comfort by maintaining power when the insert tip
encounters tenacious deposits, allowing the clinician to
effectively scale even at a decreased/lower power setting.
The Cavitron Plus System has extended the SPS technology
by expanding the Blue Zone range, providing finer resolution
to the power settings.
Advanced features that make the Cavitron Plus a wise
investment include a Tap-OnTM Wireless Foot Pedal with
Tap-On™ Technology, Turbo Mode, illuminated diagnostic
display, rinse setting, automated purge function, Steri-Mate®
detachable sterilizable handpiece, and 330˚ swivel handpiece
cable with more precise lavage water control. These
features, combined with a low power range (Blue Zone™) and
hands-free Boost Mode, are designed to deliver a positive
ultrasonic scaling experience for your patients while providing
your practice with the quality and reliability you’ve come to
expect from Cavitron brand ultrasonic systems.
The Cavitron Plus Ultrasonic Scaler is UL/ULC certified and
approved. The Cavitron Plus Ultrasonic Scaler is classified
by Underwriters Laboratories Inc. with respect to electric
shock, fire, mechanical hazards in accordance with the
IEC 60601 Standard. The Cavitron Plus Ultrasonic Scaler
complies with Part 15 of the FCC Rules. Operation is subject
to the following two conditions: 1) this device may not cause
harmful interference, and 2) this device must accept any
interference received, including interference that may cause
undesired operation. Cavitron Plus base FCC certification/
registration number: FCC ID: TF3-DPD73227323; IC:
4681B-73227323. Cavitron Plus Tap-OnTM Foot Pedal FCC
certification/registration number: FCC ID: TF3-DPD81675;
IC: 4681B-81675. The term IC before the certification/
registration number signifies that the Industry Canada
technical specifications were met.
TECHNICAL SUPPORT
For technical support and repair assistance in the U.S., call
the DENTSPLY Professional Cavitron CareSM Factory Certified
Service at 1-800-989-8826, Monday through Friday, 8:00
A.M. to 5:00 P.M. (Eastern Time). For other areas, contact
your local DENTSPLY®Professional representative.
SUPPLIES &
REPLACEMENT PARTS
To order supplies or replacement parts in the U.S., contact
your local DENTSPLY Professional Distributor or call
1-800-989-8826, Monday through Friday, 8:00 A.M. to 5:00
P.M. (Eastern Time). For other areas, contact your
local DENTSPLY Professional Representative.
* data on file

7
SECTION 1:
Indications For Use
Ultrasonic Procedures
• All general supra and subgingival scaling applications
• Periodontal debridement for all types of periodontal
diseases
• Endodontic procedures
SECTION 2:
Contraindications
• Ultrasonic Systems should not be used for restorative
dental procedures involving the condensation of
amalgam.
SECTION 3: Warnings
• Persons fitted with cardiac pacemakers, defibrillators
and other active implanted medical devices, have been
cautioned that some types of electronic equipment
might interfere with the operation of the device.
Although no instance of interference has ever been
reported to DENTSPLY, we recommend that the
handpiece and cables be kept 6 to 9 inches (15 to 23
cm) away from any device and their leads during use.
• There are a variety of pacemakers and other medically
implanted devices on the market. Clinicians should
contact the device manufacturer or the patient’s
physician for specific recommendations. This unit
complies with IEC 60601 Medical Device Standards.
• It is the responsibility of the Dental Healthcare
Professional to determine the appropriate uses of this
product and to understand:
– the health of each patient,
– the dental procedures being undertaken,
– and applicable industry and governmental agency
recommendations for infection control in dental
healthcare settings,
– requirements, and regulations for safe practice of
dentistry; and
– these Directions for Use in their entirety, including
Section 4 Precautions, Section 6 Infection Control, and
Section 10 System Care.
• The use of High Volume Saliva Evacuation to reduce
the quantity of aerosols released during treatment is
highly recommended.
• Where asepsis is required or deemed appropriate in
the best professional judgment of the Dental Healthcare
Professional, this product should not be used.
• During boil-water advisories, this product should not
be operated as an open water system (e.g. connected
to a public water system). A Dental Healthcare
Professional should disconnect the system from the
central water source. The Cavitron DualSelect system
can be attached to this unit and operated as a closed
system until the advisory is cancelled. When the
advisory is cancelled, flush all incoming waterlines
from the public water system (e.g. faucets, waterlines
and dental equipment) in accordance with the
manufacturer’s instructions for a minimum of 5 minutes.
• Prior to beginning treatment, patients should rinse with
an antimicrobial such as Chlorhexidine Gluconate 0.12%.
Rinsing with an antimicrobial reduces the chance of
infection and reduces the number of microorganisms
released in the form of aerosols during treatment.
• Per FCC Part 15.21, changes or modifications not
expressly approved by the party responsible for
compliance could void the user’s authority to operate
this equipment.
• Failure to follow recommendations for environmental
operating conditions, including input water temperature,
could result in injury to patients or users.
• Insufficient water flow could result in elevated water
and tip temperature. When operated at the input water
temperature specified in the Water Line Requirements
Section 7.1 and with sufficient water flow, the water and
tip temperature should not exceed 50o C (122o F).
SECTION 4: Precautions
4.1 System Precautions
• Do not place the system on or next to a radiator or
other heat source. Excessive heat may damage the
system’s electronics. Place the system where air is free
to circulate on all sides and beneath it.
• The system is portable, but must be handled with care
when moving.
• Equipment flushing and dental water supply system
maintenance are strongly recommended. See Section
10: System Care.
• Close manual shut-off valve on the dental office water
supply every night before leaving the office.
• The use of an in-line water filter is recommended.
• Never operate system without fluid flowing through the
handpiece.
• Always ensure that the electrical connections on the
handpiece cable and the Steri-Mate®handpiece are clean
and dry before assembling them for use.
4.2 Procedural Precautions
General
• As with all dental procedures, use universal precautions
(i.e., wear face mask, eyewear, or face shield, gloves and
protective gown).

8
Ultrasonics
• The Cavitron Plus unit works with Cavitron inserts as
a system, and was designed and tested to deliver
maximum performance for all currently available Cavitron
brand ultrasonic inserts. Companies that manufacture,
repair or modify inserts carry the sole responsibility for
proving the efficacy and performance of their products
when used as a part of this system. Users are cautioned
to understand the operating limits of their insert before
using in a clinical setting.
• Like bristles of a toothbrush, ultrasonic inserts “wear”
with use. Inserts with just 2 mm of wear lose about 50%
of their scaling efficiency. In general it is recommended
that ultrasonic inserts be discarded and replaced after
one year of use to maintain optimal efficiency and
avoid breakage. A DENTSPLY Professional Insert
Efficiency Indicator is enclosed for your use.
• If excessive wear is noted, or the insert has been
bent, reshaped or otherwise damaged, discard the insert
immediately.
• Ultrasonic insert tips that have been bent, damaged, or
reshaped are susceptible to in-use breakage and should
be discarded and replaced immediately.
• Retract the lips, cheeks and tongue to prevent contact
with the insert tip whenever it is placed in the patient’s
mouth.
SECTION 5:
Adverse Reactions
None Known.
SECTION 6:
Infection Control
6.1 General Infection Control
• For operator and patient safety, carefully practice the
infection control procedures detailed in the Infection
Control Information Booklet accompanying your
System. Additional booklets can be obtained by calling
Customer Service at 1-800-989-8826, Monday through
Friday, 8:00 A.M. to 5:00 P.M. (Eastern Time). For
areas outside the U.S., contact your local DENTSPLY
Professional representative.
• As with high speed handpieces and other dental devices,
the combination of water and ultrasonic vibration from
the Cavitron Plus Ultrasonic Scaler will create aerosols.
Following the procedural guidelines in Section 9 of this
manual can effectively control and minimize aerosol
dispersion.
6.2 Water Supply Recommendations
• It is highly recommended that all dental water supply
systems conform to applicable CDC (Centers
for Disease Control and Prevention) and ADA
(American Dental Association) standards, and that all
recommendations be followed in terms of flushing,
chemical flushing, and general infection control
procedures. See Sections 7.1 and 10.
• As a medical device, this product must be installed
in accordance with applicable local, regional, and
national regulations, including guidelines for water quality
(e.g. drinking water). As an open water system, such
regulation may require this device to be connected to
a centralized water control device. The Cavitron®
DualSelect™ Dispensing System may be installed to allow
this unit to operate as a closed water system.
SECTION 7:
Installation Instructions
Anyone installing a Cavitron Plus System should observe the
following requirements and recommendations.
7.1 Water Line Requirements
• A water supply line with user-replaceable filter is
supplied with your system. See Section 10 System Care
for replacement instructions.
• Incoming water supply line pressure to the system must
be 20 psi (138 kPa) to 40 psi (275 kPa). If your dental
water system’s supply line pressure is above 40 psi,
install a water pressure regulator on the water supply line
to your Cavitron Plus Ultrasonic Scaler.
• A manual shut-off valve on the dental water system
supply line should be used so that the water can be
completely shut-off when the office is unoccupied.
• In addition to the water filter supplied, it is recommended
that a filter in the dental water system supply line be
installed so that any particulates in the water supply will
be trapped before reaching the Cavitron system.
• After the above installations are completed on the dental
water supply system, the dental office water line should
be thoroughly flushed prior to connection to the Cavitron
system.
• Incoming water temperature to the Cavitron System
should not exceed 25˚C (77˚F). If needed a device
should be installed to maintain a temperature within this
specification, or a Cavitron DualSelect Dispensing
System attached to allow this system to be operated
as a closed water system.
7.2 Electrical Requirements
• Incoming power to the system must be 100 volts AC to
240 volts AC, single phase 50/60 Hz capable of
supplying 1.0 amps.
• The system power should be supplied through the AC
power cord provided with your system.
• WARNING:Toavoidriskofelectricshock,thisequipment
must only be connected to a supply mains with protective
earth.

9
7.3 Unpacking the System
Carefully unpack your Cavitron Plus Ultrasonic Scaler and
verify that all components and accessories are included:
1. Cavitron®Plus™System with Handpiece Cable
Assembly with swivel
2. Water Line Assembly (Blue) with Filter and Quick
Disconnect
3. Additional Water Line Filter
4. Detachable AC Power Cord
5. Cavitron®Tap-OnTM Wireless Foot Pedal
6. “AA” Batteries (4-Pack)
7. Auxiliary Cable for Tap-OnTM Foot Pedal
8. Steri-Mate® Grip (not shown)
9. Steri-Mate®Detachable Sterilizable Handpiece
10. Cavitron®Ultrasonic Inserts (quantity optional)
11. Efficiency Indicator for Cavitron Inserts
12. Literature Packet
7.4 System Installation
• The Cavitron Plus System is designed to rest on a level
surface. Be sure unit is stable and resting on four feet.
• The Cavitron Plus System should not be positioned such
that access to the power input and AC power cord are
limited.
• Placing unit in direct sunlight may discolor plastic
housing.
• The system has been equipped with a Tap-OnTM Wireless
Foot Pedal which was factory synchronized to operate
with the system’s base unit. If your office has more than
one Cavitron Plus system, it is recommended that you
mark the Tap-OnTM Foot Pedal and base unit for easy
reference as to which Tap-OnTM Foot Pedal operates with
which base unit. Should resynchronization be necessary,
follow the instructions in section 7.10.
7.5 Power Cord Connection
• Verify the Main Power ON/OFF switch, located at the
center front underside of the System, is set to the OFF
(O) position before proceeding.
• Insert the AC power cord into the power input on the
back of the System.
• Insert the pronged plug into an AC wall outlet.
7.6 Water Supply Line Connection
• Grasp the Water Supply Line (blue hose) by the end
opposite the quick-disconnect and insert it into the water
inlet connector until fully seated.
• Connect the quick-disconnect to the dental office water
supply or a Cavitron DualSelect Dispensing System.
• Inspect all connections to make certain there are no
leaks.
• To remove the water line from the Cavitron Plus System,
turn off the dental office water supply. Disconnect the
water supply line from the dental office water supply. If
a quick-disconnect connector is attached to the end of
the hose, relieve the water pressure by pressing the tip
of the connector in an appropriate container and allow
water to drain. To remove the hose from the system,
push on the outer ring of the system’s water inlet and
gently pull out the water line.
Press ring to release
water supply tube.

10
• Turn Tap-OnTM Wireless Foot Pedal over and using a
Philips screwdriver carefully remove battery cover screw
and battery cover. If applicable, remove used batteries
and install two new “AA” batteries as shown. Do not
depress Tap-OnTM Foot Pedal while installing batteries.
• The communication light will blink for approximately
two seconds to indicate the Tap-OnTM Foot Pedal’s ability
to communicate with the unit. If the light does not blink,
check the batteries. If the batteries are good and the
light doesn’t blink, a communications error may exist.
To re-establish communication with Tap-OnTM Foot Pedal
review Synchronization procedure, section 7.10.
• The remote frequency communication can be bypassed
using the auxillary Tap-OnTM Foot Pedal cable. Refer to
Section 11.2 Technical Support and Repair for further
action.
• Replace the battery cover and screw and hand tighten
cover with Philips screwdriver.
• Remove batteries if Tap-OnTM Foot Pedal is to be stored
for an extended period of time.
7.8 Tap-OnTM Foot Pedal Synchronization
The Tap-OnTM Wireless Foot Pedal supplied with your
system has been factory synchronized with the base unit.
Should a replacement Tap-OnTM Foot Pedal be necessary,
synchronization will be required prior to system operation.
Perform the following steps to synchronize the Tap-OnTM Foot
Pedal with the base unit.
1. Turn the MainPower switch located at the center
front underside of the system to the OFF (O) position.
2. Install a new set of “AA” batteries into the foot
control (see section 7.9). Leave the battery cover of
the Tap-OnTM Foot Pedal open so the red push button
is accessible.
3. Maintain a distance of no more than 10 feet (3
meters) between the base unit and Tap-OnTM Foot
Pedal during the synchronization process.
4. Remove any inserts from the handpiece and adjust
the Power Level Control out of Rinse Mode. Turn the
Main Power switch to the ON (I) position and wait for
the Diagnostic Display graphics to light (refer to
Section 8.2).
5. While all graphics are lit, press the Purge button,
located on the Diagnostic Display.
The graphics will begin to blink in a sequential
pattern, representing the synchronization mode.
This mode will last 5 to 6 seconds.
6. During this mode, press the red button located in the
battery compartment of the Tap-OnTM Foot Pedal. This
will complete the synchronization process.
7. Synchronization is successful when all graphic lights
blink at the same time.
8. To verify proper communication, press the foot
control to the Boost position (Tap-OnTM Foot Pedal
fully depressed – 2nd position) and ensure the Boost
graphic on base unit illuminates.
9. Attach battery cover and tighten the screw.
10. In the event communication cannot be established,
temporarily use the supplied Auxillary Tap-OnTM Foot
Pedal Cable to connect the Tap-OnTM Foot Pedal
directly to the unit.
Look for blinking
communications
light.
7.7 Tap-OnTM Foot Pedal Battery
Installation/Replacement
PURGE

11
Main Power ON/OFF Switch
ON/OFF Switch located at the center front
underside of the system.
Ultrasonic Power Level Control
Turn knob to select the ultrasonic power level for operation. Turning the knob clockwise
increases the distance the insert tip moves (stroke) without changing the frequency;
turning the knob counterclockwise decreases the distance the insert tip moves (stroke)
without changing the frequency.
The Blue Zone is an extended low-power range for improved patient comfort when
subgingivally scaling.
Rinse
Turn the ultrasonic power level control knob fully counterclockwise until a “click” is heard.
Rinse mode is for use during an ultrasonic scaling procedure when lavage is desired with
minimal cavitation.
Diagnostic
Display
See section
8.2 for more
details.
Blue Zone
Rinse
Steri-Mate®
Handpiece
Accepts all
Cavitron®30K™
Ultrasonic inserts
See section 8.3 for
more details.
Handpiece Holder
Securely holds the
system’s handpiece,
or cable connector
when handpiece is not
installed.
Tap-OnTM Wireless Foot
Pedal
Eliminates the need to hold down
or pump foot pedal.
See section 8.6 for more details.
SECTION 8: Cavitron Plus Ultrasonic Scaler Description
8.1 System Controls

12
Power Indicator
Illuminates (3 sec. delay)
when the Main Power ON/
OFF Control Switch is ON
(“I” position).
Low Battery
Indicator
Illuminates when
the Tap-OnTM Foot
Pedal battery power
is approaching end of
life. Replace batteries
as instructed in
section 7.9.
Service Indicator
Illuminates when the system
is not functioning properly.
This display has three
distinct modes.
•Afastblink(3blinksper
second) indicates an
improper set-up.
•Aslowblink(1blinkper
second) means the
system is operating out
of factory specifications.
•Asteadylightindicates
the system is overheating.
Refer to section 11.1 for
Troubleshooting guidelines.
Boost Indicator
Illuminates when the
Boost Mode has
been activated with
the Tap-OnTM Foot Pedal.
Turbo Mode
When pressed, extra
power is delivered to the
system, up to 25%. This
will increase the stroke
of the tip. Turbo power
remains on until the
button is pressed again
or unit is turned off.
(Purple arrows illuminate
when Turbo Mode is on).
Purge Control
Illuminates when the Purge function is activated.
To activate Purge, remove insert from the
handpiece, turn the handpiece lavage control
to maximum water flow and press the Purge
button. Water will purge through system for
2 minutes. To deactivate mode during the 2
minute cycle, press Purge button again or press
Tap-OnTM Foot Pedal.
The Purge Control is also used during the Tap-
OnTM Foot Pedal Synchronization process. See
section 7.10 for more details.
8.2 Diagnostic Display Indicators and Control

13
Lavage Control
Turn the Lavage Control to select flow rate during system
operation. Flow rate is based on a scale from 1 to 6. Turn
clockwise toward 6 to increase flow at insert tip. Turn
counter-clockwise toward 1 to decrease flow. The flow rate
through the handpiece also determines the temperature of
the lavage. Lower flow rates produce warmer lavage. Higher
water flow rates produce cooler lavage.
If the handpiece becomes warm, increase the flow rate.
With experience the Dental Healthcare Professional will be
able to determine the best flow rate setting for optimum
operating efficiency and patient comfort.
Swivel Feature
Reduces cable drag as handpiece rotates during procedures.
Steri-Mate®Grip accessory (not shown)
The Steri-Mate®Grip is designed to provide an ergonomic
and comfortable grasp of the handpiece. The grip is
sterilizable and is available in several different colors as an
accessory for your Steri-Mate Handpiece. See installation
instructions provided with the grip.
O-Ring
Provides seal for handpiece coolant.
O-ring should be replaced when worn.
Connecting Body
Transfers and amplifies mechanical
motion of stack to insert tip.
Magnetostrictive Stack
Converts energy provided by the handpiece into
mechanical oscillations used to activate the insert
tip.
Insert Marking
Manufacturer,
Date (YDDD=Single digit
year and triple digit day
of year), Frequency,
Type, Tip Lot Number
(if applicable).
Finger Grip
Insert Tip
Shape and size of tip
determines access and
adaptation. Preheated
Lavage directed to tip.
The many styles of Cavitron and Cavitron Bellissima 30K Ultrasonic Inserts are easily interchangeable for various procedures
and applications. See enclosed literature for specific information.
8.3 Handpiece / Cable
8.4 Cavitron 30K Ultrasonic Inserts

14
6. Cavitron DualSelect Dispensing system
7. Steri- MATE®Grip
8. Steri- MATE® Light Lighted Sleeve
8.6.2 User Replaceable Part Kits
1. Cavitron Insert Replacement O-ring Kits, 12/packs
Part Number 62351 (black) for plastic and soft grips
Part Number 62605 (green) for metal grips
2. Handpiece Cable O-Ring, Part Number 79357
3. Lavage (Water) Filter, 10/Pack, Part Number 90158
For detailed information, contact your local DENTSPLY
Professional Representative or authorized DENTSPLY
Professional Distributor.
SECTION 9: System Setup,
Operation and Techniques
for Use
9.1 Handpiece Setup
• This handpiece is sterilizable. Refer to Infection Control
Information Booklet for sterilization instructions prior to
using handpiece.
• Connect the Handpiece to the Cable Assembly by
aligning the electrical connections. If Cable Assembly
does not seat into the handpiece, gently rotate
the handpiece until contacts align, then fully insert
handpiece.
• Hold empty handpiece in a semi-upright position over a
sink or drain. Activate the Tap-OnTM Foot Pedal until
water exits to release any air bubbles that might
be trapped inside the handpiece. NOTE: Tap-OnTM
Technology only operates when an insert is in the
handpiece.
• Lubricate the O-ring on the insert with water before
placing it into the handpiece. Fully seat insert with a
gentle push-twist motion. DO NOT FORCE.
• Turn the Lavage Control to select flow rate during
system operation. Flow rate is placed on a scale from 1
to 6. Turn control clockwise toward 6 to increase flow
at insert tip. Turn control counter-clockwise toward 1
to decrease flow. The flow rate through the handpiece
also determines the temperature of the lavage. Lower
water flow rates produce warmer lavage. Higher flow
rates produce cooler lavage. If the handpiece becomes
9.1
Handpiece Setup
8.5 Tap-On™Technology Wireless Foot
Pedal Operation
Using Foot Pedal in Tap-OnT M Mode
For scaling procedures,Tap-OnTM Technology eliminates the
need to hold the pedal down. Tapping the foot pedal once
activates ultrasonic power or rinse mode for approximately
4 minutes. Tapping the foot pedal while in Tap-OnTM mode
disables the ultrasonic power and water flow. Boost is still
available while scaling in Tap-OnTM mode. To use boost, simply
depress the foot pedal to the second position (all the way to
the floor) to activate and hold as long as boost is desired.
Release foot pedal to return to Tap-On T M mode.
TIPS:
Tap-OnTM Technology will not run water unless an
insert is in the handpiece
A sensor in the handpiece holder will prevent
Tap-OnTM Technology from operating when the
handpiece is in the holder.
If the foot pedal is not tapped quickly, it will
function in a conventional manner.
How to Disable and Enable Tap-OnTM Technology
Tap-OnTM Technology feature can be disabled by
simultaneously holding the Purge and Turbo buttons for a
period of approximately 5 seconds. The two buttons will blink
approximately 6 times. When the buttons are released, they
will blink an additional 6 times to confirm Tap-OnTM has been
disabled.
Tap-OnTM Technology feature can be enabled by
simultaneously holding the Purge and Turbo buttons for a
period of approximately 5 seconds. The two buttons will blink
approximately 6 times to confirm Tap-OnTM has been enabled.
Using Foot Pedal without Tap-OnT M Mode
For scaling operation, the first position activates both the
ultrasonic energy and lavage at the insert tip. The second
position activates the Boost Mode. The Boost Mode (fully
depressed Tap-OnTM Foot Pedal) increases the ultrasonic
power level for quick removal of tenacious deposits without
adjusting the power level knob. To deactivate Boost Mode,
release Tap-OnTM Foot Pedal to first position.
8.6 Accessories and User
Replaceable Parts
8.6.1 Accessories
1. AC Power Cord
2. Tap-OnTM Technology Wireless Foot Pedal
3. Auxiliary Tap-OnTM Foot Pedal Power Cable
4. Cavitron Steri- MATE®Sterilizable Handpiece
5. Cavitron 30K Ultrasonic Inserts
NON-DEPRESSED DEPRESSED
1st POSITION
DEPRESSED
2nd POSITION

15
warm, increase the flow rate. With experience the Dental
Healthcare Professional will be able to determine the
best flow rate setting for optimum operating efficiency
and patient comfort.
9.2 Turbo Mode
Pressing the “Turbo” button on the scaling unit increases
the unit’s ultrasonic power up to 25%. When you need more
power for an extended period of time, simply press the Turbo
button on the display panel (arrows will illuminate to show you
are in “Turbo” mode).
DENTSPLY recommends that the clinician familiarize
themselves with the available power levels throughout the
power knob rotation in both normal and turbo modes. To do
this, simply hold your favorite insert over the sink and adjust
the power knob while observing the inserts spray pattern and
toggling between both normal and turbo modes.
9.3 Boost Mode
Boost provides a temporary increase in ultrasonic scaling
power for quick removal of tenacious calculus without
touching the unit. Boost is activated by fully depressing the
Tap-OnTM Foot Pedal to second position (all the way to the
floor). When boost is activated, the boost icon will illuminate
on the display panel. Boost remains on as long as the
clinician has the foot pedal pressed all the way down. In
order to deactivate boost, release the Tap-OnTM Foot Pedal to
first position.
9.4 Patient Positioning
For optimal access to both the upper and lower arches, the
backrest of the chair should be adjusted as for other dental
procedures. This assures patient comfort and clinician
visibility.
Have the patient turn his/her head to the right or left. Also
position chin up or down depending upon the quadrant and
surface being treated. Evacuate irrigant using either a saliva
ejector or High Volume Evacuator (HVE).
9.5 Performing Ultrasonic
Scaling Procedures
Note: Refer to the Infection Control Information
booklet supplied with your system and Section 10 of
this manual for general procedures to be followed at
the beginning of each day and between patients.
• Follow precautions listed in the General and Ultrasonic
sections of 4.2 Procedural Precautions
• The edges of Cavitron Ultrasonic Inserts are intentionally
rounded so there is minimal danger of tissue laceration
with proper ultrasonic scaling technique. Whenever the
insert tip is placed in the patient’s mouth, the lips,
cheek and tongue should be retracted to prevent
accidental (prolonged) contact with the activated tip.
• Turn Power Level Control to select ultrasonic power level
for operation. Clockwise increases system power.
Power level will increase throughout the full range of the
control. Hold the handpiece over a sink or drain. While
in Tap-OnTM mode, simply tap the Tap-OnTM Foot Pedal
to activate the system. (If Tap-OnTM mode is turned off,
press and hold the Tap-OnTM Foot Pedal down to activate
the system.) Check water spray to verify fluid is reaching
the working end of the insert tip. Adjust the water lavage
control until the water (lavage) flows with a rapid drip or
small spray. Higher water flow settings provide cooler
irrigation.
• It may be necessary to adjust lavage with the system
in “Boost” mode (Tap-OnTM Foot Pedal fully depressed)
so adequate fluid will be available to cool tip to tooth
interface.
• In general, it is suggested that a “feather-light-touch” be
used for ultrasonic scaling. The motion of the activated
tip and acoustic effects of the irrigating fluid, in most
cases, are adequate to remove even the most tenacious
calculus.
• Periodically check the Cavitron Ultrasonic Insert for wear
with the Cavitron Insert Efficiency Indicator.
• The use of a saliva ejector or High Volume Evacuator
(HVE) is recommended during all procedures.
• Set the system’s Power Level Control to the lowest
efficient power setting for the application and the
selected insert
• Keep the foot pedal near your foot to make it convenient
to access.
9.6 Patient Comfort Considerations
Reasons for sensitivity
• Incorrect tip placement. The point should never be
directed toward tooth root surfaces.
• Not keeping tip in motion on tooth. Do not allow the
insert to remain in a static position on any one area of
the tooth. Change the insert’s path of motion.
• Applying excessive pressure. Use a very light grasp
and pressure, with a soft tissue fulcrum whenever
possible, especially on exposed cementum.
If sensitivity persists, decrease power setting and/or move
from the sensitive tooth to another and then return.
and patient comfort.
Turbo Mode

16
SECTION 10: System Care
It is recommended that you perform the following
maintenance procedures.
10.1 Daily Maintenance
START-UP PROCEDURES AT THE
BEGINNING OF THE DAY:
1. Open the manual shut-off valve on the dental office
water supply system.
2. Install a sterilized Steri-Mate®Handpiece onto the
handpiece cable.
3. Set the Power Level Control to minimum and the
Lavage Control to maximum.
4. Turn the system ON.
5. Hold the sterilized handpiece (without an insert or
nozzle insert installed) over a sink or drain. Activate
the Purge Control button.
•ThePurgebuttonwilllightfortwominutesindicating
proper activation of the purge function.
•IfthePurgebuttonisactivatedwithaninsertpresent
in the handpiece, the button will blink for 3 seconds
and disable. Remove the insert from the handpiece
and press the Purge button again.
•ThePurgefunctioncanbeinterruptedatanytime
during the two minute cycle by pressing the Purge
button again or by pressing the Tap-OnTM Foot Pedal.
6. After completing the purge cycle, place a sterilized
30kHz Cavitron®Ultrasonic Insert into the handpiece
and set the Power Level Control and Lavage Control to
your preferred operating position for ultrasonic
scaling.
BETWEEN PATIENTS:
1. Remove the used Cavitron®Ultrasonic Insert. Clean
and sterilize following the Infection Control Procedures
that were enclosed with your insert.
2. Hold the handpiece over a sink or drain and activate
Purge function as described in Step 10 of the Start-Up
procedure.
3. After the purge cycle is complete, turn the System
to the OFF (0) position.
4. Remove the Steri-Mate®handpiece, clean and sterilize
following the procedures outlined in the Cavitron
Systems Infection Control Procedures booklet that was
enclosed with your system.
5. Disinfect the surfaces of the cabinet, Power Cord,
Handpiece Cable, Tap-OnTM Foot Pedal and Cable
assembly (if applicable), Water Supply and Air Supply
lines by applying an approved non-immersion type
disinfectant solution* carefully following the instructions
provided by the disinfectant solution manufacturer.
To clean the system, generously spray disinfectant
solution on a clean towel and wipe all surfaces.
Discard used towel. Dry with a clean cloth. To disinfect
the system, generously spray disinfectant on a clean
towel and wipe all surfaces. Allow disinfectant solution
to air dry. Never spray disinfectant solution directly on
the system.
6. Inspect the handpiece cable for any breaks or tears.
7. If using a closed water supply or DualSelect Dispensing
System, check for adequate fluid volumes for the next
patient.
8. When ready to use, place a sterilized Steri-Mate®
Handpiece onto the handpiece cable assembly and
insert a sterilized ultrasonic insert into the handpiece
and adjust system controls as preferred.
SHUT-DOWN PROCEDURES AT THE
END OF THE DAY:
Follow the “Between Patients” maintenance procedures,
Steps 1 through 6. In addition, it is recommended to close
the manual shut-off valve on the dental water supply system.
*NOTE: Water-based disinfection solutions are
preferred. Some alcohol-based disinfectant
solutions may be harmful and may discolor
plastic materials.
10.2 Weekly Maintenance
It is strongly recommended that this system be
disinfected by chemically flushing the waterlines with
a 1:10 Sodium Hypochlorite solution (NaOCl) at the
end of each week. This can be accomplished by
connecting this device to the Cavitron DualSelect
Dispensing System or a number of other devices
available from your local distributors. Where this device
is connected to the Cavitron DualSelect Dispensing
System, please follow the DualSelect system’s Directions
for Use manual. If connected to another device, please
follow those directions for use, keeping in mind that
a chemical flush should be performed at maximum water
flow for at least 30 seconds. The system should be left
undisturbed for 10 minutes but no more than 30 minutes
to allow the sodium hypochlorite solution to soak in the
lines. As a suggestion, it is recommended that a sign
be placed on the system stating that the SYSTEM IS
BEING DISINFECTED WITH A STRONG DISINFECTANT
AND SHOULD NOT BE USED. When ready, flush system
with clean water for at least 30 seconds or until sodium
hypochlorite odor disappears. ALL CHEMICALS MUST
BE FLUSHED FROM THE SYSTEM BEFORE IT IS READY
FOR PATIENT USE.

17
10.3 Monthly Maintenance
WATER LINE FILTER MAINTENANCE:
When the water line filter becomes discolored, the filter
should be replaced to prevent reduced water flow to the
Cavitron Plus system. A 10-pack of replacement filters is
available by ordering Part Number 90158 from your local
DENTSPLY Professional distributor.
1. Verify that the system is turned OFF.
2. Disconnect the water supply hose from the dental
office water supply. If a quick disconnect connector is
attached to the end of the hose, relieve the water
pressure by pressing the tip of the connector in an
appropriate container and drain the water.
3. Grasp the fittings on either side of the filter disk and
twist counterclockwise. Remove the filter section
from either side of the water hose.
4. Install the replacement filter onto the water hose
fittings. The filter should be positioned to match up
with the correct hose fitting.
5. Hand tighten one hose fitting onto filter in a clockwise
direction. Tighten second hose onto filter in clockwise
direction. Reconnect the water supply line, operate unit
to bleed the air and test for leaks.
SECTION 11:
Troubleshooting
Although service and repair of the Cavitron Plus Ultrasonic
Scaler should be performed by DENTSPLY personnel, the
following are some basic troubleshooting procedures that will
help avoid unnecessary service calls. Generally, check all
lines and connections to and from the System. A loose plug
or connection will often create problems. Check the settings
on the System’s controls.
11.1 Troubleshooting Guide
Symptom:
System operates: Tap-On™ Technology is not
working
1. Tap-OnTM Technology might be disabled. Refer to section
8.5.
2. Check to see if handpiece is in holder. Tap-On™
Technology is disabled when handpiece is in holder.
3. Check to see if the insert is secured inside the
handpiece. Tap-On™Technology is disabled when there
is no insert in the handpiece.
Symptom:
System will not operate: No Power ON indicator
1. Check that the Main Power Switch is in the ON (l)
position, and that the detachable Power Cord is fully
seated in the receptacle on back of System.
2. Check that the system’s power cord plug is fully seated
in an approved AC wall outlet.
3. Check that the wall outlet is functional.
Symptom:
System will not operate: Power ON Indicator is
illuminated
1. If the office has more than one Tap-OnTM Foot Pedal,
test each to ensure that the proper Tap-OnTM Foot
Pedal is being used. With a handpiece and insert
installed, depress the Tap-OnTM Foot Pedal to the
first position. The system should dispense water. If
none of the Tap-OnTM Foot Pedals operate the system,
continue to the next step.
2. Resynchronize one Tap-OnTM Foot Pedal to the system
(see Section 7.8 Tap-OnTM Foot Pedal Synchronization).
Symptom:
System operates: No water flow to insert tip or
handpiece overheats
1. Assure that handpiece lavage control is properly
adjusted.
2. Check for clogged insert. Replace insert if necessary.
3. Check that dental office water supply valves are open.
4. If the system is connected to DualSelect Dispensing
System, check that fluid level in the selected bottle is
sufficient. Make sure valves are open when using
external water source.
5. Check that the water line filter is clean. Replace
filter if needed.
Symptom:
System operates: No insert cavitation
1. Check that the Power Level Control is not in Rinse
Mode.
2. Check the insert for damage and that it is properly
installed in the handpiece.
3. Check that the handpiece is properly installed to the
cable assembly.
4. Verify that the soft nozzle grip is flush with the hard
plastic of the insert port.
5. Turn the system’s Main Power Switch to the OFF (0)
position. Wait 5 seconds and turn the system back ON.
6. If problem still exists, replace both “AA” batteries in
Tap-OnTM Foot Pedal with new “AA” batteries (Refer to
section 7.7) or connect the Auxiliary Tap-OnTM Foot
Pedal Cable.
Symptom:
System operates: Purge Mode will not function –
icon flashing
1. Check that there is no insert in the handpiece.
2. Check that handpiece is properly installed to the cable
assembly.
Symptom:
System operates: Service Indicator blinking
• FastBlinking(3blinkspersecond)
– Indicates improper set-up
1. If insert is in the handpiece, remove. Verify the
handpiece is properly seated and hold the foot
control for 2 seconds. If blinking stops, the system
is ready for use. If blinking remains, continue to the
next step.

18
2. Attach a NEW handpiece and hold Tap-OnTM Foot Pedal
for 2 seconds. If blinking stops, the system is ready
for use. Discard the old handpiece or return if within
warranty. If blinking remains, continue to the next
step.
3. Install and fully seat an insert into handpiece. Hold
Tap-OnTM Foot Pedal for 2 seconds. If blinking stops,
the system is ready for use. If blinking remains,
continue to the next step.
4. Install and fully seat a NEW insert in handpiece and
hold Tap-OnTM Foot Pedal for 2 seconds. If blinking
stops, system is ready for use. Discard old insert or
return if within warranty. If blinking remains, refer to
Section 11.2 Technical Support and Repairs to
have unit serviced as soon as possible.
• SlowBlinking(1blinkpersecond)
-- The system is not operating within factory
specifications.
1. Remove insert.
2. Turn Main Power Switch OFF, (O) position. Wait five
seconds. Turn unit ON, (I) position.
3. Operate Purge function.
4. If service indicator still blinks, refer to section 11.2
Technical Support and Repairs to have unit
serviced as soon as possible.
Symptom:
System operates: Service Indicator illuminated
1. Ensure that the base unit has adequate ventilation and
is not near a heat source (i.e. radiator, heat lamp,
sunlight or other heat producing operatory equipment).
2. Turn Main Power Switch to the OFF (O) position.
Allow system to cool for 10 minutes and turn system
to the ON (I) position. Verify light is not illuminated.
3. If light is still illuminated, refer to section 11.2
Technical Support and Repairs to have unit serviced as
soon as possible.
11.2 Technical Support and Repairs
For technical support and repair assistance call DENTSPLY
Professional Cavitron CareSM Factory Certified Service at 1-800-
989-8826 Monday through Friday, 8:00 A.M. to 5:00 P.M.
(Eastern Time). For areas outside the U.S., contact your local
DENTSPLY Professional representative.
SECTION 12:
Warranty Period
The Cavitron Plus Ultrasonic Scaler is warranted for TWO
YEARS from date of purchase. The Steri-Mate®Handpiece
enclosed with your system is warranted for SIX MONTHS from
date of purchase. Refer to the Warranty Statement Sheet
furnished with your system for full Warranty Statement and
Terms.
SECTION 13:
Specifications
Electrical Voltage Continuous (100-240 VAC)
Current 1.0 Amperes, Maximum
Phase
Single
Frequency 50/60 Hertz
Water Pressure 20 to 40 psig
(138 to 275 kPa)
Water Temperature < 25oC (77oF)
Air Pressure 65 to 100 psig
(448 to 600 kPa)
Water Flow Rate Minimum Setting (CCW) < 15 ml/min
Maximum Setting (CW) > 55 ml/min
Weight 4.4 lbs (2 Kg)
Dimensions Height: 6 in. (15,24 cm)
Width: 9.5 in. (24,13 cm)
Depth: 8 in. (20,32 cm)
Handpiece Cable length: 6.5 ft. (2.0 M)
Auxillary Footswitch Cable length: 8 ft. (2.4 M)
Water Supply Line length: 8 ft. (2.4 M)
Footswitch Protection Class IPX1. Not for operating
theatres.
Remote
Communication Frequency: 2405 to 2480 MHz
Power: < 1mW
Channels: 16
Operating Environment Temperature: 15 to
40 Deg.Celsius (59 to 104 Deg
Fahrenheit)
Relative Humidity: 30% to 75% (non-condensing)
Transport and Storage Conditions Temperature:
-40 to 70 Deg. Celsius (-40 to 158 Deg.
Fahrenheit)
Relative Humidity: 10% to 100% (non-
condensing)
Atmospheric Pressure: 500 to 1060 hPa

19
SECTION 14: Classifications
•Typeofprotectionagainstelectricshock: Class1
•Degreeofprotectionagainstelectricshock: TypeB
•Degreeofprotectionagainsttheharmfulingressofwater: Ordinary
•Modeofoperation: Continuous
•Degreeofsafetyofapplicationinthepresenceofaflammable
anaesthetic mixture with air or with oxygen or nitrous oxide: Equipment not suitable for use in the presence of
flammable anaesthetic or oxygen.
•Accordingtomedicaldevicedirective:
IIA (rule 9)
SECTION 15: Disposal of Unit
U.S. - Dispose of the system components in accordance with state and local laws.
EU - Dispose of in accordance with the Waste Electrical and Electronic Equipment Directive 2002/96/EC of the European
Parliament and the Council of the European Union.
Symbol Identification
``
AC POWER
TYPE B APPLIED PART EQUIPMENT
PROTECTIVE EARTH (GROUND)
Footswitch not for operating theatres
Protection Class- IPX1
IPX1 Classification of ingress of water
Consult Instructions for Use
O/ IAC Power Switch (0 = Off, |= On)
Manufactured By
Date of Manufacture
Footswitch
Serial Number
Batch Code / Lot Number
MEDICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL-
2601-1/60601-1, CAN/CSA C22.2 NO.601.1
13VA
IPX1
This device complies with part 15 of the FCC Rules.
Operation is subject to the following two conditions:
1) this device may not cause harmful interference, and
2) this device must accept any interference received,
including interference that may cause undesired operation.
Foot pedal: FCC ID:TF3-DPD81675
IC: 4681B-81675
Base: FCC ID:TF3-DPD73227323
IC: 4681B-73227323
Dispose of in accordance with the Waste Electrical and
Electronic Equipment Directive 2002/96/EC of the
European Parliament and the Council of the European
Union
This is a wireless device.
WEEE
LOT
This symbol is a mandatory marking for devices enter-
ing the European market to indicate conformity with
the essential health and safety requirements set out in
European Directives. The symbol may be accompanied
by a four-digit identification number of the notified
body.

20
SECTION 16: Electro-Magnetic Compatibility Precautions
Guidance And Manufacturer’s Declaration - Electromagnetic Emissions
The Ultrasonic Scaler model G136 is intended for use in the electromagnetic environment specified
below. The customer or the user of the Ultrasonic Scaler should assure that it is used in such
an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The Ultrasonic Scaler uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B The Ultrasonic Scaler is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A Not Applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not Applicable The Ultrasonic Scaler is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Guidance And Manufacturer’s Declaration – Electromagnetic Immunity
The Model G136 is intended for use in the electromagnet environment specified below. The customer or the
end user of the Model should assure that it is used in such an environment.
Immunity test IEC 60601
Test level
Compliance level Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
-
+6 kV contact
+8 kV air
-
-
-
Table of contents
Languages:
Other Cavitron Dental Equipment manuals
Popular Dental Equipment manuals by other brands

KERR
KERR SonicFill Instructions for use

Sealey
Sealey RE101.V2 instructions

BRYANT DENTAL
BRYANT DENTAL IGNIS manual

Forest Dental
Forest Dental Syringe Operator's guide

Aseptico
Aseptico AEU-27A Operation and maintenance instruction manual

Dentsply Sirona
Dentsply Sirona CEREC Primescan AC Unpacking and installation instructions

Gendex
Gendex ORTHORALIX 8500 Operator's manual

Woodpecker
Woodpecker LED.H instruction manual

iRay Technology
iRay Technology Pluto0001X user manual

BRASSELER USA
BRASSELER USA enLITE Instructions for use

Dentsply Sirona
Dentsply Sirona Endo IQ Directions for use

KaVo
KaVo ERGOcom Customer's operating and assembly instructions