CEFLA NewTom GO User manual

97050801
Rev. 09
2018-07
NewTom GO
EN

2
OPERATOR'S MANUAL
EN
ITALIANO
Contents
0. DEVICE IDENTIFICATION........................................................................................................................................4
1. INTRODUCTION AND INDICATIONS FOR USE .....................................................................................................5
1.1. DESCRIPTION OF THE MANUAL .......................................................................................................................6
1.2. GENERAL WARNINGS........................................................................................................................................7
1.3. REQUIREMENTS (NOT PROVIDED WITH THE PRODUCT) ............................................................................8
1.4. STANDARDS AND REGULATIONS ....................................................................................................................8
1.5. CLASSIFICATIONS..............................................................................................................................................8
1.6. STYLISTIC CONVENTIONS ................................................................................................................................9
1.7. GENERAL SAFETY WARNINGS.......................................................................................................................10
1.7.1. INSTALLATION CONDITIONS ...................................................................................................................10
1.7.2. CONDITIONS OF USE ...............................................................................................................................11
1.7.3. WARRANTY ................................................................................................................................................11
1.7.4. MAINTENANCE AND DISPOSAL...............................................................................................................12
1.7.5. CLEANING AND DISINFECTION ...............................................................................................................13
1.7.6. HYGIENE PROCEDURES FOR PATIENT PROTECTION ........................................................................14
1.8. SAFETY WARNINGS .........................................................................................................................................15
1.8.1. CONDITIONS OF USE ...............................................................................................................................15
1.8.2. GENERAL SAFETY ....................................................................................................................................15
1.8.3. SAFETY DURING X-RAY DEVICE MOVEMENTS ....................................................................................15
1.8.4. EMERGENCY BUTTON .............................................................................................................................16
1.8.5. EXPOSURE TO LASER RADIATION .........................................................................................................16
1.8.6. ELECTROMAGNETIC SAFETY .................................................................................................................17
1.8.7. PROTECTION AGAINST RADIATION .......................................................................................................19
1.8.8. APPLIED PARTS ........................................................................................................................................19
1.8.9. STRAY RADIATIONS..................................................................................................................................20
2. DESCRIPTION OF OPERATION ............................................................................................................................21
3. COMPONENTS .......................................................................................................................................................22
4. CONTROL PANEL ..................................................................................................................................................23
4.1. CONSOLE ONBOARD THE MACHINE .............................................................................................................23
4.2. X-RAY EMISSION REMOTE CONTROL ...........................................................................................................23
4.3. PERFORM A SIMULATION (DUMMY RUN) .....................................................................................................24
5. PERFORMING A 2D X-RAY EXAMINATION .........................................................................................................25
5.1. STARTING THE SYSTEM..................................................................................................................................25
5.2. SELECTING THE EXAMINATION FROM THE CONTROL CONSOLE ............................................................26
5.2.1. 2D EXAMINATIONS AVAILABLE ...............................................................................................................26
5.2.2. SELECTING AN EXAMINATION ................................................................................................................29
5.2.2. SELECTING AN EXAMINATION.....................................................................................................................29
5.2.3. SETTING THE PROJECTION TYPE ..........................................................................................................33
5.3. PREPARATION OF THE X-RAY EXAMINATION..............................................................................................34
5.3.1. EDENTOLOUS PATIENTS DEVICES (OPTIONAL) ..................................................................................34
5.4. PATIENT POSITIONING ....................................................................................................................................34
5.4.1. LASER TRACES .........................................................................................................................................35
5.4.2. PATIENT POSITIONING DESCRIPTION (CRANIOSTAT) ........................................................................36
5.4.3. PAN, DENT EXAMINATIONS .....................................................................................................................37
5.4.4. TMJ EXAMINATION....................................................................................................................................38
5.4.4.1. LATERAL TMJ ...................................................................................................................................38
5.4.4.2. FRONTAL TMJ ..................................................................................................................................39
5.4.5. SIN EXAMINATION.....................................................................................................................................39
5.5. ACQUISITION OF THE EXAM ...........................................................................................................................40
5.6. TYPICAL IMAGES OF THE EXAMS..................................................................................................................41
5.6.1. ADULT PANORAMIC IMAGING .................................................................................................................41
5.6.2. CHILD PANORAMIC IMAGING ..................................................................................................................41
5.6.3. TMJ EXAMINATIONS (TEMPOROMANDIBULAR JOINT) ........................................................................41
5.6.4. SIN EXAMINATIONS ..................................................................................................................................42
5.6.5. DENTITION/BITEWING EXAMINATIONS..................................................................................................42
6. 3D TOMOGRAPHIC EXAMINATION (CB3D) .........................................................................................................43
6.1. SELECTING THE EXAMINATION FROM THE CONTROL CONSOLE ............................................................43
6.2. POSITIONING THE PATIENT FOR 3D EXAMINATIONS .................................................................................49

EN
OPERATOR'S MANUAL
3
6.3. PERFORMING THE EXAMINATION .................................................................................................................49
6.4. SPECIAL CONSIDERATIONS FOR CBCT EXAMINATIONS WITH CHILDREN OR SMALL PATIENTS .......50
7. VIEWING AND SAVING..........................................................................................................................................51
8. PERIODIC CHECKS AND MAINTENANCE ...........................................................................................................52
8.1. PERIODIC INSPECTIONS TO CHECK THE IMAGE ........................................................................................52
8.1.1. X-RAY EMISSION QUALITY CONTROL....................................................................................................52
8.1.2. COLLIMATOR TEST...................................................................................................................................53
8.1.3. 2D IMAGE QUALITY CONTROLS BY MEANS OF PHANTOM.................................................................53
8.1.4. 3D IMAGE QUALITY CONTROLS..............................................................................................................54
8.2. QA PHANTOM ACQUISITION ...........................................................................................................................55
8.2.1. QA PHANTOM TROUBLESHOOTING.......................................................................................................57
9. TECHNICAL DATA..................................................................................................................................................58
9.1. ELECTRICAL CHARACTERISTICS ..................................................................................................................58
9.2. RADIOLOGICAL CHARACTERISTICS..............................................................................................................58
9.3. ISODOSE CURVES FOR 2D EXAMINATIONS.................................................................................................60
9.4. ISODOSE CURVES FOR CBCT EXAMINATIONS ...........................................................................................60
9.5. PANORAMIC SENSOR CHARACTERISTICS(PAN).........................................................................................61
9.6. 3D PANEL FEATURES ......................................................................................................................................61
9.7. LASER CHARACTERISTICS.............................................................................................................................61
9.8. DIMENSIONAL CHARACTERISTICS................................................................................................................62
9.9. ENVIRONMENT CHARACTERISTICS ..............................................................................................................65
9.10. PC REQUIREMENTS ....................................................................................................................................66
9.11. MOBILE DEVICE REQUIREMENTS .............................................................................................................67
9.12. IDENTIFICATION LABEL POSITION ............................................................................................................68
10. ERROR MESSAGES ..............................................................................................................................................69
11. USER’S LICENCE CONTRACT..............................................................................................................................70
11.1. GENERAL CONDITIONS OF THE IMAGE SOFTWARE LICENCE .............................................................70
11.1.1. USER’S LICENCE .................................................................................................................................70
11.1.2. COPYRIGHT..........................................................................................................................................70
11.1.3. USE OF THE SOFTWARE PRODUCT AND EXPRESS TERMINATION CLAUSE.............................71
11.1.4. GUARANTEE AND EXCLUSIONS FROM THE GUARANTEE ............................................................71
11.1.5. LIMITS OF RESPONSIBILITY...............................................................................................................71
11.1.6. APPLICABLE LAWS, JURISDICTION AND COURT OF JURISDICTION ...........................................71
11.1.7. WRITTEN FORM ...................................................................................................................................72
11.1.8. TRANSLATION......................................................................................................................................72
11.1.9. SOFTWARE COMPLIANCE WITH THE REGULATIONS IN FORCE..................................................72
12. INSPECTION AND MAINTENANCE.......................................................................................................................73
12.1. USER INSPECTION ......................................................................................................................................73
12.2. TECHNICAL MAINTENANCE........................................................................................................................74

4
OPERATOR'S MANUAL
EN
0. DEVICE IDENTIFICATION
This manual applies to the following models of the NewTom GO family:
-NewTom GO REF. 708J (NewTom GO 2D version);
-NewTom GO REF. 708K (NewTom GO 3D version).
Unless otherwise indicated, the instructions included in this manual apply to all models of the NewTom GO family
(hereinafter referred to as "NewTom GO").
In the following part of this manual, instructions that only apply to the device model NewTom GO REF. 708J are
identified by the description “
Only for 2D machines
”.
Instructions that only apply to the device model NewTom GO REF. 708K are identified by the icon, by the
description “
Only for 3D machines
” or by both.

EN
OPERATOR'S MANUAL
5
1. INTRODUCTION AND INDICATIONS FOR USE
For 2D machine only
:
NewTom GO is an extraoral X-ray system for digital panoramic exams. It is intended to produce orthopanoramic
images of the maxillofacial region and to carry out diagnostic examination on teeth, dental arches and other
structures in the oral cavity.
For 3D machine only
:
NewTom GO is an extraoral X-ray system for digital panoramic exams and tomographies, intended to:
(i) produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental
arches and other structures in the oral cavity;
(ii) produce tomographic images of the oral cavity and maxillofacial structures and carry out diagnostic examination
on teeth, dental arches, structures of the oral cavity and some cranial bones.
The system performs tomographic exams with the acquisition of X-ray images through a rotating sequence and the
reconstruction of a three-dimensional matrix of the examined volume, thus producing two- and three-dimensional
views of the volume itself. This technique is known as CBCT.
NewTom GO is a digital X-ray device, suitable for expert professionals, which allows to obtain dental images in a
simple and automatic way. The image is acquired by means of an X-ray detector and a constant potential X-ray
source, powered by a high-voltage and high-frequency generator. The image is then transmitted to a computer in real
time.
USA federal law restricts this device to sale by or on the order of a licensed healthcare practitioner.
NewTom GO allows the following 2D projections:
- standard or panoramic views for paediatric patients (PAN);
- complete or partial views of the teeth, selected by the user (DENT);
- front and side views of maxillary sinuses (SIN);
- side and posteroanterior views of the temporomandibular joints (TMJ), from different angles.
Furthermore, the 3D version machine also allows to acquire tomographic images.
NewTom GO is intended for use in dentistry, in the following fields:
- endodontics;
- periodontology;
- dental prosthesis;
- functional diagnosis and therapy of craniomandibular dysfunctions;
- dental surgery;
- dental implants;
- maxillofacial surgery;
- orthodontics.
Do not use on patients (children) less than approximately 104 cm in height and less than 19 kg in weight.
These height and weight measurements approximately correspond to that of an average 4 year old.
Use of equipment and exposure settings designed for adults of average size can result in excessive
radiation exposure for a smaller patient. Studies have shown that pediatric patients may be more
radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing radiation is higher), and so
unnecessary radiation exposure is of particular concern for pediatric patients.
Not for use with patients not vigilant and cooperative, since the patient must be able to understand and
follow the operator's instructions for a correct positioning.

6
OPERATOR'S MANUAL
EN
Contraindications:
- visualisation of cartilaginous structures;
- the CBCT technique (
for 3D machines only
) has a limited ability to detect soft tissues.
In order to ensure a safe use of the device, specifically in the radiography of children it is suggested to
consult the general indications described in the guidelines for dental radiographs as i.e. the ones referred on
the website Image Gently (www.imagegently.org), for dental Rx, or even in the FDA website for “Pediatric X-
Ray imaging”.
The manufacturer’s website contains a list of authorised agents.
Address for circulation of consumers within the Russian Federation:
Zenith R.S. LLC
Bldg 3, Property 15, Gorodok-17 Str., Bolshie Vyazemy Settlement, Odintsovo District
143051 Moscow Region, Russia
phone: +7 (495) 980-13-50
+7 (495) 787-68-36
e-mail: sales@zenith-rs.ru
Website: www.zenith-rs.ru
For technical service, it is possible to contact the local distributor.
1.1. DESCRIPTION OF THE MANUAL
This manual is an essential consultation tool and contains important information and instructions for the use
of the X-ray system and the relevant controls.
These instructions describe how to properly and safely use the digital X-ray system.
Carefully read and familiarise yourself with the entire contents of the manual before attempting to use the
system.
To use the software, refer to the specific manual.
The manual is only provided in electronic format and can be consulted directly on the PC screen during use.
It is advisable to keep a copy of this manual within reach with the aim of training the operators and as guide
for consultation during the use of the device. This manual also contains all the essential information for the
safety of patient, operator and device.
It is therefore advisable to read carefully the paragraphs on the safety rules.
The original text is in Italian.

EN
OPERATOR'S MANUAL
7
1.2. GENERAL WARNINGS
The digital X-ray system with its drivers and software have been developed and manufactured by CEFLA s.c. - Via
Selice Prov.le 23/A 40026 Imola (Italia), hereinafter referred to as the Manufacturer, which is the manufacturer and
distributor in compliance with the EC Medical Device Directive.
In order to use the system, the user must have a Personal Computer with a suitable software for capturing and
saving images; further information about its installation and use is included in the Software user manual. Carefully
read this Manual and the Computer and Software manuals before using the equipment.
- The contents of this publication are valuable trade secrets and must not be given to third parties, stored, copied,
reproduced, disclosed or transferred in any manner (via computer, photocopies, translations or other means)
without the prior written consent of the Manufacturer.
- The Manufacturer pursues a policy of continual improvement of its products; therefore, some specific instructions
and images contained in this manual may differ from the product purchased.
- The Manufacturer reserves the right to make changes without prior notice.
- The information, technical specifications and illustrations contained in this publication are not binding. The
Manufacturer reserves the right to make technical modifications and improvements without modifying these
instructions.
- All the registered trademarks and the product names mentioned are the property of the respective owners.
- Carefully read the USER LICENSE AGREEMENT before using the product. When installing the program, you will
explicitly be asked to accept the agreement; if you do not accept, the program cannot be installed.
Please pay particular attention to the sections in the manual where the following symbols appear:
Patient or operator safety-related warnings.
Important information on product use.
For 3D machine only.
In accordance with the privacy laws in force in several countries, all sensitive personal information must be
adequately protected. In addition, patients must sign a consent form before any personal information or
images are transmitted across networks. If required by the laws in force, dentists are obliged to protect data
using a protection password. Refer to the Microsoft® Windows operating system manual for data access
protection methods by means of password.
It is recommended to regularly (at least once a week) make a backup copy of the databases. This will allow
restoring the data in the event of damage to the hard disc of the PC or the databases themselves.

8
OPERATOR'S MANUAL
EN
1.3. REQUIREMENTS (NOT PROVIDED WITH THE PRODUCT)
For proper functioning, the device requires connection to a personal computer (PC) and the relative software. For the
minimum requirements of the PC, refer to the paragraph TECHNICAL DATA.
The PC is not included with the device. It is recommended to only use a PC compliant to the IEC 60950-
1:2007 standards for information technology devices.
1.4. STANDARDS AND REGULATIONS
This system has been designed in accordance with the following standards:
- Directive 93/42/EEC and subsequent amendments and integrations
(Dir. 2007/47/EC) - Medical Device Directive;
- Directive 2006/42/EC - Machinery Directive.
Technical standards:
IEC 60601-1:2005
IEC 60601-1-2:2007
IEC 60601-1-3:2008
IEC 60601-2-63:2012
IEC 60601-1-6:2010
IEC 62366:2007
IEC 62304:2006
IEC 60825-1:1993
The CE mark certifies the compliance of the product described below with the European Community
Medical Device Directive 93/42/EEC and subsequent amendments.
1.5. CLASSIFICATIONS
The system is classified as Class I and Type B as regards safety according to IEC 60601-1.
The system is classified as a Class IIB medical electrical X-ray device in accordance with Medical Device Directive
93/42/EEC and subsequent amendments.

EN
OPERATOR'S MANUAL
9
1.6. STYLISTIC CONVENTIONS
The following symbols may be found on the X-ray device:
Equipment compliant with directive 93/42/EEC as amended.
Notified body: IMQ spa.
Applied part of type B, according to IEC 60601-1.
Product/equipment identification code.
Product serial number.
Manufacturer.
Date of manufacture (month / year).
Consult the enclosed documentation before using the relevant part of the equipment.
It is necessary to read the user's manual before using the device.
POWER
Power switch.
I
Unit ON.
O
Unit OFF.
Ionizing radiation warning symbol.
Class 1 LASER radiation warning symbol.
Crushing hazard.
Disposal symbol in accordance with Directive 2012/19/EU.
Ukraine compliance mark.

10
OPERATOR'S MANUAL
EN
1.7. GENERAL SAFETY WARNINGS
These instructions describe how to use the system correctly. Please carefully read this manual before using
the device.
The owner or the installation site manager is obliged to verify compliance with the local regulations in force and/or
ask a Qualified Expert for advice. Pay particular attention to fulfilling the obligations of the law regarding the
protection of workers, the population and patients against radiation.
The main REFERENCE STANDARDS are listed in this manual (1.4 - Standards and Regulations).
Do not use the system for uses different from those described in the indications for use (Introduction) and do not use
it if you do not have the necessary knowledge in the dentistry and radiological field.
The law restricts and reserves the sale and use of this device exclusively to physicians, dentists or radiology
specialists.
1.7.1. INSTALLATION CONDITIONS
- The system must not be used if it shows any electrical, mechanical or radiation defect. Like for all medical
electrical systems, this device requires proper installation, use, maintenance and service with the aim of assuring
safe and efficient operation.
- The entire system must be installed by a technician authorised by the Manufacturer under supervision of a
Qualified Expert.
- The room where the system is installed must exclusively be for medical use and designed by an expert in
protection against the risks associated with exposure to radiation in accordance with the regulations in force in
the country of use.
- For Europe, the electric system in the room where the device is installed must comply with the IEC 60364-7-710
standards (requirements for electric systems in rooms used for medical purposes).
- The X-ray system requires special precautions with respect to electromagnetic compliance and must be installed
in accordance with the recommendations given in the paragraph “Electromagnetic safety” in this manual.
- The maximum dimensions reached while the unit is being installed must be taken into consideration to avoid
banging into any objects present in the room. Refer to the dimensional diagram in the service manual.
- Make sure that the operator can communicate verbally and visually with the patient during the examination.
- The system can be installed in the following configurations:
1. wall mounting;
2. on the floor surface with a static baseplate (optional).
- Installation not in compliance with the instructions provided by the Manufacturer might cause increased
electromagnetic emission of the X-ray system and reduce its immunity to disturbances.
- In particular, use a screened cable for connection of the X-ray remote control and make the connection as
specified in the technical manual.
For further details, refer to the installation template and the detailed instructions given in the service manual.

EN
OPERATOR'S MANUAL
11
1.7.2. CONDITIONS OF USE
The equipment may only be used by authorised and adequately trained staff (dentists and paramedics).
Use in compliance with national regulations for protection against ionising radiation, e.g.:
(a) Three-dimensional imaging must not be used for routine screening exams. Three-dimensional imaging
exams must be prescribed according to the patient's needs.
(b) Each exam must be justified by evidence that the benefits outweigh the risks.
(c) All patients must wear leaded aprons with thyroid collar.
(d) Before the exam, ask women of reproductive age if they are pregnant or if there is any possibility that
they might be. If there is, the patient should not undergo the exam, unless a radiologist from a trustworthy
hospital is consulted in order to evaluate, together with the patient and the operator, the benefits and the
risks associated with this procedure, taking into account the possibility of choosing a different kind of exam.
(e) The operator must keep a safe distance, protect him/herself with an adequate shielding and remain
close to the patient in the exam room only if the patient needs assistance. If the operator has to remain in
the exam room, he/she must wear a leaded apron with thyroid collar.
(f) Inform the patient about the risks associated with the exam, acquire his/her informed consent and archive
the document.
In case of claims or need of technical assistance, users in Brazil are required to contact the following email
address: servico.odontologico@cefla.it.
Users in the USA market are required to contact:
Cefla North America Inc.,
6125 Harris Technology Blvd., Charlotte, NC, 28269 United States
Phone: +1 704 598 0020, e-mail: info@ceflaamerica.com
1.7.3. WARRANTY
The Manufacturer guarantees the safety, reliability and performance of the device.
The warranty is valid only under the following terms:
-closely observe the conditions specified in the warranty certificate itself;
-the equipment is only to be used as instructed in this manual;
-equipment installation, upgrade and technical support must be performed exclusively by personnel
authorised by the Manufacturer to carry out these operations;
-do not open the device covers: installation, repairs and in general all the operations that require opening
the device must be carried out exclusively by technicians authorised by the Manufacturer;
-the equipment is to be installed in rooms that satisfy the requirements specified in the manual;
-the room where the X-ray unit is installed must be in compliance with the official directives that govern
radiation protection in the country of use.

12
OPERATOR'S MANUAL
EN
1.7.4. MAINTENANCE AND DISPOSAL
Never remove the device covers.
The device does not contain parts that can be repaired directly by the user. In the event of malfunctioning, do not
attempt to carry out any type of maintenance operation. If you find or suspect any kind of system malfunctioning, do
not attempt to carry out any type of maintenance operation and do not use the system on a patient, but directly
contact your local distributor.
The user may not carry out maintenance on any mechanical or electronic part of the x-ray system.
Opening the cases to access the internal circuits may cause device breakage and failure of the electrical safety
devices and will lead to forfeiture of the warranty.
Any maintenance, repairs and modifications of the device must be carried out only by personnel directly authorised
by the Manufacturer or by third parties expressly authorised by the Manufacturer and must be carried out according
to the laws in force and the generally accepted technical standards.
All the system components must be checked and replaced, if necessary, by qualified personnel.
For any maintenance operation, please contact the Manufacturer via the website indicated on the cover of this
manual by filling in the Information Request form.
Further information about the device regular inspection and maintenance is provided in the document “NewTom GO -
Inspection and Maintenance”.
Should you for any reason need to return the device or its parts to the Manufacturer or a Technical Service centre,
disinfect all the external parts of the device using a specific product (see the paragraph “Cleaning and disinfection”)
and preferably return it in its original packaging.
At the end of its lifetime, dispose of the device in accordance with the regulations in force. It is also advisable to
disinfect all the external parts of the device before disposal and to separate the materials for differentiated waste
collection.
In compliance with Directives 2011/65/EU and 2012/19/EU regarding restriction of the use of certain hazardous
substances in electrical and electronic equipment along with waste electrical and electronic equipment, it is forbidden
to dispose of this equipment in the municipal waste stream as unsorted municipal waste. When purchasing a new
device of an equivalent type, one for one, the device that has come to the end of its lifetime should be returned to the
distributor for disposal. As regards reuse, recycling and other forms of recovery of waste electrical and electronic
equipment, the Manufacturer carries out the functions defined by current local laws. Appropriate differentiated waste
collection for subsequent recycling treatment and environmentally friendly disposal contributes to preventing possible
negative effects on the environment and health and encourages recycling of the materials of which the device is
made up. The crossed-out bin symbol on the device indicates that the product must be collected separately from
other waste at the end of its useful life. Under local legislation, fines can be imposed if the equipment is disposed in
an illegal manner.

EN
OPERATOR'S MANUAL
13
1.7.5. CLEANING AND DISINFECTION
Cleaning is the first step necessary in any disinfection process. The physical action of rubbing a surface with
detergents and surface-active agents and rinsing with water removes a large number of microorganisms. If
a surface is not cleaned first, the disinfection process cannot be successful.
When a surface cannot be adequately cleaned, it should be protected with barriers.
The external parts of the device must be cleaned and disinfected using a product for hospital use with indications for
HIV, HBV and tuberculosis (medium-level disinfectant) specifically for small surfaces.
The various drugs and chemical products used in dentist’s surgeries may damage the painted surfaces and the
plastic parts. The tests and research conducted have demonstrated that the surfaces cannot be fully protected
against attack of all the products found on the market. It is therefore recommended to use barrier protections
whenever possible.
The aggressive effects of chemical products also depend on the time they stay in contact with the surfaces. It is
therefore important not to leave the product on the surfaces of the device for longer than the time indicated by the
manufacturer.
It is recommended to use the specific medium-level disinfectant, STER 1 PLUS (CEFLA S.C.), which is compatible
with the painted surfaces, plastic parts and unpainted metal surfaces. Alternatively, it is recommended to use
products that contain:
-96% ethanol Concentration: maximum 30 g for every 100 g of disinfectant.
-1-propanol (N-propanol, propyl alcohol, N-propyl alcohol). Concentration: maximum 20 g for every 100 g of
disinfectant.
-Combination of ethanol and propanol. Concentration: the combination of the two must be at maximum 40 g for
every 100 g of disinfectant.
-Painted surfaces and plastic parts.
- Incidin Spezial (Henkel Ecolab);
- Omnizid (Omnident);
- Plastisept (Alpro) (not tuberculocide as not an alcohol-based disinfectant);
- RelyOn Virkosept (DuPont);
- Green & Clean SK (Metasys) (not tuberculocide as not an alcohol-based disinfectant).
-Do not use products containing isopropyl alcohol (2-propanol, isopropanol).
-Do not use products containing sodium hypochlorite (bleach).
-Do not use products containing phenols.
-Whatever product you use, follow the instructions given by the manufacturer.
-Do not mix the STER 1 PLUS disinfectant with other products.
-Do not spray the product directly on the device surfaces.
For cleaning and disinfection use disposable, soft, non-abrasive
paper (do not use recycled paper) or sterile gauze.
- It is recommended to turn off the device before cleaning and
disinfecting the external parts.
- Everything used for cleaning and disinfection must be thrown
away after use.

14
OPERATOR'S MANUAL
EN
1.7.6. HYGIENE PROCEDURES FOR PATIENT PROTECTION
Single-use hygienic protections are the main means of protection against transmission of cross-infections
between patients. In order to prevent transmission of infectious diseases from patient to patient, it is
essential to always use the single-use protections. The single-use protections are classified as Class I
medical device and may not be replaced with others in a lower class.
The single-use protections must be in compliance with the ISO 10993-1 standards on biocompatibility and
approved by the control bodies where required (e.g. FDA, CE).
Always change the disposable protective cover for the bite before using it for a new patient.
The disposable protective covers must be stored in a clean dry place away from direct sunlight and UV rays.
The bite piece and the chinrest can be disinfected by immersing them in a cold sterilising liquid. For
sterilisation of these parts, follow the instructions of the supplier of the sterilising product.
Cover all the components that will come into contact with the hands of the dental staff with disposable
single-use protections, as they might be contaminated by indirect contact with the patient’s mouth. In
particular, be careful how you handle the control console of the device, the mouse and keyboard of the PC.
Prior to positioning patients for any X-ray exposure, always cover the bite block with a new clean non-sterile
plastic barrier, to avoid cross contamination.
Note for users in Canada: ask your trusted dental material distributor for any plastic barrier that is suitable in
size and is marketed in Canada according to the local regulations in force.
According to Health Canada, bite block covers are Class I devices and are distributed by authorised
establishments only, as listed in the MDEL database.

EN
OPERATOR'S MANUAL
15
1.8. SAFETY WARNINGS
1.8.1. CONDITIONS OF USE
In order to use the device in safe conditions, refer to the following paragraphs in the manual.
1.8.2. GENERAL SAFETY
-Do not forget to turn off the main switch on the equipment before leaving the surgery.
-The device is not protected against liquid penetration (Class IPX0 –common protection).
-The equipment is not suitable for use in the presence of a mixture of flammable anaesthetic gas with oxygen or
nitrous oxide.
-Portable telecommunications devices (RF) may interfere with the X-ray device; use in the vicinity of the X-ray
device should therefore be prohibited.
-This equipment must be stored properly so that it is kept in top working order at all times.
-The user must be present at all times when the equipment is turned on or ready for start-up. In particular, never
leave the equipment unattended in the presence of children or other unauthorised personnel in general.
-The Manufacturer shall not be held responsible (under civil and criminal law) for misuse, carelessness or
improper use of the equipment.
-If any person who is not an authorised technician changes the product in any way by replacing parts or
components with other ones not used by the Manufacturer, they shall assume responsibility for the product.
-Any computer, monitor, printer, mouse, keyboard and any other device connected to the X-ray device must be
compliant with ISO, IEC, EN or local standards.
-The Manufacturer is not responsible for problems or malfunction of parts and/or components not approved by
itself, not complying with the regulations and not installed by qualified technical personnel acknowledged by the
Manufacturer.
-Do not use electronic equipment not compliant with IEC 60601-1-2:2007 in proximity of life support equipment
(e.g. pacemakers or heart stimulators) and hearing aids. Before using any electronic device in health facilities,
always check that it is compatible with the other equipment present.
-The X-ray tube contains insulating mineral oil. This oil is potentially hazardous if ingested or if it comes into
contact with the skin or mucous membranes. In the event of a defect or fault, the oil may leak out. Avoid direct
contact with the oil and do not inhale its vapours.
-Do not eat, drink or smoke near the device.
1.8.3. SAFETY DURING X-RAY DEVICE MOVEMENTS
The X-ray device carries out movements in the vicinity of the patient and the operator.
During the execution of the X-rays, the operator controls the movements by holding down dedicated
buttons.
The reset procedure must be carried out before the patient accesses the device.
The operator must stand at a suitable distance from the moving parts. Movements can be stopped at any
time by pressing the emergency button.
During all the movements of the X-ray device, the operator must:
-closely watch over the patient, and if there is a risk of collision between the X-ray device and the patient,
immediately interrupt the movement by releasing the control button;
-not allow the patient to assume incorrect positions (resting the hands or other parts of the body in inappropriate
areas) or to move away from the examination area.
The operator must be careful not to interfere with the sensor movement during its servo-controlled activation.

16
OPERATOR'S MANUAL
EN
1.8.4. EMERGENCY BUTTON
The system is equipped with an emergency button to stop X-ray device operation, located under the patient support
arm in proximity of the telescopic column. A remote emergency button is connected using the dedicated connector on
the board located at the feet of the lifting column.
1 - Emergency button
These buttons must be activated in the event of danger and emergency, for example, failed interruption of radiation
from the source, in situations of evident danger to persons or when an emergency is signalled.
If the emergency button is pressed, radiation emission is immediately interrupted and all the movements of the
patient support and the rotary arm are stopped, all the linear motors run a distance of less than 10 mm and the rotary
movements a distance of less than 2 degrees, and the button remains locked in safe conditions.
After the emergency, turn the button in the direction indicated by the arrow to resume normal operation.
1.8.5. EXPOSURE TO LASER RADIATION
The system contains some class 1 LASER diodes, in compliance with IEC 60825-1.
Both the patient and the operator may be dazzled by laser traces.
-Do not look directly into the laser beam. Pay attention to avoid that the laser beam hits the patient’s eye.
-Keep a distance of at least 10 cm between the eye and the laser.
The position of the laser sources is indicated with the following symbol .

EN
OPERATOR'S MANUAL
17
1.8.6. ELECTROMAGNETIC SAFETY
Do not use electronic equipment not compliant with IEC 60601-1-2:2007 in proximity of life support equipment (e.g.
pacemakers or heart stimulators) and hearing aids.
Before using any electronic device in health facilities, always check that it is compatible with the other equipment
present, including the X-ray device.
Guidance and Manufacturer's declaration - Electromagnetic emissions
NewTom GO is designed to operate in the specified electromagnetic environment. The customer or the user of
NewTom GO must ensure its use in an electromagnetic environment with the following features:
Emission test
Conformity
Electromagnetic Environment
RF emissions
CISPR 11
Group 1
NewTom GO uses RF energy only for its internal
operation.
For this, the RF emissions are very low and do not
interfere with the electronic devices nearby.
RF emissions
CISPR 11
Class A
NewTom GO must be used only by adequately
trained personnel (dentists and paramedics).
NewTom GO may cause radio interferences or disturb
the operation of the nearby equipment. It may be
necessary to adopt countermeasures, such as re-
orienting or moving NewTom GO or shielding the
installation site.
Guidance and Manufacturer's declaration - Electromagnetic immunity
NewTom GO is designed to operate in the specified electromagnetic environment. The customer or the user of
NewTom GO must ensure its use in an electromagnetic environment with the following features:
Immunity test
IEC 60601-1-2
Test level
Level of
conformity
Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
IEC 60601-1-2
Test level
Floors must be made of wood, concrete or
ceramic tiles. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
IEC 61000-4-4
fast/burst electric
transients
2 kV for electrical lines
1 kV for input/output
lines > 3 m
IEC 60601-1-2
Test level
The power supply line quality should be that
of a typical commercial or hospital
environment.
Overvoltage
IEC 61000-4-5
1 kV differential
mode
2 kV standard mode
IEC 60601-1-2
Test level
The power supply line quality should be that
of a typical commercial or hospital
environment.
Voltage drops,
short interruptions
and voltage
change on the IEC
61000-4-11 input
electric line
0% Unfor 0.5 cycles
40% Unfor 5 cycles
70% Unfor 25 cycles
0 % Unfor 5 s
IEC 60601-1-2
Test level
The power supply line quality should be that
of a typical commercial or hospital
environment. If the NewTom GO user
requires a continuous operation also in
case of blackout, it is recommended to
power the NewTom GO with uninterruptible
power supply or batteries.
Magnetic field at
network frequency
(50/60 Hz)
IEC 61000-4-8
3 A/m
IEC 60601-1-2
Test level
The magnetic fields at network frequency
should feature levels typical of a standard
commercial or hospital environment.

18
OPERATOR'S MANUAL
EN
Guidance and Manufacturer's declaration - Electromagnetic immunity
NewTom GO is designed to operate in the specified electromagnetic environment. The customer or user of
NewTom GO must ensure that it is used in such environment.
Immunity test
IEC 60601-1-2
Test level
Level of
conformity
Electromagnetic Environment
The RF communication devices (portable
and mobile) must not be used at a distance
from NewTom GO and its components,
including cables, lower than the
recommended distance using the
corresponding equation applicable to the
transmitter frequency.
Recommended distance.
Radiated RF
EN 61000-4-3
3 V/m
From 80 MHz to 2.5
GHz
3 V/m
d =
1.2 x √P
80 MHz to 800MHz
d
=
2.3 x √P
800 MHz to 2.5GHz
Conducted RF
EN 61000-4-6
3 V
from 150 kHz to 80
MHz
3 V
d =
1.2 x √P
Where P is the maximum output power of
the transmitter in Watt (W) according to the
transmitter Manufacturer, and d is the
recommended distance in metres (m).
The field intensity of the fixed RF
transmitters, determined based on an
electromagnetic site, could be lower than
the conformity level in each frequency
interval.
Near the equipment with the following
symbol interferences can be caused:
Recommended distance between the RF portable and mobile communication devices and NewTom GO.
NewTom GO is intended for use in electromagnetic environment where RF irradiated disturbances are controlled.
The customer or the user of the NewTom GO can prevent electromagnetic interferences by ensuring a minimum
distance between RF mobile and portable (transmitter) communication devices and NewTom GO as shown below,
according to maximum power output of the communication devices.
Transmitter maximum nominal
output
(W)
Distance according the transmitter frequency
(m)
150KHz to 80MHz
d = 1.2 x √P
80KHz to 800MHz
d = 1.2 x √P
800KHz to 2.5MHz
d = 2.3 x √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters having a maximum nominal output power not listed above, the recommended distance d in metres
(m) can be determined using the corresponding equation applicable to the transmitter frequency where P is the
maximum output power of the transmitter in Watt (W) according to the transmitter Manufacturer.
Note:
(1) At 80MHz and 800MHz it is necessary to apply the distance defined for the highest frequency interval.
(2) These guidelines cannot be applicable to all situations. The electromagnetic propagation is influenced by the
absorption and reflection of structures, objects and people.

EN
OPERATOR'S MANUAL
19
1.8.7. PROTECTION AGAINST RADIATION
This system is an X-ray equipment. As such, it exposes patients and operators to risks from radiation. It
must be used in compliance with the safety standards set forth by the radiological protection regulation in
force in the country of use. Following are some of the provisions:
-Control the emission of X-rays exclusively from the control workstation, the exam room must be suitably
shielded (if required by the local regulations in force).
-Make sure that the doors of the X-ray room are closed before starting the exam.
-During the emission, only the patient should stay in the exam room. If it is necessary for the operator to
remain during the exam (e.g. when the patient is not self-sufficient and requires assistance), he/she
must wear protective equipment against scattered radiation and, in any case, no parts of his/her body
should ever be exposed to direct X-ray emission. Patients may not be assisted by pregnant women or
minors.
-During the system's initialisation procedure, nobody is allowed in the exam room and its doors must
remain closed. Authorised personnel must always be present outside the room until the projection is
completed.
-Pay attention not to release the emission button too soon. Please note that during an exposure cycle,
radiation can be emitted several times. Wait for the exposure cycle to be completed.
-Always observe the following points:
-Stay at least 2 metres away from the X-ray source during exposure. For installations in Canada, the
minimum distance is 3 metres.
-All persons that are not directly involved with the patient should stay outside the exam room, or
behind a lead or lead glass shielding during exposure.
-Make sure that the operator is able to communicate with the patient both verbally and visually.
-If required, use a dosimeter for personal monitoring.
-Full use must be made of all radiation protection devices, accessories, and procedures available to
protect the patient and operator from X-ray radiation, especially for child.
1.8.8. APPLIED PARTS
The parts that, during standard use, necessarily come into contact with the patient, so that the device may carry out
its functions correctly, are: chinrest, bite piece and hygienic protections, craniostat and handles.
The non-applied parts that may come into contact with patient are the external covers and the patient arm.

20
OPERATOR'S MANUAL
EN
1.8.9. STRAY RADIATIONS
Stray radiation measurement is highly influenced by ambient conditions, such as wall composition and positioning,
therefore, under certain circumstances, the detected values can be significantly different.
The measurement points used are at 0.5 m, 1.0 m and 2.0 m, respectively, from a central rotation axis.
The circular measurement points are determined based on the patient bite position on the machine.
1) Standard panoramic exam: stray radiation measured at the maximum usage percentage allowed by the X-ray
generator (corresponding to an average anode power of 42 W).
Distance between rotation axis and measurement point
(Circular measurement points)
Stray radiation *
0.5 m
80 µGy / h
1.0 m
26 µGy / h
2.0 m
6 µGy / h
3.0 m
3 µGy / h
2) Standard panoramic examination: stray radiation measured at the current average usage percentage or based on
4 examinations per hour.
Distance between rotation axis and measurement point
(Circular measurement points)
Stray radiation *
0.5 m
12 µGy / h
1.0 m
3 µGy / h
2.0 m
1 µGy / h
3.0 m
0.4 µGy / h
Only for 3D machines
3) 3D standard resolution 10x10: stray radiation measured at the maximum usage percentage allowed by the X-ray
generator (corresponding to an average anode power of 42 W).
Distance between rotation axis and measurement point
(Circular measurement points)
Leakage radiation *
0.5 m
890 µGy / h
1.0 m
225 µGy / h
2.0 m
60 µGy / h
3.0 m
30 µGy / h
4) 3D standard resolution 10x10: stray radiation measured at the current average usage percentage or based on 4
examinations per hour.
Distance between rotation axis and measurement point
(Circular measurement points)
Leakage radiation *
0.5 m
65 µGy / h
1.0 m
20 µGy / h
2.0 m
5 µGy / h
3.0 m
2 µGy / h
*It is the maximum value at 15 cm above the horizontal plane of intersection with the patient bite. Other values on
the vertical axis are lower than these values.
Table of contents
Other CEFLA Dental Equipment manuals
Popular Dental Equipment manuals by other brands

T&S
T&S B-2250 Installation and maintenance instructions

Renfert
Renfert Twister Translation of the original instructions for use

3M Unitek
3M Unitek AlastiK EZ3 manual

Rapid Shape
Rapid Shape RS wash operating instructions

Air Techniques
Air Techniques SensorX Installation and operating instructions

Sirona
Sirona C1+ Care and cleaning