
·Cleveland Controls
Di-".ision~of
UniControl Inc.
Model·
RFS-4150
/
AIR PRESSURE SENSING SWITCH WITH ADJUSTABLESET POINT
APPLICATION
•'
Model RFS-4150 is a general purpose prov-
ing switch designed for HVAC and Energy
Management-applicatipns. This switch can
be used to sense positive, negative,
or
dif-
ferential air pressure.
GENERAL DESCRIPTION &
OPERATION
The
plated
hotJsiog_contains_a__dlapbLagm~<L~
__
-------,
calibration spring, and a snap-acting switch.
The sample line connections located on
each side
of
the diaphragm accept
o/is",
W',
and
7'\s"
flexible tubing. Various electrical
connections are available.
MOUNTING {FIG. 1)
Select a mounting location which is free
from vibration. The Model RFS-4150 must
be mounted with the diaphragm
in
any verti-
cal plane
in
order to maintain the specified
operating set point. Avoid mounting with the
sample line connections
in
the "up" position.
The standard model
is
surface~mounted
via
th~
two round holes (0.14" dia.) ortwo
of
the
fourslots
(o/is"
wide) on the zinc-plated strap
bracket.The mounting holes and slots are
3-Va"
apart. Custom mounting configurations
are available.
Figure
1:
Mount with the diaphragm
in
any
vertical plane.
-
-~,-~
·~
\1)
AIR SAMPLING CONNECTION
{FIG. 2)
Tri-barbed sample line connectors suitable
to
accept flexible tubing are situated on either
side
of
the diaphragm as shown
in
Figure
2.
The tri-barbed connectors accept
·o/is",
Y..",
and
o/;s"
ID
flexible tubing. For sample
lines up to 10 feet
in
length,
11s"
ID
tubing
is
acceptable. For lines up to 20 feet, use
Y.."
ID tubing.
Locate the sampling probe a minimum
of
Figure 2
INLET
Positive Only
LowerNegative
HigherPositive
NLETL
Negative Only
HigherNegative
LowerPositive
--
-1;5-ductl:liameters-downstreamiromth.::.-.:H.·--+t=lctu.r:e
~---
---
------+
source. Install the sampling probe as close
to the center
of
the airstream as possible.
Refer to Figure 2 to identify the high pres-
sure inlet (H)
and
the low pressure inlet
(L).
Selectone ofthe five application options
listed below, and connect the sample lines
as recommended.
HIGH
PRESSURE
LOW
PRESSURE
0
N.C.
N.O.