Co2meter iSense Oxygen User manual

Copyright © 2014 CO2 Meter. All Rights Reserved.
User Manual
iSenseOxygen
SamplingDataLogger
CM-0185 (0-25% O2)
CM-0195 (0-95% O2)

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 2 of 15
Table of Contents
TABLEOFCONTENTS.....................................................................................2
WELCOME!....................................................................................................3
IMPORTANTSAFEGUARDS.............................................................................4
DEVICESPECIFICATIONS................................................................................5
PACKAGECONTENTS.....................................................................................6
OPTIONALACCESSORIES........................................................................................6
GASLAB®SOFTWARE...................................................................................6
MINIMUMSYSTEMREQUIREMENTS.........................................................................7
QUICKSTARTGUIDE......................................................................................9
STANDARDOPERATION.................................................................................9
POWERINGTHEUNIT............................................................................................9
LCDDISPLAY.......................................................................................................9
THEORYOFOPERATION.......................................................................................10
CALIBRATION.....................................................................................................10
CALIBRATINGUSINGMODBUSORPCCONNECTION................................................11
PERFORMINGANATMOSPHERICSPANCALIBRATION..................................12
PERFORMINGA0%O2ZEROPOINTCALIBRATION.......................................12
SUPPORT.....................................................................................................13
WARRANTY.................................................................................................13
RETURNS.....................................................................................................14
LIABILITY.....................................................................................................15
CONTACTUS...............................................................................................15

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 3 of 15
Welcome!
Thankyouforpurchasingourproduct.Thismeterallowsyouto
measureoxygenininaccessibleareas.TheiSenseOxygen
Samplingdata‐loggerisalonglife0‐25%and0‐95&oxygenlevel
transmitterthatprovidesmultipleanaloglinearoutputsofO2
concentrations.It’sdesignedtoberugged,maintenance‐free
andcanbecalibratedinnormalairofanyknownoxygen
concentration.
Ourgoalistomakehigh‐quality,low‐costgasdetectionand
indoorairqualityproductsavailabletoyou.
Pleasetakethetimetoreadthroughthismanualinordertoget
themostoutofyourmeter.Visitourwebsiteregularlytofind
moreinformationaboutthisproductandtodownloadthe
latestfreesoftware.Onceyouhaveinstalledthesoftware,you
canusethe“Checkforupdates”functionundertheHelpmenu
tosearchfornewupdates.Pleasealsopayspecialattentionto
theimportantsafeguardsshownonthepagesahead.

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 4 of 15
Important Safeguards
Toreducetheriskoffire,electricalshockand/orinjurytopersons,basic
safetyprecautionsshouldalwaysbefollowedwhenusingelectrical
appliances,includingthefollowing:
1. READALLINSTRUCTIONSBEFOREUSINGTHISDEVICE!
2. Useonlythesuppliedpowersupplytooperatetheunit.
3. Ensurethatwhensamplinginaclosedenvironmentthatthe
tubesaresecurelyfashionedtothedevice.
4. Donotoperatewithablockedoffsamplepath.
5. Donotoperatethedeviceifitismalfunctioning.
6. Donotoperatethedevicewiththecoverdetached.
SAVE THESE
INSTRUCTIONS!

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 5 of 15
Device Specifications
MeasuringRange:
25%O2 0‐95%vol.
95%O2 0‐25%vol.
Repeatability:
0‐25%CO2 ±0.02%,±1%ofmeasuredvalue
0‐95%O2 ±0.05%,±1%ofmeasuredvalue
Accuracy
25%/95%±2%fullscale
PowerSupply:
MaximumVoltage 24VDC
MinimumVoltage 20VDC
PowerConsumption 600mA@24VDC
SensorRatings:
LifeExpectancy >15years
MaintenanceInterval Nomaintenancerequired
Warm‐upTime <15min(instantmeasurements)
Outputs:
O2OutputValue4‐20mA,linearlyscaled
RS‐485 ModbusInterface
PUMP STP0.40LPM,MaxPre1atm
(M.V.150mbars)(M.P.500mbars)

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 6 of 15
Package Contents
Pleaseverifythatyourpackagecontainsthefollowingitemsbefore
usingthemeter:
MAPunits:
(1)Meter
(1)CertificateofCalibration
(1)UserManual
(1)Powersupply
(1)1‐M12,12‐Positions5MCable
(1)10‐footlength1/8”tubing
(2)MoistureTrap(CM‐0112)
(2)1/8”Barbtubingbulkheadfitting
(2)1/8”Barb10‐32fittings
(2)ParticleFilter(CM‐0118)
(2)HydrophobicFilter(CM‐0117)
Optional Accessories
Ifnotincluded,youcanpurchaseanExtremeMoistureFilterseparately.
Thesefiltersareidealforhigh‐humidityenvironmentstoallowairflow
whilekeepinghumidityout.SKU#CM‐0103

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 7 of 15
GasLab ® Software Capabilities
OurGasLab®softwarewillallowusersto:
•Manageanddownloadlogs
•ConfigureSensor
•Adjustloggingintervals
•Calibratethemeter
•AutomaticdataloggingwhenmeterispoweredON
•DataloggingsessionstatusdisplayedonLCDscreen
•Collectdatareal‐time
Please visit:
http://www.co2meter.com/pages/downloadsforyourfreedownload
ofourGasLab®software.PleasenoteGasLab®mustbeinstalledpriorto
connectingthemeter/sensortoyourcomputer.
IMPORTANT: INSTALL SOFTWARE BEFORE CONNECTING
YOUR DEVICE TO A COMPUTER
Minimum System Requirements
Toutilizeourfreesoftware,yourcomputermustmeetthefollowing
requirements:
WindowsXPSP3orhigher
Microsoft.NetFramework3.5SP1
Pentium4(ornewer)operatingat2.4Ghzorfaster
1GBofRandomAccessMemory(RAM)
Harddiskspacewithatleast20Megabytes(MB)free(200+MB
recommendedforlogsandapplicationfiles)

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 8 of 15
Softwareiscompatiblewith64‐bitoperatingsystemsandisfullytested
withWindows7.Softwareiscompatiblewith64‐bitoperatingsystems
andisfullytestedwithWindows

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 9 of 15
Quick Start Guide
Makesureyoureadthroughtheseinstructionsthoroughlybeforeusing
yourmeter.Thisguidewillhelpyoufamiliarizewithyourmeterinorder
tobeasproductiveasyouneedtobeintheleastamountoftime
possible.
Standard Operation
Theoxygenmeterisfactorycalibratedforeasyinstallation.Duringits
lifecyclecalibrationmaybecomeperiodicallynecessary.We
recommendperformingaspanO2calibrationeverymonthfor
guaranteedoperationandensuredaccuracy.Noothercalibrationis
necessaryduringthelifecycleofthesensor.
Powering the Unit
Theunitcanbepoweredbyhardwiring24VDCpowerthroughthe
MDCM‐4FP‐2MCable(included).Samplingunitscanalsobepoweredby
theincluded24VDCInternationalwallpowersupply.
LCD Display
TheLiquidCrystalDisplay(LCD)screenshowsthefollowingfeatures:
CO2inpercentageformat(##.##%)for95%

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 10 of 15
Theory of Operation
Thezirconiaoxygensensorutilizessolid‐stateelectrochemicalreactions
toproduceavoltageproportionaltothecurrentoxygen.Thisvoltageis
thendigitalizedandfilteredtoausablevalue.Theoxygensensorhas
alsobeenfactorycalibratedtomeetspecifications.
Calibration
Thecalibrationprocessvariesdependingonthetypeofunitand
whetherithasoptionaldataloggingfunctionalityornot.Allunitsare
factory‐calibratedwithmultiplereferencepointsofgas,andhavebeen
verifiedtobeaccuratewithintheirspecificfunctionalitybefore
shipment.However,iftheunitisseverelyjoltedorotherwise
mechanicallydisturbed,thesensorcandrift,requiringrecalibration.All
calibrationproceduresfollowasingle‐pointcalibrationroutinethat
effectivelyshiftsthezero‐pointoftheO2sensor.

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 11 of 15
Calibrating using MODBUS or PC Connection
Downloadthesoftwaredirectlyfromthislinkinourwebsite.
http://www.co2meter.com/pages/downloads
MoreprecisecalibrationscanbeperformedusingthePCconnection.
TheunitcanbeattachedtoaPCusinganoptionalUSBtoRS‐485cable,
availableonourwebsite,orviaamini‐USBcabletotheinternalUSB
connection.InthedesktopsoftwareapplicationO2canbecalibrated
individually.TheO2sensorcanbecalibratedto0%andthento20.9%.
Duringnormaloperation0%calibrationoftheoxygentransmitter
shouldnotbenecessary.
CalibrationcanalsobeperformedoverMODBUS.Seeadditional
documentationforMODBUSregisters.
Calibrationcanbeperformedusingeither0%CO2calibrationgas
(typicallynitrogen,Argon,etc.),orusingafreshsourceofoutdoorair,
assumedtobeapproximately20.9%O2.
Startbyattachingthecalibrationgastothemeterandconnectingthe
metertoapersonalcomputer.Openthecalibrationscreeninthe
GasLab®software.Clickthe“Calibrate”buttoninthecalibrationtabfor
thedesiredgas,locatedinthe“ConfigureSensor”screen.Aslongasthe
gasconcentrationisstable,theunitshouldinstantlyreflectthe
calibratedvalue.Thiscanbeconfirmedbywatchingthedisplay.Tosee
thecalibrationvalueinrealtime,clickinthe“CollectRealtime”button

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 12 of 15
tocapturethesevaluesbeforeopeningtheconfigurationscreen.
Real‐timecapture
Performing an Atmospheric Span Calibration
1. Allowsensortostabilizeforatleast1hour
2. Ensurethesensorisexposedtofreshairorcalibrategas20.9%O2
3. PressSpantoperformtheatmosphericcalibration
Performing a 0% O2 Zero Point Calibration
1. Allowsensortostabilizeforatleast1hour
2. ApplyN2calibrationgas
3. Wait5minutesforstablereadings
4. Zerotocalibrate

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 13 of 15
Zero or Fresh Air Calibration
ApplygasandselecttheappropriateconcentrationandClickthe
“Calibrate”button.Thesensorreadingshouldinstantlyreflectthe
calibration.
NOTE:DONOTADJUSTTHESPANVALUE.
Support
Thequickestwaytoobtaintechnicalsupportisviaemail.Pleasesendall
supportenquirestosupport@co2meter.com.Inyouremail,please
includeaclear,concisedefinitionoftheproblemandanyrelevant
troubleshootinginformationorstepstakensofar,sowecanduplicate
theproblemandquicklyrespondtoyourinquiry.
Warranty
Thisunitcomeswitha1YEAR(warrantyperiod)limitedmanufacturer’s
warranty,startingfromthedatetheunitwasshippedtothebuyer.
Duringthisperiodoftime,CO2Meter.comwarrantsourproductstobe
freefromdefectsinmaterialsandworkmanshipwhenusedfortheir
intendedpurposeandagreestofixorreplace(atourdiscretion)any
partorproductthatfailsundernormaluse.Totakeadvantageofthis
warranty,theproductmustbereturnedtoCO2Meter.comatyour
expense.If,afterexamination,wedeterminetheproductisdefective,
wewillrepairorreplaceitatnoadditionalcosttoyou.
Thiswarrantydoesnotcoveranyproductsthathavebeensubjectedto
misuse,neglect,accident,modificationsorrepairsbyyouorbyathird
party.NoemployeeorresellerofCO2Meter.com’sproductsmayalter
thiswarrantyverballyorinwriting.

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 14 of 15
Returns
Iftheproductfailsundernormaluseduringthewarrantyperiod,an
RMA(ReturnMaterialAuthorization)numbermustbeobtainedfrom
CO2Meter.com.Aftertheitemisreceived,CO2Meter.comwillrepairor
replacetheitematourdiscretion.
ToobtainanRMAnumber,pleasecallCO2Meter.comat(877)678‐
4259.WhenrequestinganRMAnumber,pleaseprovidethereasonfor
returnandoriginalordernumber.
Ifwedeterminethattheproductfailedduetoimproperuse(water
damage,dropping,tampering,electricaldamageetc.)orabuse,orifitis
beyondthewarrantyperiod,wewillinformyouofthecosttofixor
replaceyourdevice.
Ifyouarereturningyourdeviceduetoawarrantyclaim(withanRMA
number)andyoustillhavetheunitoriginalpackage,pleaseuseitto
shipyourunittous.PleasemakesuretoincludetheprovidedRMA
numberontheoutsideofthebox,preferablyontheshippinglabel.
Makesureyousecuretheunitinsidethepackageproperlytoprevent
anydamageduringtransitthatcouldvoidyourdevice’swarranty.
Finally,pleaseshipyourdevicetotheaddressshownunderthe
“ContactUs”sectionbelow.CO2Meter.comwillnot,underany
circumstances,beresponsibleforyourshipmentexpensesandno
refundwillbeissuedforshippingchargesnecessaryforyoutoshipthe
unittous.

Sample Draw/Data Logger– Manual CO2Meter.com
Rev. 27 October 2014 Pg. 15 of 15
Liability
Allliabilitiesunderthisagreementshallbelimitedtotheactualcostof
theproductpaidtoCO2Meter.com.InnoeventshallCO2Meter.combe
liableforanyincidentalorconsequentialdamages,lostprofits,lossof
time,lostsalesorlossordamagetodata,injurytopersonorpersonal
propertyoranyotherindirectdamagesastheresultofuseofour
products.
Contact Us
Weareheretohelp!
Ifthetroubleshootingguideabovedoesn’thelpyousolvingyour
problemorformoreinformation,pleasecontactususingthe
informationbelow.
(386)256‐4910(M‐F9:00am–6:00pmEST)
(866)422‐2356
www.co2meter.com
Address:
CO2Meter,Inc.
131BusinessCenterDrive
OrmondBeach,FL32174