Cooey B61T User manual

Version Number: 1.0
Fully Automatic Upper Arm
Cooey Blood Pressure Monitor
Model Number: B61T
--------------------------------------------------------------------------
USER’S MANUAL
A Special Thank You...
Thank you for choosing a blood pressure instrument. We're proud of the care and quality that goes into the
manufacture of each and every item that bears our name. Only the finest materials are used to assure you of
a timeless instrument designed for optimum performance. You'll quickly appreciate the results, for you now
own one of the finest blood pressure instruments that money can buy. With proper care and maintenance,
your automatic blood pressure monitor is sure to provide you with many years of dependable service. Please
read the following instructions and general information which will prove helpful in allowing you to enjoy your
measurement in hospitals and physician's offices throughout the world, where accuracy and dependability
are critical, professional diagnostic products are the instruments of choice.
Now you too can enjoy the benefits of engineering and quality at the comforts of your home. This feature
rich instrument was designed to simplify the measurement of blood pressure and pulse rate at home and
deliver consistent, dependable results. Your blood pressure monitor is a fully automatic digital blood pressure
measuring device for use on the upper arm. It enables very fast and reliable measurement of the systolic and
diastolic blood pressure as well as the pulse by way of the oscillometric method. This device offers clinically
proven accuracy and has been designed to be user friendly. Read this booklet thoroughly before attempting
to use your new Digital Blood Pressure Monitor.

Version Number: 1.0
Thank you for your patronage. It is indeed our pleasure to serve you.
Sincerely,
Cooey Technologies
Introduction and Intended Use
----------------------------------------------------------------------------------------------------------------
This manual is for B61T models of Cooey’s fully automatic digital blood pressure monitor.
This measuring device is for use by adults on the upper arm at home or at a typical physician’s office by a
doctor or a paramedic. It enables very fast and reliable measurement of systolic and diastolic blood pressure
as well as heart rate through the oscillometric method. This device offers clinically proven accuracy and has
been designed to be user friendly.
Before using, please read this instruction manual carefully and then keep it in a safe place. Please contact
your doctor for further questions on the subject of blood pressure and its measurement.
Warning:Not suitable for neonatal and infants.
Warning:Not suitable for pregnant patient.
This device can not be used together with hf surgical equipment.
Remember…
-----------------------------------------------------------------------------------------------------------------
• Only a healthcare professional is qualified to interpret measurements of your blood pressure or heart rate.
• This device is NOT intended to replace regular medical checkups.
• It is recommended that your physician review the procedure for using this device.
• Blood pressure readings obtained by this device should be verified before prescribing or making
adjustments to any medications used to control hypertension. Under no circumstances should YOU alter the
dosages of any drugs prescribed by your doctor.
• This monitor is intended for use by adults only. Consult with a physician before using this instrument on a
child.
•In cases of irregular heartbeat (Arrhythmia), measurements made with this instrument should
only be evaluated after consultation with your doctor.
• Familiarize yourself with the section titled "Important Information on Blood Pressure and its
Measurement". It contains important information on the dynamics of blood pressure readings
and will help you to obtain the best results.
NOTE!
• This device contains sensitive electronic components. Avoid strong electrical or electromagnetic fields in
the direct vicinity of the device (e.g. mobile telephones, microwave
ovens) during use. These can lead to erratic results.
• Do not attempt to service or repair this device yourself. Should a malfunction occur, refer to your local
distributor or the manufacturer.
Warning:
1. Too frequent measurements can cause injury to the PATIENT due to blood flow interference
2. Don’t place the cuff over wound part
3. Pressurization of the CUFF can temporarily cause loss of function of simultaneously used monitoring ME
EQUIPMENT on the same limb

Version Number: 1.0
Contraindication
---------------------------------------------------------------------------------------------------------------------
Use of this instrument on patients under dialysis therapy or on anticoagulant, antiplatelets, or steroids could
cause internal bleeding.
Warnings and Precautions
---------------------------------------------------------------------------------------------------------------------
•Warning: The device contains sensitive electronic components. Avoid strong electrical or electromagnetic
fields in the direct vicinity of the device (e.g. mobile telephones, microwave ovens). These can lead to
temporary impairment of the measuring accuracy.
•Warning: This system may fail to yield specified measurement accuracy if operated or stored in temperature
or humidity conditions outside the limits stated in the specifications section of this manual.
•Warning: The separate AC adapter which is intended to connect USB interface of Blood Pressure Monitor
has not been evaluated according to IEC 60601-1. The safety of the product should be re-appraised when it
gets power supply by a separate AC adapter.
•Warning: Turn off the power if the EQUIPMENT is not likely to be used for some time.
•Warning: The user must check that the equipment functions safely and see that it is in proper working
condition before being used.
•Warning: No modification of this equipment is allowed.
•Warning: The device is not suitable for use in the presence of flammable anesthetic mixtures with air or with
oxygen or nitrous oxide.
•Warning: This equipment shall not be serviced or maintained while in use with the patient
•Warning: The patient is the intended operator, the functions of monitoring blood pressure and pulse rate can
be safely used by the patient. The routine clean and changing batteries can be performed by the patient.
•Caution: To avoid any possibility of accidental strangulation, keep this unit away from children
and do not drape tubing around your neck.
•Caution: To avoid damaging the device, keep this unit away from children and pets.
•Caution: The standard material used for the bladder and tubing is latex-free.
•Attention: Self-measurement means control, not diagnosis or treatment. Unusual values must
always be discussed with your doctor. Under no circumstances should you alter the dosages of any drugs
prescribed by your doctor.
•Attention: The display of pulse reading is not suitable for checking the frequency of heart pacemakers!
•Attention: In cases of irregular heartbeat, measurements made with this instrument should only be
evaluated after consultation with your doctor.
Note: To obtain the greatest accuracy from your blood pressure instrument, it is recommended
that the instrument be used within the specified temperature and the relative humidity, please see the
Technical Specifications
Note: The cuff is treated as the applied part.
The user should contact the manufacturer for assistance, if needed, in setting up, using or maintaining the
device.
2. Important Information on Blood Pressure and its Measurement

Version Number: 1.0
---------------------------------------------------------------------------------------------------------------------------
2.1. How does high or low blood pressure arise?
Your level of blood pressure is determined in the circulatory center of the brain and adjusts to a
variety of situations through feedback from the nervous system. To adjust blood pressure, the
strength and speed of the heart (Pulse), as well as the width of circulatory blood vessels is altered. Blood
vessel width is controlled by fine muscles in the blood vessel walls.
Your level of arterial blood pressure changes periodically during heart activity: During the “blood ejection”
(Systole) the value is at the highest (systolic blood pressure value). At the end of the heart’s “rest period”
(Diastole) pressure is at the lowest (diastolic blood pressure value).
Blood pressure values must lie within certain normal ranges in order to prevent particular diseases.
2.2. Which values are normal?
Please refer to the diagram below(Picture-01)
Picture-01
Blood pressure is very high if your diastolic pressure is above 90 mmHg and/or your systolic
blood pressure is over 160 mmHg, while at rest. In this case, please consult your physician immediately.
Long-term values at this level endanger your health due to continual damage to the blood vessels in your
body. If your systolic blood pressure values are between 140 mmHg and 159mmHg and/or the diastolic
blood pressure values between 90 mmHg and 99mmHg, consult your physician. Regular self-checks are
necessary. If you have blood pressure values that are too low, (i.e., systolic values under 105mmHg and/or
diastolic values under 60 mmHg), consult your physician. Even with normal blood pressure values, a regular
self-check with your blood pressure monitor is recommended. You can detect possible changes in your
values early and react appropriately. If you are undergoing medical treatment to control your blood pressure,
keep a record of values along with time of day and date. Show these values to your physician. Never use the
results of your measurements to independently alter the drug doses prescribed by your physician.
Further information
• If your values are mostly normal under resting conditions but exceptionally high under conditions of
physical or psychological stress, it is possible that you are suffering from so-called “labile hypertension.”

Version Number: 1.0
Consult your doctor.
• Correctly measured diastolic blood pressure values above 120mmHg require immediate medical treatment.
2.3. What can be done if regular high or low values are obtained?
1) Consult your doctor.
2) Increased blood pressure values (various forms of hypertension) are associated with considerable health
risks over time. Arterial blood vessels in your body are endangered due to constriction caused by deposits in
the vessel walls (Arteriosclerosis). A deficient supply of blood to important organs (heart, brain, muscles) can
result from arteriosclerosis. Furthermore, the heart will become structurally damaged with increased blood
pressure values.
3) There are many different causes of high blood pressure. We differentiate between the common primary
(essential) hypertension, and secondary hypertension. The latter group can be ascribed to specific organ
malfunctions. Please consult your doctor for information about the possible origins of your own increased
blood pressure values.
4) There are measures which you can take to reduce and even prevent high blood pressure.
3. Components of your blood pressure monitor
----------------------------------------------------------------------------------------------------------------
a) Measuring unit
Features of Model B61T:
1- Voice broadcast function;
2- Bluetooth wireless conection with Smart phone;
3- Remoted by applications on IOS or Android smart phones;
4- Operated on device with one key;

Version Number: 1.0
5- With 99 groups memories;
6- Support the average measurement results;
7- Biult-in chargeable lithium ion battery;
8- Cuff and main unit are integrated as one part;
9- With charging indicator;
10- With Blood Pressure result Indicator
Note: Arm circumference should be measured with a measuring tape in the middle of the relaxed upper arm.
Do not force cuff connection into the open wound.
4. Using your Monitor for the First Time
----------------------------------------------------------------------------------------------------------------------------
4.1 Battery Power checking
Battery Power Checking
The battery is built-in chargeable Lithium ion battery
Press the “Power ON/OFF ”button, if the “Battery Indicator”become “Red”, it means the battery power is
low, and the unit need to be re-charged, the power supply should be AC 5/1A. Please recharge it and when
the “Battery Indicator” become “Green”, it means the battery is fully recharged. It can be used normally. If
battery is not fully re-charged, the indicator flashes in Red.
Assessment indicator
The indicator is set around the stop/start button. After measurement, the indicator will turn into different
colors according to the assessment. Blue means normal; Yellow means mild high blood pressure; Red
means high blood pressure.
4.2 System Settings
After the battery is fully recharged, some functions can be set before use this monitor:
a. Start/Stop
Press the Start/Stop button one time. The indicator around the button turns blue, the bluetooth is waiting for
connection. Long press the button again can start the messurement directlly, or connect device with smart
phone, it can be remote controlled by App which is installed beforehand in the smart phone. There will be a
voice “Bluetooth is connected”if the device connects with smart phone by Bluetooth successfully.
b. Setting the Volume
When the device is turned off, press and hold the Vol+ / Vol- button, you will hear the voice guide of Volume
up / Volume down to set your favorite voice volume.
c. Assessment of the memory record
When the device is turned off, press and hold the Vol+ and Vol- simultaneously, it will broadcast the average
value. Then press Vol+ / Vol- button once, it will seek for the next/previous assessment record.
When the device is turned on, press and hold the Vol+ and Vol- simultaneously, it will clear all the
assessment records.
d. Turn off the voice notice and background music.
Press the Vol+ or Vol- button during the very beginning when reading the notice, the voice notice and
background music will be turned off and start measurement immediately (It will save time if you don’t want to

Version Number: 1.0
hear the notice and background music.). After the measurement, the device will read the blood pressure and
pulse result only.
5. Measurement Procedure
--------------------------------------------------------------------------------------------------------------------
Note: You should always be seated and calm before and during measurement.
5.1. Before measurement:
• Avoid eating and smoking as well as all forms of exertion directly before measurement. These
factors influence the measurement result. Find time to relax by sitting in an armchair in a quiet
atmosphere for about ten minutes before taking a measurement.
• Remove any garment that fits closely to your upper arm.
• Always measure on the same arm (normally left).
• Always compare measurements taken at the same time of day, since blood pressure changes during the
course of the day, as much as 20-40 mmHg.
5.2. Common sources of error:
Note: Comparable blood pressure measurements always require the same conditions!
•Conditions should always be quiet.
•All efforts by the user to support the arm can increase blood pressure. Make sure you are in a
comfortable, relaxed position and do not flex any of the muscles in the measurement arm during the
measurement. Use a cushion for support if necessary.
• If the arm artery lies considerably lower or higher than the heart, an erroneously high or low blood pressure
will be measured! Each 25-30cm difference in height between your heart and the cuff results in a
measurement error of 10 mmHg!
• Cuffs that are too narrow or too short result in false measurement values. Selecting the correct cuff is
extremely important. Cuff size is dependent upon the circumference of the arm (measured in the center). The
permissible range is printed on the cuff.
Cuff works Under the pressure range 0-299mmHg
The wide range rigid cuff is : 9.1” – 14.2” (23-36 cm)
• A loose cuff or a sideways protruding air pocket causes false measurement values.
• With repeated measurements, blood accumulates in the arm, which can lead to false results. Consecutive
blood pressure measurements should be repeated after a 1 minute pause or after your arm has been held
up in order to allow the accumulated blood to flow away.
5.3. Fitting the Cuff
Please refer to picture-06
a) The cuff is pre-formed for easier use. Remove tight or bulky clothing from your upper arm.
b) Wrap the cuff around your upper left arm. The rubber tube should be on the inside of your arm extending
downward to your hand. Make certain the cuff lies approximately 0.39” to 0.79” (1 to 2 cm) above the elbow.
Important! The red mark on the edge of the cuff (Artery Mark) must lie over the artery which runs down the
inner side of the arm.
c) To secure the cuff, wrap it around your arm and press the hook and loop closure together.
d) There should be little free space between your arm and the cuff. You should be able to fit 2 fingers
between your arm and the cuff. Cuffs that don’t fit properly result in false measurement values. Measure your
arm circumference if you are not sure of proper fit.

Version Number: 1.0
e) Lay your arm on a table (palm upward) so the cuff is at the same height as your heart. Make sure the tube
is not kinked.
f) Remain seated quietly for at least two minutes before you begin the measurement.
Picture-06
5.4Measure Procedure
Refer to picture 07
1. Sit comfortably in a chair with your feet flat on the floor.
2. Stretch your arm forward on the desk and keep relaxing, make sure the palm of hand is upturned. Make
sure arm is in correct position, to avoid body movement. Sit still and do not talk or move during the
measurement.
After the cuff has been appropriately positioned on the arm and connected
to the blood pressure monitor, the measurement can begin:
a) Operate on the device
Press the Stop/Start button as the power on (blue indictor is turned on). The pump begins to inflate the cuff.
After automatically reaching an individual pressure, the pump stops and the pressure slowly falls, until the
measurement finish, the assessment result will be voice out, and the assessment indicator will turn to
different colors accordingly.
b) Operate via the App on smart phone.
NOTE:
Patient Position:
1) comfortably seated
2) legs uncrossed
3) feet flat on the floor
4) back and arm supported
5) middle of the CUFF at the level of the right atrium of the heart
Recommended Use Methods
1. Recommendation that the PATIENT relax as much as possible and not talk during the measurement
Artery

Version Number: 1.0
PROCEDURE
2. Recommendation that 5 min should elapse before the first reading is taken
3. Any reading can be affected by the measurement site, the position of the PATIENT, exercise, or the
PATIENT’S physiologic condition
4. Performance of the AUTOMATED SPHYGMOMANOMETER can be affected by extremes of
temperature, humidity and altitude
5. To stop the inflation or measurement, push the START/STOP button. The monitor will stop inflating, start
deflating, and will turn off.
6. After the monitor has detected your blood pressure and pulse rate, the cuff automatically deflates. Your
blood pressure will be voiced out or shown on the app on smart phone if operated by App.
7. The monitor will automatically turn off after two minutes no operation.
5.6 Error Indicates
The following symbol will appear on the display when measuring abnormal
Trouble removal
Problem
Check
Cause and solution
Power low
The battery indicator turns in red
Charge the device under DC 5V/1A
No power after charging
After charging, the battery indicator
never turns into green
Send the device to after sales center
No Inflation
Whether the cuff is broken or leak
Send to after sales center

Version Number: 1.0
Err and stop working
Whether move the arm when the cuff
is inflating
Keep the no moving and peaceful
Check if chatting when measuring
Keep quiet when measuring
Cuff leak
Whether the cuff wrap too loose
Wrap the cuff tightly
Whether the cuff borken
Send the device back to after sales
center
Please contact the distributor if you can not solve the problems. Do not dsassemble the unit by yourself
SYMBOL DESCRIPTIONS
The following symbols may appear in this manual, on the Digital Blood Pressure Monitor B61T, or on it's
accessories. Some of the symbols represent standards and compliances associated with the Digital Blood
Pressure Monitor B61T and its use.
Authorized Representative in the European Community
CE Mark: conforms to essential requirements of the Medical Device Directive
93/42/EEC.
Date of manufacture.
Manufacturer
SN
Specifies serial number
Type BF applied part
Direct current
DISPOSAL: Do not dispose this product as unsorted municipal waste.
Collection of such waste separately for special treatment is necessary.
Follow instructions for use
IP21
The degree of avoid ingress of water or particulate matter into ME equipment
5.7. Memory
At the end of a measurement, this monitor automatically stores each result with date and time. Each unit
stores 99 sets measurements
Viewing the stored values
When the device is turn off, press and hold the Vol+ / Vol- button, it will broadcast the average value. Then

Version Number: 1.0
press Vol+ / Vol- button once, it will seek for the assessment record of next/previous assessment records.
When the device is turn on, press and hold the Vol+ / Vol- button simultaneously, it will clear all the
assessment records.
5.8. Discontinuing a Measurement
If it is necessary to interrupt a blood pressure measurement for any reason (e.g the patient feels unwell), the
Start/Stop button can be pressed at any time. The device then immediately lowers the cuff pressure
automatically.
5.10. Using the AC Adapter
Press the ”Power on/off ”button, if the “Battery Indicator” keep in “Red”, it means the battery power is low,
and the unit need to be re-charged, the power supply should be AC 5/1A.
You may also operate this monitor using the AC adapter (output 5V DC/1A with Micro USB plug).
Use only the approved brand AC adapter to avoid damaging the unit.
a) Ensure that the AC adapter and cable are not damaged.
b) Plug the adapter cable into the AC adapter port on the right side of the blood pressure monitor.
c) Plug the adapter into your electrical outlet. When the AC adapter is connected, no battery current is
consumed.
(e.g., by accidental removal of the AC adapter from the outlet, the monitor must be reset
by removing the plug from the socket and reinserting the AC adapter connection.
6. Care and Maintenance
Wash hands after each time measurement.
If one device is used by different patients, wash hands before and after each use.
a) Do not expose the device to either extreme temperatures, humidity, dust or direct sunlight.
b) The cuff contains a sensitive air-tight bubble. Handle this cuff carefully and avoid all types of stress
through twisting or buckling.
c) Clean the device with a soft, dry cloth. Do not use gas, thinners or similar solvents. Spots on the cuff can
be removed carefully with a damp cloth and soapsuds. The cuff with bladder must not be washed in a
dishwasher, clothes washer, or submerged in water.
d) Handle the tube carefully. Do not pull on it. Do not allow the tubing to kink and keep it away from sharp
edges.
e) Do not drop the monitor or treat it roughly in any way. Avoid strong vibrations.
f) Never open the monitor! This invalidates the manufacturer’s warranty.
g) Batteries and electronic instruments must be disposed of in accordance with the locally applicable
regulations, not with domestic waste.
6.1. Accuracy test
Sensitive measuring devices must be checked for accuracy from time to time. We recommend a periodical
inspection of your unit by an authorized dealer every 1 year. Please turn to local distributor or the
manufacturer.
7. Warranty
-----------------------------------------------------------------------------------------------------------------------
Your blood pressure monitor is guaranteed for 2 years against manufacturers’ defects for the original
purchaser only, from date of purchase. The warranty does not apply to damage caused by improper
handling, accidents, professional use, not following the operating instructions or alterations made to the
instrument by third parties.
Warranty only applies to the instrument.

Version Number: 1.0
Note: According to international standards, your monitor should be checked for accuracy every
year.
8. Certifications
---------------------------------------------------------------------------------------------------------------
Device standard:
This device is manufactured to meet the European blood pressure monitors:
EN1060-1 / 1995 • EN1060-3 / 1997 • EN1060-4 / 2004
Electromagnetic compatibility:
Device fulfills the stipulations of the International standard
IEC60601-1-2
9. Technical Specifications
-------------------------------------------------------------------------------------------------------------------------
Model:B61T
Wight:252g (unit net weight)
Size: 121 (W) x 67 (L) x 30(H) mm
Accessories: 1×Main Device, 1×Users manual, 1×Warranty card
Operating Conditions: Temperature: 5℃ to 40℃; Humidity: 15% to 93% RH; Pressure altitude: 70KPa~
106Kpa
Storage And Shipping Conditions:Temperature: -25℃ to 70℃; Humidity: 10% to 93% RH; Pressure altitude:
70KPa~ 106Kpa
Measuring method: Oscillometric
Pressure sensor: Capacitive
Measuring range: DIA: 40-130mmHg; SYS: 60-230mmHg Pulse: 40 to 199 per minute
Cuff pressure display range: 0–299 mmHg
Memory: Automatically stores the last 99 measurements
Measuring resolution: 1 mmHg
Accuracy: Pressure within ± 3 mmHg / pulse ± 5 % of the reading
Power source: Built-in Lithium Ion Polymer Battery
Accessories: Wide range rigid cuff 8.7” – 14.2” (22 - 36 cm)
Automatically power off : 120 seconds
Users: Adult
Expected service life of the device and accessories: 7 years
Technical alterations reserved!
ENVIRONMENTAL REQUIREMENTS
OPERATING CONDITIONS
Temperature: 5℃ to 40℃
Humidity: 15% to 93% RH
Pressure altitude: 70KPa~ 106Kpa
STORAGE AND SHIPPING CONDITIONS
Temperature: -25℃ to 70℃

Version Number: 1.0
Humidity: 10% to 93% RH
Pressure altitude: 70KPa~ 106Kpa
CLASSIFICATION
1. Internally powered equipment; (supplied by Built-in Lithium Ion Polymer Battery)
2. Type BF applied part;
3. IP21;
4. No Sterilize requirement;
5. Not category AP / APG equipment
6. Mode of operation: continuous
EMC Declaration
Guidance and manufacturer’s declaration – electromagnetic immunity
The “Digital Blood Pressure Monitor B61T” is intended for use in the electromagnetic environment specified below. The customer
or the user of the “Digital Blood Pressure Monitor B61T” should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/output
lines
2 kV for power
supply lines
1 kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV differential
mode
2 kV common mode
1 kV differential
mode
2 kV common mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
“Digital Blood Pressure Monitor B61T” requires
continued operation during power
mains interruptions, it is recommended
that the “Digital Blood Pressure Monitor B61T” be
powered from an uninterruptible power
supply or a battery.

Version Number: 1.0
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.

Version Number: 1.0
4. EMC Declaration (Continued)
Guidance and manufacturer’s declaration – electromagnetic immunity
The “Digital Blood Pressure Monitor B61T” is intended for use in the electromagnetic environment specified below. The customer or
the user of the “Digital Blood Pressure Monitor B61T” should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment –guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the
“Digital Blood Pressure Monitor B61T”, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
d=1.2√P
d=1.2√P 80MHz to 800MHz
d=2.3 √P 800MHz to 2.5 GHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
3 V/m
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,ashould be
less than the compliance level in each frequency
range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
b
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the “Digital Blood Pressure Monitor B61T” is
used exceeds the applicable RF compliance level above, the Medical Digital Blood Pressure Monitor B61T should be observed to
verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating
the “Digital Blood Pressure Monitor B61T”.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.

Version Number: 1.0
4. EMC Declaration (Continued)
Guidance and manufacturer’s declaration – electromagnetic emissions
The “Digital Blood Pressure Monitor B61T” is intended for use in the electromagnetic environment specified below. The
customer or the user of the “Digital Blood Pressure Monitor B61T” should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment –guidance
RF emissions
CISPR 11
Group 1
The “Digital Blood Pressure Monitor B61T” uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The “Digital Blood Pressure Monitor B61T” is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Recommended separation distances between
portable and mobile RF communications equipment and the Medical Digital Blood
Pressure Monitor B61T
The “Digital Blood Pressure Monitor B61T” is intended for use in an electromagnetic environment in which radiated RF
disturbances
are controlled. The customer or the user of the Medical Digital Blood Pressure Monitor B61T can help prevent
electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
“Digital Blood Pressure Monitor B61T” as recommended below, according to the maximum output power of the
communications
equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.

Version Number: 1.0
10. Warranty Card
-----------------------------------------------------------------------------------------------------------------------

Version Number: 1.0
CONTACT INFORMATION
Digital Blood Pressure Monitor B61T is manufactured by:
Company name :Shenzhen Jamr Medical Technology Co., Ltd
Address:2nd Floor, A- building, No.2 Guiyuan Road,Guihua Community,Guanlan Town, Longhua New
District, Shenzhen China.
Authorized Indian Representative:
Cooey Technologies Pvt, Ltd.
1101,12th Main, 5th Cross,
Sundar Ram Shetty Nagar,
Bilekahalli, Bangalore – 560076.
Email: needsupport@cooey.co.in