Cooper Atkins EnviroTrak Sensorex TX100 Installation guide

pH Monitoring Installation and Operating Guide
Cooper-Atkins Corporation |
Accuracy to the Highest Degree
www.cooper-atkins.com

2

3
TABLE OF CONTENTS
For additional mounting instructions, consult the enclosed Installation and User Manual
provided in the pH Kit from Cooper-Atkins. Please feel free to contact I-Care at 888-
533-6900 x2 for assistance.
Package'Contents'.................................................................................................................................'3'
Hardware'Mounting'.............................................................................................................................'4!
A.!Mounting!..............................................................................................................................................!4!
pH'Electrode'-'Calibration'.....................................................................................................................'5!
A.!Calibrating!pH!Electrode!(7.00pH!and!4.00pH)!....................................................................................!5!
pH'Electrode'–'Clean'Electrode'Notification'.........................................................................................'6!
A.!Setting!Cleaning!Reminder!...................................................................................................................!6!
Transmitter'Setup'................................................................................................................................'7'
Software'Configuration'–'pH'Electrode'.................................................................................................'8!

4
Package Contents
Contents of the pH Kit.
A. pH Mounting Plates
B. Sensorex Installation and
Operating Manual
C. Sensorex pH Monitor
D. Analog Transmitter
a. Transmitter Box
b. Analog Transmitter wired to
Sensorex pH Monitor
E. pH Electrode and Cable
F. pH Power Plug
Note: 4.01pH and 7.0pH calibration solutions NOT INCLUDED but required for
calibrating pH electrode. Needs to be purchased seperately. Not sold by Cooper-
Atkins.
A
F
B
Da
Db
C
E

5
Hardware Mounting
A. MOUNTING:
1. First determine mounting location for the Sensorex pH Monitor probe. A
mounted example is pictured below.
2. Locate a suitable location to mount the Sensorex pH Monitor box near an AC
outlet.
3. Four different mounting brackets are supplied allowing for a wide range of
mounting options. Please see mounting directions in the Sensorex Installation
and Operating Manual.
4. Typically, the Sensorex pH Monitor box should be mounted at eye level (5ft to
6ft) for easy reading.
SAMPLE ELECTRODE MOUNTING - MONITORING WATER LINE
FOR ADDITIONAL MOUNTING INSTRUCTIONS, REFER TO THE SENSOREX
MANUAL
Note: Installation of probe is the responsibility of the facility. Please consult local
zoning codes for correct installation.
Transmitter
Power Supply
pH Monitor
pH Electrode
Water Line

6
pH Electrode - Calibration
Periodically, the pH Electrode must be calibrated. This can be done during the
electrode cleaning cycle. Follow the instructions below to calibration the pH probe.
REFER TO SENSOREX MANUAL PAGE 21 IF NEEDED
A. CALIBRATING pH ELECTRODE (7.00pH and 4.01pH):
1) Press and hold <E> to enter
setup
2) Press <E> to start
calibration process
3) The main display will show
the actual pH reading. The
small display will show the
expected buffer (7.00 or
6.86) for the first point.
Use the <UP> and
<DOWN> arrows to select
7.00. Place the electrode
into the 7.00 buffer
solution.
Press and hold <E> to start
1st Calibration
4) When the reading is stable,
Press <E> to save first point
5) Now, the small display will
show the 2nd expected buffer
(4.01, 9.18, or 10.00).
Use the <UP> and
<DOWN> arrows to select
4.01. Place the electrode
into the 4.01 buffer solution.
Press and hold <E> to start
2nd Calibration
6) When the reading is stable,
Press <E> to save second
point
7) Press <E> to save second
point and complete
calibration. “CON” will flash
confirming calibration.

7
pH Electrode – Clean Electrode Notification
A reminder to clean, and if need be, calibrate your pH electrode can be set up for 1 –
250 days. Default setting is 90 days. Follow the instructions below to change the
number of days between cleanings.
REFER TO SENSOREX MANUAL PAGE 25 IF NEEDED
A. SETTING CLEANING REMINDER:
1) Press and hold <E> to enter
setup
2) Press <UP> arrow 2 times.
“C.P.” will display on the
main display and “Clean”
and electrode icon will flash
3) Press <E> and “OFF” will
flash
4) Use the <UP> and
<DOWN> arrows to turn ON
“Clean Probe”
5) Press <E> to accept. Day
number will flash. The
default value is 90.
6) Use the <UP> and <DOWN>
arrows to change number of
days.
7) Press <E> to save number
of days and return to normal
operation.

8
Transmitter Setup
Once the pH Electrode has been mounted and configured, the reset button on the
Analog Transmitter must be pressed to initiate sending the newly configured pH values.
900Mhz Digital Analog Transmitter
1. Remove the Transmitter cover.
2. Verify that the Transmitter is wired (D) for 4-20mA input.
3. Verify that the Transmitter Jumper (C) is configured for 4-20mA input.
4. Insert the battery(s) into the Battery Slot (A).
5. Press the Reset Button (B) to initialize the Transmitter.
6. Install the Transmitter cover.
D
C

9
Software Configuration – pH Sensor
In order for the transmitters that have been installed to work they must be registered in
the software. There is a Transmitter ID Label on the exterior of the transmitter box,
another label on the transmitter’s lid and an extra label to be placed at the monitoring
point.
1. Log into EnviroTrak.
2. To register a pH Electrode, click the Configuration > EnviroTrak > Device
Registration > Sensor Registration.
3. The sensor registration information is located inside the lid or on the supplied
stickers.
4. Enter the Sensor ID (A), Security ID (B), select DIGITAL RF/Analog Input
(C)** as the transmitter type, and then select Sensorex TX100-pH Monitor (D)**
as the #1 input.
**Note: (C) Make sure you select the correct transmitter type: DIGITAL
RF/Analog Input and Sensorex TX100-Ph Monitor (D).
B
A
C

10
**Note: (D) You may need to create a new formula for transmitter type
Sensorex TX100-Ph Monitor if it does not appear in the drop-down. Please
contact I-Care for assistance.
5. Select the sensor group you wish to assign the transmitter to from the Assign To
Group (Optional**) (E) drop-down list.
**Note: If no group is selected, the sensor will only display for users using the
Default: All Sensors view.
6. Click Add (F). You will be asked if you wish to name the transmitter. Follow the
directions and rename the transmitter, or close the Set Sensors window.
You should see the following confirmation at the bottom of the Transmitter
Registration Page.
A
B
C
D
E
F

11
NOTES:

Cooper-Atkins Corporation 33 Reeds Gap Rd., Middlefield, CT 06455
www.cooper-atkins.com
27-241
V1017
Table of contents
Popular Transmitter manuals by other brands

Transmitter Solutions
Transmitter Solutions Stinger 2 390GED21V2 manual

Heystar
Heystar HS-ZSD14-RF01 Instruction book

Continental Refrigerator
Continental Refrigerator TXN5 user manual

WIKA
WIKA TIF11 operating instructions

HK Instruments
HK Instruments DPT-2W- 100-R2 user guide

Spirax Sarco
Spirax Sarco LD357BL Series Installation and maintenance instructions