CS Instruments DS 500 User manual

DS 500 Side 1 of 66
EN - English
Installation and operating instructions
Intelligent paperless recorder
DS 500

Foreword
DS 500 Side 2 of 66
I. Foreword
Dear customer,
thank you very much for deciding in favour of the DS 500. Please read this
installation and operation manual carefully before mounting and initiating the device
and follow our advice. A riskless operation and a correct functioning of the DS 500
are only guaranteed in case of careful observation of the described instructions and
notes.
Sales Office South/Geschäftsstelle Süd
Zindelsteiner Str. 15
D-78052 VS-Tannheim
Tel.: +49 (0) 7705 978 99 0
Fax: +49 (0) 7705 978 99 20
Mail: [email protected]
Web: http://www.cs-instruments.com
Sales Office North/Geschäftsstelle Nord
Am Oxer 28c
D-24955 Harrislee
Tel.: +49 (0) 461 700 20 25
Fax: +49 (0) 461 700 20 26
Mail: [email protected]
Web: http://www.cs-instruments.com

Table of contents
DS 500 Side 3 of 66
II. Table of contents
I. Foreword ....................................................................................................................... 2
II. Table of contents ....................................................................................................... 3
1Safety instructions ........................................................................................................ 5
2Application area ............................................................................................................ 6
3Technical data DS 500 .................................................................................................. 7
4Input signal .................................................................................................................... 8
5Cable cross-section ...................................................................................................... 8
6Package dimensions .................................................................................................... 9
7Assembly instructions .................................................................................................10
8Panel and wall mounting .............................................................................................11
9Connection diagrams DS 500 ......................................................................................12
9.1 DS 500 with 4 channels ........................................................................................................ 12
9.2 DS 500 with 8 channels ........................................................................................................ 12
9.3 DS 500 with 12 channels ...................................................................................................... 13
9.4 Power supply DS 500 standard version 100 - 240 VAC ..................................................... 13
9.5 X2.1 and X2.2 by standard version 100 - 240 VDC, wired ex works ................................. 13
9.6 Power supply by special version 24 VDC ........................................................................... 14
9.7 Connection 4 x alarm relay, max. 230 VAC, 6 A ................................................................. 14
9.8 Connection bus systems X4.1 and S4.1 ............................................................................. 14
9.9 Connector pin assignment for all sensors XA.1 – XA.4, XB.1 – XB.4, XC.1 – XC.4 ....... 14
10 Connection diagrams of the different sensor types ...............................................15
10.1 Connection CS dewpoint sensors, serial FA 415/FA 300 .............................................. 15
10.2 Connection for dew point- and consumption sensors, serial FA/VA 400 ................... 15
10.3 Connection pulse sensors ............................................................................................... 16
10.4 Analogue two-, three-, and four-wire current signal ...................................................... 17
10.5 Three- and four-wire power supply 0 - 1/10/30 VDC ...................................................... 18
10.6 Two-, three-, and four-wire connector pin assignments for PT100/PT1000/KTY81 .. 19
10.7 Connection with RS485 .................................................................................................... 19
11 Connect the DS 500 with a PC .................................................................................20

Table of contents
DS 500 Side 4 of 66
12 Operation DS 500 ......................................................................................................21
12.1 Main menu (Home) ............................................................................................................ 21
12.1.1 Initialization ...................................................................................................................... 21
12.1.2 Main menu after initialization ........................................................................................... 22
12.2 Settings .............................................................................................................................. 23
12.2.1 Password settings ............................................................................................................ 23
12.2.2 Sensor settings ................................................................................................................ 24
12.2.2.1 Choice of the sensor type (For example type CS-Digital sensor) ............................ 24
12.2.2.2 Name the measurement data and define the decimal places .................................. 27
12.2.2.3 Recording measurement data .................................................................................. 27
12.2.2.4 Alarm settings ........................................................................................................... 28
12.2.2.5 More-Settings (scale analogue output) .................................................................... 29
12.2.2.6 Dewpoint sensor with type CS-Digital ...................................................................... 30
12.2.2.7 Label and set the description fields .......................................................................... 31
12.2.2.8 Configuration of analogue sensors ........................................................................... 34
12.2.2.8.1 Type 0 - 1/10/30 Volt and 0/4 - 20 mA ............................................................................. 34
12.2.2.8.2 Type PT100x and KTY81 .................................................................................................. 35
12.2.2.8.3 Type Pulse (Pulse ration) .................................................................................................. 36
12.2.2.8.4 Type RS485 ...................................................................................................................... 38
12.2.2.8.5 Type No Sensor ................................................................................................................ 38
12.2.3 Logger settings (data logger) ........................................................................................... 39
12.2.4 Device settings................................................................................................................. 43
12.2.4.1 Set language ............................................................................................................ 43
12.2.4.2 Date & Time .............................................................................................................. 43
12.2.4.3 Network settings ....................................................................................................... 44
12.2.4.4 ModBus settings ....................................................................................................... 45
12.2.4.5 SD-Card .................................................................................................................... 45
12.2.4.6 Update System ......................................................................................................... 45
12.2.4.7 Factory Reset ........................................................................................................... 48
12.2.5 Report settings (optional) ................................................................................................. 49
12.3 Chart ................................................................................................................................... 51
12.4 Chart/Real time values ...................................................................................................... 56
12.5 Real time values ................................................................................................................ 58
12.6 Alarm overview .................................................................................................................. 58
12.7 Further setting options ..................................................................................................... 59
12.7.1 Set backlight .................................................................................................................... 59
12.7.2 Calibrate touch-screen ..................................................................................................... 60
12.7.3 Cleaning ........................................................................................................................... 60
12.7.4 System Status .................................................................................................................. 60
12.7.5 About DS 500................................................................................................................... 61
12.8 Consumption report with cost settings, export data and webserver .......................... 61
12.8.1 Consumption report (optional) ......................................................................................... 61
12.8.2 Cost settings (optional) .................................................................................................... 63
12.8.3 Webserver (optional) ....................................................................................................... 64
12.9 Export data ......................................................................................................................... 65

Safety instructions
DS 500 Side 5 of 66
1 Safety instructions
Please check whether this manual corresponds with the device type.
Please attend to all notes indicated in this instruction manual. It contains essential information
which has to be followed during installation, operation and maintenance. Therefore this
instruction manual has to be read categorically by the technician as well as by the responsible
user/qualified personnel before installation, initiation and maintenance.
This instruction manual has to be available at any time at the operation site of the DS 500.
Regional and national regulations respectively, have to be observed in addition to this
instruction manual if necessary.
In case of any obscurities or questions with regard to this manual or the instrument please
contact CS Instruments GmbH.
Warning!
Supply voltage!
Contact with supply voltage carrying non-insulated parts may cause an electric shock
with injury and death.
Measures:
• Note all applicable regulations for electrical installations (e. g. VDE 0100)!
• Carry out maintenance only in strainless state!
• All electric works are only allowed to be carried out by authorized qualified personnel.
Warning!
Inadmissible operating parameters!
Undercutting and exceeding respectively of limit values may cause danger to persons
and material and may lead to functional and operational disturbances.
Measures:
• Make sure that the DS 500 is only operated within the admissible limit values indicated on the
type label.
• Strict observance of the performance data of the DS 500 in connection with the application.
• Do not exceed the admissible storage and transportation temperature.
Further safety instructions:
• Attention should also be paid to the applicable national regulations and safety instructions during
installation and operation.
• The DS 500 is not allowed to be used in explosive areas.
Additional remarks:
• Do not overheat the instrument!
Attention!
Malfunctions at the DS 500!
Faulty installation and insufficient maintenance may lead to malfunctions of the DS 500
which may affect the measuring results and which may lead to misinterpretations.
r

Application area
DS 500 Side 6 of 66
2 Application area
Our long-term hands-on experience in measurement and control technology was implemented in the
new DS 500.
From recording of the measured data, automatic sensor identification, indication on a big colour
screen, alerting, storage up to remote read-out via web server, all that is possible with DS 500. By
means of the CS-Soft, software alarms can be sent via SMS or e-mail.
On the big 7' colour screen with touch panel all information is available at a glance.
The operation is very easy. All measured values, measured curves and threshold exceedings are
indicated. The progression of the curve, since the beginning of the measurement, can be viewed by
an easy slide of the finger.
Daily, weekly and monthly reports, with costs in € and counter reading in m
3
, for each consumption
sensor are completing the sophisticated system concept.
The huge difference to ordinary paperless chart recorders reveals in the easy initiation as well as in
the evaluation of the measured data. All sensors are identified directly and powered by DS 500.
Everything is matched and tuned.
Versatile:
Up to 12 sensors, incl. all CS sensors
(consumption, dewpoint, pressure, current,
KTY, PT 100, PT 1000) are identified auto-
matically by DS 500. Optional analogue
sensors (0/4 - 20 mA, 0 - 1/10/30 V, pulse)
can be configured easily and quickly.
Digital sensors can be connected via
RS 485, Modbus RTU and SDI.
Flexible:
Network-compatible and
worldwide remote data
transmission via
Ethernet, integrated
web server.
Alarm relay / fault
indication:
Up to 32 threshold
values can be
configured freely and
allocated to 4 different
alarm relay. Collective
alarms are possible.

Technical data DS 500
DS 500 Side 7 of 66
3 Technical data DS 500
Dimensions of housing
280 x 170 x 90 mm, IP 65
Connections
18 x PG 12 for sensors and supply, alarm relay
1 x RJ 45 Ethernet connection
Version panel mounting
Cutout panel 250 x 156 mm
Weight
7,3 kg
Material
Die cast metal, front screen polyester
Sensor inputs
4/8/12 sensor inputs for analogue and digital sensors
freely allocatable.
Digital CS sensors for dew point and consumption with SDI interface
FA/VA 400 Series.
Digital third-party sensors RS 485/Modbus RTU, other bus systems
realizable on request.
Analogue CS sensors for pressure, temperature, clamp-on
ammeters preconfigured.
Analogue third-party sensors 0/4 – 20 mA, 0 - 1/10/30 V, pulse,
Pt100/Pt1000, KTY.
Power supply for sensors
Output voltage: 24 VDC ± 10% galavanically isolated
Output current: 130 mA by continous opertion, peak 180mA
Maximum output current over all channels with
- one power supply: 400mA
- two power supplies (optional): 1A
Interfaces
USB stick, USB cable, Ethernet/RS 485 Modbus RTU/TCP,
SDI other bus systems on request, web server optionally
Outputs
Outputs 4 relay (max. switching voltage: 400 VAC / 300 VDC,
Switching current: min. 10mA, max. 6A) alarm management,
relay freely programmable, collective alarm.
Analogue output, pulse in case of sensors with own signal
output looped, like e. g. VA/FA Series.
Memory card
Memory size 2 GB SD memory card standard,
optionally up to 4 GB
Power supply
100 - 240 VAC/50 - 60 Hz, special version 24 VDC
Colour screen
7" touch panel TFT transmissive, graphics, curves, statistics
Accuracy
See sensor specifications
Operating temperature
0 - 50 °C
Storage temperature
-20 - 70 °C
Optionally
Web server
Optionally
Quick measurement with 10 ms sampling rate for analogue sensors,
Max/Min indication per second.
Optionally
Option "consumption report” statistics, daily/weekly/monthly report.

Input signal/Cable cross-section
DS 500 Side 8 of 66
4 Input signal
5 Cable cross-section
5.1 Power supply 100 - 240 VAC, 50 - 60 Hz, special version 24 VDC:
AWG12 – AWG24, cable cross-sections: 0,2 - 2,5 mm2
5.2 Sensor circuit points/Output signal:
AWG16 – AWG28, cable cross-sections: 0,14 - 1,5 mm2
Input signal
Current signal
(0 – 20 mA / 4 – 20 mA)
internal or external
power supply
Measuring range
0 – 20 mA / 4 – 20 mA
Resolution
0,0001 mA
Accuracy
0,003 mA 0,05 %
Input resistance
50
Voltage signal
(0 - 1V)
Measuring range
0 - 1 V
Resolution
0,05 mV
Accuracy
0,2 mV 0,05 %
Input resistance
100 k
Voltage signal
(0 - 10 V / 30 V)
Measuring range
0 - 10 V/30 V
Resolution
0,5 mV
Accuracy
2 mV 0,05 %
Input resistance
1 M
RTD
Pt100
Measuring range
-200 - 850 °C
Resolution
0,1 °C
Accuracy
0,2 °C at -100 - 400 °C
0,3 °C (further range)
RTD
Pt1000
Measuring range
-200 - 850 °C
Resolution
0,1 °C
Accuracy
0,2 °C at -100 - 400 °C
0,3 °C ( further range )
Pulse
Measuring range
minimal pulse length 100
μ
s
frequency 0 - 1 kHz
max. 30 VDC

Package dimensions
DS 500 Side 9 of 66
6 Package dimensions
Dimensions of holes for wall mounting (D = 6 mm):

Assembly instructions
DS 500 Side 10 of 66
7 Assembly instructions

Panel and wall mounting
DS 500 Side 11 of 66
8 Panel and wall mounting
Cut-out panel
Dimension of housing 270 x 170 x 90 mm
Dimension of holes for wall mounting D = 6 mm

Connection diagrams DS 500
DS 500 Side 12 of 66
9 Connection diagrams DS 500
9.1 DS 500 with 4 channels
X1.1:
power supply
X2.1, X2.2:
for internal use
X3.1 - X3.4:
alarm relay 1 – 4
X4.1, S4.1:
for CAN, RS485,
R120 configuration
XA.1 - XA.4:
sensor channels 1 – 4
9.2 DS 500 with 8 channels
X1.1:
power supply
X2.1, X2.2:
for internal use
X3.1 - X3.4:
alarm relay 1 – 4
X4.1:
for CAN, RS485,
R120 configuration
XA.1 - XA.4:
sensor channels 1 – 4
XB.1 - XB.4:
sensor channels 5 – 8

Connection diagrams DS 500
DS 500 Side 13 of 66
9.3 DS 500 with 12 channels
X1.1:
power supply
X2.1, X2.2:
for internal use
X3.1 - X3.4:
alarm relay 1 – 4
X4.1:
for CAN, RS485, R120
configuration
XA.1 - XA.4:
sensor channels 1 – 4
XB.1 - XB.4:
sensor channels 5 – 8
XC.1 - XC.4:
sensor channels 9 – 12
9.4 Power supply DS 500 standard version 100 - 240 VAC
X1.1
L1
N
PE
100 - 240 VAC, 50 - 60 Hz
9.5 X2.1 and X2.2 by standard version 100 - 240 VDC, wired ex works
X2.1, X2.2
L1´
N´
PE´
GND
U+ (24 VDC)
Only for internal use

Connection diagrams DS 500
DS 500 Side 14 of 66
9.6 Power supply by special version 24 VDC
X2.1
GND
U+ (24 VDC)
External supply 24 VDC (X 2. 2 not occupied)
Internal power supplies 100 - 240 VAC/24 VDC are not assembled.
The supply voltage 24 VDC connect directly into pin 4 and 5.
9.7 Connection 4 x alarm relay, max. 230 VAC, 6 A
X3.1 – X3.4
NO
COM
NC
X3.1: alarm relay 1
X3.2: alarm relay 2
X3.3: alarm relay 3
X3.4: alarm relay 4
NC and COM are closed by: alarm, voltage breakdown,
sensor break
9.8 Connection bus systems X4.1 and S4.1
X4.1
A
B
GND
Y
Z
S4.1
CAN S4, S5 ON
RS485 FULL S3, S6 ON
RS485 HALF S2, S7, S3 ON
TERMINATING 120R S1, S8 ON
DS 500 version with
4 channels
9.9 Connector pin assignment for all sensors XA.1 – XA.4, XB.1 – XB.4, XC.1 – XC.4
XA.1 – XA.4
XB.1 – XB.4
XC.1 – XC.4
RS485 terminating resistor ON/OFF
+ RS485
- RS485
SDI (CS-internal data transfer for all dew point and consumption
sensors)ANALOGUE IN + (current signal and voltage signal)
ANALOGUE IN – (current signal and voltage signal)
CURRENT SOURCE 500 µA
+VB, 24V DC sensor power supply
- VB-, GND sensor

Connection diagrams of the different sensor types
DS 500 Side 15 of 66
10 Connection diagrams of the different sensor types
By/With consumption sensors and dewpoint sensors from CS Instruments there is the possibility to
provide values as analogue current signal 4 – 20 mA for further working process.
The measurement/reading of the current signal for an external PLC/ZLT or external third-party display
is explained in the connection diagrams.
The following connection diagrams in Chapter 9 apply to XA.1 to XC.4!
FA serial: dewpoint sensors from CS Instruments
VA serial: consumption sensors from CS Instruments
10.1 Connection CS dewpoint sensors, serial FA 415/FA 300
FA 300
FA 415
The digital data transmission between DS 500 and the dewpoint sensors
FA 415 and FA 300 occur via the SDI bus line.
It´s possible to connect the FA 300/FA 425 alternatively as 4 – 20 mA
analogue sensor in 2-wire technology.
See chapter 9.3 analogue 4 - 20 mA sensors using the 2-wire technology.
To route the current signal of 4 - 20 mA to an external display, PLC, ZLT,
etc., please open bridge 5 and 8 and use external.
Please make sure that the circuit is closed in any case.
10.2 Connection for dew point- and consumption sensors, serial FA/VA 400
FA 410
FA 400
VA 400
VA 420
The digital data transmission between DS 500 and the sensors FA
400/410 and VA 400/420 occurs via the SDI bus line.
To route the current signal of 4 - 20 mA to an external display, PLC or
ZLT, please use PIN 5 and 8.

Connection diagrams of the different sensor types
DS 500 Side 16 of 66
10.3 Connection pulse sensors
signal level 0:
low = 0 – 0,7 VDC
signal level 1:
high = 2,5 – 30 VDC
t = 400 µs
max. frequency
(duty cycle 1:1) = 1000 Hz
input resistance:
min. 100 kilo ohm
external essential
R = 4K7
Attention:
The DS 500 is counting a
consumption unit, by
switching „power on“.
external essential
R = 4K7
It won’t work!

Connection diagrams of the different sensor types
DS 500 Side 17 of 66
10.4 Analogue two-, three-, and four-wire current signal
Sensors with 4 - 20 mA output in 2-wire technology
To route the current signal of 4 - 20 mA to an external display, PLC,
ZLT, etc., please open bridge 5 and 8 and use external.
Please make sure that the circuit is closed in any case.
(e.g. CS pressure sensors 1,6/16/40/100/250/400 bar or
temperature sensors with 4 - 20 mA)
Sensors with 0/4 - 20 mA output in 3-wire technology
To route the current signal of 4 - 20 mA to an external display, PLC,
ZLT etc. please open bridge 5 and 8 and use external.
Sensors with 0/4 - 20 mA output in 4-wire technology
To route the current signal of 4 - 20 mA to an external display, PLC or
ZLT:
Please split the cable between PIN 5 and the sensor and conduct the
current signal via PLC, ZLT direct back to the sensor, to close the
current circuit.

Connection diagrams of the different sensor types
DS 500 Side 18 of 66
10.5 Three- and four-wire power supply 0 - 1/10/30 VDC
Sensor with voltage output in 3-wire technology
Sensor with voltage output in 4-wire technology

Connection diagrams of the different sensor types
DS 500 Side 19 of 66
10.6 Two-, three-, and four-wire connector pin assignments for PT100/PT1000/KTY81
2-wire PT100/PT1000/KTY81
3-wire PT100/PT1000/KTY81
4-wire PT100/1000/KTY81
10.7 Connection with RS485
Sensor with RS485 interface

Connect the DS 500 with a PC
DS 500 Side 20 of 66
11 Connect the DS 500 with a PC
Important:
The IP addresses of PC and DS 500 must be statically assigned (DHCP off) and have to be
in the same network.
If the IP-address of the DS 500 has changed, you have to reboot!
Remark:
IP-address of the DS 500: See chapter, 12.2.4.3 Network settings
Reboot the DS 500: See chapter, 12.2.4.7 Factory Reset
The DS 500 can be connected with the PC by a crossover cable, which has a RJ45 plug on each side,
or an Ethernet cable with a crossover adapter.
Crossover cable with RJ45-plug
After connecting the DS 500 via a suitable cable to the PC, you can make graphical and tabular data
evaluations with the CS Soft Basic software.
Windows PC’s, network settings:
Windows 7:
Start Control Panel Network and Sharing Center adapter Networking Properties
Internet Protocol Version 4 (TCP/IPv4) Use the Following IP address enter IP address and
Subnet mask
After this: OK OK Close
Windows Vista:
Start Control Panel Network and Sharing Center
Network connection Networking Properties Internet Protocol Version 4 (TCP/IPv4) Use the
Following IP address enter IP address and Subnet mask
After this: OK OK Close
Windows XP:
Start Properties Control Panel Network connection Networking Properties Internet
Protocol Version 4 (TCP/IPv4) Use the Following IP address enter IP address and Subnet mask
After this: OK OK Close
Crossover-Adapter
Other manuals for DS 500
2
Table of contents
Other CS Instruments Voice Recorder manuals
Popular Voice Recorder manuals by other brands

Lumel
Lumel KD8 user manual

Panasonic
Panasonic RRUS551 - IC RECORDER Specifications

Sony
Sony ICD-MS1 - Memory Stick Ic Recorder operating instructions

Memoq
Memoq MR-710 owner's manual

Noise Engineering
Noise Engineering Mimetic Sequent manual

TS-market
TS-market EDIC-mini Tiny16 A46 Short operating instructions